Establishment of sex-specific reference intervals for automated haematology analyser-delivered research parameters in healthy Korean adults: a retrospective database review.

Kibum Jeon1, Miyoung Kim2, Junhee Han3, Jiwon Lee4, Jee-Soo Lee4, Han-Sung Kim4, Hee Jung Kang4, Young Kyung Lee4.   

Abstract

OBJECTIVES: Automated haematology analysers measure various parameters of relevance to clinical research along with routine complete blood count (CBC)-related components. We aimed to establish ethnicity-specific and sex-specific reference intervals for 26 research-specific parameters as well as 18 routinely reported components using a large cohort of healthy Korean adults. The necessity of requiring separate sex-specific reference intervals for each parameter was also examined.
DESIGN: A retrospective database review.
SETTING: Single tertiary-care hospital of approximately 375 physicians and 530 nurses. PARTICIPANTS: This study included 1383 reference individuals (840 men and 543 women). PRIMARY AND SECONDARY OUTCOME MEASURES: Following the Clinical and Laboratory Standards Institute guidelines for establishing reference intervals, routine CBCs as well as research parameters were measured using an ADVIA 2120i instrument.
RESULTS: All the routine components except for mean platelet volume and per cent lymphocytes differed significantly between men and women. Most research parameters also differed between the sexes; the exceptions were large platelets, platelet dry mass distribution width, per cent basophil saturation, per cent peroxidase saturation and per cent abnormal peroxidase absorption. Despite these differences, separate reference intervals for men and women were required only for two research-specific parameters: 'percentage high cellular haemoglobin' and 'percentage of hyperchromic red blood cells (RBCs)'.
CONCLUSION: Even though most parameters showed significant differences between men and women, none of the evaluated parameters except two RBC-related factors required separate reference intervals for each sex. © Author(s) (or their employer(s)) 2020. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Keywords:  general medicine (see internal medicine); haematology; haematopathology

Mesh:

Year:  2020        PMID: 33033084      PMCID: PMC7545637          DOI: 10.1136/bmjopen-2020-036887

Source DB:  PubMed          Journal:  BMJ Open        ISSN: 2044-6055            Impact factor:   2.692


A set of stringent criteria were used to select the healthy individuals for this study (although not all CLSI guideline criteria could be evaluated). The study is the largest of its kind to date. The study also examined the necessity of establishing sex-specific reference intervals. Our exclusion criteria did not incorporate all the rules mentioned in the CLSI guideline, but included additional rules to best exclude non-healthy individuals. We were unable to evaluate some other parameters that were found to be clinically significant in previous studies because limited data availability.

Introduction

The clinical decision-making process begins by comparing data obtained from a patient to reference values.1 To reliably interpret test results, accurate reference intervals that are derived from healthy individuals and categorised according to major covariates such as age, sex and/or ethnicity are required. An important step in establishing accurate reference intervals is selecting a sufficiently large number of healthy reference individuals representing relevant demographic groups.1–4 Achieving this requires establishing and applying a set of criteria to exclude non-healthy individuals from the reference population.1 Modern automated haematology analysers use up-to-date techniques including electrical impedance, radiofrequency conductivity, light scattering and/or cytochemistry.5 They enable the gathering not only of data that are routinely reported to patients but also various research parameters that deliver valuable clinical information.6 7 Such parameters include those that reflect the size and haemoglobinisation of red blood cells (RBCs)8 as well as the morphological parameters (cell size, cytoplasmic granularity and nuclear lobularity) of white blood cells (WBCs).5 Platelet-related parameters include the distribution width, plateletcrit (PCT), mean platelet component, mean platelet mass (MPM) and large platelet count.9 Previous representative studies found that certain research parameters have particularly useful clinical implications in practice. Kim et al showed that platelet indices including PCT, platelet component distribution width (PCDW), platelet dry mass distribution width (PMDW) and mean platelet volume (MPV) were useful for predicting the in-hospital mortality of patients suspected of having disseminated intravascular coagulation.10 Another study suggested that the reticulocyte haemoglobin content (CHr) correlated well with the Ferritin Index and was a useful marker of functional iron deficiency in blood donors.8 In a study that validated an algorithm-guided preoperative anaemia management method that raises the perioperative haemoglobin level and reduces blood transfusion volume, Enko et al used the CHr to define functional iron deficiency in their study subjects.11 Furthermore, Rocco et al detected acute promyelocytic leukaemia by combining six haematologic parameters including large unstained cells (LUCs), hyperchromic cells, per cent saturated cells, platelets, monocyte percentage and blast percentage.12 Such research parameters as well as others would be more applicable to real-world practice if reliable reference intervals are provided for each. Because there are no published research parameter reference intervals based on healthy adult Koreans to date, we aimed to establish appropriate sex-specific reference intervals for 26 research parameters as well as for 18 routinely reported complete blood count (CBC) components. We also performed a comparative analysis of all these parameters in healthy Korean men and women to investigate the necessity of establishing separate reference intervals for each sex. Importantly, we selected a large healthy population based on a set of stringent criteria to ensure high-quality test results.

Materials and methods

Selection of healthy reference individuals

An indirect sampling technique was used to select healthy reference individuals;1 the process is depicted in figure 1. The study initially recruited 8936 participants (4914 men and 4022 women) targeting those between 20 and 60 years of age (mean for total cohort: 46.0 years, range: 20.0–60.0 years; mean for men: 46.6 years, range: 20.0–60.0 years; and mean for women: 45.6 years, range: 20.0–60.0 years) who underwent regular health check-ups at Hallym University Sacred Heart Hospital, Republic of Korea, between September 2014 and April 2018 and whose CBCs were obtained using the ADVIA 2120i instrument (Siemens, Munich, Germany).
Figure 1

Schematic illustration of the selection process for reference individuals.

Schematic illustration of the selection process for reference individuals. Among these subjects, we selected only those who underwent electrocardiography, chest radiography, abdominal ultrasonography and laboratory tests that included CBC, chemistry, viral serology and urinalysis. During this process, 7264 participants (3919 men and 3345 women comprising 79.8% and 83.2% of the parent populations, respectively) were excluded owing to the lack of one or more required test results. Next, we applied exclusion criteria to extract non-healthy individuals (box 1); these were devised based on the Clinical and Laboratory Standards Institute (CLSI) EP28-A3 guideline as well as additional measures.1 Finally, 1383 reference individuals (840 men and 543 women representing 17.1% and 13.5% of their parent populations, respectively) were selected. The median age of all subjects in the final cohort was 48.0 years (range: 23.0–60.0 years); that of men was 51.0 years (range: 23.0–60.0 years), while that of women was 45.0 years (range: 24.0–60.0 years).

Laboratory

Hyperlipidaemia (low-density lipoprotein >3.362 mmol/L), diabetes mellitus (fasting blood glucose >5.55 mmol/L or glycated haemoglobin >7.8 mmol/L) Positivity for hepatitis B surface antigen and/or hepatitis C virus antibody Abnormal urinalysis results, including the presence of erythrocytes, granulocytes, glucose, protein or nitrite Serum iron level <6 μmol/L Haemoglobin concentration <90 g/L White blood cells<3.0×109/L or >12.5×109/L

Electrocardiography

Abnormal findings on electrocardiography, including ST segment elevation, pacemaker insertion or evidence of ischaemic heart disease

Imaging studies

Abnormal findings in chest radiography, including suspected thyroid disease, mediastinal tumortumour, active tuberculosis, pneumonia or lung cancer Abnormal findings in abdominal ultrasonography, including suspected alcoholic liver disease or liver cirrhosis

Sample collection and measurement of parameters

Venous whole blood (5 mL) was drawn from the cubital vein into a K2-ethylenediaminetetraacetic acid (EDTA) tube (Becton Dickinson, Franklin Lakes, New Jersey, USA) after fasting for at least 8 hours prior to phlebotomy. The tube was inverted five to six times and transported to the haematology laboratory. An ADVIA 2120i instrument was used to measure 18 routine CBC parameters and 26 research parameters, according to the manufacturer’s instructions, within 4 hours of sample collection. The 18 routine parameters included RBC count, haemoglobin, haematocrit (Hct), mean corpuscular volume (MCV), mean corpuscular haemoglobin (MCH), MCH concentration (MCHC), red cell distribution width, WBC, neutrophil count, neutrophil fraction, lymphocyte fraction, monocyte fraction, eosinophil fraction, basophil fraction, LUC fraction, platelets, MPV and platelet distribution width (PDW). We included 26 research parameters (12 RBC, 7 platelet, 3 basophil and 4 peroxisome) based on the data availability and usefulness as described in the previous studies.9 13 14 RBC parameters included cellular haemoglobin (CH), CH concentration mean, per cent high CH, per cent low CH, CH distribution width, percentage of hyperchromic RBCs (%Hyper), percentage of hypochromic RBCs, percentage of macrocytic RBCs, percentage of microcytic RBCs, RBC volume and haemoglobin concentration covariance, R count (red cell count as measured in a different channel from that used to obtain the routine parameter RBC count) and haemoglobin distribution width. Peroxidase-related parameters included the Mean Peroxidase Staining Index (MPXI), percentage of peroxidase saturation (Perox %Saturation), percentage of neutrophils with a high absorption value and percentage of abnormal peroxidase absorption (Perox %Abnormal). Basophil-related parameters included the Lobularity Index, percentage of polymorphonucleated neutrophils (which is distinct from the routinely measured neutrophil fraction) and percentage of basophil saturation (Baso %Saturation). Platelet parameters included large platelets, mean platelet component, MPM, two-dimensional platelet count, PCDW, PMDW and PCT.

Statistical analysis

Statistical analyses for the comparison of values between the two sex group, the establishment of reference intervals for each group and the necessity for separate reference intervals were performed as described in our previous study;15 namely, differences in values between the two groups were compared using the Mann-Whitney U test. P values<0.05 were considered statistically significant. The reference intervals for measured parameters were established according to a non-parametric method based on the CLSI EP28A-3C criteria for each sex group.1 The lower and upper reference limits (the values of the 2.5th and 97.5th percentiles, respectively) for each parameter were established, and the 90% CIs for the upper and lower limits of each reference interval were determined as follows; where Sy is the SD of the reference values and n is the number of values.16 Reed et al suggest a cut-off value of 1/3; that is, if the observed value of D (the absolute difference between an extreme observation and the next largest (or smallest) observation) was equal to or greater than one-third of the range R (range of all observations, including extremes), the extreme observation would be deleted. This method was used to exclude extreme outliers.1 17 To determine whether separate reference intervals for each sex group were necessary, we used the method of Harris and Boyd,1 18 who suggested criteria based on the ratio between subclass SD, a normal deviate test of means and calculation of critical decision values dependent on the sample size.19 It recommends partitioning reference intervals for different groups in the following situations: (1) the standard normal deviation (z) exceeds the critical value (z*), which is 3(naverage/120)1/2, where n is the number of individuals in the subgroup; (2) the larger standard deviation (s2) exceeds 1.5-fold the smaller standard deviation (s1), or equivalently, s2/(s2–s1) is less than 3 regardless of the z value. If these conditions are met, separate reference intervals should be calculated for each subclass assuming that the difference between the two reference intervals is likely to be of clinical importance.1 Data were analysed using the Statistical Package for Social Sciences V.24.0 (IBM Corp., Armonk, New York, USA) and MedCalc V.17.9.7 (MedCalc Software, Ostend, Belgium).

Patient/public involvement

There were no patients involved in this research.

Results

Reference intervals for 18 routine parameters

The calculated median (IQRs) and reference intervals with 90% CIs for the upper and lower limits of the reference intervals of 18 routine CBC parameters in each sex group are presented in table 1; histograms are also shown in online supplemental figure 1.
Table 1

Age-specific and sex-specific medians, IQRs and reference intervals for 18 complete blood count components obtained from healthy Korean adults

ParameterSexNMedianIQRP valueReference intervalLower 90% CIUpper 90% CI
RBC, ×1012/LM8404.94.7–5.1<0.0014.3–5.44.3 to 4.45.4 to 5.4
F5434.34.1–4.53.8–4.83.8 to 3.94.7 to 4.8
Hb, g/LM840152146–158<0.001137–167135 to 138166 to 169
F543131125–137115–144113 to 117143 to 144
Hct, L/LM8400.4420.421–0.4580.0030.397–0.4880.395 to 0.4010.483 to 0.492
F5430.3870.370–0.4010.346–0.4260.343 to 0.3490.423 to 0.428
MCV, fLM84090.688.1–93.3<0.00184.7–96.984.3 to 85.196.4 to 97.3
F54390.087.5–92.382.9–95.581.9 to 83.995.0 to 96.3
MCH, fmolM8401.9361.880–1.992<0.0011.812–2.0731.794 to 1.8182.067 to 2.091
F5431.8871.831–1.9491.719–2.0171.707 to 1.7382.005 to 2.029
MCHC, g/LM840344337–352<0.001327–361326 to 328361 to 363
F543339331–345320–356319 to 321355 to 357
RDW(proportion of 1.0)M8400.1250.123–0.1290.0050.118–0.1340.118 to 0.1190.133 to 0.135
F5430.1260.122–0.1310.117–0.1420.117 to 0.1180.140 to 0.145
WBC, ×109/LM8405.74.9–6.8<0.0013.9–8.93.7 to 4.08.6 to 9.2
F5435.24.9–6.23.6–8.23.5 to 3.77.9 to 8.8
Neutrophil count, ×109/LM8403.12.5–3.90.0011.5–6.61.5 to 1.66.2 to 7.5
F5432.92.3–3.71.6–5.91.4 to 1.75.5 to 6.8
Neutrophil (fraction)M8400.5510.493–0.607<0.0010.363–0.7290.351 to 0.3830.709 to 0.751
F5430.5690.514–0.6230.412–0.7260.384 to 0.4220.713 to 0.760
Lymphocyte (fraction)M8400.3290.279–0.3870.6620.180–0.4800.140 to 0.1900.478 to 0.500
F5430.3310.281–0.3790.188–0.4770.149 to 0.2030.458 to 0.491
Monocyte (fraction)M8400.0550.047–0.063<0.0010.034–0.0810.033 to 0.0370.079 to 0.087
F5430.0490.042–0.0580.031–0.0780.029 to 0.0330.075 to 0.082
Eosinophil (fraction)M8400.0270.017–0.042<0.0010.006–0.0880.005 to 0.0070.082 to 0.095
F5430.0210.014–0.0310.003–0.0690.004 to 0.0060.066 to 0.081
Basophil (fraction)M8400.0040.003–0.006<0.0010.001–0.0100.001 to 0.0010.010 to 0.010
F5430.0040.003–0.0050.001–0.0090.001 to 0.0010.008 to 0.011
LUC (fraction)M8400.0200.016–0.025<0.0010.012–0.0320.011 to 0.0120.031 to 0.035
F5430.0190.015–0.0230.011–0.0300.010 to 0.0120.029 to 0.032
PLT, ×109/LM840233205.0–263.0<0.001163.0–324.0161.0 to 167.0318.0 to 331.0
F543244211.5–277.5176.0–336.8169.0 to 179.0325.0 to 352.0
MPV, fLM8408.17.6–8.60.1907.1–9.57.0 to 7.19.3 to 9.6
F5438.07.6–8.57.2–9.47.1 to 7.29.2 to 9.5

Hb, haemoglobin; Hct, haematocrit; LUC, large unstained cell; MCH, mean corpuscular haemoglobin; MCHC, mean corpuscular haemoglobin concentration; MCV, mean corpuscular volume; MPV, mean platelet volume; PLT, platelets; RBC, red blood cells; RDW, red cell distribution width; WBC, white blood cell.

Age-specific and sex-specific medians, IQRs and reference intervals for 18 complete blood count components obtained from healthy Korean adults Hb, haemoglobin; Hct, haematocrit; LUC, large unstained cell; MCH, mean corpuscular haemoglobin; MCHC, mean corpuscular haemoglobin concentration; MCV, mean corpuscular volume; MPV, mean platelet volume; PLT, platelets; RBC, red blood cells; RDW, red cell distribution width; WBC, white blood cell. Among 18 routinely reported CBC parameters, 16 were significantly different in terms of median and IQR values between the 2 sexes. RBC, haemoglobin, Hct, MCV, MCH, MCHC, PDW, WBC, neutrophil count, monocyte fraction, eosinophil fraction, basophil fraction and LUC fraction were significantly higher in men than in women (all p values were <0.001 except for Hct, which was 0.003). Platelets, red cell distribution width and neutrophil fraction were significantly higher in women than in men (p<0.001, 0.005 and <0.001, respectively). There was no significant difference in MPV and lymphocyte fraction between the two groups (p>0.05). None of the 18 parameters required separate reference intervals for each sex.

Reference intervals for 26 research parameters

The calculated median (IQR) and reference intervals with 90% CIs for the upper and lower limits of reference intervals, as well as results of the analyses of whether separate reference intervals are required for each sex, are summarised in table 2; histograms are also shown in online supplemental figure 2.
Table 2

Age-specific and sex-specific medians, IQRs and reference intervals for 26 clinically relevant complete blood count parameters obtained from healthy Korean adults

ParameterSexNMedianIQRP valueReference intervalLower 90% CIUpper 90% CI
RBC parameter
 CH, fmolM8401.9181.862–1.973<0.0011.787–2.0671.775 to 1.8002.048 to 2.079
F5431.8621.806–1.9241.688–1.9981.651 to 1.7071.986 to 2.005
 CHCM, g/LM840342335–349<0.001327–359325 to 327358 to 362
F543335329–341320–351319 to 321350 to 354
 %High CH*M84038.129.0–47.7<0.00117.6–63.515.7 to 19.160.4 to 65.2
F54328.920.9–39.48.8–52.17.5 to 9.748.9 to 53.8
 %Low CHM84013.99.2–19.7<0.0014.9–30.84.4 to 5.528.4 to 32.4
F54319.312.6–26.87.3–47.96.4 to 8.044.5 to 54.4
 CHDW, fmolM8400.2230.217–0.230<0.0010.205–0.2420.205 to 0.2050.242 to 0.248
F5430.2170.211–0.2230.199–0.2420.199 to 0.2050.242 to 0.242
 %Hyper†M8401.00.6–1.7<0.0010.3–3.70.3 to 0.33.2 to 4.2
F5430.50.3–0.80.1–2.00.1 to 0.21.7 to 2.2
 %HypoM8400.30.2–0.5<0.0010.1–1.30.1 to 0.11.1 to 1.5
F5430.60.3–1.00.1–4.10.1 to 0.23.1 to 5.2
 %MacroM8400.50.3–1.1<0.0010.1–2.40.1 to 0.12.0 to 2.7
F5430.40.2–0.80.1–1.80.1 to 0.11.5 to 2.0
 %MicroM8400.30.2–0.5<0.0010.1–0.90.1 to 0.10.8 to 0.9
F5430.40.3–0.60.1–1.70.1 to 0.21.4 to 2.1
 RBC CovarM840−11.6−13.2 to −10.3<0.001−15.7 to −8.5−16.1 to −15.2−8.8 to −8.3
F543−11.0−12.2 to −9.6−14.5 to −7.4−15.0 to −14.2000−7.6 to −7.0
 R countM84046 500.544 754.5–48 335.0<0.00142 206.0–50 636.941 574.0 to 42 601.050 414.0 to 50 836.0
F54342 160.040 491.0–43 620.038 114.0–45 579.037 531.0 to 38 463.045 395.0 to 46 028.0
 HDW, g/LM8402524–27<0.00122–2922 to 2229 to 30
F5432422–2621–2821 to 2128 to 29
Peroxisome parameters
 MPXIM8400.9−1.4 to 3.00.005−4.5 to 5.6−5.2 to −4.15.2 to 5.9
F5431.3−0.7 to 3.2−3.9 to 6.6−4.5 to −3.35.9 to 7.1
 Perox %SaturationM8400.40.3–0.50.0980.2–1.00.2 to 0.20.9 to 1.0
F5430.40.3–0.50.1–1.00.1 to 0.20.9 to 1.0
 %HPXM8400.50.4–0.70.0150.2–1.30.2 to 0.31.2 to 1.3
F5430.50.3–0.60.2–1.20.2 to 0.21.2 to1.5
 Perox %AbnormalM84020.619.2–22.40.93016.8–25.516.6 to 17.125.1 to 25.7
F54320.619.0–22.517.0–26.216.9 to 17.225.5 to 27.5
Basophil parameters
 LIM8402.22.1–2.2<0.0012.0–2.32.0 to 2.02.3 to 2.3
F5432.22.2–2.32.0–2.32.0 to 2.12.3 to 2.3
 %PMNM84060.855.8–66.0<0.00147.1–73.446.2 to 48.672.2 to 74.4
F54362.657.7–67.650.6–74.849.7 to 51.873.7 to 76.4
 Baso %SaturationM8400.10.1–0.20.0600–0.30.0 to 0.00.3 to 0.3
F5430.10.1–0.20–0.30.0 to 0.00.3 to 0.4
Platelet parameter
 Large PLT, ×109/LM8404.03.0–6.00.8662.0–10.02.0 to 2.09.0 to 10.0
F5434.03.0–6.02.0–10.02.0 to 2.09.0 to 11.0
 MPC, g/LM840266250–276<0.001234–291232 to 236289 to 292
F543269258–278240–290238 to 243288 to 292
 MPM, fmolM8400.1240.118–0.1300.0040.112–0.1430.106 to 0.1120.137 to 0.143
F5430.1240.118–0.1300.112–0.1430.112 to 0.1120.143 to 0.149
 P countM84020041763.5–2268.5<0.0011396.3–2792.91356.0 to 1435.02736.0 to 2837.0
F54321801871.0–2467.01552.2–2969.81480.0 to 1600.02878.0 to 3142.0
 PCDW, g/LM8405350–57<0.00147–6147 to 4860 to 61
F5435250–5546–5946 to 4758 to 60
 PMDW, fmolM8400.0500.043–0.0560.1120.043–0.0620.043 to 0.0430.062 to 0.062
F5430.0500.043–0.0500.043–0.0620.043 to 0.0430.062 to 0.062
 PCT (proportion of 1.0)M8400.0020.002–0.002<0.0010.001–0.0030.001 to 0.0010.003 to 0.003
F5430.0020.002–0.0020.001–0.0030.001 to 0.0010.003 to 0.003

*Parameters that required a separate reference interval for each sex.

Baso % Saturation, percentage of saturated basophils; CH, cellular haemoglobin; CHCM, cellular haemoglobin concentration mean; CHDW, cell haemoglobin distribution width; HDW, haemoglobin distribution width; %high CH, percentage of high cellular haemoglobin; %HPX, percentage of neutrophils with a high absorption value; %Hyper, percentage of hyperchromic red blood cells; %Hypo, percentage of hypochromic red blood cells; LI, Lobularity Index; %Low CH, percentage of low cellular haemoglobin; %Macro, percentage of macrocytic red blood cells; %Micro, percentage of microcytic red blood cells; MPC, mean platelet component; MPM, mean platelet mass; MPXI, Mean Peroxidase Staining Index; PCDW, platelet component distribution width; P count, two-dimensional platelet count; PCT, plateletcrit; Perox % Abnormal, percentage of abnormal peroxidase absorption; Perox %Saturation, percentage of peroxidase saturation; PLT, platelets; PMDW, platelet dry mass distribution width; %PMN, percentage of polymorphonucleated neutrophils; RBC Covar, red blood cell RBC volume and haemoglobin concentration covariance; R count, red blood cell count.

Age-specific and sex-specific medians, IQRs and reference intervals for 26 clinically relevant complete blood count parameters obtained from healthy Korean adults *Parameters that required a separate reference interval for each sex. Baso % Saturation, percentage of saturated basophils; CH, cellular haemoglobin; CHCM, cellular haemoglobin concentration mean; CHDW, cell haemoglobin distribution width; HDW, haemoglobin distribution width; %high CH, percentage of high cellular haemoglobin; %HPX, percentage of neutrophils with a high absorption value; %Hyper, percentage of hyperchromic red blood cells; %Hypo, percentage of hypochromic red blood cells; LI, Lobularity Index; %Low CH, percentage of low cellular haemoglobin; %Macro, percentage of macrocytic red blood cells; %Micro, percentage of microcytic red blood cells; MPC, mean platelet component; MPM, mean platelet mass; MPXI, Mean Peroxidase Staining Index; PCDW, platelet component distribution width; P count, two-dimensional platelet count; PCT, plateletcrit; Perox % Abnormal, percentage of abnormal peroxidase absorption; Perox %Saturation, percentage of peroxidase saturation; PLT, platelets; PMDW, platelet dry mass distribution width; %PMN, percentage of polymorphonucleated neutrophils; RBC Covar, red blood cell RBC volume and haemoglobin concentration covariance; R count, red blood cell count. Among the peroxidase-related parameters, the percentage of neutrophils with a high absorption value was significantly higher in men than in women (p=0.005), while MPXI was significantly higher in women than in men (p=0.015). Perox %Saturation and Perox %Abnormal did not show any significant differences between the two groups (all p>0.05). Among the basophil-related parameters, Lobularity Index and the percentage of polymorphonucleated neutrophils were significantly higher in women than in men (all p<0.001). Baso %Saturation did not show any statistically significant difference between the two groups (p>0.05). Among RBC parameters, CH, CH concentration mean, %High, CH distribution width, %Hyper, percentage of macrocytic RBCs and red cell count were significantly higher in men than in women (all p<0.001), while per cent low CH, percentage of hypochromic RBCs, percentage of microcytic RBCs and RBC haemoglobin concentration covariance were significantly higher in women than in men (all p<0.001). Among platelet-related parameters, PCDW was significantly higher in men than in women (p<0.001), while the mean platelet component, MPM, two-dimensional platelet count and PCT were significantly higher in women than in men (all p<0.001 except for that of MPM, which was 0.004). Large platelet counts and PMDW were not significantly different between the two groups (p>0.05). Notably, separate reference intervals for the two sex groups were required for %High and %Hyper.

Discussion

We established reference intervals for 26 research-related parameters as well as 18 routinely obtained CBC components for healthy Korean adults. The healthy subjects were outpatients who voluntarily underwent medical check-ups at private facilities (at their own expense) that offer more extensive examinations than those provided under the national health check-up system. Therefore, we were able to obtain additional health status information that was helpful for selecting a healthy cohort. Our data showed that, even though many parameters were significantly different between the two sex groups, separate reference intervals were not required for most such parameters that we evaluated. When the values of 18 routine parameters were compared between the sexes, all except MPV and lymphocyte fraction showed significant differences, which was likely owing to the physiologic differences between the sexes and/or the large number of values (N) included. A notable finding was that none of the 18 routine CBC parameters required separate reference intervals for each sex even though sex-specific reference intervals are generally used for routine CBC parameters in clinical laboratories. Our results suggest that clinicians and laboratory personnel may be able to use a common set of reference intervals for both men and women. However, additional studies of populations of varying ethnicities as well as of other haematology analyser brands would help clarify this issue. Almost all of the 26 CBC research parameters we examined showed significant differences between the sexes; the exceptions included large platelets, PMDW, Baso %Saturation, Perox %Saturation and Perox %Abnormal. As mentioned above, this phenomenon might be a consequence of physiologic differences between men and women or of the large number of reference individuals used in our study. Only a few studies that established reference intervals for research parameters used the same instrument as ours; however, only 2 of the 26 parameters (per cent high CH and %Hyper) required separate reference intervals for each sex. These parameters are indicators of haemoglobin concentration and RBC size; thus, interpreting this result should take into consideration the fact that MCV and MCHC did not require separate reference intervals for each sex. Red cell counts and haemoglobin levels in women are significantly lower than those in men; this is known to be of clinical significance. We used Harris and Boyd’s method to determine whether separate reference intervals are required for men and women; however, the reasons for these well-known difference are not fully clear, and further discussion is needed. A few studies have investigated reference intervals for research parameters using the ADVIA 2120i, most of which established such reference intervals for particular parameters: Nikulshin et al reported a reference interval for MPXI in children,13 Oh et al did the same for MPXI in adults14 and Kim et al reported reference intervals for platelet parameters in a study of subjects ≥50 years of age.9 The MPIX was significantly higher in women than in men (1.3 (−0.7 to 3.2) and 0.9 (−1.4 to 3.0), respectively; p=0.015) in our study, these were comparable with the values obtained by Oh et al even though they did not directly compare the two sex groups.14 As they also did not investigate the necessity of separate reference intervals for each sex, we could not compare our findings with theirs. We also could not directly compare our results concerning platelet parameters with those of Kim et al9 because they limited their reference individuals to subjects over 50 years of age. Indeed, we were generally unable to compare our results with those of previous reference interval studies that used the same analyser because the reference individual selection processes in these studies differed from ours. Nikulshin et al excluded patients without haematological disorders and tumours; however, they did not apply any other exclusion criteria to select healthy reference individuals.13 Oh et al and Kim et al included those who underwent regular medical check-ups but likewise did not use any additional exclusion criteria when selecting their healthy individuals.9 14 These approaches are common in many studies that investigate reference intervals; however, they carry the risk of including non-healthy individuals. We selected healthy reference individuals by using a stringent set of exclusion criteria from a large parent population, and established reference intervals for 26 research parameters along with 18 routine CBC components. As such, our data ought to be reliable and easily applicable to other laboratories after a proper validation process. Our study was unique in the following aspects: (1) we selected healthy reference individuals based on stringent exclusion criteria; (2) the study was performed at a single centre, which helped minimise possible institutional or instrument-dependent bias; (3) the process of establishing reference intervals followed the CLSI guideline; and (4) we not only provided reference intervals, but also determined the necessity of sex-specific intervals for relevant parameters. A possible limitation of this study was that we did not evaluate all the exclusion criteria suggested in the relevant CLSI guideline.1 Blood donation, lactation, obesity, occupation and the use of oral contraceptives, all of which are listed as possible exclusion criteria in the CLSI guideline, may affect CBC results. However, these were not considered exclusion criteria in our study owing to the lack of relevant data;1 moreover, such factors are inferred to have a minimal impact on the CBC parameters of individuals undergoing regular health check-ups. Instead, our exclusion criteria did encompass other rules that were not directly mentioned in the CLSI guideline but could nevertheless have influenced our test results, such as endoscopy/colonoscopy data, imaging studies and urinalysis. Of note, the Reed method was used to exclude extreme outliers that could have originated from non-healthy individuals.16 Another limitation is that we were unable to evaluate some parameters that were found to be clinically significant in previous studies, such as reticulocyte-associated parameters (eg, CHr), owing to the lack of availability of the relevant data.11 The measurement of such parameters require additional reagents, which precluded their evaluation from the outset because we used an indirect, a posteriori sampling technique. An a priori approach may be helpful for providing reference intervals for such parameters. Lastly, the age range of our study population was relatively narrow, and the measured values may not have followed a normal distribution. Modelling method such as the Generalized Models for Location, Scale and Shape can help to address this issue by estimating percentiles. In summary, we established a reliable set of reference intervals for 18 routine CBC components and 26 research-specific parameters in Korean adults by selecting a large number of healthy reference individuals using stringent criteria. Our study showed that, even though most parameters showed statistically significant differences between men and women, none of the evaluated parameters, except for two that are RBC-related, required separate reference intervals for each sex. As such, our findings may have the capacity to serve as a model precedent for establishing more reliable reference intervals for research parameters during clinical laboratory testing.
  15 in total

1.  Novel cell population data from a haematology analyzer can predict timing and efficiency of stem cell transplantation.

Authors:  Vera Golubeva; Juliana Mikhalevich; Julia Novikova; Olga Tupizina; Svetlana Trofimova; Yekaterina Zueva
Journal:  Transfus Apher Sci       Date:  2013-12-20       Impact factor: 1.764

Review 2.  Estimating reliable paediatric reference intervals in clinical chemistry and haematology.

Authors:  Peter Ridefelt; Dan Hellberg; Mattias Aldrimer; Jan Gustafsson
Journal:  Acta Paediatr       Date:  2013-10-30       Impact factor: 2.299

3.  On dividing reference data into subgroups to produce separate reference ranges.

Authors:  E K Harris; J C Boyd
Journal:  Clin Chem       Date:  1990-02       Impact factor: 8.327

4.  Automated blood cell counts: state of the art.

Authors:  Mauro Buttarello; Mario Plebani
Journal:  Am J Clin Pathol       Date:  2008-07       Impact factor: 2.493

5.  Hematological indices in an older population sample: derivation of healthy reference values.

Authors:  C W Tsang; R Lazarus; W Smith; P Mitchell; J Koutts; L Burnett
Journal:  Clin Chem       Date:  1998-01       Impact factor: 8.327

6.  Biochemical marker reference values across pediatric, adult, and geriatric ages: establishment of robust pediatric and adult reference intervals on the basis of the Canadian Health Measures Survey.

Authors:  Khosrow Adeli; Victoria Higgins; Michelle Nieuwesteeg; Joshua E Raizman; Yunqi Chen; Suzy L Wong; David Blais
Journal:  Clin Chem       Date:  2015-06-04       Impact factor: 8.327

7.  Intracellular neutrophil myeloperoxidase level in pediatric patients: significant age and gender variability.

Authors:  S Nikulshin; I Tolstikova; A Bartule; D Kviluna; D Gravele; D Gardovska
Journal:  Int J Lab Hematol       Date:  2014-05-12       Impact factor: 2.877

8.  Prognostic value of platelet indices as determined by ADVIA 120 in patients suspected of having disseminated intravascular coagulation.

Authors:  H K Kim; J-E Kim; C K Ham; D S Lee; S Park; H-I Cho
Journal:  Int J Lab Hematol       Date:  2008-04       Impact factor: 2.877

9.  A comparison of complete blood count reference intervals in healthy elderly vs. younger Korean adults: a large population study.

Authors:  Eun Jin Lee; Miyoung Kim; Eunyup Lee; Kibum Jeon; Jiwon Lee; Jee-Soo Lee; Han-Sung Kim; Hee Jung Kang; Young Kyung Lee
Journal:  Clin Chem Lab Med       Date:  2019-04-24       Impact factor: 3.694

10.  The impact of an algorithm-guided management of preoperative anemia in perioperative hemoglobin level and transfusion of major orthopedic surgery patients.

Authors:  Dietmar Enko; Franz Wallner; Achim von-Goedecke; Christa Hirschmugl; Vinzenz Auersperg; Gabriele Halwachs-Baumann
Journal:  Anemia       Date:  2013-03-27
View more
  2 in total

1.  Blood metabolic and physiological profiles of Bama miniature pigs at different growth stages.

Authors:  Jiayuan Mo; Yujie Lu; Tianqi Xing; Di Xu; Kun Zhang; Shuai Zhang; Yubin Wang; Gang Yan; Ganqiu Lan; Jing Liang
Journal:  Porcine Health Manag       Date:  2022-08-08

2.  Reference intervals of complete blood count parameters for individuals aged 80 to 89 years in Guizhou, China: A STROBE-compliant retrospective study.

Authors:  Jinlie Long; Xiuhong Wang; Jianbo Yuan; Jianru Yang; Jie Zhou; Yuan Chen; Enxi Hu; Yuanzhong Zhou; Xun Min
Journal:  Medicine (Baltimore)       Date:  2022-10-07       Impact factor: 1.817

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.