| Literature DB >> 36188762 |
Grace Lavelle1,2, Meriel Norris2, Julie Flemming2, Jamie Harper2, Joan Bradley3, Helen Johnston3, Jennifer Fortune2,4, Andrea Stennett2,5, Cherry Kilbride2, Jennifer M Ryan2,4.
Abstract
Multiple wearable devices that purport to measure physical activity are widely available to consumers. While they may support increases in physical activity among people with multiple sclerosis (MS) by providing feedback on their performance, there is little information about the validity and acceptability of these devices. Providing devices that are perceived as inaccurate and difficult to use may have negative consequences for people with MS, rather than supporting participation in physical activity. The aim of this study was, therefore, to assess the validity and acceptability of commercially available devices for monitoring step-count and activity time among people with MS. Nineteen ambulatory adults with MS [mean (SD) age 52.1 (11.9) years] participated in the study. Step-count was assessed using five commercially available devices (Fitbit Alta, Fitbit Zip, Garmin Vivofit 4, Yamax Digi Walker SW200, and Letscom monitor) and an activPAL3μ while completing nine everyday activities. Step-count was also manually counted. Time in light activity, moderate-to-vigorous activity, and total activity were measured during activities using an Actigraph GT3X accelerometer. Of the 19 participants who completed the validity study, fifteen of these people also wore the five commercially available devices for three consecutive days each, and participated in a semi-structured interview regarding their perception of the acceptability of the monitors. Mean percentage error for step-count ranged from 12.1% for the Yamax SW200 to -112.3% for the Letscom. Mean step-count as manually determined differed to mean step-count measured by the Fitbit Alta (p = 0.002), Garmin vivofit 4 (p < 0.001), Letscom (p < 0.001) and the research standard device, the activPAL3μ (p < 0.001). However, 95% limits of agreement were smallest for the activPAL3μ and largest for the Fitbit Alta. Median percentage error for activity minutes was 52.9% for the Letscom and 100% for the Garmin Vivofit 4 and Fitbit Alta compared to minutes in total activity. Three inductive themes were generated from participant accounts: Interaction with device; The way the device looks and feels; Functionality. In conclusion, commercially available devices demonstrated poor criterion validity when measuring step-count and activity time in people with MS. This negatively affected the acceptability of devices, with perceived inaccuracies causing distrust and frustration. Additional considerations when designing devices for people with MS include an appropriately sized and lit display and ease of attaching and charging devices.Entities:
Keywords: acceptability; multiple sclerosis; physical activity; step-count; validity; wearable devices
Year: 2022 PMID: 36188762 PMCID: PMC9397948 DOI: 10.3389/fresc.2021.737384
Source DB: PubMed Journal: Front Rehabil Sci ISSN: 2673-6861
Description of activities.
|
|
|
|---|---|
| Deskwork | Participants sat at a desk, browsed the internet, made a phonecall, and typed for ten minutes. |
| Elevator | Participants travelled up and down three floors in an elevator. |
| Washing and drying | Participants washed and dried dishes at a sink for ten minutes |
| Stairs | Participants ascended and descended one flight of stairs. |
| Indoor walking | Participants walked indoors along a straight 10 m walkway at a self-selected speed. |
| Walking with obstacles | Participants walked indoors at a self-selected speed navigating around objects on the ground. |
| Outdoor walking | Participants walked outdoors along a predefined route on a path that included straight paths, bends, and stepping on and off curbs, at a self-selected speed. |
| Stationary cycling | Participants cycled for 10 minutes at a self-selected speed on an upright cycle ergometer. |
| Driving | Participants who arrived to their appointment by car drove a predefined loop of the university campus at 20 mph |
Participant characteristics.
|
|
|
| |
|---|---|---|---|
| Age, yr | 19 | 52.1 (11.9) | 55 (27, 72) |
| Women | 13 (68.4) | ||
| BMI, kg.m−2 | 19 | 29.4 (9.3) | 26.3 (16.9, 52.2) |
|
| |||
| Relapsing-remitting | 16 (84.2) | ||
| Secondary progressive | 2 (10.5) | ||
| Benign | 1 (5.3) | ||
| Relapse in past 3–12 months | 5 (26.3) | ||
| MS duration, yr | 19 | 14.4 (11.8) | 13 (0, 48) |
|
| |||
| 1.0–4.0 | 10 (52.6) | ||
| 4.5–5.0 | 9 (47.4) | ||
|
| |||
| No aid | 12 (63.2) | ||
| Sticks or crutches | 2 (10.5) | ||
| Combination of sticks/crutches and | 5 (26.3) | 1 (0, 5) | |
| PDDS | 19 | 1 (0, 5) | |
| Walking speed, m/s | 19 | 0.98 (0.36) | 0.85 (0.64, 2.18) |
Description of steps measured by each device and manually counted for each activity.
|
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|---|
|
| |||||||
| Median (min, max) | 0 (0, 19) | 0 (0, 0) | 0 (0, 15) | 0 (0, 6) | 0 (0, 11) | 0 (0, 0) | 0 (0, 0) |
|
| |||||||
| Mean (SD) | 10.2 (6.7) | 12.1 (6.5) | 11.6 (10.8) | 12.7 (8.4) | −1.3 (5.7) | 6.1 (3.0) | 22.9 (7.6) |
| Median (min, max) | 11 (0, 25) | 14 (0, 21) | 11 (0, 30) | 14 (0, 21) | 0 (-25, 0) | 5.5 (3, 14) | 21 (14, 48) |
|
| |||||||
| Mean (SD) | 169.8 (125.6) | 15.4 (13.4) | 143.4 (71.6) | 19.0 (13.4) | 406.6 (255.2) | 7.4 (4.5) | 54.1 (29.0) |
| Median (min, max) | 130 (24, 504) | 13 (6, 67) | 130 (59, 315) | 15 (4, 57) | 307 (0, 840) | 5.5 (2, 19) | 40 (20, 131) |
|
| |||||||
| Mean (SD) | 26.2 (11.3) | 28.4 (8.2) | 28.3 (15.3) | 26.1 (8.2) | 18.8 (19.6) | 13.9 (2.8) | 32.4 (4.6) |
| Median (min, max) | 30 (0, 41) | 31 (0, 36) | 34 (0, 45) | 31 (0, 36) | 16 (0, 43) | 14 (10, 20) | 31.5 (27, 45) |
|
| |||||||
| Mean (SD) | 32.6 (26.2) | 37.0 (7.2) | 37.5 (18.1) | 36.6 (8.8) | 32.3 (23.3) | 18.4 (6.5) | 40.3 (12.7) |
| Median (min, max) | 33 (-62, 76) | 36 (18, 56) | 39 (0, 78) | 36 (25, 67) | 37 (0, 86) | 16 (13, 35) | 36 (28, 77) |
|
| |||||||
| Mean (SD) | 274.4 (72.9) | 254.4 (50.2) | 280.8 (109.4) | 182.6 (99.4) | 271.4 (81.7) | 137.5 (45.7) | 304.1 (187.0) |
| Median (min, max) | 258.5 (219, 530) | 256 (119, 374) | 263 (144, 667) | 236 (0, 286) | 265.5 (152, 552) | 128 (107, 287) | 262 (218, 976) |
|
| |||||||
| Median (min, max) | 81 (−9, 642) | 0 (0, 544) | 0 (0, 777) | 1 (0, 154) | 0 (0, 480) | 0 (0, 0) | 0 (0, 0) |
|
| |||||||
| Median (min, max) | 29.5 (0, 144) | 2 (0, 14) | 112 (19, 236) | 40 (1, 66) | 0 (-29, 99) | 0 (0, 0) | 0 (0, 0) |
|
| |||||||
| Mean (SD) | 47.9 (25.1) | 47.3 (9.3) | 61.3 (28.6) | 42.1 (14.7) | 57.0 (26.7) | 26.2 (10.3) | 61.1 (44.8) |
| Median (min, max) | 43 (11, 123) | 49.5 (19, 59) | 55 (0, 130) | 42.5 (16, 80) | 48.5 (0, 120) | 23, (19, 56) | 48 (8, 222) |
|
| |||||||
| Mean (SD) | 745.3 (357.8) | 448.6 (249.7) | 693.5 (289.6) | 323.1 (118.7) | 832.5 (314.7) | 179.6 (83.7) | 442.5 (269.6) |
| Median (min, max) | 745 (147, 1425) | 409 (83, 956) | 730 (190, 1259) | 365 (91,491) | 965 (316, 1378) | 190.5 (45, 428) | 450 (101, 1442) |
n = 19 for all devices except for activPAL3μ (n = 16) unless stated otherwise;
n = 16 for all devices except for activPAL3μ (n = 13);
n = 16 for all devices except for activPAL3μ (n = 13) and manual count (n = 15);
n = 17 for all devices except for activPAL3μ (n = 15);
n = 14 for all devices except for activPAL3μ (n = 13);
n = 18 for all devices except for activPAL3μ (n = 15).
Percentage error between visually counted steps and steps from each monitor.
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| Fitbit Alta | −80.6 (−113.0, −48.3) | −216.7, 48.3 | 19 (100) | 18 (95) | 18 (95) |
| Fitbit Zip | −14.9 (−47.0, 17.2) | −151.6, 82.8 | 17 (89) | 14 (74) | 8 (42) |
| Garmin Vivofit 4 | −67.6 (−90.9, −44.2) | −168.4, 28.8 | 18 (95) | 18 (95) | 16 (84) |
| Yamax SW200 | 12.1 (−5.2, 29.4) | −79.5, 93.7 | 16 (84) | 13 (68) | 7 (37) |
| Letscom | −112.3 (−148,8, −75.9) | −318.8, 32.7 | 19 (100) | 19 (100) | 18 (95) |
| activPAL3μ | 59.0 (56.6, 61.3) | 54.2, 70.3 | 19 (100) | 19 (100) | 19 (100) |
Positive value indicates the device underestimates steps in comparison to manual count.
n = 16.
Difference and agreement between visually counted steps and steps from each monitor.
|
|
|
|
|
|---|---|---|---|
| Fitbit Alta | −302.84 (-483.32, −122.36) | 0.002 | −1036.8, 431.1 |
| Fitbit Zip | −6.16 (−181.05, 168.73) | 0.942 | −717.4, 705.0 |
| Garmin Vivofit 4 | −251.05 (-375.15, −126.95) | <0.001 | −755.7, 253.6 |
| Yamax SW200 Digi-walker pedometer | 119.37 (−32.46, 271.19) | 0.116 | −498.0, 736.8 |
| Letscom | −390.00 (−541.65, −238.35) | <0.001 | −1006.7, 226.7 |
| activPAL3μ | 277.31 (165.72, 388.90) | <0.001 | −133.1, 687.8 |
Mean difference calculated as manually counted total steps minus monitor total steps.
n = 16.
Description of activity time measured by Fitbit Alta, Garmin Vivofit 4, Letscom, and Actigraph GT3x.
|
|
|
|
|
| |||||
|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
| ||
| Fitbit Alta | 19 | 3.0 (10.2) | 0 (0, 43) | n/a | n/a | n/a | n/a | n/a | n/a |
| Garmin Vivofit 4 | 19 | 0 (0) | 0 (0, 0) | n/a | n/a | n/a | n/a | n/a | n/a |
| Letscom | 18 | 18.7 (5.3) | 18.5 (8, 31) | n/a | n/a | n/a | n/a | n/a | n/a |
| Actigraph GT3x+ | 18 | n/a | n/a | 34.6 (11.7) | 31.5 (17, 56) | 3.2 (3.2) | 3 (0, 11) | 37.7 (12.1) | 35 (17, 57) |
Percentage error, mean difference, and 95% limits of agreement between total activity measured by Actigraph GT3x and activity minutes from each monitor.
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
|
| |||||||
| Letscom | 17 | 52.9 | 5.6, 65.1 (59.6) | 17 (100) | 16 (94.1) | 15 (88.2) | <0.001 |
| Fitbit Alta | 18 | 100.0 | −38.7, 100.0 (138.7) | 18 (100) | 18 (100) | 18 (100) | <0.001 |
| Garmin Vivofit 4 | 18 | 100.0 | 100.0, 100.0 (0.0) | 18 (100) | 18 (100) | 18 (100) | <0.001 |
|
| |||||||
| Letscom | 17 | 48.6 | 5.3, 63.3 (58.1) | 17 (100) | 15 (88.2) | 14 (82.3) | <0.001 |
| Fitbit Alta | 18 | 100.0 | −38.7, 100.0 (138.7) | 18 (100) | 18 (100) | 18 (100) | <0.001 |
| Garmin Vivofit 4 | 18 | 100.0 | 100.0, 100.0 (0.0) | 18 (100) | 18 (100) | 18 (100) | <0.001 |
|
| |||||||
| Letscom | n/a | n/a | n/a | n/a | n/a | <0.001 | |
| Fitbit Alta | n/a | n/a | n/a | n/a | n/a | 0.052 | |
| Garmin Vivofit 4 | n/a | n/a | n/a | n/a | n/a | <0.001 | |
Positive value indicates the device underestimates steps in comparison to Actigraph GT3x
p value obtained from Wilcoxon signed rank test comparing activity time from each device to Actigraph GT3X.
Unable to calculate percentage error because some people had 0 minutes of moderate-to-vigorous physical activity as measured by the Actigraph GT3X.