| Literature DB >> 36185875 |
Toufic R Jildeh1, Charles A Su2, Matthew L Vopat2, Justin R Brown2, Johnny Huard3.
Abstract
Platelet-rich plasma (PRP) is a promising therapy treatment option for multiple orthopedic conditions, which has demonstrated expanding clinical use. With increased clinical use of PRP, there has been a greater demand for point-of-care (POC) biologic devices. For this review, publicly available information provided by the device corporations, PubMed, Medline, and Embase databases were searched for studies related to POC device function. A scoping review study design was selected to explore the breadth of knowledge in the literature regarding PRP POC devices. ProofPoint Biologics demonstrated the highest laboratory platelet increase (5.2 ± 0.28-fold) and the longest processing time (49 ± 1.4 minutes). Celling demonstrated the lowest laboratory platelet increase (2.7 ± 0.8-fold), while AcCELLerated had the fastest processing time (18 ± 1.4 minutes for PurePRP® AB60 Pure (Pure Accelerated Biologics, Tequesta, FL) and 13.5 ± 2.1 minutes for AbsolutePRP® (Emcyte Corporation, Fort Myers, FL)). Celling had the lowest cost out of the various biologic devices. There is significant variability in the technical features, cost, processing time, and centrifugation parameters of the different commercially available point-of-care devices.Entities:
Keywords: biologics; orthobiologics; platelet rich plasma; point-of-care; point-of-care biologics
Year: 2022 PMID: 36185875 PMCID: PMC9514393 DOI: 10.7759/cureus.28498
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Summary of commercially available point-of-care platelet-rich plasma processing.
*Two out of four celling biologics samples produced platelet yields below the baseline. BMC: bone marrow concentrate and PRP: platelet-rich plasma.
| Product | Advertised platelet increase (fold) | Laboratory platelet increase (fold; mean ± SD) | Laboratory processing time, minutes (mean ± SD) | Advantages | Disadvantages |
| ProofPoint | N/A | 5.2 (±0.28) | 49 (±1.4) | Platelet counts increase consistently by more than four-fold | Manual transfer of blood products using pipette system time-intensive two centrifugation steps |
| AcCELLerated | 4 | 5.2 (±0.3) | 18 (±1.4) (PurePRP® AB60 Pure, Pure Accelerated Biologics, Tequesta, FL) 13.5 (±2.1) minutes (AbsolutePRP®, Emcyte Corporation, Fort Myers, FL) | Shortest processing time offer PRP, BMC, and adipose platelet counts to increase consistently more than four-fold | |
| Arthrex | 2-3 | 4.2 (±2.0) | 15.8 (±6.4) | Decreased processing time representatives available in the operating room | Inconsistent platelet increases |
| Celling | N/A | 2.7 (±0.8)* | 29 (±5.2) | Convenient use | Variable results unable to reach four-fold increase device lost many platelets in two trials variability in results with altitude |
| Terumo | 3.62 | 4.1 (±1.9) | 24.43 (±5.56) | Capable of reaching a four-fold concentration of platelets | Significant variability with plasma concentrating longer processing times BMC lost platelets during the demo |