| Literature DB >> 36177007 |
Aravinthan Loganathan1,2, Fionnuala McMorrow2, Hui Lu2, Danyang Li2, Ben Mulhearn1,2, Neil John McHugh2, Sarah Louise Tansley1,2.
Abstract
Background: The reliable detection of myositis-specific autoantibodies (MSA) provides valuable clinical information regarding prognosis, clinical progression and diagnostic confirmation.Entities:
Keywords: ELISA - enzyme-linked immunosorbent assay; diagnostic accuracy; diagnostic test; immunoprecipitation (IP); inflammatory myositis; myositis autoantibody
Mesh:
Substances:
Year: 2022 PMID: 36177007 PMCID: PMC9514093 DOI: 10.3389/fimmu.2022.975939
Source DB: PubMed Journal: Front Immunol ISSN: 1664-3224 Impact factor: 8.786
Number of sera samples tested on ELISA for both case samples and health controls (HC) and the number of samples with a co-efficient variance of equal to or less than 20%.
| Autoantibody | +MSA Sample Number | HC Sample Number | +MSA sample with ++CV ≤20% | HC sample with ++CV ≤20% | Pre-defined positive cut-off (au) |
|---|---|---|---|---|---|
| Mi2 | 20 | 63 | 19 | 63 | 53 |
| MDA5 | 20 | 63 | 20 | 62 | 32 |
| ARS* MSA | 76 | 120 | 70 | 100 | 25 |
| Jo1 | 22 | 66 | 21 | 61 | 25 |
| EJ | 10 | 54 | 7 | 50 | 25 |
| KS | 3 | 47 | 2 | 47 | 25 |
| PL-7 | 21 | 65 | 21 | 60 | 25 |
| PL-12 | 20 | 64 | 19 | 58 | 25 |
*Aminoacyl tRNA synthetase myositis specific antibodies combined (Jo1, EJ, KS, PL7, PL12).
+Myositis specific antibody samples.
++Co-efficient of variation.
Sensitivity and specificity of autoantibodies in comparison to IP with Cohen’s K.
| Autoantibody | Sensitivity (%) (95% CI) | Specificity (%) (95% CI) | *Cohen’s κ (95% CI) ± SE |
|---|---|---|---|
|
| 94.7 (74-99.9) | 100 (94.3-100) | 0.97 (0.9-1) |
|
| 80 (56.3-94.3) | 100 (94.2-100) | 0.86 (0.7-1) |
|
| 98.6 (92.3-100) | 99.1 (95.2-100) | 0.98 (0.95-1) |
|
| 100 (83.9-100) | 100 (94.1-100) | 1 |
|
| 100 (59–100) | 100 (92.9-100) | 1 |
|
| 100 (15.8-100) | 100 (92.8-100) | 1 |
|
| 100 (83.9-100) | 100 (94–100) | 1 |
|
| 94.7 (74-99.9) | 98.3 (90.8-100) | 0.93 (0.8-1) |
*Value of Cohen’s κ and strength of agreement.
Figure 1Mi2 ELISA values for 19 Mi2 and 63 healthy control samples.
Figure 2MDA5 ELISA values for 20 MDA5 and 62 healthy control samples.
Figure 3ARS ELISA values for 21 Jo1, 7 EJ, 2 Ks, 21 PL7, 19 PL12 and 117 healthy control samples.
Figure 4Individual ELISA values for individual ARS antibodies and HC samples.
New sensitivity and specificity of autoantibodies using Youden’s Index for optimal cut-off points in our cohort of patients.
| Autoantibody | Current ELISA Ab cut-off (au) | New ELISA Ab cut-off (au) | AUC (95%CI) | Sensitivity 95% CI (current cut-off) | Sensitivity 95% CI (new cut-off) | Specificity 95% CI (current cut-off) | Specificity 95%CI (new cut-off) |
|---|---|---|---|---|---|---|---|
|
| 53 | 34.3 | 1 (0.91 – 1) | 94.7 (74-99.9) | 100 (82.4-100) | 100 (94.3-100) | 100 (96-100) |
|
| 32 | 6.7 | 0.974 (0.91-1) | 80 (56.3-94.3) | 90 (68.3-98.8) | 100 (94.2-100) | 98.4 (91.3-100) |
|
| 25 | 22.2 | 0.997 (0.973-1) | 98.6 (92.3-100) | 98.6 (92.3-100) | 99.1 (95.2-100) | 99 (94.5-100). |
|
| 25 | 20.2 | 1 (0.956-1) | 100 (83.9-100) | 100 (83.9-100) | 100 (94.1-100) | 100 (94.1-100) |
|
| 25 | 16.6 | 1 (0.937-1) | 100 (59-100) | 100 (59-100) | 100 (92.9-100) | 100 (92.9-100) |
|
| 25 | 16.6 | 1 (0.927 -1 | 100 (15.8-100) | 100 (15.8-100) | 100 (92.8-100) | 100 (92.5-100) |
|
| 25 | 16.6 | 1 (0.91-1) | 100 (83.9-100 | 100 (83.9 – 100) | 100 (94-100) | 100 (93.9-100) |
|
| 25 | 22.2 | 0.993 (0.94-1) | 94.7 (74-99.9) | 95 (75.14-99.9) | 98.3 (90.8-100) | 100 (93.7 – 100 |