| Literature DB >> 36166434 |
Emerson Aparecido Miguel1, Paulo Roberto Laranjeira2, Marina Ishii3, Terezinha de Jesus Andreoli Pinto4.
Abstract
BACKGROUND: Non-condensable gases (NCGs) are all gases that do not undergo liquefaction during the saturated steam sterilization process. During a sterilization cycle, the NCGs presence inside the chamber is one of the biggest threats to the sterilization process compromising process validation and product quality.Entities:
Mesh:
Substances:
Year: 2022 PMID: 36166434 PMCID: PMC9514632 DOI: 10.1371/journal.pone.0274924
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.752
Fig 1System representation for measuring non-condensable gases (NCG) used in the tests, adapted from EN285:2015, where: a) before the collection, b) after the measurement according to the norm.
Treatment steps and parameters of purified water and purified water for injection or water for injection considering pharmaceutical compendiums [23–26].
| Parameters | Brazilian Pharmacopeia | United States Pharmacopeia | European Pharmacopeia | Japanese Pharmacopeia | |
|---|---|---|---|---|---|
|
| Treatment | Reverse osmosis or a combination of purification techniques from drinking water | Reverse osmosis or a combination of purification techniques from drinking water | Distillation, ion exchange, reverse osmosis, or another suitable process | Distillation, ion exchange, reverse osmosis, ultrafiltration, or another suitable process |
| Conductivity (25°C) | max. 1.3 μS/cm | max. 1.3 μS/cm | max. 5.1 μS/cm | max. 1.3 μS/cm | |
| Heterotrophic bacteria | max. 100 UFC/mL | max. 100 UFC/mL | max. 100 UFC/mL | max. 100 UFC/mL | |
| Bacterial endotoxin | not included | Not included | <0,25 UE/mL | Not included | |
| Total Organic Carbon | ≤ 0.50 mg/L | ≤ 0.50 mg/L | ≤ 0.50 mg/L | ≤ 0.50 mg/L | |
|
| Treatment | Distillation or similar process from drinking or purified water | Reverse osmosis or distillation | Distillation | Distillation, reverse osmosis, ultrafiltration, or a techniques combination |
| Conductivity (25°C) | max. 1.3 μS/cm | max. 1.3 μS/cm | max. 1.3 μS/cm | max. 1.3 μS/cm | |
| Heterotrophic bacteria | max. 10 UFC/mL | max. 10 UFC/mL | max. 10 UFC/mL | max. 10 UFC/mL | |
| Bacterial endotoxin | < 0.25 EU/mL | < 0.25 EU/mL | < 0.25 EU/mL | < 0.25 EU/mL | |
| Total Organic Carbon | ≤ 0.50 mg/L | ≤ 0.50 mg/L | ≤ 0.50 mg/L | ≤ 0.50 mg/L |
Non-condensable gases concentration tests performed, approved, or failed tests (%) based on the type of water used for steam generation, from September/16 to August/21.
| Water quality | Total Tests | Approved tests | Failed tests | |||
|---|---|---|---|---|---|---|
| (n) | (%) | (n) | (%) | (n) | (%) | |
| Softened | 09 | 5.3 | 04 | 44.4 | 05 | 55.6 |
| One-step reverse osmosis | 14 | 8.2 | 08 | 57.1 | 06 | 42.9 |
| Purified Water (PW) | 68 | 40.0 | 24 | 35.3 | 44 | 64.7 |
| Water for Injection (WFI) | 79 | 46.5 | 73 | 92.4 | 06 | 7.6 |
| Total | 170 | 100.0 | 109 | 64.2 | 61 | 35.8 |
Fig 2Interval plot of NCGs concentration versus types of water used for steam generation.
Descriptive statistics (mean, standard deviation, and 95% confidence intervals) for each separate water type group analyzed (softened, reverse osmosis, purified, and WFI).
| Type of water | Tests (n) | Mean value | Standard Deviation | CI 95% |
|---|---|---|---|---|
| SW | 27 | 4.822 | 4.55 | (2.644; 6.149) |
| RO | 42 | 3.495 | 2.624 | (2.013; 4.978) |
| PW | 204 | 6.621 | 7.298 | (5.949; 7.294) |
| WFI | 237 | 1.8105 | 1.4106 | (1.1864; 2.4347) |
SW = softened water; RO = one-step reverse osmosis; PW = purified water; WFI = water for injection.
Fig 3Box-plot of distributions of NCGs concentration comparing tests performed with reverse osmosis water (RO) and water for injection.
Fig 4Histogram for NCGs concentration (Ho and 95% CI) for processes using WFI for steam generation.
Descriptive statistics (n, mean, standard deviation, and 95% confidence intervals) of process systems with a concentration of NCGs below 3.5%.
| Process System | N | Mean | Standard Deviation | 95% CI |
|---|---|---|---|---|
| Autoclave+ ESG +WFI | 78 | 1.760 | 1.213 | (0.736; 2.784) |
| Autoclave+ PSG +WFI | 78 | 2.240 | 1.314 | (1.216; 3.264) |
| Bioreactor+ PSG +WFI | 12 | 0.883 | 1.083 | (-1.727; 3.494) |
| Filling machine+ PSG +WFI | 12 | 0.600 | 0.457 | (-2.010; 3.210) |
| Reactor+ PSG +WFI | 18 | 1.306 | 0.847 | (-0.826; 3.437) |
| Tank+PSG+WFI | 15 | 0.893 | 0.859 | (-1.442; 3.228) |
PSG = Pure Steam Generator; ESG = Electric Steam Generator; PW = purified water; SW = softened water; RO = reverse osmosis water; WFI = water for injection.