A radiation safety survey on a Clinac-20 linear accelerator.

N L Powell, A Newing, M A Bullen, C Sims, S F Leaton.   

Abstract

The radiation protection problems associated with a 20 MV linear accelerator are discussed, in particular those due to the considerable neutron production accompanying the 18 MV photon beam. These include the potential neutron dose to the patient, the special shielding requirements and the neutron activation of the accelerator and its surroundings. The results of the structural protection survey are described. In particular, the methods recommended by the NCRP for predicting the neutron dose equivalent at the end of a maze are shown to be reliable. Dose rates due to activation of the treatment room and various parts of the accelerator were measured, and potential doses to staff were assessed. Neutron dose equivalents were measured in the specified two metre diameter patient plane using the standard NRPB neutron badges and also with the recently developed CR-39 dosimeters. The latter appear to give sensible results.

Entities:  

Mesh:

Year:  1987        PMID: 3615576     DOI: 10.1088/0031-9155/32/6/004

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Nationwide survey on the operational status of electron accelerators for radiation therapy in Japan.

Authors:  Ichiro Yamaguchi; Shinji Tanaka; Toshioh Fujibuchi; Tetsuo Kida; Hiroaki Nagaoka; Hiroshi Watanabe
Journal:  Radiol Phys Technol       Date:  2010-01-22

2.  Evaluation of equivalent dose from neutrons and activation products from a 15-MV X-ray LINAC.

Authors:  Isra Israngkul-Na-Ayuthaya; Sivalee Suriyapee; Phongpheath Pengvanich
Journal:  J Radiat Res       Date:  2015-08-11       Impact factor: 2.724

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.