Literature DB >> 36129421

Delta Robot Kinematic Calibration for Precise Robot-Assisted Retinal Surgery.

Boyang Xiao1, Alireza Alamdar1, Kefan Song1, Ali Ebrahimi1, Peter Gehlbach2, Russell H Taylor1, Iulian Iordachita1.   

Abstract

High precision is required for ophthalmic robotic systems. This paper presents the kinematic calibration for the delta robot which is part of the next generation of Steady-Hand Eye Robot (SHER). A linear error model is derived based on geometric error parameters. Two experiments with different ranges of workspace are conducted with laser sensors measuring displacement. The error parameters are identified and applied in the kinematics to compensate for modeling error. To achieve better accuracy, Bernstein polynomials are adopted to fit the error residuals after compensation. After the kinematic calibration process, the error residuals of the delta robot are reduced to satisfy the clinical requirements.

Entities:  

Year:  2022        PMID: 36129421      PMCID: PMC9484559          DOI: 10.1109/ismr48347.2022.9807517

Source DB:  PubMed          Journal:  Int Symp Med Robot        ISSN: 2771-9049


  4 in total

1.  Accuracy assessment and kinematic calibration of the robotic endoscopic microsurgical system.

Authors:  Paul Wilkening; Yunuscan Sevimli; Marcin Balicki; Kevin C Olds; Russell H Taylor
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2016-08

Review 2.  Microsurgical robotic system for vitreoretinal surgery.

Authors:  Yoshiki Ida; Naohiko Sugita; Takashi Ueta; Yasuhiro Tamaki; Keiji Tanimoto; Mamoru Mitsuishi
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-05-15       Impact factor: 2.924

Review 3.  Robotic Vitreoretinal Surgery.

Authors:  Roomasa Channa; Iulian Iordachita; James T Handa
Journal:  Retina       Date:  2017-07       Impact factor: 4.256

4.  Micron: an Actively Stabilized Handheld Tool for Microsurgery.

Authors:  Robert A Maclachlan; Brian C Becker; Jaime Cuevas Tabarés; Gregg W Podnar; Louis A Lobes; Cameron N Riviere
Journal:  IEEE Trans Robot       Date:  2011-11-18       Impact factor: 5.567

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.