Hamdan Alghamdi1, David Sanderson1, Lorna Carmichael1, Alan Cresswell1, L Martin1.
Abstract
During recovery phases following a nuclear or radiological incident analyses of doses received by members of the public and responders are often required. Several methods have been investigated for use at different timescales after the incident, including assessments based on measurements of materials present at the time of the incident. Common salt has previously been shown to have potential for retrospective dosimetry in the mGy dose range using laboratory instrumentation. This preliminary study investigates the use of portable instruments, with unprepared commercially sourced salt, in dose ranges below 100 μGy. Responses from pulsed IRSL and portable OSL instruments were compared. For OSL measurements, detection limits of 7 μGy have been demonstrated, with detection limits of 30-340 μGy for the other instruments investigated. Dose responses in the 0-500 μGy range were determined for the most sensitive systems, which show a linear response over this dose range with a non-zero intercept representing doses received from environmental sources since manufacture of the salt. For use as a dosimeter, methods of removing or accounting for inherited signals will be required in this low dose range. The results demonstrate that salt has considerable potential for use in retrospective dosimetry below 100 μGy, and that measurements can be conducted with portable OSL instruments.Entities:
Keywords: NaCl; emergency dosimetry; optically stimulated luminescence (OSL); photo-stimulated luminescence (PSL); portable OSL reader; retrospective dosimetry
Mesh:
Substances:
Year: 2022 PMID: 36111193 PMCID: PMC9468860 DOI: 10.3389/fpubh.2022.969829
Source DB: PubMed Journal: Front Public Health ISSN: 2296-2565
Photon counts and standard deviation for the background and 200 mGy irradiated salt samples for five sets of measurements on three instruments, with calculated sensitivities and detection limits.
|
| |||||
|---|---|---|---|---|---|
|
|
|
| |||
|
|
|
|
| ||
| Background (empty chamber) | 20 ± 34 | −10 ± 32 | −4 ± 31 | 45 ± 25 | −61 ± 26 |
| MDL (counts) | 123 | 95 | 94 | 74 | 78 |
| Pre-irradiation counts | 113 ± 72 | −4 ± 73 | −146 ± 89 | −27 ± 128 | 673 ± 160 |
| 200 mGy counts | 279,200 ± 71,200 | 605,900 ± 60,900 | 65,500 ± 21,800 | 43,300 ± 1,700 | 230,3300 ± 80,700 |
| Sensitivity (counts mGy−1) | 1,400 ± 240 | 3,030 ± 300 | 330 ± 110 | 220 ± 10 | 11,520 ± 400 |
| MDD (mGy) | 0.088 | 0.031 | 0.286 | 0.340 | 0.0067 |
Mean and standard deviation of photon counts recorded using the SUERC portable OSL reader at 470 nm, the IRSL instrument at 890 nm and the pulsed PSL instrument, from sets of ten individual sachets of salt exposed to gamma doses from 0–500 μGy.
|
|
|
| ||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
| 0 | 3,534 ± 453 | 12.8 | 1,204 ± 313 | 26.0 | 1,324 ± 102 | 7.7 |
| 20 | 3,307 ± 689 | 20.8 | 1,443 ± 300 | 20.8 | 1,256 ± 150 | 12.0 |
| 50 | 5,091 ± 460 | 9.0 | 1,843 ± 607 | 32.9 | 1,487 ± 122 | 8.2 |
| 100 | 7,323 ± 444 | 6.1 | 2,214 ± 202 | 9.1 | 1,701 ± 143 | 8.4 |
| 200 | 11,860 ± 1,086 | 9.2 | 2,388 ± 655 | 27.5 | 2,038 ± 237 | 11.7 |
| 500 | 25,202 ± 2,303 | 9.1 | 4,565 ± 687 | 15.1 | 3,649 ± 385 | 10.6 |
Figure 1Dose response from gamma irradiated samples measured using a blue OSL portable reader (blue), 890 nm IRSL (red) and the pulsed PSL (black) instruments, after subtraction of the signals from the salt which did not receive additional doses.