Diagnostic Accuracy of the Rapid Xpert HIV-1 Viral Load XC, Xpert HIV-1 Viral Load, & m-PIMA HIV-1/2 Viral Load in South African Clinics.

Jienchi Dorward1,2, Jessica Naidoo2, Pravikrishnen Moodley3, Yukteshwar Sookrajh4, Natasha Samsunder2, Fathima Sayed2, Nivashnee Naicker2, Thomas Fanshawe1, Paul K Drain5,6,7, Richard J Lessells2,8, Gail Hayward1, Christopher C Butler1, Nigel Garrett2,9.   

Abstract

BACKGROUND: We aimed to evaluate the analytic performance of 3 rapid HIV viral load assays: the novel Xpert HIV-1 VL XC (Xpert XC), Xpert HIV-1 VL (Xpert VL), and m-PIMA HIV-1/2 VL (m-PIMA).
SETTING: Two South African clinics.
METHODS: We conducted a prospective diagnostic accuracy study. Site-laboratory technicians and nurses used the Xpert XC, Xpert VL, and m-PIMA to test plasma samples from people with HIV receiving antiretroviral therapy. We compared results with the Roche cobas HIV-1 reference assay. We determined accuracy to detect viraemia at the World Health Organization (WHO) failure threshold of 1000 copies/mL on all 3 assays, and 50 and 200 copies/mL on the Xpert assays. We assessed the agreement using Bland-Altman plots.
RESULTS: We enrolled 140 participants (98 [70%] women, median age 37 years), who provided 189 paired samples at one or more timepoints. We tested 174 samples with the Xpert XC, 188 with the Xpert VL, and 128 with the m-PIMA. At 1000 copies/mL, sensitivity and specificity (95% confidence intervals) were 97% (82 to 100) and 98% (93 to 99) (Xpert XC), 100% (87 to 100) and 96% (91 to 98) (Xpert VL), and 92% (72 to 99) and 99% (93 to 100) (m-PIMA) respectively. At 50 copies/mL, sensitivity and specificity were 93% (81 to 98) and 96% (91 to 99) (Xpert XC), and 95% (84 to 99) and 95% (90 to 98) (Xpert VL) respectively. Mean bias was -0.10 (-0.54 to 0.34) log10 copies/mL (Xpert XC), 0.07 (-0.37 to 0.52) log10 copies/mL (Xpert VL), and -0.26 (-0.83 to 0.31) log10 copies/mL (m-PIMA).
CONCLUSIONS: In these South African clinics, the accuracy of all 3 assays was clinically acceptable to detect viraemia at the WHO failure threshold, whereas both Xpert assays were also accurate at detecting low-level viraemia.
Copyright © 2022 The Author(s). Published by Wolters Kluwer Health, Inc.

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 36094486      PMCID: PMC7613592          DOI: 10.1097/QAI.0000000000003037

Source DB:  PubMed          Journal:  J Acquir Immune Defic Syndr        ISSN: 1525-4135            Impact factor:   3.771


Introduction

The World Health Organization (WHO) recommends antiretroviral therapy (ART) for all people living with HIV, to suppress the HIV viral load.[1] This prevents immunosuppression and related morbidity and mortality,[2,3] as well as onward HIV transmission. Regular HIV viral load testing is a key component for monitoring the effectiveness of ART in current WHO guidelines, which recently endorsed the use of point-of-care viral load assays, where testing is conducted at the clinical site by laboratory professionals or specifically-trained healthcare workers.[4] These assays can reduce the time from blood draw to results being available, allowing clinicians to take decisions on the same day in one clinic visit, rather than having to schedule follow up appointments to discuss the results.[5,6] A randomised trial conducted by our team in South Africa, using the Xpert HIV-1 VL assay(Cepheid, Sunnyvale, USA), found that point-of-care testing improved the primary outcome of HIV viral suppression and retention in care by 13.7%.[7] Currently, there are two point-of-care HIV viral load assays that are approved by the WHO for use in low and middle-income countries. The m-PIMA HIV 1/2 VL (Abbott, Chicago, USA) received WHO pre-qualification in 2019,[8] and has been found to perform well in one laboratory-based[9] and two clinic-based evaluation studies.[10,11] The Xpert HIV-1 VL received WHO pre-qualification in 2017,[12] and has also been found to perform well in 19 studies, which have been summarised in two systematic reviews.[13,14] Only three of these studies reported conducting Xpert HIV-1 VL testing in clinics, as opposed to centralised laboratories. Because the assay was not able to detect certain recombinant HIV-1 sub-types,[15] Cepheid have recently released a next generation dual target assay, the Xpert HIV-1 Viral Load XC (HIV VL XC), that detects a wider range of HIV subtypes.[16] This assay recently received CE-IVD clearance,[17] but has not yet been evaluated in any published studies that we are aware of.

Objectives

In this study, we aimed to assess the analytic performance of the novel, dual target Xpert HIV-1 VL XC, the existing Xpert HIV-1 VL and the m-PIMA HIV 1/2 VL, when used in South African clinics, compared to an established laboratory reference assay.

Methods

Study design and participants

We conducted a prospective diagnostic accuracy evaluation as a sub-study within the Point-Of-care HIV viral load testing to Enhance Resuppression (POwER) study, and the Centre for the AIDS Program of Research in South Africa (CAPRISA) 002 Acute Infection study. POwER is a randomised feasibility study in one urban and one rural public sector, primary care clinic in KwaZulu-Natal, South Africa, and the study protocol has been published elsewhere.[18] In summary, approximately 100 non-pregnant adults aged 18 and over, receiving first-line ART and with HIV viraemia >1000 copies/mL in the past 6 weeks, are randomised to receive a second viral load test after 12 weeks by point-of-care test, or by standard laboratory based viral load testing. The primary outcome of viral suppression <50 copies/mL is assessed after 24 weeks. The CAPRISA 002 study is taking place in the same two clinics, and is a long-term cohort study of women living with HIV who are on ART.[19] For this sub-study, POwER and CAPRISA 002 participants who provided written informed consent, had additional 4 to 8mLs of venous blood drawn for point-of-care assay evaluation, taken alongside the main study blood draws at enrolment or during follow up. We initially planned to use consecutive participants enrolling in POwER and attending follow up in CAPRISA 002, but disruptions in test cartridge supply chains, assay training and staffing due to the COVID-19 pandemic meant that there were periods where various assay cartridges were not available or rarely nurses did not have time to conduct testing. We therefore used a convenience sampling strategy instead, where consenting patients enrolling or attending follow up in either study had samples tested on whichever assays were available at that time in the respective clinic. This meant that some participants in POwER had multiple samples taken for this evaluation at up to three different visits.

Test methods

Assay manufacturers provided point-of-care training to both research laboratory technicians and research nurses, as well as ongoing testing support. Initially, laboratory technicians conducted testing in a dedicated on-site point-of-care laboratory in the clinics. Once laboratory technician testing was established, research nurses conducted testing in a dedicated space in the clinic, with ongoing troubleshooting support from the manufacturer and on-site laboratory staff. For point-of-care viral load testing, 4 to 8mLs whole blood was obtained by venepuncture into an EDTA tube and centrifuged (by laboratory technicians or nurses respectively) to obtain plasma, which was transferred into the test cartridge and loaded into the respective assay platform. The m-PIMA HIV 1/2 platform is a single module instrument which requires 50 µL of plasma, and provides a quantitative viral load result between 800 to 1,000,000 copies/mL in approximately 70 minutes. The manufacturer states that the m-PIMA HIV 1/2 has a lower limit of detection of 342 (95% confidence interval [CI] 279 - 451) copies/mL for HIV-1 group M, and 228 (95% CI 187-295) copies/mL for HIV-1 group O.[20] The new Xpert HIV 1 VL XC and existing Xpert HIV 1 VL use the same GeneXpert platform (Cepheid, Sunnyvale, USA) and both require 1 mL of plasma to provide a quantitative viral load result between 40 and 10,000,000 copies/mL in approximately 90 minutes. The lower limit of detection as stated by the manufacturer is 13.6 (95% CI 11.1-15.6) copies/mL for the Xpert HIV-1 VL XC,[16] and 15.3 (95% CI 13.5-17.0) copies/mL for the Xpert HIV-1 VL.[21] We used research use only cartridges of the same version as the now commercially released Xpert XC, and single module GXP I and four module GXP IV instruments. If a point-of-care assay resulted in an error or an invalid result, the test was repeated using remaining plasma, if available. Point-of-care viral load results from this evaluation were not provided to patients and were not used for clinical decision making. For reference laboratory testing, a nurse drew whole blood in an EDTA tube at the same timepoint, which was centrifuged to plasma and stored at -80°C in the CAPRISA bio-repository. These samples were then defrosted and tested in two batches (study mid-point and study end) on the cobas® HIV-1 reference assay using the cobas 6800 platform (Roche, Basel, Switzerland) at the Inkosi Albert Luthuli Hospital National Health Laboratory Service in Durban, KwaZulu-Natal. We chose this assay because it is one of the standard viral load assays which has been widely validated and used in the South African National Health Laboratory Service. Staff performing laboratory reference testing were blinded to results of the point-of-care testing and vice versa.

Analysis

For the Xpert assays, we evaluated sensitivity and specificity to detect viraemia at thresholds of 50 copies/mL, 200 copies/mL and 1000 copies/mL. Because the m-PIMA HIV 1/2 VL does not provide a quantitative viral load below 800 copies/mL, we only evaluated the 1000 copies/mL threshold. We evaluated accuracy at these thresholds as they are used in national and international guidelines to define viraemia and treatment failure.[4,22-24] We assessed agreement with the laboratory reference assay using Bland-Altman plots, and association using Pearson correlation coefficients, restricted to paired samples with a quantifiable viral load on both the point of care and reference laboratory assay. We conducted sensitivity analyses using bootstrap methods (see supplementary material for methods) to account for repeat testing in the same participants at different time points, which could introduce clustering of errors. For example, if a participant had a viral sub-type that was not detected by a particular assay, inaccurate results would be more common among repeat tests in that participant, compared to tests in the general sample. We estimated that with the resources and time available, we could test approximately 150 samples on each assay. However, the COVID-19 pandemic interrupted cartridge supply chains for the m-PIMA HIV 1/2 VL, meaning fewer tests were possible for this assay. Because not all samples were tested on all point-of-care assays, we did not directly compare the performance of the point-of-care assays against each other.

Ethical approval

The POwER and CAPRISA 002 protocols, including use of samples for point-of-care viral load assay evaluation, were approved by the University of KwaZulu-Natal Biomedical Research Ethics Committee (BREC/00000836/2019 and E013/04 respectively), and POwER has also been approved by the University of Oxford Tropical Research Ethics Committee (OxTREC 66-19) and is registered with the Pan-African Clinical Trials Registry (PACTR202001785886049).

Results

Participants

Between June 26, 2020 and September 15, 2021, 140 participants provided paired samples, at one or more timepoints, for testing in this evaluation. Testing was conducted in the rural clinic for 25 participants, and at the urban clinic for 115 participants. 98 (70%) were women, median age was 37 years (interquartile range [IQR] 33-43), median time on ART was 4.6 years (IQR 2.9-6.0) and median CD4 count was 569 cells/µL (IQR 356-792).

Test results

Overall, 189 samples were tested on at least one point-of-care viral load assay. 174 samples were tested using the Xpert HIV-1 VL XC, 188 samples using the Xpert HIV-1 VL, and 128 samples using the m-PIMA HIV 1/2 VL (Table 1). Nurses conducted testing for 54/174 (31%) of Xpert HIV-1 VL XC samples, 26/188 (14%) Xpert HIV-1 VL samples, and 23/128 (18%) m-PIMA HIV 1/2 VL samples. Initial errors were produced for eight (4.6%) Xpert HIV-1 VL XC samples, two (1.1%) Xpert HIV-1 VL samples, and three (2.5%) m-PIMA HIV 1/2 VL samples. In some cases, tests were successfully repeated, meaning that final results were not available due to assay errors for eight (4.6%) Xpert HIV-1 VL XC samples, one (0.5%) Xpert HIV-1 VL sample, and two (1.7%) m-PIMA HIV 1/2 VL samples (see Table 1 for total number of errors and Table S1 for error codes). Four of the Xpert HIV-1 VL XC errors occurred at the beginning of the nurse testing process; this was the first assay that the nurses were trained to use.
Table 1

Samples tested on point-of-care viral load assays

Xpert HIV-1 VL XCXpert HIV-1 VLm-PIMA HIV 1/2 VL
Samples from one timepoint per participant84 (48.3)97 (51.6)67 (56.8)
Samples from two timepoints per participant70 (40.2)78 (41.5)46 (39.0)
Samples from 3 timepoints per participant12 (6.9)12 (6.4)3 (2.5)
No result due to assay errors [a]8 (4.6) [b]1 (0.5) [c]2 (1.7) [d]
Total174188118

See Table S1 for error codes

Eight samples (4.6%) produced an error. Four were retested, with the repeat test also producing an error meaning no result was available. Four were not retested meaning no result was available

Two samples (1.1%) produced an error, and both were retested. One repeat test also produced an error meaning no result was available, the other produced a valid result.

Three samples (2.5%) produced an error, and all were retested. Two repeat tests also produced an error meaning no result was available, the other produced a valid result.

Sensitivity and specificity

Sensitivity and specificity at 1000 copies/mL, compared to the Cobas 6800, were 97% and 98% (Xpert XC), 100% and 96% (Xpert VL), and 92% and 99% (m-PIMA) respectively. For the Xpert assays, sensitivity and specificity at 50 copies/mL were 93% and 96% (Xpert XC), and 95% and 95% (Xpert VL) respectively (Table 2). The m-PIMA was not assessed at 50 and 200 copies/mL because the lower limit of quantitation is 800 copies/mL.
Table 2

Raw numbers, sensitivity, and specificity (95% confidence intervals) of three point-of-care viral load assays at different viral load thresholds

Roche Cobas 6800Roche Cobas 6800Roche Cobas 6800
<50 ≥50 Total <50 ≥50 Total <50 ≥50 Total
Xpert HIV-1 VL XC <50 1084112 Xpert HIV-1 VL <50 1253128 m-PIMA HIV 1/2 VL <50 ---
≥50 45054 ≥50 65359 ≥50 ---
Total 11254166 Total 13156187 Total ---
Sensitivity: 0.93 (0.81, 0.98)Specificity: 0.96 (0.91, 0.99)Sensitivity: 0.95 (0.84, 0.99) Specificity: 0.95 (0.90, 0.98)Sensitivity: NASpecificity: NA
<200 ≥200 Total <200 ≥200 Total <200 ≥200 Total
Xpert HIV-1 VL XC <200 1164120 Xpert HIV-1 VL <200 1372139 m-PIMA HIV 1/2 VL <200 ---
≥200 34346 ≥200 14748 ≥200 ---
Total 11947166 Total 13849187 Total ---
Sensitivity: 0.91 (0.79, 0.97)Specificity: 0.97 (0.92, 0.99)Sensitivity: 0.96 (0.85, 0.99)Specificity: 0.99 (0.95, 1.00)Sensitivity: NASpecificity: NA
<1000 ≥1000 Total <1000 ≥1000 Total <1000 ≥1000 Total
Xpert HIV-1 VL XC <1000 1311132 Xpert HIV-1 VL <1000 1480148 m-PIMA HIV 1/2 VL <1000 90292
≥1000 33134 ≥1000 63339 ≥1000 12324
Total 13432166 Total 15433187 Total 9125116
Sensitivity: 0.97 (0.82, 1.00)Specificity: 0.98 (0.93, 0.99)Sensitivity: 1.00 (0.87, 1.00)Specificity: 0.96 (0.91, 0.98)Sensitivity: 0.92 (0.72-0.99)Specificity: 0.99 (0.93-1.00)

Correlation and agreement

When restricted to paired samples quantifiable at ≥40 copies/mL, the Xpert HIV-1 VL XC had a mean bias of -0.10 log10 copies/mL (95% limits of agreement -0.54 to 0.34,Fig 1A) and a Pearson correlation co-efficient of 0.97 (95% CI 0.95 to 0.99, p < 0.001, r2 = 0.95, Fig 2A), while the Xpert HIV-1 VL had a mean bias of +0.07 log10 copies/mL (95% limits of agreement -0.37 to 0.52, Fig 1B) and Pearson correlation co-efficient of 0.98 (95% CI 0.96-0.99, p <0.001, r2 = 0.95, Fig 2B). For the m-PIMA HIV 1/2 VL, when restricted to paired samples quantifiable at ≥800 copies/mL, the mean bias was -0.26 log10 copies/mL (95% limits of agreement -0.83 to 0.31, Fig 1C) and the Pearson correlation co-efficient was 0.93 (95% CI 0.85-0.97, p <0.001, r2 = 0.87, Fig 2C).
Fig 1

Bland Altman plots of Xpert HIV-1 VL XC, Xpert HIV-1 VL & m-PIMA HIV 1/2 vs Cobas 6800

Restricted to paired samples quantifiable on both the point-of-care and laboratory reference assay. The point estimate of the mean difference (bias) between the point-of-care and laboratory reference assay was within +/- 0.3 log10 copies/mL for all three assays.

Fig 2

Correlation of Xpert HIV-1 VL XC, Xpert HIV-1 VL & m-PIMA HIV 1/2 VL vs Cobas 6800

Restricted to paired samples quantifiable on both the point-of-care and laboratory reference assay

Nurse-led testing

Accuracy, correlation and agreement of the assays were comparable when conducted by nurses and laboratory technicians, but numbers of nurse tested samples were small, particularly for the Xpert HIV-1 VL and the m-PIMA HIV-1/2 VL (Tables S2 and S3).

Sensitivity analyses

In sensitivity analyses using bootstrap methods to account for some participants having samples drawn for evaluation at repeated timepoints, results were relatively similar (Table S4).

Discussion

Summary

We found that the novel Xpert HIV-1 VL XC, the Xpert HIV-1 VL and the m-PIMA HIV-1/2 VL had clinically acceptable performance when used in a South African clinic. At the WHO threshold for treatment failure of 1000 copies/mL, both the Xpert HIV-1 VL XC and Xpert HIV-1 VL had >95% sensitivity and specificity, while the m-PIMA HIV-1/2 VL had sensitivity of 92% and specificity of 99%. Both Xpert assays were also accurate at detecting low level viraemia.

Related literature

Our findings are similar to other studies of these three point-of-care assays, the majority of which have been conducted in centralised laboratory settings.

Xpert HIV-1 VL XC

Our study is the first to evaluate the Xpert HIV-1 VL XC in a clinical setting. In a centralised laboratory-based study, 513 fresh, frozen and diluted samples, were tested on the Xpert HIV-1 VL XC and compared with the Alinity m HIV-1 reference assay (Abbott, Chicago, USA).[25] The Xpert HIV-1 VL XC had a mean bias of +0.10 log10 copies/mL (95% CI -0.54 to 0.35), compared to the reference assay, and r2 was 0.9788. Taken together, these results suggest that the Xpert HIV-1 VL XC performs well when used in centralised laboratory and clinic-based settings.

Xpert HIV-1 VL

For the Xpert HIV-1 VL, our findings were similar to a meta-analysis of 13 studies including 3790 patients,[14] which found a mean bias of 0.04 (95% CI: 0.02-0.07) copies/mL, and r2 of 0.941 (P<0.001). Sensitivity and specificity was >95% at 200 and 1000 copies/mL, but was not evaluated at 50 copies/mL. Instead, a threshold of ‘detectable’ was used (40 copies/mL on the Xpert HIV-1 VL assay versus 20, 25 and 40 copies/mL in the different comparator assays used in the studies), with sensitivity of 93.3% (88.2–96.3), and much lower specificity at 80.6% (64.6–90.4). Specificity was particularly poor (69.3%) when compared to the Abbott RealTime HIV-1 VL assay. We used the 50 copies/mL threshold as this is used in South African guidelines.[22] Our finding that both the Xpert HIV-1 VL XC and Xpert HIV-1 VL performed well at this threshold contrasts slightly with the systematic review findings, but suggests that these assays are appropriate for use in the South African setting.

m-PIMA HIV-1/2 VL

Three published studies have evaluated the m-PIMA HIV-1/2 VL assay. Two of these were clinic-based; one multi-site study in Mozambique compared 686 samples tested by nurses with the m-PIMA HIV-1/2 VL versus the Cobas AmpliPrep/Cobas TaqMan 96 HIV-1 v2 (Roche) assay,[10] while another multi-site study from Kenya compared 555 samples tested by laboratory technicians on the m-PIMA HIV-1/2 VL in clinics, with the RealTime HIV-1 assay (Abbott).[11] The final study was a centralised laboratory evaluation comparing 413 frozen samples tested on the m-PIMA HIV -1/2 VL with results from fresh samples tested on the RealTime HIV-1 assay. At 1000 copies/mL, sensitivity and specificity were above 95% in all studies, except for the laboratory evaluation where specificity was 76.9 % (95% CI: 69.8–83.1). The correlation coefficients were 0.92, 0.92 and 0.93 respectively, with mean bias of -0.20, +0.16 and +0.26 log10 copies/mL respectively versus the laboratory reference assay. The WHO prequalification report found a bias of -0.21 log10 copies/mL versus the COBAS AmpliPrep/COBAS TaqMan HIV-1 Test, v2.0, with sensitivity of 95.1% and specificity of 99% at 1000 copies/mL8. Overall, these findings, particularly when using Roche reference laboratory assays, are similar to our results, although reasons for the low specificity in the laboratory study are not clear, but could be related to the use of frozen samples for m-PIMA testing.

Nurse testing

There is only one published study which has evaluated nurse-led quantitative point-of-care viral load testing, with good results using the m-PIMA HIV1/2 viral load in Mozambique, as described above.[10] Our data suggest that the Xpert HIV-1 VL XC performs well when used by nurses, and results with the Xpert HIV-1 VL and m-PIMA HIV-1/2 VL were relatively similar, although numbers were small. However, our study was not explicitly designed to compare nurse versus laboratory testing, and research nurses had support available to them from on-site laboratory technicians, which does not reflect conditions across South African primary care. Nurses and/or midwives have accurately used qualitative viral load assays (that use the GeneXpert and m-PIMA platforms) for early infant diagnosis of HIV,[26,27] and there is increasing interest in task-shifting point-of-care HIV diagnostics to non-laboratory staff in low and middle income countries.[28-30] As these assays require plasma, venous blood samples had to be centrifuged by nurses in the study clinic. This required specific training and may not be feasible for nurses in many settings, which could limit use at the clinical point-of-care. A whole blood fingerstick assay would circumvent this problem, and point-of-care early infant diagnosis assays, which also use whole blood, have demonstrated clear benefits in clinical outcomes.[26,31] This highlights the potential benefits of providing same day results using point-of-care technologies.

Strengths and limitations

Strengths of our study include the assessment of clinic-based testing in a setting with high HIV prevalence, and the inclusion of multiple point-of-care assays. As well as assessing correlation and mean bias, we focussed our analysis on sensitivity and specificity to detect viraemia at viral load thresholds that are directly relevant for patient management in South Africa and internationally. Most viral load testing was conducted by laboratory technicians, but we included some nurse led viral load testing, which has not been widely evaluated previously. Supply chain and operational disruptions due to the COVID-19 pandemic meant that our sample size was smaller than anticipated for some assays, and we were unable to test samples on all three assays concurrently. Instead, we allowed paired samples from repeat timepoints in the same participant, and conducted sensitivity analyses to account for potential clustering, with similar results. Nevertheless, the sample size of this study is relatively small, meaning confidence intervals for sensitivity estimates are wide for all assays at all viraemia thresholds, and for sensitivity and specificity estimates for nurse testing. Reference laboratory testing was conducted retrospectively using frozen plasma samples, in accordance with manufacturer’s instructions, which should not have affected results.[32]

Implications

A previous cost-effectiveness modelling study found that a sensitivity and specificity of 90% or more at a threshold of 1000 copies/mL would be acceptable for a point-of-care viral load assay in Southern Africa,[33] while mean bias should ideally be within 0.3 log10 copies/mL of reference laboratory assays.[6,34] In our evaluation, all three evaluated assays fell within these criteria. Balancing assay analytic performance and other characteristics that influence clinical outcomes, such as ease of use, turn-around time and result availability will depend on the context of where the assay is being implemented. We found that the Xpert HIV-1 VL XC performed well at both the 50 copies/mL and 1000 copies/mL thresholds, which are the two key thresholds for ART management in South Africa, suggesting that the clinical performance is appropriate for this setting. The m-PIMA HIV-1/2 VL is not able to quantitate viral loads <800 copies/mL, making it less suited to assessment of low level viraemia. More recent evidence has found that low level viraemia among people receiving efavirenz-based ART regimens predicts subsequent treatment failure,[35] although whether this remains true on dolutegravir based regimens is not yet known.[36] South African guidelines currently use 50 copies/mL to define viraemia, and determine whether a patient is eligible for down-referral into streamlined ART delivery programmes,[37,38] or whether they need enhanced adherence counselling to encourage resuppression,[37] meaning the m-PIMA is less suited for use in these situations. However, the m-PIMA has a simple interface, shorter run time, and may be more suited to implementation in clinic-based settings, particularly in settings where access to viral load testing is poor. Further work is needed to confirm the performance of these assays when used by nurses, and we are conducting qualitative work with nurses and other healthcare workers, to explore feasibility and acceptability of implementing nurse testing in clinical settings. Using these assays to provide same-day, point-of-care results is crucial to achieving the reported benefits of expedited clinical decision-making, and reduced clinic visits and patient travel costs.[39]

Conclusions

The diagnostic accuracy of the Xpert HIV-1 VL XC, Xpert HIV-1 VL and m-PIMA HIV-1/2 VL assays was clinically acceptable when testing was conducted in these South African clinics. The Xpert assays have the added benefit of accurately detecting low level viraemia. However, further clinical studies are needed to determine if these assays can be implemented in routine primary healthcare settings, and whether there are operational and clinical benefits to patients and providers when used at the clinical point-of-care.
  22 in total

1.  The performance of a new point-of-care HIV virus load technology to identify patients failing antiretroviral treatment.

Authors:  Diana Mariani; Marcelo C V M de Azevedo; Isabelle Vasconcellos; Luiz Ribeiro; Cassia Alves; Orlando C Ferreira; Amilcar Tanuri
Journal:  J Clin Virol       Date:  2019-11-09       Impact factor: 3.168

2.  Dolutegravir for first-line antiretroviral therapy in low-income and middle-income countries: uncertainties and opportunities for implementation and research.

Authors:  Jienchi Dorward; Richard Lessells; Paul K Drain; Kogieleum Naidoo; Tulio de Oliveira; Yogan Pillay; Salim S Abdool Karim; Nigel Garrett
Journal:  Lancet HIV       Date:  2018-06-05       Impact factor: 12.767

3.  Point-of-care HIV viral load testing combined with task shifting to improve treatment outcomes (STREAM): findings from an open-label, non-inferiority, randomised controlled trial.

Authors:  Paul K Drain; Jienchi Dorward; Lauren R Violette; Justice Quame-Amaglo; Katherine K Thomas; Natasha Samsunder; Hope Ngobese; Koleka Mlisana; Pravikrishnen Moodley; Deborah Donnell; Ruanne V Barnabas; Kogieleum Naidoo; Salim S Abdool Karim; Connie Celum; Nigel Garrett
Journal:  Lancet HIV       Date:  2020-02-24       Impact factor: 12.767

4.  Performance of a True Point-of-Care Assay for HIV-1/2 Viral Load Measurement at Antenatal and Postpartum Services.

Authors:  Bindiya Meggi; Timothy Bollinger; Alcina Zitha; Chishamisso Mudenyanga; Adolfo Vubil; Dadirayi Mutsaka; Carina Nhachigule; Nedio Mabunda; Osvaldo Loquiha; Arne Kroidl; Ilesh V Jani
Journal:  J Acquir Immune Defic Syndr       Date:  2021-05-01       Impact factor: 3.731

5.  Initiation of Antiretroviral Therapy in Early Asymptomatic HIV Infection.

Authors:  Jens D Lundgren; Abdel G Babiker; Fred Gordin; Sean Emery; Birgit Grund; Shweta Sharma; Anchalee Avihingsanon; David A Cooper; Gerd Fätkenheuer; Josep M Llibre; Jean-Michel Molina; Paula Munderi; Mauro Schechter; Robin Wood; Karin L Klingman; Simon Collins; H Clifford Lane; Andrew N Phillips; James D Neaton
Journal:  N Engl J Med       Date:  2015-07-20       Impact factor: 91.245

6.  Comparison between Roche and Xpert in HIV-1 RNA quantitation: A high concordance between the two techniques except for a CRF02_AG subtype variant with high viral load titters detected by Roche but undetected by Xpert.

Authors:  Boaz Avidor; Natalia Matus; Shirley Girshengorn; Svetlana Achsanov; Simona Gielman; Irene Zeldis; Inbal Schweitzer; Amos Adler; Dan Turner
Journal:  J Clin Virol       Date:  2017-05-19       Impact factor: 3.168

7.  Effect of point-of-care early infant diagnosis on antiretroviral therapy initiation and retention of patients.

Authors:  Ilesh V Jani; Bindiya Meggi; Osvaldo Loquiha; Ocean Tobaiwa; Chishamiso Mudenyanga; Alcina Zitha; Dadirayi Mutsaka; Nedio Mabunda; Adolfo Vubil; Timothy Bollinger; Lara Vojnov; Trevor F Peter
Journal:  AIDS       Date:  2018-07-17       Impact factor: 4.177

8.  The missed potential of CD4 and viral load testing to improve clinical outcomes for people living with HIV in lower-resource settings.

Authors:  Peter D Ehrenkranz; Solange L Baptiste; Helen Bygrave; Tom Ellman; Naoko Doi; Anna Grimsrud; Andreas Jahn; Thokozani Kalua; Rose Kolola Nyirenda; Michael O Odo; Pascale Ondoa; Lara Vojnov; Charles B Holmes
Journal:  PLoS Med       Date:  2019-05-29       Impact factor: 11.069

9.  Establishing a cohort at high risk of HIV infection in South Africa: challenges and experiences of the CAPRISA 002 acute infection study.

Authors:  Francois van Loggerenberg; Koleka Mlisana; Carolyn Williamson; Sara C Auld; Lynn Morris; Clive M Gray; Quarraisha Abdool Karim; Anneke Grobler; Nomampondo Barnabas; Itua Iriogbe; Salim S Abdool Karim
Journal:  PLoS One       Date:  2008-04-16       Impact factor: 3.240

10.  Understanding how community antiretroviral delivery influences engagement in HIV care: a qualitative assessment of the Centralised Chronic Medication Dispensing and Distribution programme in South Africa.

Authors:  Jienchi Dorward; Lindani Msimango; Andrew Gibbs; Hlengiwe Shozi; Sarah Tonkin-Crine; Gail Hayward; Christopher C Butler; Hope Ngobese; Paul K Drain; Nigel Garrett
Journal:  BMJ Open       Date:  2020-05-15       Impact factor: 2.692

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.