| Literature DB >> 36072575 |
Ulf Borg1, Jeffrey Aviano2, Milan Ginani2, Kun Li2.
Abstract
Purpose: Non-invasive ventilation (NIV) may reduce intubation rates and be especially beneficial in the care of preterm infants, in addition to other care modalities. Currently, ventilators do not display the pressure at the nares but the pressure in the ventilator tubing system. There are several nasal cannulas available for use to deliver NIV. The purpose of this study was to compare the inspiratory pressure on the ventilator to the measured pressure delivered at the nares using three cannula brands (Manufacturer A, Fisher & Paykel; Manufacturer B, Neotech RAM; and Manufacturer C, Hudson RCI). Patients andEntities:
Keywords: 3D print; airway pressure; model; nasal cannula; non-invasive ventilation; ventilator
Year: 2022 PMID: 36072575 PMCID: PMC9444233 DOI: 10.2147/MDER.S374418
Source DB: PubMed Journal: Med Devices (Auckl) ISSN: 1179-1470
Figure 1Nasal adaptors with varied nasal diameters to accommodate different sizes of nasal prongs.
Characteristics of Nasal Cannulas Used in Testing
| Manufacturer | Product Code | Nasal Prong Outer Diameter (mm) | Septal Space (mm) | Inner Diameter for 3D Printed Nares (mm) |
|---|---|---|---|---|
| A | 3020 | 3 | 2 | 3.3 |
| 3520 | 3.5 | 2 | 3.85 | |
| 4030 | 4 | 3 | 4.4 | |
| 4540 | 4.5 | 4 | 4.95 | |
| B | N4900 | 3 | 2.5 | 3.6 |
| N4901 | 3 | 4.25 | 3.6 | |
| N4902 | 3.5 | 4.75 | 4.2 | |
| N4903 | 4 | 5 | 4.8 | |
| C | No.0 | 3.33 | N/Aa | 4 |
| No.1 | 4 | N/Aa | 4.4 | |
| No.2 | 4.33 | N/Aa | 5.2 |
Notes: aInformation not provided in manufacturer instructions for use.
Figure 2(A) Schematic for nasal cannula testing (B) Photo of nasal cannula testing setup with 3D printed nares on the lower right side.
Low and High Ventilator Pressure and Measured Pressures at Nares
| Manufacturer | Product Code | Low Pressure | High Pressure | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Inspiratory Ventilator Pressure (cm H2O) | Average Measured Pressure at Nares ± Standard Deviation (cm H2O) | Average Pressure Difference (cm H2O) | p- value | Inspiratory Ventilator Pressure (cm H2O) | Average Measured Pressure at Nares ± Standard Deviation (cm H2O) | Average Pressure Difference | p-value | ||
| A | 3020 | 15 | 14.3 ± 0.7 | 0.7 | 0.05 | 40 | 38.4 ± 1.5 | 1.6 | 0.05 |
| 3520 | 15 | 14.5 ± 1.0 | 0.5 | 0.31 | 40 | 38.2 ± 2.1 | 1.8 | 0.09 | |
| 4030 | 15 | 14.9 ± 0.4 | 0.1 | 0.53 | 40 | 38.6 ± 1.1 | 1.4 | 0.03 | |
| 4540 | 15 | 15.1 ± 0.4 | −0.1 | 0.57 | 40 | 38.9 ± 1.1 | 1.1 | 0.06 | |
| B | N4900 | 15 | 10.2 ± 1.8 | 4.8 | 0.001 | 40 | 28.6 ± 4.8 | 11.4 | 0.002 |
| N4901 | 15 | 14.1 ± 0.8 | 0.9 | 0.0496 | 40 | 38.0 ± 2.3 | 2.0 | 0.1 | |
| N4902 | 15 | 13.1 ± 0.9 | 1.9 | 0.004 | 40 | 35.4 ± 3.4 | 4.6 | 0.02 | |
| N4903 | 15 | 13.3 ± 1.6 | 1.7 | 0.05 | 40 | 36.8 ± 3.5 | 3.2 | 0.08 | |
| C | No. 0 | 15 | 15.0 ± 0.7 | 0.0 | 0.9 | 40 | 38.2 ± 2.0 | 1.8 | 0.08 |
| No. 1 | 15 | 13.7 ± 1.7 | 1.3 | 0.1 | 40 | 33.9 ± 6.4 | 6.1 | 0.07 | |
| No. 2 | 15 | 15.4 ± 0.7 | −0.4 | 0.2 | 40 | 38.9 ± 1.8 | 1.1 | 0.2 | |
Average Ventilator Pressure and Measured Pressures for Cannulas When Used Across Eight Ventilator Settings
| Manufacturer | Product Code | Number of Measurements | Average Inspiratory Ventilator Pressure (cm H2O) | Average Measured Pressure at Nares ± Standard Deviation (cm H2O) | Average Pressure Difference (Inspiratory - Measured) (cm H2O) | p-value |
|---|---|---|---|---|---|---|
| A | 3020 | 48 | 25 | 24.3 ± 9.4 | 0.7 | <0.001 |
| 3520 | 48 | 25 | 24.3 ± 9.2 | 0.7 | 0.003 | |
| 4030 | 48 | 25 | 24.8 ± 9.1 | 0.2 | 0.12 | |
| 4540 | 48 | 25 | 25.3 ± 9.3 | −0.3 | 0.22 | |
| B | N4900 | 48 | 25 | 17.5 ± 7.8 | 7.5 | <0.001 |
| N4901 | 48 | 25 | 23.8 ± 9.3 | 1.2 | <0.001 | |
| N4902 | 48 | 25 | 22.2 ± 8.8 | 2.8 | <0.001 | |
| N4903 | 48 | 25 | 22.8 ± 9.1 | 2.2 | <0.001 | |
| C | No. 0 | 48 | 25 | 24.3 ± 9.6 | 0.7 | 0.18 |
| No. 1 | 48 | 25 | 22.2 ± 8.7 | 2.8 | <0.001 | |
| No. 2 | 48 | 25 | 25.2 ± 9.1 | −0.2 | 0.27 |
Figure 3Effect of inspiratory time on the delivered pressure for the smallest cannula size for each manufacturer. Multiple inspiratory times, along with different PEEP and inspiratory pressure ventilator settings were utilized, ranging from 5–15 cm H2O PEEP and 10–25 cm H2O inspiratory pressure. Green bars denote the total inspiratory pressure on the ventilator, for comparison to the delivered pressure. Significant differences between the inspiratory and delivered pressure for each cannula and ventilator setting are denoted as: * p<0.05; ** p<0.01; *** p<0.001.