Literature DB >> 34002956

Role of ventilator and nasal interface in pressure transmission during neonatal intermittent positive pressure ventilation: A bench study.

Camilla Rigotti1, Emanuela Zannin1,2, Giulia Dognini1, Raffaele Dellacà2, Maria L Ventura1.   

Abstract

We aimed at evaluating pressure transmission and stability during non-synchronized neonatal nasal intermittent positive pressure ventilation (NIPPV) delivered using five mechanical ventilators and three nasal interfaces. An artificial nose-throat model was connected to a mechanical analog of the infant respiratory system and a breath generator. Ventilation was administrated via a nasal mask (NM), short bi-nasal prongs (SBN), or RAM® cannula. We applied positive end-expiratory pressures (PEEP) of 5 and 10 cmH2 O, inspiratory pressures (PIP) of 15 and 30 cmH2 O, inspiratory times of 0.23, 0.42, and 0.57 s. Measurements were performed with leaks of 0, 1.5, and 4 L/min. The pressure was measured at the airways opening (PAW ) and the glottis (PGL ). The difference between set and delivered pressures (PAW ) was less than ±1 cmH2 O for all ventilators. We documented a significant difference between PAW and PGL in the presence of leaks. With 4 L/min leaks, PEEP dropped by 43%, 49%, and 63% with NM, SBP, and RAM® cannula, respectively; PIP dropped by 58%, 64%, and 74%. On average, the SD of PEEP fluctuations was ±0.60 and ±2.50 cmH2 O for PAW and PGL ; the breath-by-breath SD of PIP was ±0.77 and ±2.06 cmH2 O. During NIPPV, the PIP and PEEP transmission to the glottis is markedly lower than the set values and highly variable. The impact of leaks and nasal interface is much more significant than the differences in ventilators' performance on the efficacy of pressure transmission and stability of non-synchronized ventilator-generated NIPPV.
© 2021 Wiley Periodicals LLC.

Entities:  

Keywords:  infant; mechanical ventilators; nasal interface; noninvasive ventilation; premature

Year:  2021        PMID: 34002956     DOI: 10.1002/ppul.25449

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  1 in total

1.  Evaluation of Common Nasal Cannulas in Neonatal Noninvasive Ventilation (NIV) Using a Novel Neonatal Nasal Model.

Authors:  Ulf Borg; Jeffrey Aviano; Milan Ginani; Kun Li
Journal:  Med Devices (Auckl)       Date:  2022-09-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.