Cátia Marques1, Liangwen Liu2, Kyle D Duncan1, Ingela Lanekoff1. 1. Department of Chemistry─BMC, Uppsala University, 75123 Uppsala, Sweden. 2. Department of Medical Cell Biology, Uppsala University, 75123 Uppsala, Sweden.
Abstract
Targeted and nontargeted metabolomics has the potential to evaluate and detect global metabolite changes in biological systems. Direct infusion mass spectrometric analysis enables detection of all ionizable small molecules, thus simultaneously providing information on both metabolites and lipids in chemically complex samples. However, to unravel the heterogeneity of the metabolic status of cells in culture and tissue a low number of cells per sample should be analyzed with high sensitivity, which requires low sample volumes. Here, we present the design and characterization of the direct infusion probe, DIP. The DIP is simple to build and position directly in front of a mass spectrometer for rapid metabolomics of chemically complex biological samples using pneumatically assisted electrospray ionization at 1 μL/min flow rate. The resulting data is acquired in a square wave profile with minimal carryover between samples that enhances throughput and enables several minutes of uniform MS signal from 5 μL sample volumes. The DIP was applied to study the intracellular metabolism of insulin secreting INS-1 cells and the results show that exposure to 20 mM glucose for 15 min significantly alters the abundance of several small metabolites, amino acids, and lipids.
Targeted and nontargeted metabolomics has the potential to evaluate and detect global metabolite changes in biological systems. Direct infusion mass spectrometric analysis enables detection of all ionizable small molecules, thus simultaneously providing information on both metabolites and lipids in chemically complex samples. However, to unravel the heterogeneity of the metabolic status of cells in culture and tissue a low number of cells per sample should be analyzed with high sensitivity, which requires low sample volumes. Here, we present the design and characterization of the direct infusion probe, DIP. The DIP is simple to build and position directly in front of a mass spectrometer for rapid metabolomics of chemically complex biological samples using pneumatically assisted electrospray ionization at 1 μL/min flow rate. The resulting data is acquired in a square wave profile with minimal carryover between samples that enhances throughput and enables several minutes of uniform MS signal from 5 μL sample volumes. The DIP was applied to study the intracellular metabolism of insulin secreting INS-1 cells and the results show that exposure to 20 mM glucose for 15 min significantly alters the abundance of several small metabolites, amino acids, and lipids.
Metabolomics is the comprehensive
analysis of small molecules involved in metabolic pathways that drive
the biochemistry and function of our cells. Profiling an organism’s
dynamic metabolome can lead to objective disease diagnosis,[1,2] personalized medicine,[3,4] and realization of fundamental
molecular mechanisms involved in complex biological processes.[5,6] Important applications such as these have catalyzed rapid technology
development. For example, there have been significant advancements
to achieve higher throughput NMR-based metabolomics, which are intrinsically
quantitative with metabolite concentrations proportional to the number
of nuclei.[7,8] However, compared to NMR, the sensitivity
and selectivity of mass spectrometry (MS) based metabolomics techniques
is imperative for using smaller sample volumes and acquiring a wider
metabolite coverage. Thus, MS has evolved into the most universally
applied instrumentation for modern metabolomics.Conventional
MS-based metabolomics workflows involve several steps:
(i) metabolite extraction and clean up from biofluids, cells, or tissue;[9,10] (ii) online separation (e.g., liquid chromatography); (iii) and
MS detection. Depending on the targeted metabolite compound classes,
mass spectrometry can be coupled with either liquid chromatography
(LC),[11−14] gas chromatography (GC),[15,16] or capillary electrophoresis
(CE)[17−19] for metabolite separation and annotation. While comprehensive
separation techniques provide the widest metabolome coverage, achievable
sample duty cycles can limit large sample cohort studies and therefore
biomedical conclusions.To increase sample throughput and feasibility
for larger cohort
studies there is a drive for establishing methods where metabolites
extracted from cell and tissue samples are more rapidly profiled and
quantified. Higher throughput LC-MS and CE-MS methods have been developed
with separation times <2 and 4 min, respectively.[20,21] Nevertheless, direct infusion MS methods are the fastest reported
methods to date, with sample duty cycles of only a few seconds.[22,23] In addition to increasing sample throughput, there has been considerable
effort made to develop direct MS technology for profiling metabolites
from volume limited biological samples. For example, ambient ionization
techniques such as desorption electrospray ionization (DESI), direct
analysis in real time (DART), and liquid microjunction surface sampling
probe (LMJ-SSP) have been used to profile metabolites directly from
biological surfaces.[24] Additionally, nanospray
based direct infusion MS methods have enabled metabolite profiling
in submicroliter volumes.[25−27]One of the simplest direct
MS metabolomics methods reported in
the literature is flow injection analysis, with duty cycles <1
min.[28] Yet, flow injection suffers from
longitudinal diffusion from the external pump that limits the number
of samples injected online between solvent plugs and causes nonuniform
sample elution profiles that limit sensitivity from volume limited
injections due to diffusion. To reduce the need for an external pump,
eliminate longitudinal diffusion, and instead achieve square wave
sample profiles, several probes for direct infusion MS were developed
since the late 1990s.[29] These probes take
advantage of the Venturi effect where a flow of gas over the capillary
outlet enables direct transport of a sample from its container to
the mass spectrometer inlet for subsequent ESI.[29−34] Depending on the capillary dimensions, which generally have an inner
diameter between 100 and 250 μm, flow rates between 5 and 10
μL/min and up to 80 μL/min have been reported. Self-aspirating
ESI probes have been constructed both without an applied electrospray
voltage[30,31] and with an applied electrospray voltage
placed either in the ESI source[29,33] or in the sample solution.[32,34] However, for the analysis of low abundant metabolites the use of
voltage for ESI generates higher ionization efficiency and greater
metabolite coverage.Herein, we describe the design, optimization,
and characterization
of the direct infusion probe (DIP) coupled to MS. The DIP is self-aspirating
by using the Venturi effect, and it is made of a stainless-steel capillary
that allows for direct application of the high voltage for enhanced
ionization. Direct aspiration through the probe for ionization enables
analysis of minute sample volumes ≤5 μL and can provide
metabolomics of cell or tissue samples down to 20 cells/μL.
As a proof-of-principle study, DIP-MS was employed to measure the
metabolic response in insulin secreting INS-1 cells after 15 min stimulation
with high glucose with simultaneous nontargeted and quantitative targeted
mode. The results show significant alterations in several metabolic
pathways including both metabolites and lipids. The overall device
simplicity makes DIP-MS a new and exciting tool for higher-throughput
direct infusion metabolomics.
Experimental Section
Chemicals and Prepared Solutions
Solvents for extraction
were HLPC-grade methanol (MeOH) (Fisher Scientific), Milli-Q water
(H2O) from Milli-Q Plus, and formic acid (Merck). For quantification
the internal standards oleic acid-d9 (FA 18:1-d9), LPC 19:0, PC 11:0/11:0, GABA-d2, Acetylcholine-d9, Glucose-d2, and Cell Free Amino Acid Mixture-15N (Sigma-Aldrich) were prepared in MeOH:H2O (9:1)
with 0.1% formic acid (Tables S1 and S2). INS-1 cells were exposed to a solution contained 138 mM NaCl,
5.6 mM KCl, 1.2 mM MgCl2, 2.6 mM CaCl2, and
1, 3, or 20 mM d-glucose at pH 7.4 and room temperature.
Biological Samples
Frozen intact rat brain tissue (Innovative
Research Inc., Novi, MI, USA) was cut into small pieces and added
to 35 mL MeOH, generating a metabolite extract of 34.61 mg rat brain
tissue per mL. For cell dissociation, the sample was sonicated using
an ultrasonic processor (VCX 130, Sonics and Materials, Inc.) for
2 cycles at 50% for a total of 5 min, pulses for 2 s with 1 s pause
in between for each cycle. Finally, the sample was centrifuged at
2000 g for 6 min to remove residual tissue debris and the supernatant
collected and stored at −20 °C until use.Immortalized
human embryonic kidney 293 cells (HEK293) were cultured in a Dulbecco’s
Modified Eagle Medium (DMEM) medium (Gibco) supplemented with 10 v/v
% fetal bovine serum (FBS), 1 v/v % nonessential amino acids and 1
v/v % penicillin/streptomycin. Approximately, 1 million cells were
collected (Bürker–Türk hemocytometer), washed
three times by centrifugation (200 g for 5 min), and resuspension
in 1 mL of saline solution at 37 °C. Metabolites from the cell
pellet were extracted with 1 mL of MeOH:H2O (9:1) plus
0.1% formic acid, sonicated for 20 min, and centrifuged at 13 000 g for 20 min to remove cellular debris to eliminate potential
clogging of the capillary. Following, a working solution containing
25 000 cells/mL was prepared and further diluted in triplicates
to the final cell densities of 500, 400, 300, 200, 100, 50, and 20
cells/μL, and the solvent was evaporated (SpeedVac Vacuum Concentrator,
Thermo Fisher Scientific) for 1.5 h. After addition of 5 μL
of internal standard solution (Table S1), the samples were stored at −80 °C.Rat insulinoma
cell line INS-1 clone 832/13 cells were maintained
in RPMI 1640 (Invitrogen) containing 10 mM glucose and supplemented
with 10% fetal bovine serum, streptomycin (100 μg/mL), penicillin
(100 μg/mL), Na-pyruvate (1 mM), l-glutamine (2 mM),
and 2-mercaptoethanol (50 μM). Approximately 1 million cells
were transferred to 24-well for ∼20 h before exposure to 3
mM glucose for 1 h in 37 °C and then either 1 mM glucose or 20
mM glucose for 15 min. Following that, the cells were quickly rinsed
three times with saline solution and lysed with 100 μL of MeOH:H2O (9:1) with 0.1% formic acid and internal standards (Table S1). Finally, the samples were sonicated
and centrifuged as described above and immediately analyzed.
Mass Spectrometric Parameters
A QExactive Basic (Thermo
Fisher Scientific, Bremen, Germany) was used to acquire data in positive
and negative modes (for INS-1 cells data) between m/z 70–1000 using a mass resolution of 140 000
(m/Δm at m/z 200). Data presented was acquired in positive
mode unless stated otherwise. Detailed information about the experimental
parameters used can be found in Table S3.
Direct Infusion Probe (DIP)
The DIP probe was constructed
using a 7 cm stainless-steel capillary (320:50 μm OD:ID) (MS
Ekspert, Gdańsk, Poland), a PEEK sample tee (0.050 in. through
hole) (Upchurch Scientific, Oak Harbor, USA) and Teflon sleeves with
1/16-in. OD and 0.5 and 0.75 mm ID (VICI Valco, USA). The sample vial
and probe were positioned using a MIM micromanipulator (Quarter Research
and Development, Bend, OR). Once the probe position was optimized,
it was held in place using an in-house designed and 3D-printed support
attached to the mass spectrometer inlet. Nitrogen gas was supplied
at ∼5 bar to generate the self-aspirating Venturi pump.
Data Analysis
The acquired data were analyzed with
the software MZmine 2 (parameters used are detailed in Note S1).[35] Carryover
was calculated by taking the intensity of analyte A in scan 10 or
20 of the new sample that does not contain A, and dividing it with
the average intensity detected of A in the original sample in percentage
(eq )Other calculations such as,
limits of detection (LOD) or concentrations, were based on the integrated
signal over 1 min. Metlin and Human metabolome database (HMDB) were
used to perform putative assignment of detected peaks via accurate
mass (<5 ppm accuracy) for protonated ([M + H]+) and
sodiated ([M + Na]+) adducts. Figures were plotted using
Origin 2016 (Thermo Fisher Scientific, Bremen, Germany). Statistical
differences for the INS-1 treated cells were evaluated by performing
two-tailed unpaired homoscedastic Student’s t test. Metabolic pathway networks and Volcano plot for putatively
assigned metabolites were obtained through Metaboanalyst.[36]
Results and Discussion
Direct Infusion Probe (DIP) Design and Optimization
The DIP was designed to reproducibly sample and ionize minute sample
volumes for direct infusion metabolomics. Our design combines application
of the high voltage directly on a stainless-steel capillary with a
nebulizer constructed from a PEEK sample tee where nitrogen gas is
supplied to the top of the tee (Figure ). The nebulizer allows for self-aspiration through
the Venturi self-pumping effect and enhanced solvent desolvation compared
to conventional nanoelectrospray sources. Compared to previous miniaturized
probes for direct infusion,[32,34] the DIP has a minimalistic
design that incorporates high voltage for efficient ESI directly to
flow/ionization capillary. The minimal probe volume of 0.14 μL
enables rapid transfer of minute amounts of material directly from
the vial to the mass spectrometer inlet for pneumatically assisted
electrospray ionization.
Figure 1
Design of the direct infusion probe (DIP). (A)
Schematic overview
of the probe in front of the mass spectrometer inlet and (B) photograph
of the assembled probe with a 200 μL centrifuge tube.
Design of the direct infusion probe (DIP). (A)
Schematic overview
of the probe in front of the mass spectrometer inlet and (B) photograph
of the assembled probe with a 200 μL centrifuge tube.For characterization and optimization of the DIP,
several key parameters
were evaluated using standard solutions. These parameters include
probe-to-MS distance, electrospray angle toward the MS inlet, applied
nitrogen pressure for self-aspiration, applied electrospray potential,
and MS inlet transfer capillary temperature. Good and reproducible
data was acquired when the tip of the probe was 2.5 mm away and at
a 45° angle directly in front of the mass spectrometer inlet
(Figure S1). Furthermore, the optimal conditions
were found using a nitrogen pressure of 5 bar (Figure S2), a MS inlet transfer capillary temperature of 300
°C (Figure S3), and an ESI voltage
of 1.9 kV (Figure S4). These conditions
resulted in a flow rate of 1 μL/min, which was measured by repeated
1 min sampling of solution with known density with weighing steps
between each sampling event. The achievable flow rate is attributed
to the inner diameter of the stainless-steel capillary in combination
with the nitrogen pressure applied to the PEEK tee.[37] During optimization, it was found that the position in
front of the mass spectrometer inlet was important for reproducibility.
A holder was therefore 3D-printed to easily position and fix the DIP
in front of the mass spectrometer inlet, which minimized accidental
movements and maximized reproducible experimental setups. Overall,
the DIP probe provides a simple tool for metabolite profiling from
very small sample volumes.
Evaluation of Carryover and Repeatability
The DIP has
the capacity for rapid sampling of only 5 μL samples and provides
square wave sample profiles without longitudinal diffusion. The square
wave profile clearly separates samples and simplifies data extraction,
analysis, and metabolite annotation. Given that the DIP has a capillary
volume of 0.14 μL, the solution within the capillary of the
DIP should theoretically be fully exchanged after 8 s with a flow
rate of 1 μL/min. The carryover between samples was experimentally
determined by switching between standard solutions containing nonendogenous
compounds and endogenous compounds from rat brain tissue extract.
Thus, either endogenous molecules from the brain extract or standards
from the standard solution should be detected. Results for glutamate
are depicted in red (endogenous) or blue (15N labeled)
in the square wave extracted ion chronograms (EIC) in Figure A (additional data in Figure S5). Note that the spike at the end of
each sampling event is due to a temporarily increased flow of the
residual sample inside the capillary.[30] Despite consecutive sampling, the carryover was considered minimal
after 10 scans, which corresponds to ∼8.5 s. At this time,
the carryover was 2.5% on average, although it was further decreased
to ∼1.1% after 20 scans, ∼17 s (Table S4, eq ). This shows that the carryover between samples is minimal even
without washing steps between, which is similar to other reports.[27] However, the first 10 scans should not be considered
in data analysis due to ongoing sample exchange in the capillary.
Furthermore, washing between samples is recommended for high concentration
samples or high salt loads.[38] The rapid
exchange of solution in the capillary, minimal dead volume, and minimal
carryover between samples has the potential for using the DIP for
rapid analysis of low volume samples.
Figure 2
Carryover and comparison between established
sources. (A) Extracted
ion chronograms of glutamate-15N (m/z 149.0575) in standard solution (blue) and glutamate (m/z 148.0604) in rat brain extract (red)
with a rolling average of 5. (B and C) Fold change of relative LOD
for selected metabolite standards between m/z 75 and 230 plotted for DIP compared to (B) ESI and (C)
HESI. Positive values indicated better LOD with the DIP. Protonated
adducts are black and sodiated adducts orange.
Carryover and comparison between established
sources. (A) Extracted
ion chronograms of glutamate-15N (m/z 149.0575) in standard solution (blue) and glutamate (m/z 148.0604) in rat brain extract (red)
with a rolling average of 5. (B and C) Fold change of relative LOD
for selected metabolite standards between m/z 75 and 230 plotted for DIP compared to (B) ESI and (C)
HESI. Positive values indicated better LOD with the DIP. Protonated
adducts are black and sodiated adducts orange.In addition to minimal carryover, high repeatability
is desirable
for comparative metabolite profiling of minute biological samples.
The repeatability of DIP was evaluated by analyzing a standard solution
four times within the same day and on three consecutive days to determine
intra- and interday repeatabilities, respectively. The intraday RSDs
of 24 individual standards as protonated or sodiated species was found
to be 4.5% on average, and down to 0.9% at the lowest (Table S5). These data demonstrate that the DIP
is a reliable method for direct infusion metabolomics and is comparable
with previous reports.[39] The interday RSD
values were higher than the intraday, with 31.2% on average, and 8.3%
at the lowest (Table S5). Interday repeatability
incorporates the variability introduced by physical day-to-day reassembly
of the DIP, causing slight variations in probe positioning and applied
gas pressure that ultimately affect flow rate and ionization. We anticipate
that variability in ionization is the major component contributing
to the larger interday RSD. This is corroborated by the decreased
average interday RSD of 20.0% after normalization to the total ion
current (Table S5), which compares well
with previous reports for commercial ESI sources at 18.8%[39] and is within the RSD of <20% recommended
RSDs from FDA.[40] Overall, these results
demonstrate adequate interday repeatability for the DIP with experimental
reassembly, although it is advisable to analyze samples within the
same batch using the same probe setup to minimize measurement variations.
Comparison to Commercially Available ESI Sources
Along
with providing measurements with high repeatability, adequate sensitivity
is essential for direct infusion metabolomics. The sensitivity of
the DIP was compared with commercially available electrospray ionization
sources ESI and heated ESI (HESI) using their individual optimized
settings and the same mass spectrometer (Table S3) and standard solutions of 25 biologically relevant metabolite
and lipid species. Following analysis, 1 min of acquired data with
stable TIC signal was extracted for each experiment and the LOD for
each standard was calculated according to eq , using the slope, S, of
the calibration curve versus the integrated signal (Table S6) and standard deviation of the linear regression
line residuals, σ, (Table S7).[41]The sensitivity was evaluated by the
LOD fold change, which was determined by subtracting the LOD obtained
with the DIP from the LOD obtained with the respective ESI or HESI
source, followed by dividing with the LOD of the DIP. A positive fold
change value for a detected standard thereby represents a lower LOD
for the DIP, suggesting that a higher sensitivity is afforded to that
specific ion. The resulting LOD fold changes for DIP compared to ESI
and HESI are plotted in Figure B and C, respectively, for standards detected between m/z 75 and 230 (Table S8). Notably, there are only minor differences in LOD fold-change
between the ionization sources, which indicates their functional similarity.
However, the fold change is on average 0.21- and 0.28-times better
for DIP compared to ESI and HESI, respectively. Furthermore, DIP presents
better LODs for 30 and 42 out of 47 detected standard peaks compared
with ESI and HESI, respectively. However, the lower flow rate of the
DIP (1 μL/min) compared to the ESI and HESI (5 μL/min)
generally provides overall longer ion accumulation times and less
intense ions per spectra (Table ). Despite the differences in ion sources, the results
show that DIP is fully comparable with the established sources despite
its simplicity and low flow rate.
Table 1
Ion Accumulation Times (IT) and TIC
Values for 20 Consecutive Scans of a Sample of Rat Brain Extract of
0.086 mg/mL Acquired with DIP, ESI, and HESI Sourcesa
IT (ms) ± RSD (%)
TIC ± RSD (%)
DIP
181.15 ± 3.02
1.17 × 107 ± 3.78
ESI
1.41 ± 8.61
1.54 × 109 ± 7.71
HESI
1.17 ± 6.30
1.84 × 109 ± 5.93
The flow rate of DIP is 1 μL/min
and the flow rate of ESI and HESI is 5 μL/min.
The flow rate of DIP is 1 μL/min
and the flow rate of ESI and HESI is 5 μL/min.Biological samples are chemically complex and can
have a high salt
content. The DIP was found to be tolerant to samples with high salt
content and to provide a wide range of detected metabolites and lipids
in a single scan from an extract of rat brain (Figure ). The single scan, without any averaged
microscans, in Figure corresponds to the extract of 0.8 ng of rat brain tissue, considering
analysis of 0.23 nL rat brain extract of 3.46 mg/mL (flow rate of
1 μL/min and duty cycle of ∼14 ms). The mass spectrum
consists of a plethora of ions from small polar metabolites to larger
hydrophobic lipids, including 19 of the 20 amino acids, neurotransmitters,
fatty acids, osmoregulators, sterols, and several phospholipid species,
where only a fraction are putatively annotated in Figure A–C. A full list of
218 putatively annotated features from which 120 are metabolites in
positive ion mode is provided in Table S9. Although, the low flow rate of the DIP overall reduces the number
of metabolites detected compared to ESI and HESI, the DIP retains
the high diversity of ions detected at one time from a complex sample
while only using a minute amount of material.
Figure 3
Single scan, without
any averaged microscans, high resolution spectrum
in positive ion mode from a methanolic rat brain extract analyzed
with DIP-MS. (A) Full spectrum from 70 to 1000 Da, (B) same spectrum
zoomed into m/z 100–206 and with several metabolites
annotated, and (C) same spectrum zoomed into m/z
700–900 with several annotated lipids. A list of additional
putatively annotated peaks is provided in Table S9.
Single scan, without
any averaged microscans, high resolution spectrum
in positive ion mode from a methanolic rat brain extract analyzed
with DIP-MS. (A) Full spectrum from 70 to 1000 Da, (B) same spectrum
zoomed into m/z 100–206 and with several metabolites
annotated, and (C) same spectrum zoomed into m/z
700–900 with several annotated lipids. A list of additional
putatively annotated peaks is provided in Table S9.
Metabolite Detection in Low Cell Number Samples
Most
metabolomics studies require tens of microliters of sample and at
least one million cells. The DIP was designed to reduce the sample
volume and minimize number of cells while retaining a high metabolite
coverage. These features are especially important for analysis of
a limited number of cells, such as unique samples isolated using fluorescence
activated cell sorting (FACS) or primary cells in cultures. To evaluate
metabolite detection of samples with low cell densities, 5 μL
samples containing methanolic extracts of 20–500 HEK293 cells/μL
were analyzed. Selected endogenous molecules were quantified using
internal standards according to eq , where the endogenous concentration ([endogenous])
was determined from the integrated signal, over 1 min, of the endogenous
molecule (Areaendogenous) to the internal standard (AreaIS) and multiplied by the concentration of the internal standard
([IS]).The resulting data show increasing
intensities correlating with increasing number of cells/μL.
Furthermore, the data suggests that as few as 20 cells/μL generates
information from both lipids and metabolites, such as amino acids
and glucose, at concentration down to 0.1 μM (Figure ). The most abundant metabolite
of HEK cells quantified here is glutamate followed by aspartic acid
and alanine, while phenylalanine has the lowest concentration (Figures , S6, and S7). Notably, the most
abundant lipid, PC 34:1, has a similar concentration as glucose and
the concentration of the second most abundant lipid, PC 32:1, is much
lower. Despite the potential to analyze down to 20 cells/μL,
it is advisable to use more than 100 cells/μL for a more extensive
metabolite profiling. Using an instrument to sort the cells (e.g.,
FACS) and accurately know the number of cells would be the ideal situation
when single cell analysis is required. Nevertheless, the opportunity
to use samples with low cell numbers provides new possibilities for
isolation and chemical profiling of a small number of unique cells.
Figure 4
Quantification
of endogenous metabolites and lipids in samples
with low cell densities. Concentrations of endogenous (A) small metabolites
and (B) phospholipids from 5 μL samples with 20 to 500 cells/μL.
Note that the Y-axes are for species depicted as
squares (left) and circles (right). Error bars represent one standard
deviation of triplicates of sample preparation. n = 3 per cell density, which equals 21 samples in total.
Quantification
of endogenous metabolites and lipids in samples
with low cell densities. Concentrations of endogenous (A) small metabolites
and (B) phospholipids from 5 μL samples with 20 to 500 cells/μL.
Note that the Y-axes are for species depicted as
squares (left) and circles (right). Error bars represent one standard
deviation of triplicates of sample preparation. n = 3 per cell density, which equals 21 samples in total.
Metabolite Alterations in Glucose Stimulated INS-1 Cells
Insulin-secreting INS-1 cells are commonly used to study the release
of insulin after exposure to glucose.[42−44] The DIP was used for
a combined nontargeted and quantitative targeted metabolome profiling
of cultured INS-1 cells exposed to either 1 or 20 mM of glucose for
15 min. The metabolites were putatively annotated based on accurate
mass for nontargeted statistical analysis and the results are presented
as a volcano plot in Figure A. The results reveal the dynamics of the metabolome and shows
significant increases and decreases in numerous metabolites between
the two treatment conditions (Table S10). For example, while aspartic acid is 0.4-times decreased in the
20 mM glucose treated cells, sorbitol is only observed in the cells
exposed to high glucose environment, which is relevant since glucose
is reduced to sorbitol in the polyol pathway. The global metabolic
network map in Figure B depicts the relation between the significantly altered metabolites,
which includes metabolic pathways of amino acid metabolism, in particular
alanine, aspartate, and glutamate metabolism and arginine and proline
metabolism (Table S11). Additionally, lipid
metabolism, including fatty acid biosynthesis and glycerophospholipids
metabolism are identified together with the citric acid cycle. These
pathways are important to normal cell function and to the regulation
of insulin secretion.[45,46] The nontargeted metabolomics
results are in agreement with previously published reports stating
that several metabolites are altered in response to high glucose concentrations.[45,46] In addition to analysis in positive mode, we performed the same
experiments in negative mode and the significant metabolite alterations
are detected in the similar metabolic pathways (Figure S8 and Tables S12 and S13). Thus, the DIP provides relevant data for exploring intracellular
metabolism dynamics from a low number of cells within a short analysis
time.
Figure 5
Nontargeted and quantitative targeted metabolite profiling of INS-1
cells exposed to either low (1 mM) or high (20 mM) glucose. (A) Nontargeted
data analysis: Volcano plot originated from the global median normalization
of a total of 190 features putatively assigned. A total of 56 metabolites
were considered significant (p < 0.05) according
to Student’s t test analysis and fold change
higher than 1.04. Red and blue correspond to up and down regulated
features, respectively, and in gray are the nonsignificant features.
(B) Network: The global metabolomic network highlighting in black
dots some of the significantly altered metabolites. (C) Targeted data
analysis: Concentration of metabolites in INS-1 cell exposed to either
low (light blue) or high (dark blue) glucose. All results are significantly
different (p < 0.05) between the treatment groups
according to two-tailed unpaired homoscedastic Student’s t test analysis. All results refer to the sodiated adducts
of each metabolite, except for valine and proline where the protonated
adducts are used. Error bars represent one standard deviation of n = 10 of sample preparation from the same batch.
Nontargeted and quantitative targeted metabolite profiling of INS-1
cells exposed to either low (1 mM) or high (20 mM) glucose. (A) Nontargeted
data analysis: Volcano plot originated from the global median normalization
of a total of 190 features putatively assigned. A total of 56 metabolites
were considered significant (p < 0.05) according
to Student’s t test analysis and fold change
higher than 1.04. Red and blue correspond to up and down regulated
features, respectively, and in gray are the nonsignificant features.
(B) Network: The global metabolomic network highlighting in black
dots some of the significantly altered metabolites. (C) Targeted data
analysis: Concentration of metabolites in INS-1 cell exposed to either
low (light blue) or high (dark blue) glucose. All results are significantly
different (p < 0.05) between the treatment groups
according to two-tailed unpaired homoscedastic Student’s t test analysis. All results refer to the sodiated adducts
of each metabolite, except for valine and proline where the protonated
adducts are used. Error bars represent one standard deviation of n = 10 of sample preparation from the same batch.In addition to nontargeted analysis, quantitative
results from
targeted metabolites were simultaneously acquired with isotopically
labeled internal standards to obtain quantitative comparisons. The
results show that exposure to 20 mM glucose causes statistically significant
(P < 0.05) changes in the concentration of wide
range of metabolites (Figure C). In addition to a higher observed concentration of glucose
in the 20 mM glucose treated cells, glutamate, FA 18:1 (oleic acid),
and alanine were also increased in cells exposed to high glucose in
accordance with previous studies.[47−52] In particular, glutamate acts as a link between glucose metabolism
and insulin release justifying its increase in the presence of high
glucose environment.[53] The increased amount
of GABA correlates with the hypothesis that GABA can affect insulin-secretion
regulation through the β-cell GABA shunt in response to glucose
stimulation.[51,52,54] Furthermore, since the DIP simultaneously ionizes all material in
the cell extract, our data also reveals alterations in PC and LPC
species after 15 min of glucose exposure (Figure C). Such results have previously only been
reported after long-term culture of INS-1 cells in glucotoxic conditions.[55] These results highlight the capability of the
DIP to provide a snapshot with wide molecular coverage of ongoing
metabolic events in biological systems. Based on the simplicity of
the DIP device in both construction and use, the potential for high
throughput analysis, the minimal sample preparation required, and
the high metabolite coverage, we anticipate that the DIP will be an
important tool for future targeted and nontargeted metabolite profiling
of cells and tissue studies where minimal volumes are desired.
Conclusions
We have presented a probe for direct infusion
MS, the DIP, that
is simple, robust, and can provide fast analysis of chemically complex
biological samples, such as cells and tissue. The probe readily provides
minutes of data from samples volumes as low as 5 μL and with
a cell density down to 20 cells/μL. This opens up the possibility
for new high-throughput metabolomics studies of unique FACS sorted
cells, primary cells of limited amounts, or smaller volumes of any
type of samples. The DIP is tolerant to high salt loads and chemically
complex samples of a range of concentrations and has minimal carryover,
thereby reducing the need for laborious and time intensive sample
preparation after extraction of metabolites from cells or tissue.
Thus, DIP presents itself as a promising alternative to conventional
direct infusion mass spectrometry for metabolite profiling. With the
potential to combine and simultaneously acquire both nontargeted metabolite
profiling and quantitative targeted analysis of selected metabolites,
the metabolome can be characterized in both qualitatively and quantitatively.
Additionally, the minimal sample preparation and preselection of molecular
species simultaneously provides data on a wide range of metabolite
and lipid classes. Overall, we anticipate that the minimalist design,
method simplicity, and high-quality data generated from minute biological
samples makes DIP a valuable tool for the wider metabolomics community.
Authors: James T Yurkovich; Daniel C Zielinski; Laurence Yang; Giuseppe Paglia; Ottar Rolfsson; Ólafur E Sigurjónsson; Jared T Broddrick; Aarash Bordbar; Kristine Wichuk; Sigurður Brynjólfsson; Sirus Palsson; Sveinn Gudmundsson; Bernhard O Palsson Journal: J Biol Chem Date: 2017-10-13 Impact factor: 5.157