| Literature DB >> 36061865 |
Madhu Puri1, Harsimran Kaur Brar1, Evanka Madan1, Rajesh Srinivasan2, Kapil Rawat2, Sai Siva Gorthi2, Geeta Kumari3, Raj Sah3, Sashi Bhusan Ojha4, Subhendu Panigrahi5, Gunanidhi Dhangadamajhi4, Rohini Muthuswami1, Shailja Singh3, Rentala Madhubala1.
Abstract
LAMP diagnosis of malaria is simple and cost-effective with acceptable sensitivity and specificity as compared to standard diagnostic modules such as microscopy, RDTs and nested PCR, and thus its deployment for onsite screening of malaria in resource-limited regions is under consideration. However, the requirement of an electricity-operated dry bath and bulky read-out unit is still a major concern. In an effort to simplify this limitation, we have developed a portable LAMP device and fluorescence readout unit which can be used in the rapid point-of-care diagnosis of malaria. We have developed a point-of-care diagnostic LAMP device that is easy to operate by a mobile application, and the results can be quantified with a fluorescent readout unit. The diagnostic performance of the device was evaluated in 90 P. falciparum-infected clinical isolates stored at 4°C for 6-7 years and 10 freshly collected isolates from healthy volunteers. The LOD and quantitative ability of LAMP in estimating parasitemia levels were revealed with laboratory-grown P. falciparum strain (3D7). The LAMP assay performed in our device was exclusive for P. falciparum detection with sensitivity and specificity determined to be 98.89% and 100%, respectively, in clinical isolates. The LOD was documented to be 1 parasite/µl at the cut-off ADC value of 20. Parasite density estimated from ADC values showed concordance with microscopically determined parasite density of the cultured P. falciparum 3D7 strain. The LAMP assay performed in our device provides a possible portable platform for its deployment in the point-of-care diagnosis of malaria. Further validation of the quantitative ability of the assay with freshly collected or properly stored clinical samples of known parasitemia is necessary for field applicability.Entities:
Keywords: ADC value; LAMP; LOD; diagnosis; parasite density; plasmodium falciparum
Mesh:
Year: 2022 PMID: 36061865 PMCID: PMC9437306 DOI: 10.3389/fcimb.2022.961832
Source DB: PubMed Journal: Front Cell Infect Microbiol ISSN: 2235-2988 Impact factor: 6.073
List of LAMP primers used.
| Primer | Sequence |
|---|---|
| Forward inner primer (FIP) | 5’ AGCTGGAATTACCGCGGCTG GGTTCCTAGAGAAACAATTGG 3’ |
| Backward inner primer (BIP) | 5’ TGTTGCAGTTAAAACGTTCGTAGCCCAAACCAGTTTAAATGAAAC 3’ |
| F3 | 5’ TGTAATTGGAATGATAGGAATTTA 3’ |
| B3c | 5’ GAAAACCTTATTTTGAACAAAGC 3’ |
| LPF | 5’ GCACCAGACTTGCCCT 3’ |
| LPB | 5’ TTGAATATTAAAGAA 3’ |
Figure 1Components of the portable LAMP device and fluorescence readout unit.(A) The LAMP device has 16 sample wells, a heating thermal block and a lid which features a heating element to prevent condensation during the amplification process. Three indicator lights in the front panel: red to show the operation status of system power, orange to show the Bluetooth connection, and green to display the amplification reaction process, are present. The reaction settings are set using a mobile application that connects the LAMP device to a smartphone. The app has a simple graphical user interface to set the temperature and time for the reaction and shows the real-time temperature of the block. (B) The fluorescence readout unit has an integrated LED light source, optical filters (excitation and emission filters suitable for SYBR Gold), a sample vial holder, a photodiode, and an electronic driving circuit. The samples are mixed with SYBR Gold and placed in the sample holder. When the measurement is initiated the LED and photodiode are synchronized to read the fluorescence emission from the sample. The system measures the fluorescence levels at an interval of 3 seconds (1 second on and 2 seconds off). The data is logged into a file where it can be used for further analysis (Puri et al., 2021).
Figure 2Exclusivity of the LAMP reaction performed in the device.The exclusivity of the LAMP reaction performed in the device was determined. One ng of genomic DNA from each of L. donovani Bob, P. vivax and P. falciparum strains was used in a LAMP reaction along with P. falciparum LAMP primers for 100 minutes, as described in Materials and Methods. (A) The confirmation of LAMP amplification was done by SYBR Gold detection. Diluted SYBR Gold nucleic acid stain was added to the LAMP products, and the color change was detected by visual examination. –: non-template control. (B) The quantification of SYBR Gold fluorescence was done by measuring the fluorescence intensity of the samples by a detection device, designated as ADC values. The mean + SEM of 10 ADC values are plotted for each sample. (C) Electrophoresis of LAMP products on a 2.5% agarose gel. M: 100 bp DNA ladder.
Figure 3Limit of detection of P. falciparum genomic DNA amplified in the LAMP device.DNA isolated from P. falciparum cultures containing 0.000001 to 10% parasitemia was used in a LAMP reaction for 100 minutes, as described in Materials and Methods. As a control, DNA isolated from a sample containing zero parasitemia, i.e. only blood, was also tested in the LAMP reaction. (A) The confirmation of LAMP amplification was done by SYBR Gold detection. Diluted SYBR Gold nucleic acid stain was added to the LAMP products, and the color change was detected by visual examination. –: non-template control. (B) The quantification of SYBR Gold fluorescence was done by measuring the fluorescence intensity of the samples by a detection device, designated as ADC values. The mean + SEM of 10 ADC values are plotted for each sample. (C) Electrophoresis of LAMP products on a 2.5% agarose gel. M: 100 bp DNA ladder.
Comparison of parasitemia estimated from ADC value with parasitemia determined by microscopy.
| Parasitemia (%) | ADC value | Parasitemia estimated from ADC value |
|---|---|---|
| 0.5 | 50 | 0.479 |
| 1 | 52 | 0.954 |
| 1.5 | 53 | 1.347 |
| 2 | 54 | 1.902 |
| 5 | 57 | 5.35 |
Figure 4LAMP amplification of P. falciparum DNA from patient samples.One hundred pg of DNA from 90 patient samples were used in a LAMP reaction for 100 minutes, as described in Materials and Methods. (A) The confirmation of LAMP amplification was done by SYBR Gold detection. Diluted SYBR Gold nucleic acid stain was added to the LAMP products, and the color change was detected by visual examination. –: non-template control. (B) The quantification of SYBR Gold fluorescence was done by measuring the fluorescence intensity of the samples by a detection device, designated as ADC values. The mean + SEM of 10 ADC values are plotted for each sample. (C) Electrophoresis of LAMP products on a 2.5% agarose gel. M: 100 bp DNA ladder.
Comparison of diagnostic test evaluation parameters of ITS-1 PCR and LAMP assay.
| Parameters | QBC | RDT | PCR | LAMP assay |
|---|---|---|---|---|
| True positives | 90 | 40 | 90 | 89 |
| False negatives | 0 | 0 | 0 | 1 |
| True negatives | 10 | 10 | 10 | 10 |
| False positives | 0 | 0 | 0 | 0 |
| Sensitivity (%) (95% CI) | 100.00 | 100.00 | 100.00 | 98.9 |
| Specificity (%) (95% CI) | 100 | 100 | 100 | 100 |
| Accuracy (%) (95% CI) | 100 | 100 | 100 | 99.01 |
Confidence intervals for sensitivity are exact Clopper-Pearson confidence intervals. Confidence intervals for predictive values are the standard logit confidence intervals given by (Mercaldo et al., 2007).