| Literature DB >> 36042395 |
Fang Xie1, Yu-Kun Luo2, Yu Lan1, Xiao-Qi Tian1, Ya-Qiong Zhu1, Zhuang Jin1, Ying Zhang1, Ming-Bo Zhang1, Qing Song1, Yan Zhang1.
Abstract
BACKGROUND: To assess the diagnostic efficacy of the computer-aided ultrasonic diagnosis system (CAD system) in differentiating benign and malignant thyroid nodules.Entities:
Keywords: Computer-aided diagnosis; Diagnostic efficacy; Thyroid nodule; Ultrasonography
Mesh:
Year: 2022 PMID: 36042395 PMCID: PMC9425995 DOI: 10.1186/s12880-022-00874-7
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 2.795
Fig. 1Automatic identification and quantization of ultrasound features of malignant thyroid nodules by the Am CAD-UT Detection System
Fig. 2Automatic identification and quantization of ultrasound features of benign thyroid nodules by the AmCAD-UT Detection System
General information of 296 thyroid nodules
| Pathology (n = 296) | ||
|---|---|---|
| Malignant: 151 (51%) | Benign: 145 (49%) | |
| Gender | ||
| Male | 23 (7.8%) | 56 (18.9%) |
| Female | 128 (43.2%) | 89 (30.1%) |
| Age | ||
| < 55 | 119 (40.2%) | 89 (30.1%) |
| ≥ 55 | 32 (10.8%) | 56 (18.9%) |
| Nodule size | ||
| ≤ 1 cm | 82 (27.7%) | 76 (25.7%) |
| 1–2 cm | 56 (18.9%) | 50 (16.9%) |
| ≥ 2 cm | 13 (4.4%) | 19 (6.4%) |
Comparison of diagnostic efficacy of four different methods on 296 thyroid nodules
| Group | Sensitivity% | Specificity% | Positive predictive value% | Negative predictive value% | Accuracy % | AUC (95% CI) |
|---|---|---|---|---|---|---|
| CAD system | 83.4 | 62.1 | 69.6 | 78.3 | 73.0 | 0.728 (0.673–0.777) |
| Junior physicians | 72.2 | 66.9 | 69.4 | 69.8 | 69.6 | 0.695 (0.640–0.747) |
| Senior physicians | 85.4 | 82.1 | 83.2 | 84.4 | 83.8 | 0.858 (0.808–0.909) |
| CAD system combined with junior physicians | 84.1 | 79.3 | 80.9 | 82.7 | 81.8 | 0.842 (0.790–0.894) |
Consistency between four diagnostic methods and pathological results
| CAD system versus pathology | Junior physicians versus pathology | Senior physicians versus pathology | CAD system combined with junior physicians versus pathology | |
|---|---|---|---|---|
| Kappa | 0.457 | 0.391 | 0.675 | 0.635 |
| 0.000 | 0.000 | 0.000 | 0.000 |
Comparison of diagnostic efficacy of CAD system on thyroid nodules of different sizes
| Group | Sensitivity% | Specificity % | Positive predictive value% | Negative predictive value% | Accuracy % | AUC (95% CI) |
|---|---|---|---|---|---|---|
| ≤ 1 cm | 78.0 | 63.1 | 69.6 | 72.7 | 70.9 | 0.706 (0.628–0.776) |
| 1–2 cm | 89.2 | 62 | 72.4 | 83.8 | 76.4 | 0.756 (0.663–0.835) |
| ≥ 2 cm | 92.3 | 57.9 | 60.0 | 91.7 | 71.9 | 0.751 (0.567–0.886) |