Detection ability of corneal biomechanical parameters for early diagnosis of ectasia.

Mohammad-Reza Sedaghat1, Hamed Momeni-Moghaddam2, Javad Heravian3, Atiyeh Ansari3, Helia Shayanfar1, Majid Moshirfar4,5,6.   

Abstract

PURPOSE: To assess the detection ability of corneal biomechanical parameters for early diagnosis of ectasia.
METHODS: This retrospective descriptive-analytical study included 134 normal eyes (control group) from 134 healthy subjects and 128 eyes with asymmetric contralateral corneal ectasia with normal topography (ACE-NT, study group) from 128 subjects with definite keratoconus in the opposite eye. Placido-disk-based corneal topography with TMS-4, Scheimpflug corneal tomography with Pentacam HR, and corneal biomechanical assessment with Corvis ST and ocular response analyzer (ORA) were performed. A general linear model was used to compare Corvis ST and ORA biomechanical parameters between groups, while central corneal thickness (CCT) and biomechanically corrected intraocular pressure (bIOP) were considered covariates. Receiving operator sensitivity curve (ROC) analysis was used to determine the cut-off point with the highest sensitivity and specificity along with the area under the curve (AUC) for each parameter. RESULT: All parameters of Corvis ST and ORA showed a statistically significant difference between the two groups except for the first (P = 0.865) and second (P = 0.226) applanation lengths, and deformation amplitude (P = 0.936). The discriminative analysis of corneal biomechanical showed that the highest accuracy for the classic, new, and combined parameters of Corvis ST was related to HCR (AUC: 0.766), IR & DAR (0.846), and TBI (0.966), respectively. Using ORA, the corneal resistance factor (0.866) had a higher detection ability than corneal hysteresis (0.826).
CONCLUSIONS: TBI has the best accuracy and the highest effect size for differential diagnosis of normal from ACE-NT eyes with a cut-off point of 0.24.
© 2022. The Author(s), under exclusive licence to The Royal College of Ophthalmologists.

Entities:  

Year:  2022        PMID: 36038724     DOI: 10.1038/s41433-022-02218-9

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   4.456


  37 in total

1.  Risk factors for corneal ectasia after LASIK.

Authors:  Khalid F Tabbara; Amgad A Kotb
Journal:  Ophthalmology       Date:  2006-09       Impact factor: 12.079

2.  Corneal biomechanical evaluation in healthy thin corneas compared with matched keratoconus cases.

Authors:  Bruno Machado Fontes; Renato Ambrósio; Guillermo Coca Velarde; Walton Nosé
Journal:  Arq Bras Oftalmol       Date:  2011 Jan-Feb       Impact factor: 0.872

3.  Assessment of ocular biomechanics using dynamic ultra high-speed Scheimpflug imaging in keratoconic and normal eyes.

Authors:  Lei Tian; Match W L Ko; Li-ke Wang; Jia-ying Zhang; Tian-jie Li; Yi-fei Huang; Yong-ping Zheng
Journal:  J Refract Surg       Date:  2014-10-07       Impact factor: 3.573

4.  A statistical approach to classification of keratoconus.

Authors:  Murat Ucar; Hasan Basri Cakmak; Baha Sen
Journal:  Int J Ophthalmol       Date:  2016-09-18       Impact factor: 1.779

5.  Detection of Keratoconus With a New Biomechanical Index.

Authors:  Riccardo Vinciguerra; Renato Ambrósio; Ahmed Elsheikh; Cynthia J Roberts; Bernardo Lopes; Emanuela Morenghi; Claudio Azzolini; Paolo Vinciguerra
Journal:  J Refract Surg       Date:  2016-12-01       Impact factor: 3.573

Review 6.  Biomechanics of corneal ectasia and biomechanical treatments.

Authors:  Cynthia J Roberts; William J Dupps
Journal:  J Cataract Refract Surg       Date:  2014-04-26       Impact factor: 3.351

7.  Sensitivity and specificity of posterior corneal elevation measured by Pentacam in discriminating keratoconus/subclinical keratoconus.

Authors:  Ugo de Sanctis; Carlotta Loiacono; Lorenzo Richiardi; Davide Turco; Bernardo Mutani; Federico M Grignolo
Journal:  Ophthalmology       Date:  2008-04-11       Impact factor: 12.079

8.  Subclinical keratoconus detection by pattern analysis of corneal and epithelial thickness maps with optical coherence tomography.

Authors:  Yan Li; Winston Chamberlain; Ou Tan; Robert Brass; Jack L Weiss; David Huang
Journal:  J Cataract Refract Surg       Date:  2016-02       Impact factor: 3.351

9.  Evaluation of Scheimpflug Tomography Parameters in Subclinical Keratoconus, Clinical Keratoconus and Normal Caucasian Eyes.

Authors:  Samira Huseynli; Farah Abdulaliyeva
Journal:  Turk J Ophthalmol       Date:  2018-06-28

10.  Corneal biomechanical parameters in keratoconus eyes with abnormal elevation on the back corneal surface only versus both back and front surfaces.

Authors:  Mohammad-Reza Sedaghat; Hamed Momeni-Moghaddam; Cynthia J Roberts; Nasim Maddah; Renato Ambrósio; Seyed Rafi Hosseini
Journal:  Sci Rep       Date:  2021-06-07       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.