| Literature DB >> 36035320 |
Angela C Rutledge1,2, Anna Johnston1,3, Ronald A Booth1,4, Kika Veljkovic1,5, Dana Bailey1,6, Hilde Vandenberghe7,8, Gayle Waite6,7, Lynn C Allen7,9, Andrew Don-Wauchope1,5, Pak Cheung Chan1,10, Julia Stemp3,7, Pamela Edmond1,2, Victor Leung1,11, Berna Aslan3,12.
Abstract
Objectives: Verifying new reagent or calibrator lots is crucial for maintaining consistent test performance. The Institute for Quality Management in Healthcare (IQMH) conducted a patterns-of-practice survey and follow-up case study to collect information on lot verification practices in Ontario.Entities:
Keywords: CLSI, Clinical and Laboratory Standards Institute; Calibrator; ELISA, enzyme-linked immunosorbent assay; GC, gas chromatography; HPLC, high-performance liquid chromatography; IQMH, Institute for Quality Management in Healthcare; ISO, International Organization for Standardization; LC-MS/MS, liquid chromatography tandem mass spectrometry; Lot number; PT, proficiency testing; QC, quality control; RIA, radioimmunoassay; Reagent; TSH, thyroid-stimulating hormone; Verification
Year: 2022 PMID: 36035320 PMCID: PMC9399155 DOI: 10.1016/j.plabm.2022.e00300
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Fig. 1The case study.
Fig. 2Acceptance criteria provided to laboratories as options.
Test systems and examples provided to participants. Participants were asked to respond whether they performed verification of new lots of reagent or calibrator for all tests, some tests, or no tests within each test system category. RIA, radioimmunoassay; ELISA, enzyme-linked immunosorbent assay; HPLC, high-performance liquid chromatography; GC, gas chromatography; LC-MS/MS, liquid chromatography tandem mass spectrometry.
| Test Systems | Examples |
|---|---|
| Automated chemistry | Main analyzer |
| Automated immunoassay | |
| Semi-automated quantitative instruments | Blood gas analyzer, point-of-care testing devices |
| Semi-quantitative instruments | Electrophoresis, antibody titre, urine dipsticks |
| Qualitative or point-of-care testing assays | Drugs of abuse, pregnancy test |
| Commercial test kits | RIA, ELISA, western blot |
| In-house developed assays | HPLC, GC, LC-MS/MS |
| Other systems (to be specified in the section below this question on the questionnaire) |
Fig. 3Summary of responses from participating laboratories about reagent or calibrator lot verification practices according to test system. Results are expressed as a percentage of total responses (n-value shown in the figure) for each test system type from laboratories that performed such testing on-site.
Fig. 4Summary of the reasons provided for not conducting reagent or calibrator lot verifications. Responses were from 74 laboratories about reagents and 101 laboratories about calibrators and are expressed as a percentage of all responses within the reagent or calibrator category.
Fig. 5Reasons for not performing reagent or calibrator lot verification for all tests within a methodology category. Results are expressed as a percentage of total responses (n-value shown in the figure) for each test system.
Fig. 6Participant responses regarding types of material used in reagent or calibrator lot verifications for various analytical systems. Results are expressed as a percentage of total responses (n-value shown in the figure) for each test system.
Fig. 7Participant responses regarding number of patient or commercial material samples used in reagent or calibrator lot verifications for automated chemistry or immunoassay tests. A) average number of patient material samples used in verification of new reagent or calibrator lots. Results are expressed as a percentage of total responses (169 for reagent, 123 for calibrator). B) average number of commercial material samples used in verification of new reagent or calibrator lots. Results are expressed as a percentage of total responses (172 for reagent, 129 for calibrator).
Fig. 8Criteria used to determine acceptance or rejection of a new lot number. A total of 592 criteria were provided by 174 laboratories.
Fig. 9Actions that would be taken when accepting a new lot number that had failed previously-established acceptance criteria. A total of 246 actions were selected by 150 laboratories.
The number of laboratories that applied the QC and/or patient result acceptance criteria included in the case study and the verdicts from those criteria. The total number of laboratories that selected at least one criterion used in their decision was 119. The data are presented as a heat map, with higher numbers of responses being red and fewer numbers of responses being blue.