Commutability limitations influence quality control results with different reagent lots.

W Greg Miller1, Aybala Erek, Tina D Cunningham, Olajumoke Oladipo, Mitchell G Scott, Robert E Johnson.   

Abstract

BACKGROUND: Good laboratory practice includes verifying that each new lot of reagents is suitable for use before it is put into service. Noncommutability of quality control (QC) samples with clinical patient samples may preclude their use to verify consistency of results for patient samples between different reagent lots.
METHODS: Patient sample results and QC data were obtained from reagent lot change verification records for 18 QC materials, 661 reagent lot changes, 1483 reagent lot change-QC events, 82 analytes, and 7 instrument platforms. The significance of between-lot differences in the results for QC samples compared with those for patient samples was assessed by a modified 2-sample t test adjusted for heterogeneity of QC and patient sample measurement variances.
RESULTS: Overall, 40.9% of reagent lot change-QC events had a significant difference (P < 0.05) between results for QC samples compared with results for patient samples between 2 reagent lots. For QC results with differences <1.0 SD interval (83.1% of total), 37.7% were significantly different from the changes observed for patient samples. For QC results with differences ≥1.0 SD interval (16.9% of total), 57.0% were significantly different from those for patient samples.
CONCLUSIONS: Occurrence of noncommutable results for QC materials was frequent enough that the QC results could not be used to verify consistency of results for patient samples when changing lots of reagents.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21097677     DOI: 10.1373/clinchem.2010.148106

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  15 in total

Review 1.  Lot-to-Lot Variation.

Authors:  Simon Thompson; Douglas Chesher
Journal:  Clin Biochem Rev       Date:  2018-05

2.  Ligand binding assay critical reagents and their stability: recommendations and best practices from the Global Bioanalysis Consortium Harmonization Team.

Authors:  Lindsay E King; Esme Farley; Mami Imazato; Jeannine Keefe; Masood Khan; Mark Ma; K Susanne Pihl; Priya Sriraman
Journal:  AAPS J       Date:  2014-04-01       Impact factor: 4.009

Review 3.  Bias Assessment of General Chemistry Analytes using Commutable Samples.

Authors:  Gus Koerbin; Jillian R Tate; Julie Ryan; Graham Rd Jones; Ken A Sikaris; David Kanowski; Maxine Reed; Janice Gill; George Koumantakis; Tina Yen; Andrew St John; Peter E Hickman; Aaron Simpson; Peter Graham
Journal:  Clin Biochem Rev       Date:  2014-11

4.  CUSUM-Logistic Regression analysis for the rapid detection of errors in clinical laboratory test results.

Authors:  Maureen L Sampson; Verena Gounden; Hendrik E van Deventer; Alan T Remaley
Journal:  Clin Biochem       Date:  2015-10-30       Impact factor: 3.281

5.  Adjusting Quality Control Chart Limits for WBC, RBC, Hb, and PLT Counts to Reduce Daily Control Risks in Hospital Laboratory.

Authors:  Chen-Mao Liao; Chih-Ming Lin; Chin-Chia Kuo; Ming-Shu Chen; Chun-Yang Huang; Ching-Yuan Lin
Journal:  Risk Manag Healthc Policy       Date:  2020-12-17

Review 6.  Interpretation of EQA results and EQA-based trouble shooting.

Authors:  Gunn Berit Berge Kristensen; Piet Meijer
Journal:  Biochem Med (Zagreb)       Date:  2017-02-15       Impact factor: 2.313

7.  Internal quality assurance of HIL indices on Roche Cobas c702.

Authors:  Giuseppe Lippi; Janne Cadamuro; Elisa Danese; Matteo Gelati; Martina Montagnana; Alexander von Meyer; Gian Luca Salvagno; Ana-Maria Simundic
Journal:  PLoS One       Date:  2018-07-06       Impact factor: 3.240

8.  Optimization and validation of patient-based real-time quality control procedure using moving average and average of normals with multi-rules for TT3, TT4, FT3, FT3, and TSH on three analyzers.

Authors:  Chao Song; Jun Zhou; Jun Xia; Deli Ye; Qian Chen; Weixing Li
Journal:  J Clin Lab Anal       Date:  2020-05-03       Impact factor: 2.352

9.  Analysis of short-term variation and long-term drift during reagent kit lot change in an NABL accredited clinical biochemistry laboratory.

Authors:  Vivek Ambade; Pratibha Misra; Yaongamphi Vashum; Mukul Sharma; Bhasker Mukherjee; Kapil Bhatia; Manoj Puliyath; Ponnaiah Rasu; Prakash Berthwal Indra; Madathan Kandi Sibin
Journal:  J Med Biochem       Date:  2021-01-26       Impact factor: 3.402

10.  Failure of internal quality control in detecting significant reagent lot shift in serum creatinine measurement.

Authors:  Xiaoting Chen; Jia Wang; Wei Zhang; Erfu Xie; Bingfeng Zhang; Hua-Guo Xu
Journal:  J Clin Lab Anal       Date:  2019-08-02       Impact factor: 2.352