Western blotting is a widely used technique for molecular-weight-resolved analysis of proteins and their posttranslational modifications, but high-throughput implementations of the standard slab gel arrangement are scarce. The previously developed Microwestern requires a piezoelectric pipetting instrument, which is not available for many labs. Here, we report the Mesowestern blot, which uses a 3D-printable gel casting mold to enable high-throughput Western blotting without piezoelectric pipetting and is compatible with the standard sample preparation and small (∼1 μL) sample sizes. The main tradeoffs are reduced molecular weight resolution and higher sample-to-sample CV, making it suitable for qualitative screening applications. The casted polyacrylamide gel contains 336, ∼0.5 μL micropipette-loadable sample wells arranged within a standard microplate footprint. Polyacrylamide % can be altered to change molecular weight resolution profiles. Proof-of-concept experiments using both infrared-fluorescent molecular weight protein ladder and cell lysate (RIPA buffer) demonstrate that the protein loaded in Mesowestern gels is amenable to the standard Western blotting steps. The main difference between Mesowestern and traditional Western is that semidry horizontal instead of immersed vertical gel electrophoresis is used. The linear range of detection is at least 32-fold, and at least ∼500 attomols of β-actin can be detected (∼29 ng of total protein from mammalian cell lysates: ∼100-300 cells). Because the gel mold is 3D-printable, users with access to additive manufacturing cores have significant design freedom for custom layouts. We expect that the technique could be easily adopted by any typical cell and molecular biology laboratory already performing Western blots.
Western blotting is a widely used technique for molecular-weight-resolved analysis of proteins and their posttranslational modifications, but high-throughput implementations of the standard slab gel arrangement are scarce. The previously developed Microwestern requires a piezoelectric pipetting instrument, which is not available for many labs. Here, we report the Mesowestern blot, which uses a 3D-printable gel casting mold to enable high-throughput Western blotting without piezoelectric pipetting and is compatible with the standard sample preparation and small (∼1 μL) sample sizes. The main tradeoffs are reduced molecular weight resolution and higher sample-to-sample CV, making it suitable for qualitative screening applications. The casted polyacrylamide gel contains 336, ∼0.5 μL micropipette-loadable sample wells arranged within a standard microplate footprint. Polyacrylamide % can be altered to change molecular weight resolution profiles. Proof-of-concept experiments using both infrared-fluorescent molecular weight protein ladder and cell lysate (RIPA buffer) demonstrate that the protein loaded in Mesowestern gels is amenable to the standard Western blotting steps. The main difference between Mesowestern and traditional Western is that semidry horizontal instead of immersed vertical gel electrophoresis is used. The linear range of detection is at least 32-fold, and at least ∼500 attomols of β-actin can be detected (∼29 ng of total protein from mammalian cell lysates: ∼100-300 cells). Because the gel mold is 3D-printable, users with access to additive manufacturing cores have significant design freedom for custom layouts. We expect that the technique could be easily adopted by any typical cell and molecular biology laboratory already performing Western blots.
The Western blot has been a staple of
molecular biology research
for decades since its first description in 1979.[1] It uses vertical immersed tank-based polyacrylamide gel
electrophoresis (PAGE) to separate proteins by molecular weight, followed
by transfer to a nitrocellulose or poly(vinylidene fluoride) (PVDF)
membrane, and finally the application of antibodies to sensitively
detect levels of proteins, posttranslational modifications, and even
protein complexes.[2−4] Detection modalities include the enzyme-mediated
generation of colorimetric molecules or light, or direct conjugation
of fluorescent molecules to antibodies,[5,6] which, when
combined with carefully designed experiments, can be quantitative.[7−9] Western blotting is still widely ingrained in biomedical research
as a protein analytic tool, even perhaps the most used technique in
protein-related publications in the last 10 years.[10] In fact, the use of Western blotting, despite falling “out
of fashion”, seems stable according to publication metrics.[10]Although Western blot usage remains high,
there are notable limitations.
Reliance on antibodies for detection is increasingly criticized,[11,12] although the separation of proteins by molecular weight is a strong
indicator of antibody validity not typically available to other antibody-based
technologies, and Western blotting is often used as a confirmatory
assay to bolster the support generated by other protein assays. Multiplexing
is limited to a handful of analytes per gel, which can be increased
slightly by stripping antibodies from the membrane and reprobing with
new antibodies,[5,13] cutting the membrane into targeted
molecular weight range strips for incubation of each with different
antibodies,[14,15] or orthogonal detection methods.[16,17] Lastly, traditional Westerns are limited by throughput and sample
size; typical gels contain only ∼10 wells for the analysis
of 10 samples simultaneously, and each sample usually requires ∼10
ug of total protein content from cell or tissue lysates. It is this
latter limitation of throughput and sample size that we focus on in
this paper with the Mesowestern blot.Before describing the
Mesowestern blot, it is instructive to review
the myriad of other related protein analytic technologies that address
the shortcomings of the Western blot. Reverse-phase protein arrays
(RPPAs) use lysates similar to Western blotting but greatly increase
multiplexing by spotting lysates on chips so that hundreds of antibodies
can be used simultaneously.[18,19] However, lysates are
not separated by molecular weight, which causes increased stringency
for antibody quality; in fact, antibodies are often validated for
use in RPPA by Western blot. Luminex xMAP technology,[20] although technologically distinct from RPPA as it uses
barcoded, antibody-conjugated beads, also offers increased multiplexing
from cell lysates but does not separate proteins by molecular weight.
Enzyme-linked immunosorbent assay (ELISA) has been established even
longer than the Western blot and uses two antibodies, one to capture
the analyte from a lysate and the other to detect the captured analyte,
with detection modalities similar to Western blots.[21,22] Although ELISA does not separate analytes by molecular weight, the
use of two different antibodies for the same target can, in some cases,
compensate for specificity issues with one, although obviously the
need for two antibodies can be a drawback itself. ELISA enables high-throughput
implementation in multiwell plates for the simultaneous analysis of
hundreds of samples. Mass spectrometry-based proteomics is antibody-free
and can analyze virtually any protein present in a lysate so long
as it is ionizable.[23−27] However, specific posttranslational modifications are not always
observable.[5] Moreover, findings from mass
spectrometry experiments often require orthogonal validation with
antibody-based techniques such as a Western blot.[28] Thus, there remains space for a more high-throughput Western
blot for analytes that are not amenable to mass spectrometry or when
increased specificity is needed for antibodies. Moreover, Western
blotting is likely to remain broadly useful as a complementary and
confirmatory assay.There have been advances in Western blotting
itself that have improved
on the aforementioned limitations. Single-cell Western blotting using
PAGE (Protein Simple) has been developed, greatly reducing sample
size requirements.[29,30] Other innovative Protein Simple
apparati use capillary electrophoresis rather than a slab gel to allow
analysis of up to 96 samples in a single loading (12 simultaneously)
in a streamlined manner.[31−33] Digiwest combines Luminex technology
with Western blotting by completing electrophoresis and transfer but
then cutting the membrane into molecular weight sections to be analyzed
by separate spectrally distinct beads.[34] This provides the multiplexing capabilities of Luminex with the
molecular weight separation of Western. The Microwestern blot[35−37] uses a piezoelectric pipetting apparatus to spot nL amounts of lysate
onto a typical-sized slab gel, followed by semidry horizontal electrophoresis
(as opposed to tank-based), and finally, a gasket system for incubating
different parts of the resultant membrane with up to 96 different
antibodies. Thus, the Microwestern addresses both throughput and,
to some extent, multiplexing limitations. However, the piezoelectric
pipetting apparatus is not available to many labs.[35,37] There is not yet a slab gel-based Western technology that addresses
throughput and sample size limitations that is micropipette-loadable.Here, we present the Mesowestern blot that, similar to the Microwestern,
allows for high-throughput analysis of hundreds of small samples in
a typical-sized slab gel but does not require piezoelectric pipetting
because it is micropipette-loadable. To do this, we designed and 3D-printed
a gel casting mold that produces a polyacrylamide gel with 336, ∼0.5
μL sample wells arranged with 8 rows by 42 columns in a microplate
footprint. The main tradeoff is molecular weight resolution because
samples have less distance to migrate. However, the format is flexible
because the cast is 3D-printed and gel acrylamide % can be adjusted.
Proof-of-concept experiments using both infrared-fluorescent molecular
weight ladder and cell lysates demonstrate that proteins loaded in
Mesowestern gels are amenable to the standard Western blotting steps
of gel electrophoresis followed by the transfer to a membrane for
imaging. These experiments also show another main tradeoff that sample-to-sample
CV is high, making the technique more suited for qualitative screening
applications. The main difference from Western blotting is horizontal
electrophoresis as opposed to tank-based electrophoresis, and, as
mentioned above, reduced molecular weight resolution. Because the
gel mold is 3D-printable, users with access to institutional additive
manufacturing cores (which are relatively commonplace) have significant
design freedom for custom layouts. We expect that the technique could
be easily adopted by any typical cell and molecular biology laboratory
already performing Western blots.
Results
Mesowestern Process
The Mesowestern process (Figure A) begins with casting
a 1.2 mm thick polyacrylamide slab gel in the 3D-printed mold (Figure ). The mold itself
consists of two pieces, the “top” and “bottom”.
The top contains the ports in which the unpolymerized gel is loaded,
and the bottom contains the impressions of the microwells into which
lysates will be loaded after casting. Each microwell negative is roughly
a trapezoidal prism (due to the nature of the 3D printer used) that
is 0.5 mm in height and is slightly longer in the direction perpendicular
to voltage (we found this results in better gel entry and less band
dispersion compared to equal lengths or longer in the other direction).
This gives a total volume of a little over 0.5 μL. The entire
mold has dimensions of a microplate (∼9 by 13 cm). It contains
eight rows of microwell negatives, with each row containing 42 columns,
for a total of 336 microwells per gel. Between rows, there is ∼9
mm for proteins to separate and ∼2 mm between microwells in
the same row. All of these features are malleable by simply changing
the CAD file for 3D printing.
Figure 1
Mesowestern process and example results. A.
First, the gel is polymerized
in the gel casting mold, and after polymerization, it is removed so
that samples can be loaded as desired. Electrophoresis takes place
on a horizontal apparatus, which is the main difference between Mesowestern
and regular Western. After electrophoresis, the Mesowestern and regular
Western workflows are identical, with transfer to membrane (tank-based
is shown), scanning/visualization, and analysis. B. An example Mesowestern
PVDF membrane where the molecular weight ladder was loaded in semiregular
patterns for illustrative purposes. The entire membrane scan is roughly
of microplate dimensions, and one “block” of ladder
is highlighted. A Mesowestern lane is only approximately 9 mm but
resolves molecular weights between 125 and 25 kDa reasonably well
(at 9.5% acrylamide used here). Electrophoresis was carried out at
100 V for ∼2 h.
Figure 2
Gel casting mold. The mold consists of two pieces, which
we refer
to as “top” and “bottom”. The top contains
the loading port for the unpolymerized gel solution, whereas the bottom
contains the raised regions, which become wells in the Mesowestern
gel. The gel dimensions are approximately 9 cm by 13 cm in width and
length and is 1.2 mm thick. Each well is a rectangle that is 1 mm
by 1.2 mm in width and length and is 0.5 mm deep. Wells are spaced
1.8 mm apart and have 8.7 mm to run in their lane before the next
well is reached.
Mesowestern process and example results. A.
First, the gel is polymerized
in the gel casting mold, and after polymerization, it is removed so
that samples can be loaded as desired. Electrophoresis takes place
on a horizontal apparatus, which is the main difference between Mesowestern
and regular Western. After electrophoresis, the Mesowestern and regular
Western workflows are identical, with transfer to membrane (tank-based
is shown), scanning/visualization, and analysis. B. An example Mesowestern
PVDF membrane where the molecular weight ladder was loaded in semiregular
patterns for illustrative purposes. The entire membrane scan is roughly
of microplate dimensions, and one “block” of ladder
is highlighted. A Mesowestern lane is only approximately 9 mm but
resolves molecular weights between 125 and 25 kDa reasonably well
(at 9.5% acrylamide used here). Electrophoresis was carried out at
100 V for ∼2 h.Gel casting mold. The mold consists of two pieces, which
we refer
to as “top” and “bottom”. The top contains
the loading port for the unpolymerized gel solution, whereas the bottom
contains the raised regions, which become wells in the Mesowestern
gel. The gel dimensions are approximately 9 cm by 13 cm in width and
length and is 1.2 mm thick. Each well is a rectangle that is 1 mm
by 1.2 mm in width and length and is 0.5 mm deep. Wells are spaced
1.8 mm apart and have 8.7 mm to run in their lane before the next
well is reached.During casting, the mold stands upright and is
held tight by household
C-clamps, while the freshly prepared unpolymerized gel solution (see
the Methods section) is loaded into the casting
device from the top, similar to traditional gel casting between glass
plates. After polymerization (∼30 min), the mold top and bottom
are taken apart and the gel can be carefully removed for the loading
of samples in the microwells via micropipette. After sample loading,
horizontal electrophoresis separates proteins by molecular weight.
This step is the biggest difference from the traditional Western blotting,
which typically uses immersed tank vertical electrophoresis. Following
electrophoresis, the workflow is generally indistinguishable from
traditional Western blotting. Tank-based (or semidry) transfer can
be employed to move the separated proteins to a nitrocellulose or
PVDF membrane, membranes are incubated with antibodies (with block/wash
steps), and finally scanned for the visualization of bands (we use
LI-COR infrared fluorescence in this work).As a simple demonstration,
we loaded molecular weight ladder in
semiregular patterns throughout a 9.5% acrylamide Mesowestern gel
(Figure B). Although
the distance each sample has to run (∼9 mm) is much smaller
than the standard Western blot, and there is no “stacking gel”
available, protein separation is reasonably uniform throughout the
gel and molecular weight standards within the ladder are distinctly
observable between 25 and ∼125 kDa and to some extent at 260
kDa with lesser resolution. However, the tradeoff between Mesowestern
and traditional Western—reduced molecular weight resolution—is
clear. Overall, this pipeline establishes a Mesowestern workflow that
is highly similar to traditional Western but is much higher throughput
with smaller sample sizes.
Comparison of Macroscale Slab Western Technologies
After establishing the basic Mesowestern workflow, it is instructive
to revisit the similarities and differences between that, the traditional
Western, and the Microwestern,[36] the only
other high-throughput slab Western technology (Figure and Table ). At the stage of sample preparation, Mesowestern
and Western are identical, whereas Microwestern has multiple differences.
Sodium dodecyl sulfate (SDS) and DTT are in the Microwestern lysis
buffer, as opposed (primarily) to being added after lysis and during
the preparation of samples to be loaded for Western and Mesowestern.
This causes differences in the total protein content assays that can
be used. In Microwestern, sonication steps to clear lysate (Covaris[36] or VialTweeter[37] as
opposed to centrifugation) and spin column-based sample concentration[36,37] are used.
Figure 3
Comparison of Mesowestern to Microwestern and regular Western.
There are significant differences between the Microwestern, the most
comparable high-throughput slab Western, and the Mesowestern. In terms
of processing and workflow, the Mesowestern is very similar to traditional
Western. The main differences are that the gel is cast with the 3D-printed
mold, rather than between two glass plates, that much smaller sample
volumes are required, and that horizontal electrophoresis is employed.
Horizontal electrophoresis is the main point of similarity between
Microwestern and Mesowestern. The reliance of Microwestern on a piezoelectric
pipetting apparatus creates several upstream problems, including the
fact that there are no wells in a Microwestern gel, that lysis buffer
is nonstandard and lysates must be sonicated and concentrated. After
electrophoresis, the workflows of all three processes are the same.
Table 1
Comparison of Each Stage of the Western,
Microwestern, and Mesowestern Methods
step
Microwestern
array[35−37]
Mesowestern blot
Western blot
1. lysate preparation
∼20
μL per sample volume
∼1 μL per sample
volume
∼10 μL per sample volume
requires sonication
standard
lysis buffers
standard lysis buffers
requires high concentration (∼10 mg/mL),
necessitating
spin columns
hundreds
of samples can be “spotted” with piezoelectric
pipetting
hundreds of ∼1 μL samples can
be loaded with micropipettes
tens of ∼10 μL
samples can be loaded with micropipettes
4. protein separation
horizontal electrophoresis
horizontal electrophoresis
vertical (tank) electrophoresis
5. transfer
wet or semidry transfer
wet or semidry transfer
wet or semidry transfer
6. antibody incubation
incubate different
sections or the entire membrane with antibodies
incubate
different sections or the entire membrane with antibodies
incubate the entire membrane with antibodies.
7. analysis and quantification
lower-molecular-weight
resolution due to smaller lanes
lower-molecular-weight
resolution due to smaller lanes
distinct bands with
greater-molecular-weight resolution due
to larger lanes/stacking.
Comparison of Mesowestern to Microwestern and regular Western.
There are significant differences between the Microwestern, the most
comparable high-throughput slab Western, and the Mesowestern. In terms
of processing and workflow, the Mesowestern is very similar to traditional
Western. The main differences are that the gel is cast with the 3D-printed
mold, rather than between two glass plates, that much smaller sample
volumes are required, and that horizontal electrophoresis is employed.
Horizontal electrophoresis is the main point of similarity between
Microwestern and Mesowestern. The reliance of Microwestern on a piezoelectric
pipetting apparatus creates several upstream problems, including the
fact that there are no wells in a Microwestern gel, that lysis buffer
is nonstandard and lysates must be sonicated and concentrated. After
electrophoresis, the workflows of all three processes are the same.Both Microwestern and Western gels are cast with glass
plates,
whereas the Mesowestern uses the previously described 3D-printed mold.
There are no wells in the Microwestern gel, whereas hundreds of small
wells are built into the Mesowestern gel cast, and wells are introduced
into a Western gel via a low polyacrylamide % stacking gel, which
promotes subsequent sample focusing and improves band dispersion.[38] Also, gradient acrylamide % gels are available
for Western,[39] but not for Microwestern
or Mesowestern.A primary difference with Microwestern is the
piezoelectric pipetting-based
“spotting” of samples, which then enter the well-less
gel through adsorption or diffusion,[36] as
compared to Mesowestern and Western where both use micropipettes to
load samples into wells. The hundreds of Mesowestern gel wells hold
∼0.5 μL of lysate each, whereas the ∼10 Western
gel wells hold ∼10–40 μL of lysate each. During
piezoelectric pipetting, ∼10–20 μL of sample is
needed in the microwell plate that serves as the sample source and
also in the apparatus tubing to ensure robust spotting function.[37]After samples are loaded, both the Microwestern
and Mesowestern
use semidry horizontal electrophoresis to separate proteins by molecular
weight.[36,37] Western typically uses immersed tank-based
vertical electrophoresis. After gel electrophoresis, there are a few
differences between all three techniques with regard to transfer and
preparation for antibody incubation/imaging. The Microwestern often
uses a membrane gasket system to enable incubation with up to 96 different
antibodies. The main differences also manifest with the molecular
weight resolution, which is reduced for Microwestern and Mesowestern
because there is no stacking gel and there is less distance for separation.We conclude that the Mesowestern offers benefits compared to Microwestern,
primarily the elimination of the reliance on piezoelectric pipetting.
In comparison to the Western, the Mesowestern offers over 10-fold
higher throughput with ∼10-fold decreased sample volume requirements
(we investigate sensitivity further below).
Controlling Molecular Weight Resolution by Varying Acrylamide
Composition
Given the inherently lower-molecular-weight resolution
of the Mesowestern as compared to that of the regular Western, we
asked whether the acrylamide proportion could be varied in Mesowestern
gels to enable more targeted separation of different molecular weight
ranges. This is routinely done in regular Westerns. Therefore, we
cast gels with 6, 9.5, 12, and 18% acrylamide compositions (see the Methods section) and evaluated the molecular weight
separation of a ladder standard containing reference bands at 160,
90, 50, 30, and 15 kDa (Figures and S1). Lower % gels should
resolve higher-molecular-weight proteins more effectively and vice
versa. Peaks in the corresponding intensity profiles corroborate visualized
bands. We define a resolvable peak as a high point in line intensity
profiles between two clearly identifiable troughs and dispersion as
the trough-to-trough distance. Migration distance is the distance
from the end of the well to the high point of a resolvable peak. To
compare quantitatively across different gels, we focus on the 50 kDa
band, which was resolvable in each (Table ). In general, the migration distance and
dispersion decrease as % acrylamide increases. In 6% gel, 160, 90,
and 50 kDa standards are resolvable. In 9.5% gel, the 160, 90, 50,
and 30 kDa standards are resolvable. The 50 kDa band has lower dispersion
in the 9.5% gel as compared to that in the 6% gel. In the 12% gel,
the 50 and 30 kDa standards are resolvable. In the 18% gel, the 50,
30, and 15 kDa standards are resolvable. We conclude that varying
acrylamide proportion in Mesowestern gels over ranges typically used
in regular Western can resolve different molecular weight ranges,
albeit with lower-molecular-weight resolution as compared to regular
Western.
Figure 4
Acrylamide composition effects on molecular weight resolution.
Representative examples from molecular weight ladder runs from four
gels at different compositions of acrylamide (denoted by percent)
are shown. Electrophoresis was carried out at 100 V for ∼2
h. Full scans are shown in Figure S1. Line
intensity scans are overlaid for each image and were obtained from
Image Studio (LI-COR). In general, as expected, higher acrylamide
composition resolves lower molecular weights more robustly, at the
expense of resolving higher molecular weights. When arrows are not
shown, the peaks were not deemed resolvable.
Table 2
Electrophoresis Metrics for the 50
kDa Marker in Gels with Varying Acrylamide Compositiona
gel acrylamide (%)
migration
distance (mm)
dispersion (mm)
6
2.7 ± 0.05
1.02 ± 0.02
9.5
1.97 ± 0.05
0.85 ± 0.03
12
1.58 ± 0.04
0.86 ± 0.03
18
1.06 ± 0.03
0.42 ± 0.02
LI-COR Image Studio Lite was used
to draw rectangular regions of interest around the band area of interest,
and from line profiles, pixel lengths from the well to the peak (migration
distance) or from peak trough to trough (dispersion) were measured
in pixels. Conversion to mm was done using the imaging resolution
of 84 um. Uncertainty is the standard error of the mean from 10 different
lanes (full scans in Supporting Figure 1).
Acrylamide composition effects on molecular weight resolution.
Representative examples from molecular weight ladder runs from four
gels at different compositions of acrylamide (denoted by percent)
are shown. Electrophoresis was carried out at 100 V for ∼2
h. Full scans are shown in Figure S1. Line
intensity scans are overlaid for each image and were obtained from
Image Studio (LI-COR). In general, as expected, higher acrylamide
composition resolves lower molecular weights more robustly, at the
expense of resolving higher molecular weights. When arrows are not
shown, the peaks were not deemed resolvable.LI-COR Image Studio Lite was used
to draw rectangular regions of interest around the band area of interest,
and from line profiles, pixel lengths from the well to the peak (migration
distance) or from peak trough to trough (dispersion) were measured
in pixels. Conversion to mm was done using the imaging resolution
of 84 um. Uncertainty is the standard error of the mean from 10 different
lanes (full scans in Supporting Figure 1).
Reproducibility across a Mesowestern Gel
Having established
the basic Mesowestern workflow using ladder-based standards, we wanted
to evaluate the performance using cell lysates and antibodies. The
first question we had was related to the reproducibility across a
gel. We specifically focused here on a “quarter gel”,
which we found often useful, as it still provides high-throughput
capability but with reduced labor input. We loaded 0.5 μL of
lysate from exponentially growing MCF10A cells into each well of a
quarter gel, along with some regularly spaced molecular weight ladder,
and then blotted for β-actin using LI-COR infrared fluorescence
detection (Figure ). The experiment yielded bands at the expected molecular weight
(∼42 kDa), with a few anomalies not atypical from regular Western.
There was a noticeable variability in the band intensities, which
was found to follow a near-normal curve, with a CV of 33% (Figure B). Certainly, heterogeneities
in electrophoresis and transfer to membrane could play a role, but
we could also not rule out a substantial contribution from small volume
manual pipetting. We conclude that the Mesowestern can be used to
analyze cell lysates analogously to regular Western. Comparing values
from direct quantification of bands in different parts of the membrane
may be imprecise, and require normalization to additional controls.
The high CV indicates that Mesowestern is best suited for qualitative
screening applications at this stage.
Figure 5
Lysates analyzed for reproducibility across
a quarter gel. A. Exponentially
growing MCF10A cells were harvested and lysed as described in the
Methods section. The same lysate sample was loaded into each well
of the pictured Mesowestern blot. The total protein concentration
of the sample was 4.0 mg/mL, and 0.5 μL per well was loaded
into the gel. After electrophoresis (100 V for ∼2 h) and transfer,
the PVDF membrane was incubated with anti-β-actin antibodies
(1:1000) and a secondary antibody for detection (see the Methods section). B. We quantified each band (N = 52) in the blot image from A using Image Studio and
analyzed the distribution by z-score analysis as
pictured. The distribution is approximately normal, with very little
variation outside of two standard deviations and a CV of 33%.
Lysates analyzed for reproducibility across
a quarter gel. A. Exponentially
growing MCF10A cells were harvested and lysed as described in the
Methods section. The same lysate sample was loaded into each well
of the pictured Mesowestern blot. The total protein concentration
of the sample was 4.0 mg/mL, and 0.5 μL per well was loaded
into the gel. After electrophoresis (100 V for ∼2 h) and transfer,
the PVDF membrane was incubated with anti-β-actin antibodies
(1:1000) and a secondary antibody for detection (see the Methods section). B. We quantified each band (N = 52) in the blot image from A using Image Studio and
analyzed the distribution by z-score analysis as
pictured. The distribution is approximately normal, with very little
variation outside of two standard deviations and a CV of 33%.
Dual-Color Imaging Allows the Reduction of Variation by Normalization
to a Loading Control
One feature of LI-COR-based infrared
fluorescence approaches is a natural two-color imaging scheme, which
in this case could provide an internal loading control signal for
each well with which to improve quantitative comparison from sample
to sample. To test this, we again used a quarter gel loaded with cell
lysates from exponentially growing MCF10A cells (Figure ). As a loading control, we
blotted for α-tubulin, and as an example of a target that may
be of interest for quantification, we blotted for doubly phosphorylated
ERK1/2 (p-MAPK), a central signal transduction protein. As before,
bands were clearly visible at the expected molecular weights (Figure A,B). We note here
that the individual ERK1 and ERK2 bands (42 and 44 kDa) are not resolvable
by Mesowestern. We quantified the bands and found a reasonable correlation
between their intensities (Figure C-R2 = 0.69). To evaluate
whether normalizing (i.e., dividing) the p-MAPK signal by the α-tubulin
signal improved the reliability of the p-MAPK signal, we compared
the CV for each set of values (Figure D). Such normalization reduced the CV for the p-MAPK
signal, although it was still high at ∼30% as above. We conclude
that two-color imaging with internal loading controls may improve
comparability across samples in a Mesowestern gel. However, as above,
the high CV indicates that the Mesowestern at this point is best suited
for qualitative screening applications.
Figure 6
Dual-color Mesowestern
blotting for loading control normalization.
A–B. Lysate from exponentially growing MCF10A cells was diluted
to 2 mg/mL and loaded into a 9.5% gel at 0.5 μL/well. After
electrophoresis (100 V, ∼2 h) and transfer, the PVDF membrane
was incubated with anti-α-tubulin and anti-p-MAPK antibodies
and then different secondary antibodies for the detection of each
at different wavelengths. The membrane images depict the same membrane
but at different wavelengths to detect A. α-tubulin and B. p-MAPK
separately. C. Quantified bands (N = 55) were plotted
to examine the expected correlation between the two signals from the
same lysate. D. The variation across the gel of the independent signals
and of the p-MAPK signal normalized by the α-tubulin loading
control. Dividing by the loading control signal improves the %CV.
Dual-color Mesowestern
blotting for loading control normalization.
A–B. Lysate from exponentially growing MCF10A cells was diluted
to 2 mg/mL and loaded into a 9.5% gel at 0.5 μL/well. After
electrophoresis (100 V, ∼2 h) and transfer, the PVDF membrane
was incubated with anti-α-tubulin and anti-p-MAPK antibodies
and then different secondary antibodies for the detection of each
at different wavelengths. The membrane images depict the same membrane
but at different wavelengths to detect A. α-tubulin and B. p-MAPK
separately. C. Quantified bands (N = 55) were plotted
to examine the expected correlation between the two signals from the
same lysate. D. The variation across the gel of the independent signals
and of the p-MAPK signal normalized by the α-tubulin loading
control. Dividing by the loading control signal improves the %CV.
Direct Comparison of Mesowestern and Western
We finally
wanted to investigate sensitivity and linear range for Mesowestern
in direct comparison to Western. While this will invariably be dependent
on the epitope of interest, its abundance in the cell lysate, and
the antibody being used, we started by investigating this for β-actin.
Specifically, we performed a 6-point, 2-fold serial dilution of lysate
from exponentially growing U87 cells and replicated this dilution
curve 6 times on a portion of a Mesowestern gel (Figure A—we used a mold with
slightly larger ∼1 μL wells for this experiment; replicates
in Figure S2). Simultaneously, 20 μL
of the same lysates were loaded into a regular Western, and the same
blocking and antibody solutions were used to compare Western to Mesowestern
side by side (Figure B,C; replicates in Figure S2). Bands were
observable from each technology in each sample. We had expected the
lowest concentration lysate to be below the limit of detection for
Mesowestern, but to our surprise, this was not the case. In both cases,
the linear range of detection was at least ∼33-fold, with R2 ∼ 0.99 (Figure D). The lowest lysate concentration loaded
in Mesowestern is at least equivalent to 29 ng of total protein, which
is approximately 100–300 cells (0.1–0.3 ng of protein
yield/cell).[40−42] Based on the estimates for absolute expression levels
of β-actin in mammalian cells (∼106 copies/cell),[43] this is at least ∼500 attomol sensitivity,
but as mentioned above, the serial dilution curve did not find the
lower limit of detection. For regular Western, a 15 attomol limit
of detection using the same infrared fluorescence modality has been
reported for transferrin (no such data could be found for β-actin).[44] More broadly, the manufacturers report limits
of detection for various modalities ranging from 500 fg to 500 pg
of protein, which is consistent with ∼10 attomol for proteins
of typical molecular weight ranges.[45]
Figure 7
Comparison
of Mesowestern and regular Western. A–C. Cell
lysate from exponentially growing U87 cells was prepared at a range
of protein concentrations (twofold serial dilution) and subjected
to Mesowestern and Western analyses, as indicated. Full scans are
shown in Figure S2. The PVDF membrane was
incubated with anti-β-actin antibodies and a secondary antibody
for detection. D, E. The signal derived from the image analysis of
each band (N = 78 bands for Mesowestern from 2 blots
and 14 dilution curves; N = 18 bands for Western
from 3 blots/dilution curves) was plotted versus the known amount
of total protein mass loaded. For Mesowestern (D), data were normalized
such that 1 mg/mL lysate corresponds to a relative signal intensity
of 1. For Western, data were normalized such that the maximum normalized
signal was 1. Error bars are the standard error of the mean for each
sample (N ≥ 3).
Comparison
of Mesowestern and regular Western. A–C. Cell
lysate from exponentially growing U87 cells was prepared at a range
of protein concentrations (twofold serial dilution) and subjected
to Mesowestern and Western analyses, as indicated. Full scans are
shown in Figure S2. The PVDF membrane was
incubated with anti-β-actin antibodies and a secondary antibody
for detection. D, E. The signal derived from the image analysis of
each band (N = 78 bands for Mesowestern from 2 blots
and 14 dilution curves; N = 18 bands for Western
from 3 blots/dilution curves) was plotted versus the known amount
of total protein mass loaded. For Mesowestern (D), data were normalized
such that 1 mg/mL lysate corresponds to a relative signal intensity
of 1. For Western, data were normalized such that the maximum normalized
signal was 1. Error bars are the standard error of the mean for each
sample (N ≥ 3).
Discussion
Here, we have described the development
and first functional testing
of a high-throughput, small sample size Western blotting protocol
called Mesowestern. As compared to a Western blot, it enables at least
10-fold greater sample throughput with at least 1/10 the amount of
lysate per sample and only requires horizontal as opposed to the traditionally
employed vertical electrophoresis. The main tradeoff is the reduced
molecular weight resolution because samples have a shorter separation
distance and no stacking gel. Evaluation of sample-to-sample variability
indicated that another limitation is the high CV, making this technique
currently best suited for qualitative screening applications. As compared
to the Microwestern, it eliminates the need for piezoelectric pipetting
by using a 3D-printed gel mold that makes micropipette-loadable gels.
We have demonstrated that the molecular weight resolution of a Mesowestern
gel can be adjusted in expected ways by changing the gel acrylamide
composition. We have explored the limits of detection and linear range
of the Mesowestern, which was found to be largely similar to regular
Western within the investigated ranges, although the lower limit of
detection was not found with the given concentrations. Overall, the
Mesowestern is a promising technology that could be readily adopted
by molecular biology labs having interest in more high-throughput
Western blotting with small sample sizes, particularly for qualitative
screening applications.Although we have used a single mold
design here, the layout can
be quite easily modified as users desire for their particular needs.
Although most individual labs may not have the necessary 3D printing
equipment, access to additive manufacturing facilities is relatively
common and the printing itself is fast, so we anticipate that custom
molds will be straightforward to implement. For example, some users
may want fewer wells but to be able to load more sample volume per
well. Others may want more separation space available to each well.
Yet others may wish to make an even larger gel, much larger than a
microwell plate footprint (compatible with the downstream horizontal
electrophoresis). All such variations are straightforward and possible
depending on the needs that arise. Lastly, the commercial availability
of gels could increase access.One difference between the Mesowestern
and regular Western is a
“stacking” portion in a regular Western gel.[38,46] The stacking gel has a low acrylamide composition (∼5%) with
the purpose of allowing all of the proteins from the cell lysate to
easily enter the subsequent “resolving” gel at the same
time, which allows for decreased band dispersion and thus better molecular
weight resolution. In the Mesowestern, there is only a resolving gel.
In our applications so far, we indeed have noted that the resulting
bands have higher dispersion than Western bands, although they were
certainly identifiable at the expected molecular weight ranges. However,
given the fact that Mesowestern inherently has lower-molecular-weight
resolution due to less distance for proteins to migrate, future innovations
incorporating stacking portions would be a welcome development. This
is quite challenging, however, as the unpolymerized gel is loaded
from a single entry port, making it difficult to isolate spatial regions
where wells reside and stacking gels would be appropriate. In that
regard, the ability to cast polyacrylamide gradient[39] Mesowestern gels would also be welcome but similarly challenging.Lastly, although the Mesowestern makes significant advances with
regard to throughput and sample size, we have not demonstrated here
the multiplexing capabilities offered by Microwestern.[35,36] The Microwestern achieves high multiplexing (e.g., 96 antibody pairs
at once) by placing the resultant membrane in a microwell plate-sized
gasketing apparatus, which allows different antibodies on different
parts of the membrane. There is no barrier to applying such an approach
to the Mesowestern, so we expect that similar multiplexing can be
done, albeit of course at the cost of being less high throughput,
since wells must contain repeated patterns of the same lysates to
be then blotted by different antibodies.In conclusion, we have
demonstrated here a new approach to Western
blotting called the Mesowestern that increases throughput greater
than 10-fold with greatly reduced sample size requirements. Given
the tradeoffs of molecular weight resolution and sample-to-sample
CV, qualitative screening applications are likely most suitable. Most
notably, the Mesowestern is straightforward to implement in typical
cell and molecular biology labs having a few dissimilarities from
Western blotting. While the Western blot may be viewed by some as “old”
and “irrelevant”, it does in fact remain as one of the
most widely used assays in biomedical science,[10] and this is unlikely to change due to its popular use as
a sensitive and specific confirmatory assay modality. Thus, improvements
to Western such as the Mesowestern we developed here are still expected
to have a widespread impact.
Methods
Printing the Mold
Molds were printed in the Clemson
Additive Manufacturing Lab with the Stratasys Connex 350 and Veroclear
as the material (Stratasys, OBJ-03271-RGD810). Following printing,
a self-forming valve packing (Danco, #80794) was inlayed into the
outer edge of the well perforation unit (bottom). Schematic files
are available upon collaborative request.
Casting a Gel
Gel casting was completed through a process
of silanization of surfaces coming into contact with the gel, clamping
to ensure a tight leak-proof fit, and serological pipetting of unpolymerized
solution into the mold. Briefly, a 2.5% v/v silane solution was prepared
by combining 1.25 mL of dichlorodimethylsilane (Sigma-Aldrich, #40140)
and 48.75 mL of 100% ethanol (Fisher, #04-355-22) in a 50 mL conical
tube (Fisher, 14-432-22). We then applied 250 μL of the silane
solution to the interior surfaces of both the top and bottom mold
pieces, gently spread it across the surface by rocking, and wicked
excess with a kimwipe. After assembling the top and bottom pieces
together, four C-clamps (Irwin #1901235) were tightened onto the assembly
at the designated corner locations (indented circles). At this point,
the assembly is ready for loading.A 9.5% gel solution was prepared
by combining 47.5 mL of 30% bis/acrylamide solution 29:1 (BIO-RAD,
#161-0156) with 41 mL of MilliQ water, 30 mL of glycerol (Sigma-Aldrich,
#G5516-500 mL), 30 mL of 5× tris-acetate buffer (recipe as follows),
and 1.5 mL of 10% sodium dodecyl sulfate (SDS) (Fisher, #BP 2436)
together. Preparation of the 5× tris-acetate buffer was completed
by dissolving 145.4 g of tris base (BIO-RAD, 161-0719) in 700 mL of
MilliQ water (pH expected between 11.0 and 11.4). The pH was adjusted
by adding 65 mL of glacial acetic acid (Sigma-Aldrich, #320099) and
allowing the solution to sit overnight. Then, 0.5 mL of glacial acetic
acid was pipetted into the solution and allowed to sit for an hour
at room temperature. This was repeated until the solution reached
pH 6.9. Finally, the volume of the solution was brought up to 1 L
with MilliQ water and stored at 4 °C.Polymerizing gel
solution was made by combining 15 mL of 9.5% gel
solution with 133 μL of 10% ammonium persulfate solution (APS)
and 13.3 μL of TEMED (BIO-RAD #161-0700) into a beaker under
a fume hood. The 10% APS solution was prepared by dissolving 0.2 g
of ammonium persulfate (BIO-RAD #161-0800) into 2 mL of MilliQ water.
Quickly after preparation, 15 mL of gel solution was dispensed by
serological pipetting into the mold assembly via the loading port
(Figure ). The assembly
was kept still under the fume hood for 30 min at room temperature
to achieve full polymerization.To remove the gel from the mold,
first the C-clamps were removed.
Then, the top and bottom mold pieces were carefully separated using
a gel releaser (BIO-RAD, #165330) on the lateral protrusions, followed
by carefully moving the releaser around the internal face of the top.
After splitting the top and bottom pieces, the gel was removed by
inverting the mold so that the gel is facing thick blotter paper (BIO-RAD
#1703958) that is presoaked in running buffer (see below). The blotter
paper was approximately 5 cm larger than the gel on the top and bottom
and about 1 cm larger than the gel on each side. The gel is slowly
peeled away from a corner using the gel releaser until gravity facilitates
the remaining gel to gently fall onto the soaked blotter paper support.
The gel can be used immediately or be stored in a sealed bag at 4
°C for several months (at least).
Cell Culture
MCF10A cells (from LINCS Consortium and
STR verified internally) were cultured in DMEM/F12 (Gibco #11330032)
medium containing 5% (v/v) of horse serum (Gibco #16050122), 20 ng/mL
of EGF (PeproTech #AF-100-15), 0.5 mg/mL of hydrocortisone (Sigma-Aldrich
#H-0888), 10 μg/mL of insulin (Sigma-Aldrich #I-1882), 100 ng/mL
of cholera toxin (Sigma-Aldrich #C-8052), and 2 mM of l-glutamine
(Corning #25-005-CI). U87 cells (from ATCC and STR verified internally)
were cultured in DMEM (Gibco#10313021) medium containing 10% (v/v)
fetal bovine serum (Corning#35-011-CV) and L-glutamine (2 mM) (Corning
#25-005-CI). The cells are kept at 37 °C in 5% CO2 in a humidified incubator. To maintain subconfluency, the cells
are passaged every 2–3 days, washing once with phosphate-buffered
saline (PBS), lifting with 0.25% trypsin (Corning #25-053-CI), and
reseeding in full growth media.
Lysate and Sample Preparation
Cells growing in full
growth media were collected, counted, and seeded (150,000 cells/well)
in tissue culture-treated 6-well plates (Corning # 08-772-1B). The
cells were kept at 37 °C in 5% CO2 in a humidified
incubator for ∼48 h. The plates were removed from the incubator,
and media in the wells was aspirated. The wells were washed with ice-cold
PBS once and placed on ice. Freshly prepared, ice-cold RIPA buffer
(110 μL, 50 mM tris, pH 7–8 (Acros Organics #14050-0010),
150 mM NaCl (Fluka #71383), 0.1% SDS (v/v from 10% stock, Fisher #46040CI),
0.5% sodium deoxycholate (g/mL Alfa Aesar, J62288), 1% Triton-X-100
(v/v, Fisher, BP151) with protease and phosphatase inhibitors (1 μg/mL
aprotinin (MP Biomedicals #0219115801), 1 μg/mL leupeptin (MP
Biochemicals #0215155301), 1 μg/mL pepstatin A (MP Biochemicals
#0219536801), 10 mM β-glycerophosphate (Santa Cruz Biotechnology
#sc203323), and 1 mM sodium orthovanadate (Sigma-Aldrich # S6508))
were added into each well. The plates were kept on a rocker (slow)
in the cold room for 15 min. Then, the lysates were scraped off with
a cell scraper (Stellar Scientific TC-CS-25), and 100 μL lysate
from each well was transferred into labeled Eppendorf tubes on ice.
Each tube was vortexed three times for ∼30 s to homogenize
cell debris. Next, the tubes were centrifuged at 4 °C for 15
min at ∼21,000g (max speed). Finally, 80 μL
of the supernatant from each tube was transferred into a new Eppendorf
tube, being careful not to disturb the debris pellet. Lysates were
stored at −20 °C for short-term storage and transferred
to −80 °C for long-term storage.
Protein Quantification
Total protein quantification
of lysates was done using either the BCA-Pierce 660 Assay (Thermo
Scientific #23225) or Pierce Rapid Gold BCA (Thermo Scientific A53225),
and BSA stock (Thermo Scientific #23209) was used as a reference according
to the manufacturer’s protocol. In short, 10 μL of the
lysate sample or BSA standards were loaded into 96-well plates (Corning
#3370), in triplicate. Then, the BCA Protein Assay Reagent was loaded
into each nonempty well. The plate was covered with the lid and incubated
at room temperature for 5 min. The absorbance readings at 660 or 480
nm were obtained in a plate reader (BioTek #Epoch2). The average of
blank wells was subtracted from each reading to calculate blank-corrected
averages for each condition. The standard curve is fitted to a line
using blank-corrected mean values of each standard condition versus
its BSA concentration. The protein concentration in each sample was
calculated using the standard curve formula.
Sample Preparation
Lysate stocks are thawed on ice
(if applicable). Then, a 5× sample buffer was prepared (5 mL
of glycerol (Sigma-Aldrich #G5516), 0.5 mL of 10% SDS (Fisher #BP
2436), 0.01 g of bromophenol blue (Calbiochem #2830), 2.1 mL of 5×
tris-acetate buffer (as above), 0.5 mL of β-mercaptoethanol
(Sigma-Aldrich #M6250), then the total volume brought to 10 mL with
MilliQ water). This was mixed with lysates in a 1:4 (v/v) ratio. Next,
the tubes were heated at 95 °C for 5 min in a dry heating block
and then briefly spun in benchtop microcentrifuge before loading (below).
Loading the Gel
Following the release of the gel onto
the soaked blotter paper, the assembly was placed down on a flat surface
with the wells facing up. If folds and stretching of the gel are evident,
light rolling was used to flatten. A p2 micropipette with 10 μL
tips was used to load the prepared lysates and/or molecular weight
ladder (LI-COR, 928-60000) into wells as desired. We have found that
wells less than 2 mm away from the gel boundaries may be subject to
inconsistent electrophoresis and transfer and therefore avoid them
when possible. Care was taken not to adjust the gel on the blotter
paper after any loading and also to transport the gel with a spatula
support underneath.
Horizontal Electrophoresis
Horizontal electrophoresis
was carried out using the Flatbed Professional (Gel Company Store,
FC-EDCProf-2836). The apparatus was maintained at 10 °C during
electrophoresis. First, ∼10 mL of cooled running buffer was
poured onto the center of the apparatus, followed by transfer of the
blotter paper/loaded gel by spatula onto this buffer. Running buffer
was made by combining 20 mL of 5× tris-acetate buffer (see above)
with 29.5 mL of MilliQ water and 0.5 mL of 10% SDS. The gel should
be oriented to have the red bar at the bottom, where the proteins
will migrate toward. Additionally, the wells should be aligned with
the apparatus gridlines, and excess running buffer should be wiped
up with no buffer accumulated outside of the blotter paper. Then,
the anode and cathode wires were placed over the blotter paper, about
3 cm from the gel. Finally, the glass plate was placed on top of the
anode and cathode and the lid was closed. Electrophoresis was conducted
at 100 V for ∼2 h, although each run should be individually
monitored. Samples should be visible as blue dots in the gel after
∼30 min, and ideally, the run should be stopped when it reaches
the top edge of the next well. After 30 min, we paused the run, lifted
the blotter paper and gel with a spatula, and rehydrated by placing
another 10 mL of cool running buffer as previously.
Transfer to Membrane
Transfer buffer was prepared by
first making 10× tris-glycine buffer (600 mL of MilliQ water
with 30.3 g of tris base (BIO-RAD #161-0719) and 144 g of glycine
(VWR #0167), then MilliQ water was added to a final volume of 800
mL). Transfer buffer (∼2 L, 1×) was made by taking 160
mL of 10× tris-glycine buffer, adding MilliQ water up to a final
volume of 1600 mL, and finally, adding ∼400 mL of methanol
(Fisher #A412-in a fume hood) to 2 L. Transfer buffer is stored at
4 °C.For quarter gels, we used a Mini Trans-Blot Cell
(BIO-RAD, 1703930), and for full gels, we used a Criterion Blotter
(BIO-RAD, 1704070). We have successfully used both nitrocellulose
(GE, 10600002) and PVDF (BIO-RAD, 1620264) membranes for Mesowestern,
and in this work, the data come from PVDF. In our experience, low
fluorescence PVDF membranes tend to provide a better signal to noise
due to their increased ability to bind low abundance proteins, although
answering such questions definitively was not the purpose of this
manuscript. For PVDF, the membrane was prewet with methanol prior
to subsequent use and never allowed to dry out.To prepare the
gel and membrane for transfer, cold transfer buffer
was poured into a pyrex dish to a depth of ∼3 cm. Blotter paper,
cut to the size of the transfer cassette but larger than the gel,
was placed into the pyrex dish to soak. After soaking, the blotter
paper was placed on a clean, flat benchtop. Then, the gel was allowed
to soak in the same transfer buffer for ∼15 min, making sure
to keep track of which side of the gel has the well indentations.
The gel was then placed onto the soaked blotter paper, with the wells
facing down on the paper. A spatula was always used to transport the
gel. The gel was then gently rolled flat, and air pockets were removed
using a roller (BIO-RAD, 1651279). The membrane was cut to the same
size as the gel, being careful never to touch the membrane except
with clean tools. After wetting with methanol (if PVDF is used), the
membrane was then placed to soak in transfer buffer. Forceps were
used to gently place the membrane onto the gel. If the membrane is
not aligned, we did not move it, rather, we got a new membrane. Then,
the membrane was rolled as previously. A second piece of transfer
buffer-soaked blotter paper was then placed on top of the membrane
in line with the first piece of blotter paper and rolled as previously.
Finally, a spatula was used to lift the “sandwich” onto
a fiber pad (BIO-RAD, 1703933), and another fiber pad was placed on
top. This fiber pad-surrounded sandwich was moved to the transfer
cassette, making sure that the side of the sandwich closest to the
membrane was on the clear/positive side of the cassette (BIO-RAD,
1703931). This also means that the side of the sandwich closest to
the gel is on the black/negative side of the cassette. The cassette
was then placed into the transfer apparatus (negative to negative/black
to black, positive to positive/clear to red). If desired, a second
sandwich was made and placed into the apparatus.With the cassettes
in the transfer apparatus, a cold transfer buffer
was added until it reached the indicated volume line. The apparatus
was moved to a 4 °C room, and then transfer was carried out at
30 V for 16 h (usually overnight). After the transfer, the membrane
was removed with clean forceps and was placed in a clean incubation
box (Li-Cor, 929-97201), with the side of the membrane that was in
contact with the gel facing up.
Antibody Incubation
First, TBS and TBST buffers were
prepared. Briefly, 10× TBS was made by dissolving a 24 g tris
base (BIO-RAD #161-0719) and 88 g NaCl (CAS 7647-14-5) in 1 L of MilliQ
water. The pH was monitored with continuous magnetic stirring while
adding HCl dropwise to bring the pH to 7.6. To make 1× TBS, 50
mL of 10× TBS was added to 450 mL of MilliQ water and stored
at 4 °C (stable for several months). To make 1× TBST, 2.5
mL of 10% Tween 20 (BIO-RAD #161-0781) was added to 500 mL of 1×
TBS and similarly stored at 4 °C.All membrane incubations
were done in the dark (sealed black box or covered in aluminum foil).
The membrane was incubated first in ∼20 mL of blocking buffer
(1 g BSA (Fisher, BP1600) in 20 mL of 1× TBS) for at least 30
min at room temperature with gentle rocking. After blocking, the blocking
buffer was removed, and the membrane was directly incubated with a
primary antibody solution (10 mL blocking buffer, 50 μL of 10%
Tween 20, v/v dilution of primary antibody to the desired working
concentration) for at least 2 h at room temperature or overnight at
4 °C, all with gentle rocking. After primary antibody incubation,
the membrane was washed with ∼10 mL of 1× TBST three times,
5 min for nitrocellulose, and four times, 15 min for PVDF. After washing,
the secondary antibody solution (10 mL 1× TBST with 1:20,000
v/v; see below) was added to the membrane and incubated with gentle
rocking for 1 h at room temperature. After incubation, the secondary
antibody solution was discarded, and the membrane was washed as previously
with 1× TBST. After the last TBST wash, a final TBS wash was
done. The membrane was then scanned with the side that was facing
up (closest to gel during transfer) now facing down on the clean surface
of a LI-COR Odyssey infrared fluorescence scanning instrument (LI-COR
model number 9140).Antibodies were obtained from and used with
working concentrations
as follows: p-MAPK (Cell Signaling, #4370S, 1:1000), α-tubulin
(Novus, #NB100-690, 1:1000), β-actin (Figure : LI-COR #926-42212, 1:1000; Figure : Cell Signaling, #3700, 1:1000),
antirabbit (800CW LI-COR #926-32211, 1:20,000), and antimouse (680LT
LI-COR #925-68070, 1:20,000).
Imaging and Quantification
Placement of the membrane
on the scanning surface was set in Image Studio. Both 700 nm and 800
nm wavelength channels were set to be scanned. Resolution was set
to generally 42 μm (some exceptions for speed at times), and
the focus offset was set to 0.0 mm. After the membrane finished scanning,
the image and the associated zip file were exported from the Li-Cor
Odyssey scanner and imported into Image Studio Lite for analysis.
In Image Studio, boxes were drawn around protein bands and the “signal”
metric generated by the software was used as the quantification.[7,37]
Authors: David C Handler; Dana Pascovici; Mehdi Mirzaei; Vivek Gupta; Ghasem Hosseini Salekdeh; Paul A Haynes Journal: Proteomics Date: 2018-11-25 Impact factor: 3.984
Authors: Mathias Uhlen; Anita Bandrowski; Steven Carr; Aled Edwards; Jan Ellenberg; Emma Lundberg; David L Rimm; Henry Rodriguez; Tara Hiltke; Michael Snyder; Tadashi Yamamoto Journal: Nat Methods Date: 2016-09-05 Impact factor: 28.547
Authors: Andrea Degasperi; Marc R Birtwistle; Natalia Volinsky; Jens Rauch; Walter Kolch; Boris N Kholodenko Journal: PLoS One Date: 2014-01-27 Impact factor: 3.240