Kyoung Min Yoo1, Ray T Chen1,2. 1. Department of Electrical and Computer Engineering, The University of Texas at Austin, 10100 Burnet Road, Austin, Texas 78758, United States. 2. Omega Optics Inc., 8500 Shoal Creek Boulevard, Building 4, Suite 200, Austin, Texas 78757, United States.
Abstract
On-chip broadband optical spectrometers that cover the entire tissue transparency window (λ = 650-1050 nm) with high resolution are highly demanded for miniaturized biosensing and bioimaging applications. The standard spatial heterodyne Fourier transform spectrometer (SHFTS) requires a large number of Mach-Zehnder interferometer (MZI) arrays to obtain a broad spectral bandwidth while maintaining high resolution. Here, we propose a novel type of SHFTS integrated with a subwavelength grating coupler (SWGC) for the dual-polarization bandpass sampling on the Si3N4 platform to solve the intrinsic trade-off limitation between the bandwidth and resolution of the SHFTS without having an outrageous number of MZI arrays or adding additional active photonic components. By applying the bandpass sampling theorem, the continuous broadband input spectrum is divided into multiple narrow-band channels through tuning the phase-matching condition of the SWGC with different polarization and coupling angles. Thereby, it is able to reconstruct each band separately far beyond the Nyquist criterion without aliasing error or degrading the resolution. We experimentally demonstrated the broadband spectrum retrieval results with the overall bandwidth coverage of 400 nm, bridging the wavelengths from 650 to 1050 nm, with a resolution of 2-5 nm. The bandpass sampling SHFTS is designed to have 32 linearly unbalanced MZIs with the maximum optical path length difference of 93 μm within an overall footprint size of 4.7 mm × 0.65 mm, and the coupling angles of SWGC are varied from 0° to 32° to cover the entire tissue transparency window.
On-chip broadband optical spectrometers that cover the entire tissue transparency window (λ = 650-1050 nm) with high resolution are highly demanded for miniaturized biosensing and bioimaging applications. The standard spatial heterodyne Fourier transform spectrometer (SHFTS) requires a large number of Mach-Zehnder interferometer (MZI) arrays to obtain a broad spectral bandwidth while maintaining high resolution. Here, we propose a novel type of SHFTS integrated with a subwavelength grating coupler (SWGC) for the dual-polarization bandpass sampling on the Si3N4 platform to solve the intrinsic trade-off limitation between the bandwidth and resolution of the SHFTS without having an outrageous number of MZI arrays or adding additional active photonic components. By applying the bandpass sampling theorem, the continuous broadband input spectrum is divided into multiple narrow-band channels through tuning the phase-matching condition of the SWGC with different polarization and coupling angles. Thereby, it is able to reconstruct each band separately far beyond the Nyquist criterion without aliasing error or degrading the resolution. We experimentally demonstrated the broadband spectrum retrieval results with the overall bandwidth coverage of 400 nm, bridging the wavelengths from 650 to 1050 nm, with a resolution of 2-5 nm. The bandpass sampling SHFTS is designed to have 32 linearly unbalanced MZIs with the maximum optical path length difference of 93 μm within an overall footprint size of 4.7 mm × 0.65 mm, and the coupling angles of SWGC are varied from 0° to 32° to cover the entire tissue transparency window.
Owing to unique “fingerprint”
signatures of molecular
absorption and Raman spectra of the individual molecules, the optical
spectrometer has been known as an indispensable tool to analyze the
optical spectrum for a wide range of applications including biological
and chemical analysis, medical diagnosis, and environmental and planetary
monitoring, to name a few. Particularly, the near-infrared wavelength
is beneficial for adoption in mammalian tissues because of its low
absorbance, autofluorescence, and lower light-scattering loss from
various mammalian biocells including hemoglobin, melanin, fat, lipids,
water, and others, when compared to the shorter wavelengths.[1−3] Also, a noninvasive in vivo diffuse optical characterization of
human tissues using optical spectroscopy to assess mean absorption
and reduced scattering spectra in an NIR tissue transparency window
(∼650–1050 nm) opens a new possibility of monitoring
various vital signatures of bioanalytes.[2,4,5] Conventional spectrometers consist of movable mirrors,
and free-space optic components are typically bulky and expensive.
Moreover, they require sensitive and precise beam alignments, which
place constraints on their applications in particular platforms accompanying
environmental fluctuations, such as airborne or hand-held devices.
However, the development of on-chip spectrometers based on photonic
integrated circuits (PICs) has shown great promise in the sense that
they can offer low-cost, portable, and robust spectroscopy, along
with low power consumption and high reliability.[6] In recent decades, a myriad of on-chip spectrometer devices
have been demonstrated based on different operating schemes, such
as dispersive optics using arrayed waveguide gratings (AWGs),[7−10] echelle grating,[11] metasurface elements,[12] arrayed narrow-band filters,[13] computational spectral reconstruction-based systems,[14] and Fourier transform spectroscopy (FTS). Dispersive
spectrometers, often called grating spectrometers or scanning spectroscopy,
splits the wavelengths of input light into separate spectral ranges
and collects each wavelength individually. A lot of grating-based
devices have been studied and reported with sub-nanometer resolution
in the visible (VIS) to near-infrared (NIR) range,[7−11,37,38] but the gratings or slits on a dispersive device limit the amount
of energy reaching the detector and the scan speed of spectroscopy
because the individual wavelengths across the bandwidth have to be
measured separately. Meanwhile, FTS is a technique that measures the
spectrum with the interference of light instead of dispersion, so
it does not separate energy into individual wavelengths for measuring
the spectrum, offering advantages including high optical throughput
and a multiplexing advantage, and, in turn, a larger signal-to-noise
ratio (SNR) and faster data collection speed compared to the grating-based
dispersive counterparts. Several on-chip FTS operation schemes were
proposed, including stationary-wave integrated Fourier transform (SWIFT)[15,16] spectrometers, microring resonator (MRR)-assisted Fourier transform
(RAFT)[17] spectrometers, and spatial heterodyne
FTSs (SHFTSs).[18−21] In the SWIFT scheme, the interference patterns created from the
interference of two adjacent propagating waves through parallel waveguides
are diffracted out-of-plane and monitored by the external detectors
such as an array of photodetectors (PDs).[15] To achieve the full interferogram completely, the pitch size of
the PD arrays should be smaller than that of the interference pattern
to avoid subsampling errors.[15] Consequently,
the resolution and bandwidth of the SWIFT spectrometer are highly
limited by the minimum pitch size of the PD array, which puts a significant
constraint on practical applications. To overcome this Nyquist–Shannon
criterion, recent research reported a broadband SWIFT spectrometer
integrated with a lithium niobate (LN) electro-optics modulator to
retrieve the fully sampled interferogram without any moving components.[22]The SHFTS consists of an array of unbalanced
Mach–Zehnder
interferometers (MZIs) with linearly increasing optical path delays
(OPDs).[18,19] In this concept, the output powers of each
MZI configure each point of the spatial interferogram, which can be
captured independently by a linear on-chip PD array, allowing the
acquisition of the entire interferogram in a single capture without
any moving parts or external analyzers. However, the spectral resolution
and bandwidth of SHFTS are closely related to the number of MZIs and
maximum OPDs, which always requires a balance to meet the limited
chip-scale and detecting conditions. In other words, in a standard
SHFTS configuration, there is a trade-off between the resolution and
bandwidth; hence, achieving a fine resolution with broadband operation
requires an unrealistically large number of MZI arrays, which increases
the size of the device footprint significantly. For example, we previously
demonstrated a standard SHFTS device with 24 MZI array for a resolution
of 5 nm with a bandwidth of 60 nm centered at λo =
900 nm in the silicon nitride (Si3N4) platform.[23] However, to cover the VIS to NIR tissue transparency
window, it requires broadband operation with a ∼400 nm bandwidth.
To increase the bandwidth while maintaining a 5 nm resolution, it
requires a more than 160 MZI array, which is impractical to be implemented
as an on-chip device in that the amount of energy reaching each MZI
will be significantly limited after dividing it into 160 arrays as
well as a large device footprint. To overcome this restriction, several
approaches were proposed such as thermally tuned MZI structures,[24,25] RAFT,[17] and applying the compressive-sensing
scheme.[26] Especially, RAFT consists of
a thermally tunable MZI cascaded with a thermally tunable MRR and
demonstrated spectrum retrieval results from one output channel with
a resolution of 0.47 nm, bypassing the Rayleigh criterion of the standard
MZI-based FTS device by implementing a high-resolution MRR filter.[17] In spite of the advantages of the thermo-optic
effect contributing to the small-footprint and high-resolution performance,
it takes a longer time to collect each point of the thermally induced
interferogram separately with corresponding incremental changes of
temperature (ΔTstep), controlled
by the thermoelectric temperature controller (TEC) to achieve a uniform
sampling of the thermo-optical-induced OPD, which also makes the overall
system configuration complex compared to the standard SHFTS.In this paper, instead of adding additional active photonic components,
we propose and experimentally demonstrate an alternative approach
using a bandpass sampling theorem with standard SHFTS configuration
integrated with a subwavelength grating coupler (SWGC) to achieve
the overall bandwidth covering the entire NIR tissue transparency
window (650 to 1050 nm) with <5 nm resolution. To build the on-chip
SHFTS device for NIR wavelength, we chose the Si3N4 platform due to its low material absorption loss over the
wide spectral range from 400 nm up to 2.3 μm as well as the
low phase noise and error due to the lower refractive index contrast
between the silicon dioxide (SiO2) cladding (nSiO ≈ 1.46 @ λ = 900 nm) and
the Si3N4 core (nSi ≈ 2.01 @ λ = 900 nm)[27] compared to the silicon waveguide.[28] Also, the refractive index of Si3N4 varies depending on the deposition techniques and the
quality of films. Typically, low-pressure chemical vapor deposition
(LPCVD)-deposited Si3N4 has a higher refractive
index than the plasma-enhanced CVD (PECVD)-deposited Si3N4, so we decided to use LPCVD-deposited Si3N4 and applied corresponding refractive index values for
the device design and simulation.[28] Finally,
we have developed and fabricated the first prototype SHFTS chip on
a CMOS-compatible Si3N4 platform and demonstrated
experimental results.
Results
Concept and Principle
The theory and principle of the
standard SHFTS have been demonstrated by Florjańczyk et al.[18] previously. As briefly mentioned in the Introduction, SHFTS consists of an array of unbalanced
MZIs with linearly increasing OPDs with a constant increment across
the array configuring the spatial interferogram. For a given single-input
source, the phase change from each MZI is converted into an intensity
change based on interferometric schemes. The input spectrum can be
retrieved through the discrete Fourier transform (DFT), which can
be written as[18]Here, Pin is the
input power, N is the number of MZIs, σ is
the wavenumber, and σ̅ = σ – σmin is the shifted wavenumber, where σmin represents the minimum wavenumber at the Littrow condition;
at the Littrow condition, the phase delays in different MZIs are integer
multiples of 2π, so the output powers of each MZI (Pout) are constant. Δx is the maximum
interferometric delay, that is Δx = neffΔLmax,
where neff is the effective index of the
strip waveguide and ΔLmax is the
maximum path delay of the most unbalanced MZI. The spatial interferogram F(x) is discretized
at N equally spaced OPD values x (0 ≤ x ≤ Δx), which
is defined as x = neffΔL, where ΔL is the path length difference of the ith unbalanced
MZI. The input power Pin is constant for
all the MZIs, and Pout represents the output power
of the ith MZI with the coupling and loss coefficients
of the MZI components As and Bs. As the wavenumber of monochromatic input σ changes
from the Littrow wavenumber, the Pout distribution
becomes periodic, and different wavenumbers create different periodic
patterns. Subsequently, a polychromatic input signal, which can be
considered as a superposition of monochromatic constituents, creates
a corresponding spatial interferogram pattern formed by a superposition
of the respective periodic Pout fringes from
monochromatic input. The resolution of spectrometers, represented
by the wavenumber resolution δσ, is determined by the
maximum interferometric delay Δx. To resolve
two monochromatic signals separated by δσ at the most
unbalanced MZI with an OPD of δσ, the phase change from
respective interferograms of σ and σ + δσ
should differ by one fringe (2π); that is Δφ = 2π(σ
+ δσ)Δx – 2πσΔx = 2π; in turn,[18]Thus, the maximum path delay of the MZI array
(ΔLmax) can be designated as follows:where λo is the center wavelength
and δλ is the wavelength resolution. The number of MZIs
in the array (N) is equivalent to the discrete sampling
points in the spatial interferogram. Based on the Nyquist–Shannon
sampling theorem, the minimum sampling points (Nmin) required to fully reconstruct the input spectrum (pin(σ̅)) within the band-limit range
is determined as follows:where Δσ and Δλ are
the wavenumber and wavelength bandwidth of the spectrometer. Based
on the sampling theory, the wavenumbers that are equally distributed
above and below the Littrow wavenumbers produce the same interference
fringe patterns, and the input spectrum reconstructed by the DFT of
the spatial interferogram creates the wavenumber-shifted replicas
of the original transform pin(σ̅)
above and below the band limit as shown in Figure S1a. Consequently, when the input spectrum contains the signals
outside of the band limit exceeding the bandwidth, the reconstruction
aliasing error due to the overlap of these copies (flipped images)
makes the retrieved spectrum indistinguishable, as described in Figure S1b, and this intrinsic constraint puts
a limitation on retrieving the broadband continuum input signals from
the Fourier transform system.In signal processing, a technique
that is known as bandpass sampling or undersampling has been used
to reconstruct the signal with a sampling rate below the minimum Nyquist
rate (eq ) using a bandpass
filter.[29,30] In this paper, we exploit this concept in
the SHFTS configuration to solve the trade-off condition between the
bandwidth and resolution by employing the bandpass filter to divide
the broadband-continuous spectrum into multiple narrow-band channels.
Thereby, it is able to reconstruct each band without aliasing error,
rather than retrieving the signals all together in a single band.
As a consequence, we were able to achieve the broad overall bandwidth
coverage (Δλoverall) without degrading the
resolution.The grating coupler (GC) is one of the essential
components to
build PIC chips for the fiber-to-chip coupling.[31,32] In general, the narrow-bandwidth operation, polarization (transverse-electric
(TE) and transverse-magnetic (TM)), and angular dependencies of GC
are considered as intrinsic disadvantages of employing the broadband
coupler. However, these properties are very beneficial for the bandpass-sampling
scheme, since it can be implemented as a coupler as well as a polarization-selective
bandpass filter; in addition, the coupling wavelength can be shifted
by the coupling-angle (θ) tuning based on the phase-matching
condition,[33] which makes it a tunable-bandpass
filter.Figure demonstrates
the overall design and operation principle of a Si3N4 bandpass sampling SHFTS integrated with an SWGC that covers
the entire NIR tissue transparency window (650–1050 nm). The
SHFTS consists of an array of MZIs with a single SWGC input port (Figure b). The broadband
input signal is coupled from the fiber to the waveguide through the
SWGC (Figure a), and
the input spectrum is bandpass-filtered based on the SWGC’s
phase-matching condition, which can be tuned by the coupling angle
(θ) with different polarizations. Namely, the broadband input
spectrum is bandpass sampled into multiple narrow bands by different
phase-matching conditions, which completely divide the coupling wavelengths
into M number of discrete narrow-band channels of
the SHFTS. The coupled light is equally divided into linearly unbalanced
MZIs through the cascaded splitters, and the intensities of each MZI
output powers are measured, forming a spatial interferogram for each
sampled band (Figure c). Then, each sampled input spectrum can be fully recovered using eq without overlapped aliases
(antialiasing) (Figure d).
Figure 1
Schematic illustration of the operation principle of a bandpass
sampling SHFTS integrated with an SWGC. (a) Schematic illustration
of the SWGC operation; the SWGC serves not only as a coupler but also
as a tunable-bandpass filter to divide the broadband spectrum into
multiple narrow-band channels by changing the coupling angle with
different polarization. (b) Si3N4 SHFTS chip
with an array of N-MZIs with linearly increasing OPD; the broadband
spectrum is divided into M-spectra based on the Nyquist rate and can
be reconstructed separately. (c) The spatial interferograms of each
discrete narrow band consist of output powers from the MZI array;
phase change from each MZI is converted into the intensity change.
(d) The sampled narrow-band spectra can be reconstructed by Fourier
transform.
Schematic illustration of the operation principle of a bandpass
sampling SHFTS integrated with an SWGC. (a) Schematic illustration
of the SWGC operation; the SWGC serves not only as a coupler but also
as a tunable-bandpass filter to divide the broadband spectrum into
multiple narrow-band channels by changing the coupling angle with
different polarization. (b) Si3N4 SHFTS chip
with an array of N-MZIs with linearly increasing OPD; the broadband
spectrum is divided into M-spectra based on the Nyquist rate and can
be reconstructed separately. (c) The spatial interferograms of each
discrete narrow band consist of output powers from the MZI array;
phase change from each MZI is converted into the intensity change.
(d) The sampled narrow-band spectra can be reconstructed by Fourier
transform.To build the on-chip SHFTS with low-loss operation
in the NIR tissue
transparency window (∼650–1050 nm), we designed and
optimized the Si3N4 passive components including
the strip waveguide, multimode interferometer (MMI) splitter and combiner,
and the SWGC using the Lumerical simulation tool. In the design of
the prototype, we used the SWGCs for the MZI output ports to measure
the output powers, which can be further integrated with a PD array
by virtue of the modern PIC technology,[34] providing much more compact and stable devices. In the following
sections, we will demonstrate the simulation and experimental results
of the optimized passive components and the first prototype of the
SHFTS device. The optimal design and simulation results of the Si3N4 strip waveguide (Figure S2) and MMI splitter/combiner (Figure S5) are described in the Supporting Information.
Subwavelength Grating Coupler
A conventional GC with
shallow etched gratings requires additional alignment lithography
steps, which make the whole fabrication process complex. To alleviate
the fabrication complexity, several through-etched GC designs using
subwavelength grating (SWG) structures to reduce the Fresnel back
reflection have been reported in various platforms.[35,36] In this work, by implementing the SWG structures between the major
Si3N4 gratings, we were able to get a through-etched
structure that can be patterned and etched altogether with other passive
components without additional alignment patterning steps.Also,
it is known that the central coupling wavelength λo of the grating coupler is determined as in the following equation
based on the phase-matching condition.[33]where neff is
the effective index of the grating, nc is the refractive index of the cladding, θ is the coupling
angle, and Λ is the grating period. Therefore, the coupling
wavelength λo can be shifted by θ tuning based
on eq . Here, we designed
the Si3N4 SWGC integrated with an SHFTS chip,
serving not only as a coupler but also as a tunable bandpass filter
for the bandpass sampling of the broadband input signals. The schematic
illustration of the SWGC is shown in Figure a and b.
Figure 2
SWGC design and simulation results. (a)
3D schematic illustration
of the Si3N4 SWGC and (b) zoomed-in image of
the grating structures. (c) Coupling efficiencies (S-parameters) of the optimized SWGC with θ = 11° for the
fundamental TE and TM modes as a function of the wavelength. (d) XYZ
plot showing the simulation result of the SWGC TE mode coupling wavelength
(λo) shift by tuning the coupling angle from 0°
to 31°; x-axis: wavelength, y-axis: coupling efficiency, z-axis: coupling angle.
(e) Optical microscope and (f) SEM images of the fabricated SWGC.
SWGC design and simulation results. (a)
3D schematic illustration
of the Si3N4 SWGC and (b) zoomed-in image of
the grating structures. (c) Coupling efficiencies (S-parameters) of the optimized SWGC with θ = 11° for the
fundamental TE and TM modes as a function of the wavelength. (d) XYZ
plot showing the simulation result of the SWGC TE mode coupling wavelength
(λo) shift by tuning the coupling angle from 0°
to 31°; x-axis: wavelength, y-axis: coupling efficiency, z-axis: coupling angle.
(e) Optical microscope and (f) SEM images of the fabricated SWGC.A simulation model (Figure S3a) has
been created using 2D and 3D finite-difference time-domain (FDTD)
simulation to optimize the device for the fundamental TE mode with
small insertion loss as well as small back reflection to the waveguide.
Design parameters including the grating period (Λ), the width
of the grating (wg), and the width of
the subwavelength grating (wswg) were
optimized with the built-in particle swarm algorithm in the simulation
tool. Furthermore, the thickness of the SiO2 bottom cladding
was optimized as hcore = 2.8 μm
to get the maximum coupling efficiency as shown in Figure S3c. The refractive index profile and E-field of the optimized SWGC structure are shown in Figure S3b and d, showing that the light is coupled from the
fiber to the waveguide through the SWG structures. Figure c shows the coupling efficiency
of the optimized SWGC with θ = 11° for both TE and TM modes
as a function of the wavelength, extracted from the S-parameters of the input and output ports. The coupling efficiency
unit is normalized as a maximum of 1, indicating that 100% of light
is coupled from the fiber to the waveguide, and the reciprocal S-parameters indicated that the coupling efficiency from
the fiber to the waveguide is identical with the opposite direction.
Moreover, we can see the coupling selectivity between the TE mode
and TM mode, and the SWGC has a ∼3 dB coupling loss with a
TE mode at λo = 900 nm, along with the 3 dB bandwidth
of ∼50 nm. Finally, we swept the coupling angle from θ
= 0° to 31° and monitored the λo shifts
from 1030 to 650 nm, and the results in the XYZ plot and XY plot are
shown in Figure d
and Figure S4. The maximum coupling efficiency
of ∼65% was achieved at λo = 650 nm with θ
= 29°, and the center wavelength shifts to a longer wavelength
up to 1030 nm along with the 3 dB bandwidth increases from ∼30
nm to 70 nm as the coupling angle decreases to surface normal. On
the basis of the optimal design and simulation results, we fabricated
the device and inspected the dimension deviation by scanning electron
microscope (SEM) due to the fabrication error as in Table S2, and the corresponding optical microscope and SEM
images can be found in Figure e,f and Figure S8.Since
these deviations of the device dimensions are critical to
the phase-matching condition of SWGC and in turn the coupling efficiencies
and wavelengths, we carefully examined the simulation results with
the real-fabricated dimensions that are shown in Figure S3e. Comparing with the ideal simulation results, we
observed a slight red-shift of the overall spectrum. Moreover, we
found that the SWGC cannot effectively couple some of the wavelength
ranges with the TE mode, especially between the λ ≅ 700
and 800 nm. We presumed that this is due to the vertical destructive
interference between the Si3N4 and SiO2 layers. To address this issue and couple the entire wavelength range,
we utilized the TM mode, which has a different phase-matching condition
from the TE mode. Figure S3f shows the
TM mode simulation results using the same SWGC structures. Although
the SWGC structure is optimized for the TE mode, the SWGC can couple
the TM mode even better than the TE mode especially for the shorter
wavelengths, which allows bridging the TE mode-forbidden wavelengths
between 700 and 800 nm. Accordingly, the entire wavelength from 650
to 1050 nm can be bandpass filtered and coupled into the SHFTS with
corresponding polarization and coupling angle (θ) conditions.
Bandpass Sampling SHFTS Design
Using the optimized
Si3N4 components described above, we designed
and analyzed the bandpass sampling SHFTS. Based on eqs and 4, the
number of MZI arrays (N) and the maximum OPDs (ΔLmax) are designed to specify the bandwidth (Δλ)
and the resolution (δλ) of the SHFTS. First, the bandwidth
of the SHFTS is designed to be wider than the 3 dB bandwidths of the
SWGC, preventing the overlap of sampled aliases. In other words, the
bandwidths of each narrow-band channels of SHFTS must be able to fully
cover the coupled wavelengths from the SWGC. Here, we designed the
SHFTS with N = 32 and ΔLmax = 93 μm (Figure a) to have δλ < 5 nm and Δλ
= 30–80 nm at the wavelength range of λ = 650–1050
nm. The total wavelength window (650–1050 nm) is divided into
eight discrete channels based on the Nyquist rate and can be reconstructed
separately. The optical microscope image of the fabricated device
in Figure b shows
part of an MZI array. In order to better examine the resolution and
bandwidth of the SHFTS, we built the interconnect simulation model
of the SHFTS using the optimized components, and the λo of single peak input signals were swept from 647 to 1060 nm and
the signals were reconstructed from the SHFTS to verify the minimum
resolvable wavelength detuning in each channel. The spatial interferograms
of the output power from 32 MZIs were measured by optical power meter,
and the input signals were retrieved by using eq in MATLAB. The overall configurations of
the narrow-band channels including the corresponding bandwidths and
resolutions are shown in Table , and the reconstructed spectra with minimum resolvable wavelength
detuning from bands 1 and 7 are shown in Figure c and d, respectively.
Figure 3
SHFTS comprising 32 MZI
arrays integrated with SWGC inputs and
outputs. (a) Final outlook of the SHFTS device; the overall size is
around 4.7 mm × 0.65 mm. (b) Optical microscope image of the
fabricated device. (c) Simulation results of the SHFTS; reconstructed
spectrum in band 1 and (d) band 7 with minimum resolvable wavelength
detuning.
Table 1
Discrete Narrow-Band Channels of a
Bandpass Sampling SHFTS
band
start wavelength
[nm]
end wavelength
[nm]
bandwidth
[nm]
resolution
[nm]
1
647
680
33
2.06
2
680
715
35
2.19
3
715
755
40
2.50
4
755
800
45
2.81
5
800
855
55
3.44
6
855
915
60
3.75
7
915
985
70
4.38
8
985
1060
75
4.69
SHFTS comprising 32 MZI
arrays integrated with SWGC inputs and
outputs. (a) Final outlook of the SHFTS device; the overall size is
around 4.7 mm × 0.65 mm. (b) Optical microscope image of the
fabricated device. (c) Simulation results of the SHFTS; reconstructed
spectrum in band 1 and (d) band 7 with minimum resolvable wavelength
detuning.
Experimental Results
By virtue of a through-etched
SWGC design, we were able to pattern and etch the whole device in
one lithography and etching step without any additional alignment
lithography process. The final footprints of the fabricated chip including
the loss characterization patterns and bandpass sampling SHFTS device
are shown in Figure S7.Then, we
built the measurement setup to test the fabricated device. The schematic
diagram and real picture of the measurement setup are shown in Figure S9. We used the broadband supercontinuum
light source (NKT Photonics SuperK laser) to examine the performances
of the fabricated SWGC and bandpass sampling SHFTS.First of
all, to verify that the broadband light is coupled to
the PM-SMF successfully, we measured the fiber-coupled SuperK laser’s
spectrum by the optical spectrum analyzer (OSA) directly. Figure S10 shows the pictures of the SuperK laser
coupling setup and the OSA measured optical spectrum from the fiber,
and we examined that the broadband light is coupled to the PM-SMF
successfully with the maximum total optical power of ∼130 mW.
Using the measurement setup and the loss characterization devices
consisting of different lengths of waveguides, the propagation losses
of the Si3N4 strip waveguides were measured
as ∼2.5 dB/cm by the cut-back method (Figure S11). The power loss from the MMI splitter is wavelength and
polarization dependent, and the results are shown in Figure S5d and e.Then, the output spectra of the SWGC
were measured with different
coupling angles to test the coupling efficiencies and the wavelength
shifting. The input and output coupling angles were equally controlled
from 0° to 32°, as shown in the pictures in Figure a, and the output fiber was
connected to the OSA to measure the spectrum directly.
Figure 4
SWGC output measurement
results. (a) Pictures of the fiber to SWGC
coupling setup with different coupling angles; the enlarged picture
shows the red-color guided light through the waveguide, coupled to
the output fiber. (b) XYZ plot of the coupling efficiencies using
the TE mode; x-axis: wavelength, y-axis: coupling angle, z-axis: normalized coupling
efficiency. (c) XY plot of the coupling efficiencies using the TM
mode; blue line: θ = 18°, red line: θ = 12°.
SWGC output measurement
results. (a) Pictures of the fiber to SWGC
coupling setup with different coupling angles; the enlarged picture
shows the red-color guided light through the waveguide, coupled to
the output fiber. (b) XYZ plot of the coupling efficiencies using
the TE mode; x-axis: wavelength, y-axis: coupling angle, z-axis: normalized coupling
efficiency. (c) XY plot of the coupling efficiencies using the TM
mode; blue line: θ = 18°, red line: θ = 12°.Figure b and c
show the SWGC spectrum measurement results by the OSA with TE and
TM modes. With the fundamental TE mode (Figure b), the SWGC shows 4.5–8.5 dB losses,
and the center wavelength λo shifts from 650 nm to
1000 nm with the corresponding 3 dB bandwidth changing from 30 nm
to 70 nm as the coupling angle changes from 32° to 0°. More
detailed coupling loss data based on the coupling angle are shown
in Figure S12. The TM mode coupling efficiencies
with the two different coupling angles θ = 18° and 12°
were also measured for λo ≅ 740 and 835 nm,
as shown in Figure c. On the basis of these results, we verified the coupling wavelength
shifting range and efficiencies of the SWGC experimentally, covering
the entire tissue transparency window (λ = 650–1050 nm).
As a consequence, the SuperK broadband input signal can be bandpass
filtered and divided into the multiple narrow-band channels shown
in Table through
the SWGC and coupled into the waveguide.Finally, using the
same measurement setup, we coupled the SuperK
laser light into the bandpass sampling SHFTS and measured the output
powers of each MZI’s interferogram using different SWGC coupling
angles to reconstruct the input spectrum. The single input signal
is equally divided into 32 MZI channels by the cascaded MMI splitters,
and the output powers of each MZI are measured separately by the optical
power meter. The optical images of the light-coupled SHFTS chip can
be found in Figure S13. We have measured
the optical powers using eight different coupling conditions, which
are θ = 32°, 25°, 20°, 14°, 4°, and
0° with TE polarization and θ = 18° and 12° with
TM polarization to retrieve the entire tissue transparency wavelength
range from 650 to 1050 nm. The experimental output power measurement
results from 32 MZIs using different SWGC coupling conditions are
shown in Figure S14. After we collected
the output powers from each MZI, we reconstructed the optical spectrum
of the bandpass-sampled signals by the MATLAB code based on the DFT
equation (eq ), and
the results are shown in Figure . To validate the spectrum retrieval accuracy, the
FTS retrieved results (red solid lines) were compared with the direct
OSA measurement results (black dotted lines). The retrieved spectra
are well matched with the direct OSA measurement results, but the
discrepancies are mainly due to the optical phase errors induced from
the etching surface and sidewall roughness and detecting noise from
the optical power meter. Figure S15 shows
the overall retrieved spectra from λ = 650–1050 nm by
the superposition, and the intensity normalization is applied based
on the coupling and loss values shown in Figures S5d,e and S12.
Figure 5
Retrieved spectra from the bandpass sampling SHFTS (red
solid lines)
and direct measured spectra from the OSA (black dotted lines) with
different SWGC coupling conditions of (a) TE mode θ = 32°,
(b) TE mode θ = 25°, (c) TM mode θ = 18°, (d)
TE mode θ = 20°, (e) TM mode θ = 12°, (f) TE
mode θ = 14°, (g) TE mode θ = 4°, and (h) TE
mode θ = 0°.
Retrieved spectra from the bandpass sampling SHFTS (red
solid lines)
and direct measured spectra from the OSA (black dotted lines) with
different SWGC coupling conditions of (a) TE mode θ = 32°,
(b) TE mode θ = 25°, (c) TM mode θ = 18°, (d)
TE mode θ = 20°, (e) TM mode θ = 12°, (f) TE
mode θ = 14°, (g) TE mode θ = 4°, and (h) TE
mode θ = 0°.In conclusion, we experimentally retrieved the
input signals from
λ = 650 nm to 1050 nm with eight discrete narrow-band channels
by the bandpass sampling SHFTS through SWGC coupling.
Discussion
Our proof-of-concept experiment demonstrates
the broadband spectrum
retrieval performance using the broadband supercontinuum light source
and bandpass sampling SHFTS, representing the overall spectrometer
bandwidth coverage (Δλoverall) of 400 nm without
compromising the spectrometer resolution. To the best of our knowledge,
this is the first demonstration of the broadband FTS covering the
entire tissue transparency window on a silicon nitride platform without
adding additional active photonic components for electrical or thermal
enhancements.Previously reported miniaturized spectrometers
at VIS–NIR
wavelength ranges (Table ) either were limited to the narrow bandwidth coverage or
had poor resolution, or the measurement system requires a too large
and complicated configuration,[11] which
cannot be implemented in hand-held or portable applications. In comparison
to these spectrometers including thermally tunable FTS devices, the
bandpass sampling FTS concept stands out in the sense that it provides
broadband coverage while maintaining a fine resolution and relatively
small size of the device footprint, without external active photonic
components to introduce additional thermally or electrically induced
OPDs. Moreover, the bandpass wavelength ranges are determined by the
coupling angle of the SWGC, and each discrete channel can be retrieved
in a single capture with a fixed coupling angle, which is beneficial
to the real-time biosensing applications especially when we only need
to scan a specific window.
Table 2
Comparison of Miniaturized Spectrometers
at the VIS–NIR Wavelength Range (650–1050 nm)
schemes
principles
platforms
resolution
[nm]
bandwidth
[nm]
range [nm]
no. channel
footprint
[mm2]
dispersive
optics
waveguide
grating diffraction[37]
SiON
40
800
300–1100
35
MRR
assisted AWG[7]
SiN
0.75
52.5
831–883
70
1.44
AWG[8]
SiN
1.5
60
830–890
40
0.62
AWG[9]
SiN
0.5
4
757–762
8
2.8
AWG[10]
SiON
5.5
215
740–955
41
340
planar
echelle gratings
and single photon detection[11]
SiN/NbNa
7
1400
600–2000
200
∼30
folded metasurface[12]
SiO2/Si/Au
1.2
100
760–860
7
narrow-band filters
linear variable optical
filter[13]
multilayered films (SiO2/TiO2)
2.2
170
570–740
100
reconstructive
PC slabs[14]
SOS
1
200
550–750
36
0.044
Fourier transform
SWIFT[15]
SiN
6
100
800–900
0.1
bandpass
sampling SHFTS
(this work)
SiN
2–5
400
650–1050
32
3
Cryogenic low-noise amplifiers are
required for the measurement.
Cryogenic low-noise amplifiers are
required for the measurement.Despite the advantages of the bandpass sampling SHFTS
described
above, our prototype result leaves much room for further improvement
for the monolithic integrated circuits and chances for biosensing
applications, which require the integration of the on-chip light source
and PD array. Specifically, the current prototype device’s
scanning speed is significantly limited by the fiber alignment and
angle tuning to cover the entire bandwidth. In other words, each single
separate band can be retrieved in a single capture with a fixed coupling
angle, but scanning eight separate bands requires the coupling angle
tuning and alignment with the grating couple for each band, which
can be a slow speed operation. Moreover, to exploit the on-chip integrated
broadband light source into the bandpass sampling SHFTS scheme, a
proper method that can substitute the SWGC and fiber angle tuning
in this work should be devised as an on-chip tunable-bandpass filter.
Several studies have reported on-chip bandpass filter structures using
a single-channel optical bandpass filter based on plasmonic nanocavities,[39] gratings,[40,41] and cascaded ring resonators[42] or MZI structures,[43] which can be potentially integrated with the SHFTS configuration.
In the Supporting Information, we discuss
the possibilities and challenges of the integration with the on-chip
broadband light source and PD array, which provide guidelines to further
improve the performances toward the fully integrated portable spectrometers
for lab-on-a-chip Raman or absorption spectroscopy systems.
Conclusion
In this paper, we designed and experimentally
demonstrated the
bandpass sampling SHFTS integrated with an SWGC on the Si3N4 platform to achieve broad bandwidth coverage (Δλoverall = 400 nm) without compromising the spectrometer resolution
(δλ ≈ 2–5 nm) in the VIS–NIR tissue
transparency window (Δλ = 650–1050 nm). Unlike
the standard SHFTS operating scheme, the bandpass sampling theorem
is applied by tuning the coupling angle of the SWGC, substituting
the tunable-bandpass filter in our design. Namely, the SWGCs were
used not only as a fiber to waveguide coupler but also as an antialiasing
filter, dividing the continuous broadband spectrum into multiple narrow-band
channels. Then, the original broadband spectrum can be reconstructed
by the superposition of each retrieved narrow-band spectra. We optimized
the low-loss passive components to build the MZI structures, including
the strip waveguide, MMIs, and SWGC, and the SHFTS is designed to
have 32 linearly unbalanced MZIs with the maximum OPD (ΔLmax) of 93 μm, with a total footprint
size of ∼4.7 mm × 0.65 mm. To the best of our knowledge,
this is the first experimental demonstration of the broadband FTS
covering the entire tissue transparency window on a silicon nitride
platform without adding additional active photonic components.
Methods
Device Simulation and Optimization
The simulation software
Lumerical MODE and FDTD were used to analyze and optimize the passive
components. The internal optimization tool with the particle-swarm-algorithm
was used to generate a number of structures with different dimensions
in specific ranges and get the maximum coupling efficiency or minimum
losses as needed. Also, the S-parameters from the
input and output ports of SWGC and MMIs were calculated from the FDTD
simulation to monitor the light transmission characteristics, and
the S-parameters from the optimized passive components
were used to build the SHFTS model in the interconnected photonic
circuit simulation.
Si3N4 SHFTS Device Fabrication
The Si3N4-on-SiO2 wafers were prepared
with a 220 nm thick LPCVD-grown Si3N4 on a 2.8
μm thick SiO2 bottom cladding on a silicon substrate
(Figure S6a) from Rogue Valley Microdevices
Inc. Then, a ∼400 nm thick e-beam resist (ZEP-520A) is deposited
on top of the Si3N4 layer by spin-coating. The
patterning is done by a JEOL e-beam (JBX-6000FS) lithography tool,
followed by developing in n-amyl acetate for 2 min
and rinsing in isopropyl alcohol (IPA) (Figure S6b). Following this, the pattern is transferred to the Si3N4 layer by reactive-ion etching (RIE) (Figure S6c). Finally, the remaining resist and
polymers are cleaned using removal PG followed by cycles of acetone/IPA
postprocess treatment (Figure S6d).
Measurement Setup
The unpolarized broadband supercontinuum
light source (NKT Photonics SuperK Versa), which covers the wavelength
range from 550 to 1750 nm, is coupled to the polarization-maintaining
single-mode fiber (Thorlab PM780-HP, PM630-HP) using a Glan-Thompson
linear polarizer (Newport 5525) and microscope objective lens (100×)
for TE or TM mode transmission. The fibers are mounted on the goniometer
stages to control the coupling angles for both input and output fibers.
The device is placed on the flat stage, and the input/output fibers
are aligned with the SWGCs to couple the light to the waveguide, monitored
by the optical microscope. Finally, the output fiber is connected
to the optical power meter (Newport 2936-C) to measure the output
power.
Authors: Pascual Muñoz; Gloria Micó; Luis A Bru; Daniel Pastor; Daniel Pérez; José David Doménech; Juan Fernández; Rocío Baños; Bernardo Gargallo; Rubén Alemany; Ana M Sánchez; Josep M Cirera; Roser Mas; Carlos Domínguez Journal: Sensors (Basel) Date: 2017-09-12 Impact factor: 3.576