| Literature DB >> 35983293 |
Sachila Niroshani1,2, Tokiko Nakamura1,3, Nikaidou Michiru1, Toru Negishi1.
Abstract
Background: Dual-energy contrast-enhanced mammography (DECEM) is an advanced breast imaging technique of digital mammography. Purpose: To assess the total radiation dose received from complete DECEM using different combinations of exposure parameters for low- and high-energy images. Materials and methods: A dedicated phantom with three different concentrations of iodine inserts was used. Each iodine insert was 10 mm in diameter and concentration of 1.0 mgI/cm3, 2.0 mgI/cm3, and 4.0 mgI/cm3. The phantom was exposed at varying kVp levels. Mean glandular dose (MGD) was estimated. Contrast to noise ratio (CNR) and figure of merit (FOM) of the iodine inserts were used to assess the image quality.Entities:
Keywords: Dual-energy contrast-enhanced mammography; high energy; low energy; radiation dose
Year: 2022 PMID: 35983293 PMCID: PMC9379970 DOI: 10.1177/20584601221117251
Source DB: PubMed Journal: Acta Radiol Open
Exposure factors used for LE and HE imaging of 50 mm thickness 50% fibroglandular–50% adipose phantom.
| LE | HE | ||
|---|---|---|---|
| kVp | mAs | kVp | mAs |
| 28 | 63 | 45 | 20 |
| 30 | 56 | 47 | 18 |
| 32 | 42 | 49 | 16 |
Total MGD values of recombined images according to different phantom compositions.
| Recombined image | MGD (mGy) | ||
|---|---|---|---|
| LE + HE (kVp) | 50% glandular | 100% adipose | 100% glandular |
| 28 + 45 | 1.79 | 1.45 | 2.56 |
| 28 + 47 | 1.77 | 1.31 | 2.51 |
| 28 + 49 | 1.75 | 1.29 | 2.46 |
| 30 + 45 | 1.76 | 1.24 | 2.49 |
| 30 + 47 | 1.74 | 1.10 | 2.44 |
| 30 + 49 | 1.72 | 1.08 | 2.39 |
| 32 + 45 | 1.74 | 1.22 | 2.46 |
| 32 + 47 | 1.72 | 1.08 | 2.41 |
| 32 + 49 | 1.70 | 1.06 | 2.36 |
Calculated CNR values of three iodine inserts with different concentrations of recombined images.
| Tube voltage combination for recombined image (LE image + HE image) kVp | CNR | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 100% adipose | 50% adipose–50% fibroglandular | 100% fibroglandular | |||||||
| Iodine concentration (mgI/cm3) | |||||||||
| 1.0 | 2.0 | 4.0 | 1.0 | 2.0 | 4.0 | 1.0 | 2.0 | 4.0 | |
| 28 + 45 | 0.45 | 0.88 | 1.63 | 1.06 | 1.42 | 2.40 | 0.84 | 1.45 | 2.59 |
| 28 + 47 | −0.04 | 0.60 | 1.14 | 0.74 | 1.25 | 2.23 | 0.74 | 1.43 | 2.61 |
| 28 + 49 | 0.08 | 0.66 | 1.36 | 0.54 | 1.07 | 2.10 | 0.53 | 1.28 | 2.42 |
| 30 + 45 | 1.36 | 1.68 | 2.59 | 0.63 | 0.96 | 1.91 | 0.62 | 1.20 | 2.34 |
| 30 + 47 | 0.78 | 1.33 | 1.97 | 0.31 | 0.81 | 1.70 | 0.42 | 1.09 | 2.26 |
| 30 + 49 | 0.86 | 1.35 | 2.17 | 0.17 | 0.68 | 1.65 | 0.34 | 1.04 | 2.22 |
| 32 + 45 | 1.10 | 1.40 | 2.30 | 0.79 | 1.10 | 2.05 | 0.52 | 1.07 | 2.16 |
| 32 + 47 | 0.56 | 1.10 | 1.75 | 0.50 | 0.98 | 1.89 | 0.43 | 1.04 | 2.15 |
| 32 + 49 | 0.63 | 1.11 | 1.92 | 0.30 | 0.80 | 1.77 | 0.30 | 0.98 | 2.08 |
CNR of iodine inserts of a 50% fibroglandular–50% adipose 50 mm thickness phantom-LE images.
| kVp (LE image) | CNR | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 100% adipose | 50% adipose–50% fibroglandular | 100% fibroglandular | |||||||
| Iodine concentration (mgI/cm3) | |||||||||
| 1.0 | 2.0 | 4.0 | 1.0 | 2.0 | 4.0 | 1.0 | 2.0 | 4.0 | |
| 28 | 5.69 | 6.29 | 6.81 | 5.36 | 5.95 | 6.42 | 5.35 | 5.68 | 6.34 |
| 30 | 4.82 | 5.13 | 5.65 | 5.66 | 6.07 | 6.56 | 5.26 | 5.55 | 6.07 |
| 32 | 4.74 | 5.09 | 5.59 | 5.22 | 5.61 | 6.18 | 4.97 | 5.16 | 5.87 |
Different combinations of kVp used to create recombined images in the study.
| Recombined image |
|---|
| LE + HE (kVp) |
| 28 + 45 |
| 28 + 47 |
| 28 + 49 |
| 30 + 45 |
| 30 + 47 |
| 30 + 49 |
| 32 + 45 |
| 32 + 47 |
| 32 + 49 |
MGD values of LE and HE images of 50 mm phantom with different compositions.
| LE | HE | ||||||
|---|---|---|---|---|---|---|---|
| kVp | MGD (mGy) | kVp | MGD (mGy) | ||||
| 50% glandular | 100% adipose | 100% glandular | 50% glandular | 100% adipose | 100% glandular | ||
| 28 | 0.91 | 0.77 | 1.29 | 45 | 0.88 | 0.68 | 1.27 |
| 30 | 0.88 | 0.56 | 1.22 | 47 | 0.86 | 0.54 | 1.22 |
| 32 | 0.86 | 0.54 | 1.19 | 49 | 0.84 | 0.52 | 1.17 |
Figure 1.Total MGD of complete dual-energy contrast-enhanced mammography for 50 mm thickness, 50% fibroglandular–50% adipose phantom at different kVp combinations.
Figure 2.Figure of merit (FOM) values of 4.0 mgI/cm3 iodine inserts of a 50 mm thickness (a) 50% fibroglandular–50% adipose, (b) 100% adipose, and (c) 100% glandular phantom recombined images.
Figure 3.Contrast to noise ratio (CNR) as a function of the concentration of iodine inserts of the recombined image (a) 28 kVp LE + 49 kVp HE optimized voltage combination and (b) 28 kVp LE + 45 kVp HE voltage combination which gives the highest CNR.