The Relationship Between Different Bench Test Methodologies and Accuracy of Insulin Infusion Pumps: A Systematic Literature Review.

Fernanda Silva Tenorio1, Luiz Eduardo Galvão Martins2, Glória Sthefany Pereira da Silva2, Camila Bertini Martins1, Ana Lucia Domingues Neves1, Tatiana Sousa Cunha3.   

Abstract

Diabetes technology has rapidly evolved, and insulin infusion pumps (IIPs) have gained worldwide acceptance in diabetes care. The safety of medical equipment is highly discussed, imposing complex challenges in its use. The accuracy of IIPs can be determined through laboratory tests, generally following the IEC 60601-2-24 protocol. Studies have evaluated the accuracy and precision of IIPs, and there are discrepant results. So, we conducted a Systematic Literature Review to assess the methodologies used to evaluate the accuracy of IIPs, organizing the findings in a compiled perspective. The methodology was based on Kitchenham and Biolchini guidelines, and when possible it was carried out the Bayesian meta-analyses to compare the accuracy of IIPs. Most studies used the microgravimetric technique to evaluate the device accuracy, and some proposed adaptations for the standard protocol. The variation of results was recurrent, and the establishment of a protocol, especially to evaluate patch pumps, is necessary. The present study gives enough data to understand the scenario of the IIPs evaluation, as well as the different protocols that can be explored for its evaluation. This highlights the need for a reliable, practical, and low-cost methodology to assist the evaluation of IIPs.
© 2022. The Author(s) under exclusive licence to Biomedical Engineering Society.

Entities:  

Keywords:  Accuracy; Bayesian meta-analysis; Insulin infusion pumps; Methodologies; Systematic Literature Review

Mesh:

Substances:

Year:  2022        PMID: 35953739     DOI: 10.1007/s10439-022-03027-7

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   4.219


  1 in total

1.  Accuracy of a Low-Cost Continuous Subcutaneous Insulin Infusion Pump Prototype: In Vitro Study Using Combined Methodologies.

Authors:  Fernanda Silva Tenorio; Luiz Eduardo Galvão Martins; Tatiana Sousa Cunha
Journal:  Ann Biomed Eng       Date:  2021-01-25       Impact factor: 3.934

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.