Implantable Cardioverter-defibrillator Magnetic Interference by the iPhone 12: Is It Clinically Significant?

Harshil Patel1, Cameron Whitler1, Nathan Foster1, Christopher Bradley1, Dipak Shah1, Christian Machado1.   

Abstract

The iPhone 12 series (Apple, Inc., Cupertino, CA, USA) contains a circular array of magnets around a central charging coil (compatible with "Magsafe" technology). The device was recently reported to have magnetic interference with implantable cardioverter-defibrillators (ICDs). We sought to test the electromagnetic interference of the iPhone 12 in inhibiting life-saving therapies of ICDs in clinical settings. After obtaining written informed consent, an iPhone 12 was placed over the device generators of 17 patients in the ICD clinic. Device interrogation was performed immediately before and after placing the iPhone over the ICD generator to evaluate for any inhibition of device therapies. To emulate a real-world scenario, the iPhone 12 was not placed directly over the skin above the device generator but instead was positioned over the patients' clothes. None of the device interrogations revealed interruption of device therapies due to the iPhone. We concluded that, despite the iPhone having shown in vitro interference of ICD functioning, its effects are not clinically relevant in vivo. Larger studies need to be performed to confirm this finding and guide safety recommendations regarding the use of iPhones containing magnets by patients with implanted ICDs. Copyright:
© 2022 Innovations in Cardiac Rhythm Management.

Entities:  

Keywords:  ICD; magnetic interference; sudden cardiac arrest

Year:  2022        PMID: 35949649      PMCID: PMC9359423          DOI: 10.19102/icrm.2022.130702

Source DB:  PubMed          Journal:  J Innov Card Rhythm Manag        ISSN: 2156-3977


Introduction

The iPhone 12 series (Apple, Inc., Cupertino, CA, USA), containing a circular array of magnets around a central charging coil (compatible with the “MagSafe” technology), was recently reported to cause magnetic interference with an implantable cardioverter-defibrillator (ICD).[1] The MagSafe technology, which enables wireless charging functionality, consists of an array of magnets capable of generating a magnetic field of around 0.005 tesla.[2] Studies that have reported interference with implantable devices by the iPhone 12 were not performed on ambulatory patients in a clinical setting. We present a case series of 17 patients with ICDs in whom testing of their device for interference from an iPhone 12 was carried out in an ICD clinic.

Methods

This prospective study included 17 patients, all of whom were randomly selected. The study was conducted at the hospital device clinic after obtaining ethics committee approval. Written informed consent was obtained, after which each patient underwent a baseline device interrogation. An iPhone 12 in working condition with a fully charged battery, active WiFi, and active cellular data/calling services was used for conducting the study. The iPhone 12, without any phone cover, was placed over the ICD pocket above the clothing. The iPhone was held in 3 separate positions: directly above the device pocket and superior and inferior to the pocket, for 20 s in each position. Device interrogation was repeated to determine any change in settings or functionality.

Results

None of the 17 patients had any demonstrable change in the settings or magnet response events upon repeat interrogation (.

Discussion

The magnet response in ICDs is the inhibition of tachyarrhythmia therapies, whereas the magnet response is asynchronous pacing in pacemakers.[3] The MagSafe technology by Apple enables wireless charging capabilities of the iPhone. The charging coil is surrounded by an array of magnets. When the iPhone is placed on a charging pad with an inbuilt magnet, it is aligned with relative precision to allow wireless charging. Recent studies by Greenberg et al. and Nadeem et al. demonstrated inhibition of lifesaving therapies by the Apple iPhone 12 Pro Max.[1,2] After these reports were published, the iPhone manufacturer (Apple) released a statement detailing precautionary measures for iPhone users with a cardiac implantable electronic device (CIED) in place. Recommendations were also made to keep the iPhone 12 at a distance of ≥6 in (15 cm) from the CIED as a safety measure to prevent inhibition of lifesaving therapies.[4] While the iPhone 12 Pro Max interfered with the signals when placed directly over the skin with a CIED generator underneath, our study could not replicate these findings when the phone was placed over clothing (rather than on the skin). The rationale of placing the device over the clothing was to mimic a more practical scenario, such as placing the phone in a shirt pocket. The results of a recent study by Held et al. are similar to those of ours.[5] In their case series of 12 patients, 4 different phones as well as an iPhone 12 case with an embedded magnet were used to evaluate the interference of implanted device signals. None of the 12 devices showed any interference in their signals. These results raise pertinent questions regarding the clinical significance of recommendations of keeping the phone 6 in away from the device generator. Studies with measurable clinical outcomes like mortality from cardiac arrest due to the inhibition of tachycardia therapies would certainly help answer these questions. However, an ethical dilemma might arise when designing and implementing such a protocol. This would particularly be the case when the risk of inhibition of device therapies can be avoided by keeping the phone away from the device. The fact that studies have shown conflicting results on the risk of phones with magnets interfering with CIEDs’ functionality suggests that more data are certainly needed with multiple variables.[1,2,5] The precise distance of the phone from the device, the thickness of the clothing, and the thickness of subcutaneous tissue over the generator are some of the factors that could have a significant impact on the conflicting results.
Table 1:

Device and Demographic Information of Study Participants

Device Information
Patient Demographics
Patient #ICD ManufacturerDevice ModelDescriptionBMI (kg/m2)SexAge (Years)Pacemaker Dependence?
1Boston Scientific (Natick, MA, USA)Energen ICD E140Single chamber38.3Female61No
2Boston ScientificTeligen 100 E110Dual chamber39.2Male50No
3Boston ScientificEnergen VR E141Single chamber25Male64No
4St. Jude Medical (St. Paul, MN, USA)Ellipse VR 1411-36QSingle chamber44.1Male53No
5St. Jude MedicalUnify Assura 3357-40QBiventricular23.3Female82No
6St. Jude MedicalUnify Assura 3357-40QBiventricular22.9Male36No
7St. Jude MedicalQuadra Assura 3365-40QBiventricular28Female73No
8Medtronic (Minneapolis, MN, USA)Visia AF MRI VR DVFB1D4Single chamber26Male67No
9St. Jude MedicalEllipse VR 1411-36QSingle chamber24Male62No
10MedtronicClaria MRI CRT-D DTMA1Q1Biventricular28Male78No
11St. Jude MedicalEllipse VR 1411-36Single chamber27Female84No
12St. Jude Medical2411-36QDual chamber28Male80No
13Boston ScientificDynagen D150Single chamber35Male47No
14St. Jude MedicalUnify Assura 3357-40CBiventricular28Female49No
15St. Jude MedicalEllipse VR 1411-36CSingle chamber35Male76No
16St. Jude MedicalEllipse VR 1411-36Single chamber22Male83No
17MedtronicClaria MRI CRT-D DTMA1Q1Biventricular32Female55No

Abbreviations: BMI, body mass index; ICD, implantable cardioverter-defibrillator.

  4 in total

1.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

2.  Letter to the Editor-Lifesaving therapy inhibition by phones containing magnets.

Authors:  Joshua C Greenberg; Mahmoud R Altawil; Gurjit Singh
Journal:  Heart Rhythm       Date:  2021-01-05       Impact factor: 6.343

3.  To the Editor-Interference of smartphones containing magnets and cardiac implantable electronic devices-Is this common?

Authors:  Elizabeth P Held; Michael Shehata; Charles D Swerdlow; Roopinder K Sandhu
Journal:  Heart Rhythm       Date:  2021-03-09       Impact factor: 6.343

4.  Magnetic Interference on Cardiac Implantable Electronic Devices From Apple iPhone MagSafe Technology.

Authors:  Fahd Nadeem; Arismendy Nunez Garcia; Cao Thach Tran; Michael Wu
Journal:  J Am Heart Assoc       Date:  2021-06-02       Impact factor: 5.501

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.