The iPhone 12 series (Apple, Inc., Cupertino, CA, USA) contains a circular array of magnets around a central charging coil (compatible with "Magsafe" technology). The device was recently reported to have magnetic interference with implantable cardioverter-defibrillators (ICDs). We sought to test the electromagnetic interference of the iPhone 12 in inhibiting life-saving therapies of ICDs in clinical settings. After obtaining written informed consent, an iPhone 12 was placed over the device generators of 17 patients in the ICD clinic. Device interrogation was performed immediately before and after placing the iPhone over the ICD generator to evaluate for any inhibition of device therapies. To emulate a real-world scenario, the iPhone 12 was not placed directly over the skin above the device generator but instead was positioned over the patients' clothes. None of the device interrogations revealed interruption of device therapies due to the iPhone. We concluded that, despite the iPhone having shown in vitro interference of ICD functioning, its effects are not clinically relevant in vivo. Larger studies need to be performed to confirm this finding and guide safety recommendations regarding the use of iPhones containing magnets by patients with implanted ICDs. Copyright:Entities:
Keywords: ICD; magnetic interference; sudden cardiac arrest
Year: 2022 PMID: 35949649 PMCID: PMC9359423 DOI: 10.19102/icrm.2022.130702
Source DB: PubMed Journal: J Innov Card Rhythm Manag ISSN: 2156-3977
Device and Demographic Information of Study Participants
| Device Information | Patient Demographics | ||||||
|---|---|---|---|---|---|---|---|
| Patient # | ICD Manufacturer | Device Model | Description | BMI (kg/m2) | Sex | Age (Years) | Pacemaker Dependence? |
| 1 | Boston Scientific (Natick, MA, USA) | Energen ICD E140 | Single chamber | 38.3 | Female | 61 | No |
| 2 | Boston Scientific | Teligen 100 E110 | Dual chamber | 39.2 | Male | 50 | No |
| 3 | Boston Scientific | Energen VR E141 | Single chamber | 25 | Male | 64 | No |
| 4 | St. Jude Medical (St. Paul, MN, USA) | Ellipse VR 1411-36Q | Single chamber | 44.1 | Male | 53 | No |
| 5 | St. Jude Medical | Unify Assura 3357-40Q | Biventricular | 23.3 | Female | 82 | No |
| 6 | St. Jude Medical | Unify Assura 3357-40Q | Biventricular | 22.9 | Male | 36 | No |
| 7 | St. Jude Medical | Quadra Assura 3365-40Q | Biventricular | 28 | Female | 73 | No |
| 8 | Medtronic (Minneapolis, MN, USA) | Visia AF MRI VR DVFB1D4 | Single chamber | 26 | Male | 67 | No |
| 9 | St. Jude Medical | Ellipse VR 1411-36Q | Single chamber | 24 | Male | 62 | No |
| 10 | Medtronic | Claria MRI CRT-D DTMA1Q1 | Biventricular | 28 | Male | 78 | No |
| 11 | St. Jude Medical | Ellipse VR 1411-36 | Single chamber | 27 | Female | 84 | No |
| 12 | St. Jude Medical | 2411-36Q | Dual chamber | 28 | Male | 80 | No |
| 13 | Boston Scientific | Dynagen D150 | Single chamber | 35 | Male | 47 | No |
| 14 | St. Jude Medical | Unify Assura 3357-40C | Biventricular | 28 | Female | 49 | No |
| 15 | St. Jude Medical | Ellipse VR 1411-36C | Single chamber | 35 | Male | 76 | No |
| 16 | St. Jude Medical | Ellipse VR 1411-36 | Single chamber | 22 | Male | 83 | No |
| 17 | Medtronic | Claria MRI CRT-D DTMA1Q1 | Biventricular | 32 | Female | 55 | No |
Abbreviations: BMI, body mass index; ICD, implantable cardioverter-defibrillator.