Yong Li1, Jianxi Guo1, Yanfang Zhang1, Jian Kong1. 1. Department of Interventional Radiology, Shenzhen People's Hospital (Second Clinical Medical College of Jinan University, First Affiliated Hospital of Southern University of Science and Technology), Shenzhen, China.
Abstract
Objective: This study aimed to explore the incidence of abnormal catheter positioning and the effectiveness and safety of intravascular adjustment or removal of abnormally positioned catheters through percutaneous punctures. Materials and methods: A retrospective analysis was conducted on 58 patients with abnormal catheter positioning, treated between January 2009 and June 2019. Intravascular adjustment of the migrated catheters and removal of the fractured catheters were performed through percutaneous puncture using a pigtail catheter, cobra catheter, and gooseneck snare. Results: Of the 58 cases, there were 23 cases of catheter migration and 35 cases of catheter fracture. The incidence of abnormal catheter positioning was 3.0%, corresponding to 1.2% migrations and 1.8% fractures. Among the 23 cases of migration, 1 case did not require adjustment and another underwent unsuccessful adjustment. The rate of successful adjustment of migrated catheters was 91.3%, whereas the rate of successful removal of fractured catheters was 100%. No surgery-related complications were observed either immediately or during the 1-month follow-up period. Conclusions: This study showed that the incidence of abnormal catheter positioning is low. Intravascular techniques used for the adjustment or removal of abnormally placed catheters are safe, efficient, and minimally invasive.
Objective: This study aimed to explore the incidence of abnormal catheter positioning and the effectiveness and safety of intravascular adjustment or removal of abnormally positioned catheters through percutaneous punctures. Materials and methods: A retrospective analysis was conducted on 58 patients with abnormal catheter positioning, treated between January 2009 and June 2019. Intravascular adjustment of the migrated catheters and removal of the fractured catheters were performed through percutaneous puncture using a pigtail catheter, cobra catheter, and gooseneck snare. Results: Of the 58 cases, there were 23 cases of catheter migration and 35 cases of catheter fracture. The incidence of abnormal catheter positioning was 3.0%, corresponding to 1.2% migrations and 1.8% fractures. Among the 23 cases of migration, 1 case did not require adjustment and another underwent unsuccessful adjustment. The rate of successful adjustment of migrated catheters was 91.3%, whereas the rate of successful removal of fractured catheters was 100%. No surgery-related complications were observed either immediately or during the 1-month follow-up period. Conclusions: This study showed that the incidence of abnormal catheter positioning is low. Intravascular techniques used for the adjustment or removal of abnormally placed catheters are safe, efficient, and minimally invasive.
The Port-a-cath system (PCS) is a long-term indwelling venous access device that can be implanted subcutaneously. It avoids unnecessary pain and repeated punctures and allows the administration of medicines directly to the central venous system. The most common complications of PCS are venous thrombosis, infection, and abnormal catheter positioning., Abnormal positioning includes migration, and fracture of catheters. Percutaneous puncture is a safe, effective, and minimally invasive method of intravascular adjustment or removal of abnormally positioned catheters. This report describes the clinical use of cobra catheters, pigtail catheters, and retrieval devices for intravascular adjustment or removal of abnormally positioned catheters through percutaneous punctures.
Materials and methods
Patients
A retrospective analysis was conducted on 58 patients with tumors who were noted to have abnormal catheter positioning and underwent intravascular treatment through percutaneous punctures in the Interventional Radiology Department of Shenzhen People’s Hospital between January 2009 and June 2019. The diagnosis of abnormal catheter positioning was confirmed in all patients using preoperative plain chest radiography or computed tomography (CT) images. Percutaneous puncture was then performed for intravascular adjustment of the migrated catheters or removal of the fractured catheters. All interventional procedures were performed using digital subtraction angiography (DSA) in interventional operating rooms. All patients were hospitalized and signed an informed consent form prior to study enrollment. The present study was reviewed and approved by the Ethics Committee of Shenzhen People’s Hospital, China (NO. LL-KY-2020443).
Perioperative management
Color ultrasonography of the subclavian vein and internal jugular vein, D-dimer test, routine blood test, and coagulation function examination were performed on all patients before surgery. During surgery, real-time monitoring of the blood pressure, heart rate, and electrocardiogram indices was performed. Data on total blood count, prothrombin time (PT), and partial thromboplastin time (PTT) were obtained within three days after surgery.
Operative technique
Local anesthesia was given in the supine position by administering 2% lidocaine hydrochloride injection to the patient’s right inguinal region. The femoral vein was then punctured using the Seldinger technique, and guidewires and vascular sheaths were introduced under fluoroscopy guidance. In this study, 6 F sheaths were used to adjust the catheter position, whereas 8 F or larger sheaths were used to remove fractured catheters.A cobra catheter (Cordis, Miami Lakes, FL, USA) or pigtail catheter (Cordis, Miami Lakes, FL, USA) was used to relocate migrated catheters to the superior vena cava by pulling on the junction of the subclavian vein and the internal jugular vein. The tail of the pigtail catheter was cut into a semicircular shape using sterile scalpels to hook the catheter. A 6 F gooseneck snare (Microvena, White Bear Lake, MN, USA) was used to hook the free end of the catheter and reposition it to the superior vena cava when the use of both the cobra and pigtail catheters had failed.The gooseneck snare was also used to retrieve fractured catheters. Exposure of the free end was achieved by using the tail of the pigtail or cobra catheter to hook the middle part of the fractured catheter when it was difficult to see. The gooseneck snare was then used to hook the free end and pull the catheter downward to the vascular sheath. The entire system would be subsequently removed.Intraoperative angiography was performed to assess the presence of thrombosis and electrocardiographic monitoring was performed to record immediate operation-related complications. Patients with catheter migration underwent chest radiography every month after surgery and were followed up for three consecutive months. Patients with catheter fracture were followed-up through telephone interviews for one month after surgery.
Results
A total of 1716 venous access systems were implanted in the hospital between January 2009 and June 2019. A total of 58 cases of abnormal positioning were detected, consisting of 23 cases of migrated catheters and 35 cases of fractured catheters. Of the 58 cases, 52 patients with abnormally positioned catheters underwent implantation in this hospital, whereas the other 6 underwent the procedure in other hospitals. Therefore, the incidence of abnormal catheter positioning in this hospital was 3.0% (52/1716). The detailed patient data are presented in Table 1, Table 2. The patients were comprised of 25 men and 33 women, and the average age of the patients was 55.59 ± 14.09 years old (range: 31–82 years). The average time from implantation to diagnosis of abnormal positioning was 549.38 ± 312.72 days (range: 10–1401 days), whereas the average time from diagnosis to adjustment or removal was 18.05 ± 12.23 days (range: 1–45 days).
Table 1
Clinical data of patients with catheter migration.
No.
Tumor type
Symptom
Gender/Age
Time span from implantation to diagnosis (Days)
Time span from detection to adjustment (Days)
Surgical device
Operation time (Minute)
PCS type
1
Breast cancer
Injection difficulty
Female/34
158
2
Pig-tail Catheter
15
Bard
2
Colon cancer
Injection difficulty
Male/46
1013
7
GooseNeck Snare
20
Bard
3
Breast cancer
Injection difficulty
Female/53
621
13
GooseNeck Snare
47
Bard
4
Lung cancer
Injection difficulty
Male/67
737
11
GooseNeck Snare
25
MedCOMP
5
Liver cancer
No
Female/31
526
/
/
/
B. Braun
6
Breast cancer
No
Female/52
538
37
Pig-tail Catheter
14
Bard
7
Breast cancer
Injection difficulty
Female/47
813
9
GooseNeck Snare
21
Bard
8
Gastric cancer
Injection difficulty
Male/61
256
18
Cobra Catheter
11
MedCOMP
9
Rectal cancer
Dizziness, headache
Male/51
196
21
Pig-tail Catheter
14
Bard
10
Breast cancer
Injection difficulty
Female/48
457
11
Pig-tail Catheter
13
Bard
11
Breast cancer
No
Female/37
518
37
Pig-tail Catheter
11
Bard
12
Colon cancer
Distending pain in subclavian region
Male/46
10
7
GooseNeck Snare
20
Bard
13
Breast cancer
Injection difficulty
Female/53
216
13
GooseNeck Snare
45
Bard
14
Lung cancer
Injection difficulty
Female/63
378
11
GooseNeck Snare
25
MedCOMP
15
Liver cancer
No
Male/38
729
41
Cobra Catheter
16
B. Braun
16
Breast cancer
No
Female/52
428
12
Pig-tail Catheter
13
Bard
17
Breast cancer
Injection difficulty
Female/47
823
9
GooseNeck Snare
22
Bard
18
Gastric cancer
Discomfort in shoulders and neck
Male/65
636
18
Cobra Catheter
17
MedCOMP
19
Rectal cancer
Dizziness, headache
Male/61
91
12
Pig-tail Catheter
12
Bard
20
Endometrial cancer
Injection difficulty
Female/67
454
17
Pig-tail Catheter
13
Bard
21
Rectal cancer
No
Male/54
411
38
Pig-tail Catheter
16
Bard
22
Lung cancer
Injection difficulty
Male/71
345
14
Pig-tail Catheter
9
Bard
23
Breast cancer
Injection difficulty
Female/46
84
22
Cobra Catheter
21
Bard
Table 2
Clinical data of patients with catheter fracture.
No.
Tumor type
Symptom
Gender/Age
Time span from implantation to diagnosis (Days)
Time span from detection to adjustment (Days)
Cause
Site
Surgical device
Length of fractured catheter (cm)
Operation time (Minute)
PCS type
1
Lung cancer
Injection difficulty
Female/82
586
16
Detached anastomosis
Superior vena cava
GooseNeck Snare
17
25
Bard
2
Rectal cancer
No
Male/79
379
9
Detached anastomosis
Superior vena cava
GooseNeck Snare
18
20
Bard
3
Colon cancer
Injection difficulty
Female/76
1401
2
Detached anastomosis
Pulmonary artery
Pig-tail Catheter & GooseNeck Snare
15
42
Bard
4
Lung cancer
Chest tightness
Male/71
754
27
POS
Right ventricle
Pig-tail Catheter & GooseNeck Snare
9
37
B. Braun
5
Breast cancer
Injection difficulty
Female/65
44
44
Detached anastomosis
Superior vena cava
GooseNeck Snare
18
24
Bard
6
Breast cancer
No
Female/59
398
29
Detached anastomosis
Inferior vena cava
GooseNeck Snare
16
15
Bard
7
Colon cancer
Injection difficulty
Female/61
579
17
POS
Pulmonary artery
Pig-tail Catheter & GooseNeck Snare
12
28
MedCOMP
8
Breast cancer
Chest tightness
Female/38
816
1
POS
Right atrium
GooseNeck Snare
9
26
Bard
9
Colon cancer
No
Male/34
925
34
Detached anastomosis
Superior vena cava
GooseNeck Snare
20
13
Bard
10
Lung cancer
Injection difficulty
Female/81
147
41
Detached anastomosis
Superior vena cava
GooseNeck Snare
20
25
Bard
11
Gastric cancer
No
Female/72
271
23
Detached anastomosis
Superior vena cava
GooseNeck Snare
18
23
Bard
12
Breast cancer
Chest tightness
Female/38
572
4
POS
Pulmonary artery
EN Snare Loop
13
40
Bard
13
Breast cancer
Chest tightness
Female/47
428
3
POS
Right ventricle
GooseNeck Snare
9
37
MedCOMP
14
Colon cancer
Injection difficulty
Male/59
883
15
Detached anastomosis
Superior vena cava
GooseNeck Snare
18
27
Bard
15
Lung cancer
No
Female/69
1247
28
Detached anastomosis
Right atrium
GooseNeck Snare
17
12
Bard
16
Rectal cancer
Chest tightness
Male/37
481
5
POS
Right ventricle
Pig-tail Catheter & GooseNeck Snare
12
28
Bard
17
Gastric cancer
Injection difficulty
Male/71
773
8
POS
Right atrium
GooseNeck Snare
9
29
MedCOMP
18
Endometrial cancer
No
Female/51
694
45
Detached anastomosis
Superior vena cava
Pig-tail Catheter & GooseNeck Snare
15
13
Bard
19
Gastric cancer
Injection difficulty
Male/53
571
13
POS
Right atrium
GooseNeck Snare
17
20
Bard
20
Lung cancer
No
Male/69
493
29
Detached anastomosis
Superior vena cava
GooseNeck Snare
16
24
Bard
21
Colon cancer
Injection difficulty
Female/76
664
16
POS
Pulmonary artery
Pig-tail Catheter & GooseNeck Snare
10
40
Bard
22
Lung cancer
Chest tightness
Male/63
451
6
Detached anastomosis
Right ventricle
GooseNeck Snare
17
37
Bard
23
Breast cancer
Injection difficulty
Female/35
582
9
POS
Superior vena cava
GooseNeck Snare
8
18
Bard
24
Breast cancer
No
Female/42
618
25
Detached anastomosis
Right atrium
GooseNeck Snare
19
13
Bard
25
Breast cancer
Chest tightness
Female/55
872
4
POS
Right ventricle
Pig-tail Catheter & GooseNeck Snare
11
38
Bard
26
Lung cancer
No
Male/71
346
33
Detached anastomosis
Right atrium
GooseNeck Snare
18
27
Bard
27
Endometrial cancer
No
Female/67
157
24
Detached anastomosis
Superior vena cava
GooseNeck Snare
20
16
Bard
28
Rectal cancer
Injection difficulty
Male/32
1074
18
Detached anastomosis
Superior vena cava
GooseNeck Snare
20
24
Bard
29
Gastric cancer
Injection difficulty
Male/49
1317
11
Detached anastomosis
Right atrium
GooseNeck Snare
16
26
Bard
30
Lung cancer
Chest tightness
Female/78
229
5
POS
Right atrium
Pig-tail Catheter & GooseNeck Snare
11
47
Bard
31
Breast cancer
Injection difficulty
Female/52
819
13
Detached anastomosis
Right ventricle
GooseNeck Snare
19
38
Bard
32
Rectal cancer
Injection difficulty
Male/68
248
20
Detached anastomosis
Superior vena cava
GooseNeck Snare
18
27
Bard
33
Breast cancer
No
Female/34
837
45
Detached anastomosis
Inferior vena cava
GooseNeck Snare
17
12
Bard
34
Breast cancer
Injection difficulty
Female/47
357
16
POS
Right ventricle
Pig-tail Catheter & GooseNeck Snare
12
28
Bard
35
Colon cancer
Injection difficulty
Male/53
413
11
POS
Right atrium
GooseNeck Snare
9
29
Bard
POS:pinch-off syndrome, catheter fracture in the subclavian region caused by continuous pressure from the clavicle and the first rib.
Clinical data of patients with catheter migration.Clinical data of patients with catheter fracture.POS:pinch-off syndrome, catheter fracture in the subclavian region caused by continuous pressure from the clavicle and the first rib.The 23 cases of migrated catheters were comprised of 5 cases of elastic retraction and 18 cases of tip migration. Among the 18 patients who suffered tip migration, and 2 received implantation in other hospitals; therefore, the incidence of migration in our hospital was 1.2% (21/1716). Of the 5 cases of elastic retraction, routine chest radiography of 1 case showed that the proximal end of the catheter was twisted and the tip was in the upper segment of the superior vena cava. However, as the PCS was functioning effectively, adjustments were not performed for this patient. The other 4 cases of elastic retraction presented with difficulties in injection, and the radiograph showed that the segment of the catheter in the subclavian vein was folded and the tip was located in the subclavian vein. Repositioning was successfully performed in 1 case through a pigtail catheter and in another 2 cases using a gooseneck snare. In the last case, the entire port system was removed percutaneously, and a new PCS was implanted in the contralateral subclavian vein, after relocation using both a pigtail catheter and gooseneck snare failed. Meanwhile, in all 18 patients diagnosed with tip migration, the tips of the catheters were displaced into the internal jugular vein. Clinical manifestations included injection difficulty in 9 cases, dizziness and headache in 2, distending pain in the subclavian region in 1, shoulder and neck discomfort in 1, and no symptoms in 5. All migrated catheters were adjusted intravascularly through percutaneous punctures, and the average operation time was 19.09 ± 9.82 min. Repositioning was successfully performed in 9 patients with only a homemade pigtail catheter (Fig. 1A–F), in 4 patients with only a cobra catheter, and in 5 patients with a gooseneck snare after the use of both pigtail and Cobra catheters failed. Among the 5 patients who underwent repositioning using a gooseneck snare, 1 presented with fibrin sheath adhesion between the distal end of the catheter tip and the wall of the internal jugular vein. The distal end of the catheter was not exposed after relocation, and both the pigtail and cobra catheters failed in retrieval. The gooseneck snare initially failed to hook the distal end. Repeated pulling with a pigtail catheter exposed the free end, which was then successfully hooked by the gooseneck snare and repositioned after several attempts. The success rate of percutaneous vascular adjustment was 91.3% (21/23).
Fig. 1
Surgery procedure for a 61-year-old male patient with rectal cancer, diagnosed with catheter migration 3 months after PCS implantation.
A, B: Cutting the tail of a 5 F Pig-tail Catheter into a semi-circular shape with a sharp knife.
C: X-ray scan showing that the catheter tip was displaced to the right internal jugular vein.
D, E: X-ray scan showing insertion of the homemade Pig-tail Catheter through the inferior vena cava, and hooking of the displaced catheter with its tail.
F: X-ray scan showing that the catheter tip was relocated to the superior vena cava.
Surgery procedure for a 61-year-old male patient with rectal cancer, diagnosed with catheter migration 3 months after PCS implantation.A, B: Cutting the tail of a 5 F Pig-tail Catheter into a semi-circular shape with a sharp knife.C: X-ray scan showing that the catheter tip was displaced to the right internal jugular vein.D, E: X-ray scan showing insertion of the homemade Pig-tail Catheter through the inferior vena cava, and hooking of the displaced catheter with its tail.F: X-ray scan showing that the catheter tip was relocated to the superior vena cava.The fracture site was the superior vena cava in 13 cases, the right atrium in 9, the right ventricle in 7, the pulmonary artery in 4, and the inferior vena cava in 2. Clinical manifestations included injection difficulty in 16 cases, chest tightness in 8, and no symptoms in 11. Among the 35 patients with catheter fractures, 14 (40%) developed pinch-off syndrome (POS, catheter fracture in the subclavian region caused by continuous pressure from the clavicle and the first rib), whereas the other 21 cases (60%) developed disconnection between the port and catheter. Four of the fracture patients received implantation in other hospitals; therefore, the incidence rate of fractures in our hospital was 1.8% (31/1716). All ruptured catheters were removed by percutaneous puncture with a 100% success rate (35/35). The average operation time was 26.51 ± 9.46 min. The mean length of the removed catheters was 15 ± 4 cm (range: 8–20 cm). Both the gooseneck snare and pigtail catheter were used in 9 cases, as the pigtail catheter was needed to expose the free end. Meanwhile, in 25 patients, the gooseneck snare was the only device used to remove the fractured catheters (Fig. 2A–C). The distal end of the fractured catheter was near the right atrium wall in 1 patient and the sheath of the gooseneck snare could not reach the proximal end of the free catheter. Therefore, a 6 F JR4.0 angiographic catheter and a 6 F gooseneck snare were used together to hook the free end, enabling successful removal. An EN snare loop (Angiotech Medical Device Technologies, South Jordan, Utah, USA) was used to treat 1 patient. Entry of the external sheath of the loop to the pulmonary trunk was challenging because the fracture was in the middle of the catheter (at the junction between the first rib and the clavicle). The distal end of the catheter was located in the middle branches of the right pulmonary artery, and the proximal end was located in the right main pulmonary artery close to the bifurcation site and attached to the vessel wall. After several unsuccessful attempts, a 6 F JL4.0 guide catheter and an EN snare loop were jointly used, and the fractured catheter was successfully hooked (Fig. 3A–E). All PCSs with fractured catheters were removed, and new systems were implanted in the contralateral subclavian region in patients who still needed central venous access.
Fig. 2
Surgery procedure of a 63-year-old male patient with lung cancer, diagnosed with catheter fracture 15 months after PCS implantation.
A: Chest radiograph showing catheter fracture with one free end in the right ventricle.
B: Hooked of the free end with GooseNeck Sanre.
C: Fracture site was at the anastomosis of the port and the catheter.
Fig. 3
Surgery procedure for a 38-year-old female patient with breast cancer, diagnosed with catheter fracture 19 months after PCS implantation.
A: CT scan showing that the distal end of the catheter was in the middle branch of the right main pulmonary artery, whereas the proximal end was in the right main pulmonary artery.
B: Chest X-ray scan showing the fracture site was in the middle of the catheter at the conjunction of the first right rib and the clavicle.
C: Hooking of the fractured catheter using EN Snare Loop.
D: Removal of the fractured catheter from the femoral vein.
E: Location of the fracture in the middle part of the catheter.
Surgery procedure of a 63-year-old male patient with lung cancer, diagnosed with catheter fracture 15 months after PCS implantation.A: Chest radiograph showing catheter fracture with one free end in the right ventricle.B: Hooked of the free end with GooseNeck Sanre.C: Fracture site was at the anastomosis of the port and the catheter.Surgery procedure for a 38-year-old female patient with breast cancer, diagnosed with catheter fracture 19 months after PCS implantation.A: CT scan showing that the distal end of the catheter was in the middle branch of the right main pulmonary artery, whereas the proximal end was in the right main pulmonary artery.B: Chest X-ray scan showing the fracture site was in the middle of the catheter at the conjunction of the first right rib and the clavicle.C: Hooking of the fractured catheter using EN Snare Loop.D: Removal of the fractured catheter from the femoral vein.E: Location of the fracture in the middle part of the catheter.All the patients were treated using the right femoral venous approach. The detailed patient data are presented in Table 1, Table 2. The 23 patients with catheter migration did not present with surgery-related complications. Among the 35 cases of catheter fracture, 3 developed transient atrial arrhythmias or ventricular arrhythmias during surgery; however, the complications were not severe, and there was no need for intervention with drugs. No surgery-related complications were observed either immediately or during the 1-month follow-up period.
Discussion
Subcutaneous implantation of a PCS was introduced by Niederhuber in 1982. Currently, it is widely used in the clinical treatment of patients requiring long-term intravenous infusion. Studies have reported that approximately 5 million PCSs are implanted every year in the United States. The interventional radiology center of our hospital has implanted 1716 PCSs into subclavian veins since its establishment in 2009. This figure has been increasing yearly.Spontaneous catheter migration is rare. The incidence rate of catheter migration within the follow-up period of 5–9 months is 0.9%–1.8%., The incidence of catheter migration in the current study was 1.2% (21/1716), which is consistent with the previous findings. The ideal site of the catheter tip is between the lower third of the superior vena cava and the upper third of the right atrium. The PCS should not be used when the tip becomes displaced to small veins, the brachiocephalic vein, and the subclavian vein, as this increases the risk of venous thrombosis. Additionally, the risk of injection into the veins is high, leading to severe complications.,11, 12, 13 Therefore, it is necessary to adjust the position of the catheter or remove the PCS when migration occurs.A previously reported technique of high-pressure injection of normal saline was initially used in the current hospital to reposition migrated catheters through irrigation. However, this technique is ineffective, and can lead to rupture and separation of the connection between the port and the catheter. Adjustment with catheters or gooseneck snares is more effective and feasible than reimplantation or replacement of PCS. Gebauer et al. reported the use of catheters and a gooseneck snare in adjusting migrated catheters through the femoral vein with a 93% success rate (27/29). In the current study, the success rate of percutaneous vascular adjustment was 91.3% (21/23), which is similar to that reported in previous studies.The gooseneck snare is used when the free end of the migrated catheter tip is exposed so that it can be hooked and relocated to the superior vena cava, which is similar to the removal of fractured catheters. Bessoud et al. reported that the success rates of other surgical tools were very low if the gooseneck snare had failed to adjust the position of the migrated catheter. In 1 patient in this study, the catheter tip was located in the internal jugular vein, and the formation of fibrin sheaths led to a firm attachment of the distal end to the vein. The migrated catheter was successfully hooked using a pigtail catheter. However, pulling multiple times did not change the position of the catheter, which can be attributed to the soft tail end of the guide catheter or endothelialization between the migrated catheter and the vessel wall. Successful adjustment was achieved by hooking the freed end of the tip using a gooseneck snare.The cause of spontaneous catheter displacement is not fully understood. Potential causes include a short indwelling catheter, vigorous movement and pulling of the shoulders and upper arms, change in chest pressure resulting from severe cough and frequent vomiting, and migration and pulling off the subcutaneous port.,, Therefore, the solution involves minimizing the number of high-risk factors during PCS implantation.Catheter fracture was initially reported to cause the formation of embolisms in the distal end in 1954, and several clinical studies reported that the incidence of catheter fracture ranges from 0 to 4.1%,20, 21, 22, 23 whereas the death rate or rate of severe complications was approximately 71%., Therefore, intervention should be carried out immediately if a fracture occurs, and intravascular removal of fractured catheters through percutaneous puncture is the conventional technique. The success rate of PCS catheter removal by percutaneous puncture is 90–97.8%., In the current study, the incidence of catheter fracture was 1.8% (31/1716), which is consistent with previous reports. Notably, all fractured catheters were successfully removed through percutaneous punctures, with a 100% success rate (35/35), and no surgery-related complications were observed.Several interventional devices are currently used for intravascular removal. The most commonly used device in previous studies was the gooseneck snare., In the present study, removal of fractured catheters was performed in 25 patients using the gooseneck snare, while the EN snare loop was used in only 1 patient. There was at least 1 free end in 26 patients, which made it easier for the 2 devices to hook the fractured catheter. Nine patients presented with fractured catheters that did not have free ends or the free ends were not exposed. In these cases, the gooseneck snare alone failed to hook the free ends of the fractured catheter; therefore, a pigtail catheter and a gooseneck snare were jointly used. The tail of the pigtail catheter was first used to hook the fractured catheter and expose the free ends by pulling them, and then the gooseneck snare was used to successfully hook the free ends. During the removal of the fractured catheters in 2 cases, it was observed that the fixed angle made hooking of the free ends challenging owing to the rigid external sheath of the snare. The external casing is replaced by an angiographic or guiding catheter with a stable angle, and the catheter is continuously checked to successfully remove the broken catheter.The position of fractured foreign bodies in the cardiovascular system depends on gravity, their entry path, length, stiffness, and blood flow pattern in the blood vessels and heart cavities.,, Bessoud et al. reported that the most common site of embolisms was the pulmonary artery. Cheng et al. reported that most embolisms were located between the inferior vena cava and the right atrium or between the right atrium and the superior vena cava, which is consistent with the findings of Koseoglu et al. and Liu et al. In the current study, the most common location of fractured catheters was at the junction of the superior vena cava and the right atrium.Possible causes of catheter fracture include the use of improper equipment leading to damage to the catheter, use of small syringes leading to increased pressure inside the catheter, continuous pressure on the catheter between the clavicle and the first rib (POS), improper connection between the port and the catheter, deformity of the connection between the port and the catheter, and improper positioning of the catheter.,, In the current study, there were 14 cases of POS (14/35, 40%) and 21 cases of detachment of the proximal end from the port (21/35, 60%). Among the 35 patients with catheter fractures, 31 were implanted in our hospital, including 27 patients for whom the Bard Access System (27/1401, 1.9%) (BD, Franklin Lakes, NJ, USA) was used, 1 patient for whom the B. Braun Access System(B. Braun Medical Inc.,Bethlehem, PA,USA) was used (1/63, 1.6%), and 3 patients for whom the MedCOMP Access System(Medical Components, Inc.,Harleysville, PA,USA) (3/252, 1.2%) was used. Lin et al. reported that the risk of fracture may be related to POS and PCS design. Shoulder movement exerts a pulling force on the catheter (“pinching” the catheter), leading to continuous pulling or compression of the connection between the catheter and the port, resulting in pressure on the catheter or fracture of the segment being pulled. Liu et al. reported that the subclavian vein should be punctured at the junction of the middle and outer thirds of the clavicle to avoid POS. Furthermore, the port and the catheter should be firmly connected to avoid an angle between the 2 PCS parts or deformation of their connection. Moreover, follow-up chest radiography should be performed routinely and over a long follow-up period, and the system should be removed immediately if it is no longer in use.
Conclusions
The incidence of abnormal catheter positioning is low. Once an abnormality is detected, the use of the system should be stopped immediately, and adjustment or removal should be performed with effective methods. Intravascular techniques for catheter adjustment or removal are safe, efficient, and minimally invasive.
Author statement
Yong Li: Conceptualization, Methodology, Software, Data curation, Writing- Original draft preparation, Visualization, Investigation. Jianxi Guo: Conceptualization, Methodology, Software, Data curation, Writing- Original draft preparation, Visualization, Investigation, Writing- Reviewing and Editing. Yanfang Zhang: Validation, Writing- Reviewing and Editing. Jian Kong: Conceptualization, Methodology, Data curation, Writing- Original draft preparation, Visualization, Investigation, Supervision, Validation, Writing- Reviewing and Editing.
Conflicts of interest and source of funding
There are no conflicts of interest. It has received honoraria from Key-Area Research and Development Program of Guangdong Province, China (2020B010165004).