Xingen Wu1, Andrew McDonald1, Sui Shen1, Dennis Stanley1, Richard Popple1, Samuel Marcrom1, Robert Kim1.
Abstract
Purpose This study aimed to explore the relationship between applicator surface dose and 5 mm-depth dose and to optimize both locations simultaneously for three most used cylinder sizes (2.5, 3.0, and 3.5 cm in diameter) in treating patients with endometrial adenocarcinoma. Materials and methods A total of 216 plans were created for each dose level and applicator size. For each dose level, four plans were created with single or double prescription doses. For plans with double prescription doses, the dose constraints were applied to all those points on the surface and 5 mm depth and optimize the two sites simultaneously. Results A dose table between surface and 5 mm depth and its fifth order polynomial mapping functions were established for each applicator size, so any prescribed dose at one site can find the prescription dose on the other site in optimization on both locations. For plans with a 5 mm-depth prescription, the maximum dose on the surface can be reduced from 145% to 133% if the surface prescription dose is also used; for plans with surface dose prescription, the minimum dose and mean dose can be improved by 2% if 5 mm-depth dose prescription is also used in optimization. Conclusion Dose table and their mapping functions between surface prescription dose and their corresponding 5 mm-depth doses were created. A new optimization method that uses two prescription doses on both surface and 5 mm-depth sites was proposed to reduce the hot dose on the surface and improve the cold dose at 5 mm depth.Entities:
Keywords: dose optimization; dose prescription; endometrial adenocarcinoma; hdr brachytherapy; vaginal cylinder
Year: 2022 PMID: 35911294 PMCID: PMC9312305 DOI: 10.7759/cureus.26303
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Figure 1Single applicator and three surface reference lines (color: tomato) and three 5 mm-depth reference lines (color: cyan)
A prescription dose table between surface and 5 mm depth for three cylinder applicator sizes
| 2.5 cm cylinder | 3.0 cm cylinder | 3.5 cm cylinder | |||
| 5 mm depth (cGy) | Surface (cGy) | 5 mm depth (cGy) | Surface (cGy) | 5 mm depth (cGy) | Surface (cGy) |
| 300 | 500 | 340 | 500 | 360 | 500 |
| 400 | 660 | 400 | 600 | 400 | 550 |
| 430 | 700 | 430 | 640 | 430 | 590 |
| 500 | 800 | 500 | 730 | 500 | 710 |
| 550 | 880 | 550 | 830 | 550 | 810 |
| 600 | 950 | 600 | 910 | 600 | 890 |
| 700 | 1100 | 700 | 1050 | 700 | 1000 |
| 800 | 1250 | 800 | 1200 | 800 | 1150 |
| 850 | 1350 | 850 | 1300 | 850 | 1250 |
Coefficients for the fifth polynomial mapping functions
| Coefficients | From 5 mm to surface | From surface to 5 mm | ||||
| 2.5 cm | 3.0 cm | 3.5 cm | 2.5 cm | 3.0 cm | 3.5 cm | |
| a | 0.004975 | 0.007740 | -0.004073 | -0.000316 | -0.000461 | 0.000448 |
| b | -0.140437 | -0.219375 | 0.158500 | 0.014380 | 0.019668 | -0.023634 |
| c | 1.550614 | 2.427377 | -2.295163 | -0.256824 | -0.326899 | 0.472499 |
| d | -8.351723 | -13.101861 | 15.633410 | 2.245143 | 2.645898 | -4.488110 |
| e | 23.370814 | 36.031033 | -48.875896 | -8.909707 | -9.778320 | 20.930508 |
| f | -21.646311 | -35.645704 | 61.292976 | 15.522264 | 16.149188 | -34.560913 |
Figure 2Six dose mapping functions between surface dose and 5 mm depth dose for three different cylinder applicator sizes
(a) 2.5 cm cylinder dose mapping from surface to 5 mm depth; (b) 2.5 cm cylinder dose mapping from 5 mm depth to surface; (c) 3.0 cm cylinder dose mapping from surface to 5 mm depth; (d) 3.0 cm cylinder dose mapping from 5 mm depth to surface; (e) 3.5 cm cylinder dose mapping from surface to 5 mm depth; (f) 3.5 cm cylinder dose mapping from 5 mm depth to surface.
Figure 3Dose distribution comparison among the four plans
(a) P1_5mm; (b) P2_5mm&Sur; (c) P3_Sur; (d) P4_Sur&5mm. The two prescription doses are 2100 cGy (700 cGy per fraction) at 5 mm depth and 3150 cGy (1050 cGy per fraction) on the surface and the cylinder size is 3 cm.
P1_5mm: Plan #1 prescribed only at 5 mm depth; P2_5mm&Sur: Plan #2 prescribed at 5 mm depth and on the surface simultaneously; P3_Sur: Plan #3 prescribed only on the surface; P4_Sur&5mm: Plan #4 prescribed on the surface and at 5 mm depth simultaneously.
Dosimetry comparison of surface dose and 5 mm depth dose from 108 plans along four prescription methods
P1_5mm: Plan #1 prescribed only at 5 mm depth; P2_5mm&Sur: Plan #2 prescribed at 5 mm depth and on the surface simultaneously; P3_Sur: Plan #3 prescribed only on the surface; P4_Sur&5mm: Plan #4 prescribed on the surface and at 5 mm depth simultaneously.
σ: standard deviation
| Surface dose | 5 mm depth dose | |||||
| Maximum (%)±σ | Minimum (%)±σ | Mean (%)±σ | Maximum (%)±σ | Minimum (%)±σ | Mean (%)±σ | |
| P1_5mm | 145.24±5.33 | 93.16±2.92 | 108.28±4.63 | 112.16±0.58 | 92.23±0.74 | 100.66±0.19 |
| P2_5mm&Sur | 133.40±3.81 | 94.32±2.45 | 105.70±3.53 | 105.53±1.66 | 90.83±0.84 | 98.80±1.22 |
| P3_Sur | 113.99±2.66 | 89.59±1.59 | 100.36±0.65 | 101.65±3.04 | 86.44±3.12 | 94.65±3.48 |
| P4_Sur&5mm | 117.16±3.15 | 92.08±2.68 | 101.64±1.24 | 101.85±2.08 | 87.94±2.49 | 95.82±2.74 |
Figure 4Dose profiles on the surface lines and at 5 mm depth lines sorted from highest to lowest in the four plans: 600 cGy was prescribed to 5 mm depth and 950 cGy to the surface with cylinder size 2.5 cm in diameter
P1_5mm: Plan #1 prescribed only at 5 mm depth; P2_5mm&Sur: Plan #2 prescribed at 5 mm depth and on the surface simultaneously; P3_Sur: Plan #3 prescribed only on the surface; P4_Sur&5mm: Plan #4 prescribed on the surface and at 5 mm depth simultaneously