| Literature DB >> 35898835 |
Kazuya Miyaguchi1,2, Masafumi Mizuide1, Yuki Tanisaka1, Akashi Fujita1, Ryuhei Jinushi1, Katsuda Hiromune1, Tomoya Ogawa1, Yoichi Saito1, Tomoaki Tashima1, Yumi Mashimo1, Hiroyuki Imaeda2, Shomei Ryozawa1.
Abstract
Objectives: Understanding the exact morphology of the bile duct opening is important for determining the success of bile duct cannulation. Texture and color enhancement imaging (TXI) has been reported to enhance slight changes in color tone and structure that are difficult to see with white light imaging. This study investigated whether TXI mode1 could improve papillary recognition by trainees inexperienced in endoscopic retrograde cholangiopancreatography.Entities:
Keywords: biliary cannulation; endoscopists; papilla of Vater; texture and color enhancement imaging; white light imaging
Year: 2022 PMID: 35898835 PMCID: PMC9307722 DOI: 10.1002/deo2.125
Source DB: PubMed Journal: DEN open ISSN: 2692-4609
FIGURE 1Algorithm for texture and color enhancement imaging. This algorithm outputs two types of texture and color enhancement images: mode2, which enhances texture and brightness, and mode1, which enhances texture, brightness, and color tone.
FIGURE 2Enrollment flow chart We evaluated the procedure of biliary cannulation according to the Inomata classification (six types of evaluation) and divided papilla into two groups to determine their distinctive qualities: Separate type group and Septal type group. Separate type groups: Papilla₋I ( ) and Papilla‐G ( ). Septal type groups: Papilla‐A ( ), Papilla‐U ( ) Papilla‐V ( ), and Papilla‐LO ( .
FIGURE 3Separate type groups (Papilla‐I, G) Upper row (I: white light imaging, i: texture and color enhancement imaging mode1); Papilla‐I Lower row (G: white light imaging, g: texture and color enhancement imaging mode1) ; Papilla‐G
FIGURE 4Septal type groups (Papilla‐LO, A) Upper row (LO: white light imaging; lo: texture and color enhancement imaging mode1); Papilla‐LO Lower row (A: white light imaging; a: texture and color enhancement imaging mode1); Papilla‐A
FIGURE 5Septal type groups (Papilla‐U, V) Upper row (U: white light imaging; lo, u: texture and color enhancement imaging mode1); Papilla‐U Lower row (V: white light imaging; v: texture and colour enhancement imaging mode1); Papilla‐V
Agreement among the four raters in the trainee
| Inter‐subject agreement | |||||||
|---|---|---|---|---|---|---|---|
| Evaluation | Mode | Endoscopists | n | Kappa | 95% CI | ||
| Two groups of papilla of Vater | |||||||
| WLI | |||||||
| Trainee | 4 | 0.346 | 0.202 | , | 0.489 | ||
| TXI | |||||||
| Trainee | 4 | 0.754 | 0.610 | , | 0.898 | ||
Kappa: Fleiss' kappa; ‡95 % CI: 95 % confidence interval
2 groups of papilla: 0_[I, G]; 1_[A, V, U, LO]
WLI: White‐light imaging
TXI: Texture and color enhancement imaging
(a) Inter‐method comparison of agreement rates for two‐groups evaluation
| Match rate | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| WLI | TXI |
| |||||||
| % | 95% CI | % | 95% CI | WLI vs. TXI | |||||
| Trainee A | 76.67 | 57.72 | , | 90.07 | 96.67 | 82.78 | , | 99.92 |
|
| Trainee B | 86.67 | 69.28 | , | 96.24 | 93.33 | 77.93 | , | 99.18 | 0.625 |
| Trainee C | 80.00 | 61.43 | , | 92.29 | 96.67 | 82.78 | , | 99.92 | 0.125 |
| Trainee D | 83.33 | 65.28 | , | 94.36 | 93.33 | 77.93 | , | 99.18 | 0.375 |
p‐value: McNemar test
WLI: White‐light imaging
TXI: Texture and color enhancement imaging
95% CI: 95% confidence interval
Match rate: Percentage of judgments in agreement
n = 30 (Cases in which uniform answers were obtained from the experts)
(b) Comparison of kappa coefficients for two‐groups evaluation between methods
| Kappa coefficient | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| WLI | TXI | ||||||||
| κ | 95% CI | κ | 95% CI |
| |||||
| Trainee A | 0.34 | −0.02 | , | 0.70 | 0.92 | 0.78 | , | 1.00 |
|
| Trainee B | 0.68 | 0.40 | , | 0.97 | 0.86 | 0.67 | , | 1.00 | 0.320 |
| Trainee C | 0.45 | 0.10 | , | 0.81 | 0.92 | 0.78 | , | 1.00 |
|
| Trainee D | 0.53 | 0.20 | , | 0.86 | 0.85 | 0.66 | , | 1.00 | 0.147 |
p‐value: Kappa coefficient comparison test
WLI: White ‐light imaging
TXI: Texture and color enhancement imaging
κ: Cohen's kappa coefficient
95% CI: 95% confidence interval
n = 30 (Cases in which uniform answers were obtained from experts)
Endoscopic retrograde cholangiopancreatography (ERCP) with texture and color enhancement imaging (TXI)
| Total | ||||||
|---|---|---|---|---|---|---|
|
|
| % | 95% CI | |||
| Cannulation success rate | 31 | 30 | 96.77 | 83.30 | , | 99.92 |
| No trainee change rate | 31 | 20 | 64.52 | 45.37 | , | 80.77 |
Abbreviations: CI, confidence interval; ERCP, endoscopic retrograde cholangiopancreatography; TXI, texture and color enhancement imaging.