| Literature DB >> 35877619 |
Yazan Al-Hayek1,2, Kelly Spuur1, Rob Davidson3, Christopher Hayre1, Xiaoming Zheng1.
Abstract
BACKGROUND: This study investigates the effects of vertical off-centring, localiser direction, tube voltage, and phantom positioning (supine and prone) on computed tomography (CT) numbers and radiation dose.Entities:
Keywords: CT number accuracy; localizer direction; phantom positioning; radiation dose; tube voltage; vertical off-centring
Year: 2022 PMID: 35877619 PMCID: PMC9316438 DOI: 10.3390/jimaging8070175
Source DB: PubMed Journal: J Imaging ISSN: 2313-433X
Figure 1The phantom trunk is pictured supine with axial CT slices of the upper thorax (A), mid thorax (B), and abdomen (C). Two ROIs (yellow circle) were used to calculate the mean CT number measurements.
The absolute computed tomography (CT) number difference of the ROI between the 100 mm off-centring points above and below the gantry iso-centre for 0° and 180° localisers and supine and prone phantom positions.
| 80 kVp | 120 kVp | 140 kVp | |||||
|---|---|---|---|---|---|---|---|
| S.ROI a | I.ROI b | S.ROI | I.ROI | S.ROI | I.ROI | ||
| 0° localiser/Supine position | Upper Thorax | 29 | 10.8 | 4.3 | 33.6 | 0.5 | 27.9 |
| Mid Thorax | 6.1 | 28.3 | 3.5 | 26.0 | 4.2 | 29.1 | |
| Abdomen | 25.6 | 23.8 | 22.5 | 21.7 | 18.9 | 25.0 | |
| 180° localiser/Supine position | Upper Thorax | 8.7 | 27.6 | 1.4 | 23.0 | 1.5 | 21.0 |
| Mid Thorax | 2.8 | 43.3 | 2.5 | 31.4 | 1.6 | 29.3 | |
| Abdomen | 14.3 | 23.5 | 12.0 | 24.8 | 12.4 | 23.0 | |
| 0° localiser/Prone position | Upper Thorax | 17.7 | 0.9 | 6 | 4.0 | 19.5 | 7.4 |
| Mid Thorax | 33.1 | 0.1 | 23.4 | 0.1 | 21.0 | 10.4 | |
| Abdomen | 22.5 | 30.8 | 18.7 | 23.3 | 17.4 | 28.5 | |
| 180° localiser/Prone position | Upper Thorax | 10.9 | 7.0 | 17.2 | 2.9 | 12.2 | 4.0 |
| Mid Thorax | 15.9 | 15.4 | 14.8 | 13.9 | 13.5 | 9.9 | |
| Abdomen | 14.6 | 29.0 | 11.7 | 27.2 | 12.4 | 32.1 | |
a Superior ROI, b Inferior ROI.
Figure 2The CT number change as a function of CT table vertical off-centring for the superior and inferior ROIs for the abdominal section at 120 kVp.
Figure 3The difference in the mean CT number between the superior and inferior ROIs as a function of vertical off-centring for the upper thorax, mid thorax and abdominal phantom sections, at 120 kVp and 0° localiser.
Figure 4The difference in the mean CT number in the abdominal section between the superior and inferior ROIs as a function of vertical off-centring, phantom position, and localiser direction, at 120 kVp.
Figure 5The CT number change of superior and inferior ROIs in the abdominal scans as a function of phantom vertical off-centring using different tube voltages.
Figure 6The effect of CT table off-centring on CTDIvol with different phantom positions and localiser directions at 120 kVp.