Effect of CT Localizer Radiographs on Radiation Dose Associated With Automatic Tube Current Modulation: A Multivendor Study.

Fabio Paolicchi1, Luca Bastiani2, Jacopo Negri3, Davide Caramella4.   

Abstract

OBJECTIVES: To assess the influence of the CT localizer radiograph on the automatic tube current modulation system of 7 CT scanners produced by 4 different CT manufacturers.
METHODS: The influence of the localizer orientation, table height, tube current and tube potential values on the radiation dose of the related CT scan were evaluated. Images were acquired by using an anthropomorphic phantom positioned in the CT gantry isocenter as well as from -6 cm to +6 cm vertically to the isocenter.
RESULTS: Vertical movement of the CT table height affected the radiation dose in all scanners using anterior-posterior or a posterior-anterior localizer orientation albeit differently, depending on the manufacturer; only in 1/7 scanner no influence was observed. The latero-lateral localizer orientation proved to be more effective in limiting the influence of the vertical miscentering in all scanners. Changing localizer's tube voltage influenced the scan radiation dose in scanners produced by two manufacturers, while no significant effect was observed in scanners produced by the other two manufacturers. No significant dose variation was observed in 6/7 scanners when changing the localizer's tube current.
CONCLUSION: Localizer radiograph shows a significant influence on the radiation exposure but with different outcomes depending on the manufacturer of the CT scanner. Radiologists and radiographers should have a thorough understanding of these differences to assure patients the best examination in terms of radiation dose and image quality.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2019        PMID: 30704769     DOI: 10.1067/j.cpradiol.2018.12.010

Source DB:  PubMed          Journal:  Curr Probl Diagn Radiol        ISSN: 0363-0188


  4 in total

1.  Explore the Value of Dual Source Computer Tomography Automatic Tube Current Regulation in Reducing the Radiation Dose of CTA in Lower Extremity Vessels.

Authors:  Xin Hu; Yi Yu
Journal:  Front Surg       Date:  2022-05-03

2.  0° vs. 180° CT localiser: The effect of vertical off-centring, phantom positioning and tube voltage on dose optimisation in multidetector computed tomography.

Authors:  Yazan Al-Hayek; Xiaoming Zheng; Rob Davidson; Christopher Hayre; Dana Al-Mousa; Campbell Finlay; Kelly Spuur
Journal:  J Med Radiat Sci       Date:  2021-08-17

3.  The Impacts of Vertical Off-Centring, Localiser Direction, Phantom Positioning and Tube Voltage on CT Number Accuracy: An Experimental Study.

Authors:  Yazan Al-Hayek; Kelly Spuur; Rob Davidson; Christopher Hayre; Xiaoming Zheng
Journal:  J Imaging       Date:  2022-06-21

Review 4.  Best practice for the nuclear medicine technologist in CT-based attenuation correction and calcium score for nuclear cardiology.

Authors:  Luca Camoni; Andrea Santos; Marieclaire Attard; Marius Ovidiu Mada; Agata Karolina Pietrzak; Sonja Rac; Sebastijan Rep; Christelle Terwinghe; Pedro Fragoso Costa
Journal:  Eur J Hybrid Imaging       Date:  2020-07-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.