| Literature DB >> 35876743 |
Rebecca Rojansky1, Seshi R Sompuram2, Ellen Gomulia1, Yasodha Natkunam1, Megan L Troxell1, Sebastian Fernandez-Pol1.
Abstract
Assessment of automated immunohistochemical staining platform performance is largely limited to the visual evaluation of individual slides by trained personnel. Quantitative assessment of stain intensity is not typically performed. Here we describe our experience with 2 quantitative strategies that were instrumental in root cause investigations performed to identify the sources of suboptimal staining quality (decreased stain intensity and increased variability). In addition, these tools were utilized as adjuncts in validation of a new immunohistochemical staining instrument. The novel methods utilized in the investigation include quantitative assessment of whole slide images (WSI) and commercially available quantitative calibrators. Over the course of ~13 months, these methods helped to identify and verify correction of 2 sources of suboptimal staining. One root cause of suboptimal staining was insufficient/variable power delivery from our building's electrical circuit. This led us to use uninterruptible power managers for all automated immunostainer instruments, which restored expected stain intensity and consistency. Later, we encountered one instrument that, despite passing all vendor quality control checks and not showing error alerts was suspected of yielding suboptimal stain quality. WSI analysis and quantitative calibrators provided a clear evidence that proved critical in confirming the pathologists' visual impressions. This led to the replacement of the instrument, which was then validated using a combination of standard validation metrics supplemented by WSI analysis and quantitative calibrators. These root cause analyses document 2 variables that are critical in producing optimal immunohistochemical stain results and also provide real-world examples of how the application of quantitative tools to measure automated immunohistochemical stain output can provide a greater objectivity when assessing immunohistochemical stain quality.Entities:
Mesh:
Substances:
Year: 2022 PMID: 35876743 PMCID: PMC9345521 DOI: 10.1097/PAI.0000000000001045
Source DB: PubMed Journal: Appl Immunohistochem Mol Morphol ISSN: 1533-4058
FIGURE 1Timeline of events described in this report. Time is not to scale. ALK indicates anaplastic lymphoma kinase; UPM, uninterruptible power manager.
Summary of Variables Tested During Troubleshooting and Their Impact on Stain Intensity and Consistency
| Variables Evaluated/Tested | Impact |
|---|---|
| Antibody lot | No clear improvement |
| Antibody concentration | Mild increase in intensity with higher concentration |
| New slide trays (to control for tray warping) | No clear improvement |
| New covertiles | No clear improvement |
| Instrument calibration | No clear improvement |
| Preventative maintenance performed | No clear improvement |
| UPM | Improved intensity and decreased variability |
| Additional electrical circuit installed | Stain quality no longer impacted by UPM under test conditions |
UPM indicates uninterruptible power manager.
FIGURE 2Impact of uninterruptible power managers (UPMs) on stain intensity and variability. Digital image analysis of ALK-stained control tissue supports the hypothesis that, before electrical circuit upgrades, in the absence of UPMs immunohistochemical stain intensity was reduced. The “Average of Max Intensity” represents the average of the maximum pixel intensity detected for each segmented cell. Each point represents the results of analyzing one tissue control (with one tissue control per slide). The no-UPM condition was performed as a dedicated test run (no patient samples). The UPM runs were cases that were run in the process of routine clinical care with on slide control material that was identical or similar to the control material used in the no-UPM condition. Details of the digital image analysis are provided in the Materials and Methods section. ALK indicates anaplastic lymphoma kinase.
FIGURE 3Whole slide images analysis and quantitative calibrators confirm persistent instrument malfunction. A, Whole slide image analysis shows that ALK stain intensity was lower for instrument #108 even when operating with a UPM. B and C, PD-L1 quantitative calibrators confirm that one instrument (designated #108) was yielding suboptimal staining (staining was performed with UPM in place). The E1L3N antibody was used as described in the Materials and Methods. B, Shows microscopic images are of calibrator beads stained on instrument 108 (defective instrument) and a comparator instrument 109. The small beads in both images are optical reference microbeads that are used for image quantification. The color of the reference microbeads is constant and not affected by staining. The larger circles are glass beads with a known number of PD-L1 peptides per bead. In these images, beads with 643,000 peptides per bead are shown. The E1L3N antibody–staining protocol shows dimmer staining when run on the defective instrument 108 (left image) when compared with a comparator instrument 109 (right image). The stain intensity seen with instrument 109 was visually indistinguishable from the other comparator instruments 195 and 458. C, Shows the analytic response curves of PD-L1 bead calibrators run on the 5 instruments in our laboratory. The legend indicates the instruments run, which are designated as 109, 108, 458, 211, and 195. The analytic response curve for instrument 108 clearly differs from those of the comparator instrument. D, The new instrument #111 shows E1L3N staining that is indistinguishable from comparator instrument #195. UPM did not impact staining intensity at the time of this run, which was performed after electrical circuit upgrades. The legend indicates instrument number (111 or 195) followed by tray position (eg, 1-7, 2-4, etc.) and then indicates the presence or absence of UPM. ALK indicates anaplastic lymphoma kinase.
FIGURE 4Within run variability seen with the whole slide image analysis method (A) and with bead control (E1L3N antibody with PD-L1 control beads) (B). All runs shown are after upgrade of the electrical circuits. In (A), whole slide image analysis of ALK-stained tissue run on instrument #195 shows that there is no consistent difference in stain intensity with or without UPM after the electrical circuit upgrade. Slide position 2-1 with UPM showed clearly a decreased staining by digital image analysis and visual evaluation confirmed lack of staining of the control tissue. B, Medium-level control beads were spotted onto glass slides as described in the Materials and Methods. Runs performed on instrument #111 with and without UPM in place are shown. The control beads showed reduced signal on the slide placed in position 3-10 (indicated by the red arrow). The instrument indicated an error for this slide. A repeat run did not result in a similar error (data not shown). Similar results were seen with high-level control beads that were spotted on the same slides (data not shown). ALK indicates anaplastic lymphoma kinase.
Comparison of Different Methods of Assessing Immunohistochemical Stain Results
| Visual Assessment of Control Tissue (Current Method Used by All Laboratories) | Digital Image Analysis of Control Tissue | Control Microbeads (High and Low Levels) | Calibrator Microbeads | |
|---|---|---|---|---|
| Summary of workflow | Identify appropriate cases from archives | Identify appropriate cases from archives | Purchase microbead controls | Purchase calibrator slides |
| Locate and pull blocks | Locate and pull blocks | Manually pipette controls instead of mounting tissue controls | Process calibrators on immunostainer | |
| Block cutting | Block cutting | Process slides with patient samples on immunostainer | Evaluate calibrators visually or with digital image analysis (currently performed by the vendor) | |
| Stain slides | Stain slides | Stained slides reviewed by pathologist | ||
| Stained slides reviewed by pathologist | Whole slide images | |||
| Digital image analysis (quantification of stain intensity) | ||||
| Analyte concentration known? | No | No | Yes | Yes |
| Current uses | Detect gross assay problems or instrument failure | None | Detect gross assay problems or instrument failure) | Used to establish limit of detection |
| Potential future uses | Same as above | Fine discrimination of partial IHC assay problems | Routine demonstration that the correct IHC stain was performed and was performed with satisfactory results (expected intensity) | Determine LOD for ensuring interinstitutional alignment of assay performance |
| Levey-Jennings analysis | With added programs for image analysis and routine quantitation, global comparisons within runs, between runs, and between instruments to identify subtle changes in IHC stain performance that may be go unnoticed by visual assessment | |||
| Degree of quantitation of intensity | Limited, typical scale is to 0 to 3+ | Precise but relative quantification | Precise but relative quantification | Intensity quantification used for calculating LOD and dynamic range |
| Sensitivity to detect subtle decrements and variability in stain intensity | Lowest | Moderate | Low if visually assessed | Highest |
| Moderate if with digital image analysis | ||||
| Assess limit of detection? | No | No | No | Yes |
| Currently available for routine use | Yes | No (requires custom workflow) | Dependent on availability from vendor | Dependent on availability from vendor |
IHC indicates immunohistochemistry; LOD, limit of detection.