| Literature DB >> 35875283 |
Yasar Pazir1, Fatih Yanaral1, Ufuk Caglar1, Mazhar Ortac2, Omer Sarilar1, Faruk Ozgor1.
Abstract
Introduction The aim of this study was to evaluate patient and partner satisfaction, device reliability, and complications in patients who underwent two-piece inflatable penile prosthesis (IPP) implantation. Patients and methods The data of 22 patients who underwent two-piece inflatable penile prosthesis implantation in our department between 2015 and 2018 were retrospectively analyzed, and a detailed review of all clinical reports was performed. Phone or face-to-face interviews were undertaken to assess the satisfaction rates of the patients and their partners using the modified Erectile Dysfunction Inventory of Treatment Satisfaction (EDITS) questionnaire. Results The mean patient age and partner age were 57.95 ± 6.16 and 58.12 ± 6.66 years, respectively. The mean erectile dysfunction (ED) period was 5.33 ± 2.16 years, and the etiologies of erectile dysfunction were radical pelvic surgery (41%), diabetes mellitus (37%), and vascular disorders (22%). The mean operative time and postoperative hospital stay were 102 ± 29 minutes and 1.8 ± 0.66 days, respectively. Over a mean follow-up period of 29.04 ± 14.48 months, two (9%) cases underwent revision surgery due to mechanical device failure in one and infection in the other. The overall patient and partner satisfaction rates were 73% and 59%, respectively. Conclusions The two-piece inflatable penile prosthesis is an effective, reliable, and user-friendly prosthesis with acceptable complication and revision rates and provides a high level of patient and partner satisfaction in selected and fully informed patients.Entities:
Keywords: erectile dysfunction; partner satisfaction; patient satisfaction; penile prosthesis; two-piece inflatable penile prosthesis
Year: 2022 PMID: 35875283 PMCID: PMC9298674 DOI: 10.7759/cureus.26097
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Domains of the non-validated questionnaire administered to the patients
| Questions | Answers |
| 1. Did your prosthesis cause a decrease in penile length? | Yes/no |
| 2. Would you recommend the same procedure to another patient? | Yes/no |
| 3. Do you have satisfying feelings during sexual intercourse? | Yes/no |
| 4. Do you achieve orgasms? | Yes/no |
Clinical characteristics of the patients
*Continuous variables are presented as mean ± SD.
| Variables* | |
| Patients (number) | 22 |
| Patient age (years) | 57.95 ± 6.16 |
| Partner age (years) | 58.12 ± 6.66 |
| Body mass index (kg/m2) | 27 ± 2.62 |
| Duration of erectile dysfunction (years) | 5.33 ± 2.16 |
| Etiologies of erectile dysfunction (%) | |
| Radical pelvic surgery | 41% |
| Diabetes mellitus | 37% |
| Vascular disorders | 22% |
| Penile Doppler US findings (%) | |
| Arterial insufficiency | 55% |
| Veno-occlusive dysfunction | 27% |
| Combined pathologies | 18% |
| Operative time (minutes) | 102 ± 29 |
| Length of implanted cylinders and rear tip extenders (cm) | 18.1 ± 1.68 |
| Hospital stay (days) | 1.8 ± 0.66 |
| Follow-up period (months) | 29.04 ± 14.48 |
| Complications (number (%)) | 2 (9%) |
| Device infection | 1 (4.5%) |
| Mechanical failure | 1 (4.5%) |
Results obtained using the modified Erectile Dysfunction Inventory of Treatment Satisfaction (EDITS) for Ambicor penile prostheses
| Questions | Answers | Number (%) |
| 1. Overall, how satisfied are you with your penile prosthesis? | Very satisfied | 14 (64%) |
| Partially satisfied | 2 (9%) | |
| Dissatisfied | 6 (27%) | |
| 2. To what extent did penile prosthesis meet your expectations? | Completely | 13 (59%) |
| Partially | 3 (14%) | |
| Not at all | 6 (27%) | |
| 3. Is the penile prosthesis appropriate for continuous use? | Very appropriate | 11 (50%) |
| Partially appropriate | 7 (32%) | |
| Inappropriate | 4 (18%) | |
| 4. Is it easy for you to use the penile prosthesis? | Very easy | 11 (50%) |
| Partially easy | 7 (32%) | |
| Not easy | 4 (18%) | |
| 5. How confident do you feel about your competence in sexual intercourse? | Very confident | 13 (59%) |
| Partially confident | 3 (14%) | |
| Not confident | 6 (27%) | |
| 6. Are you generally satisfied with your partner’s penile prosthesis? | Very satisfied | 9 (41%) |
| Partially satisfied | 4 (18%) | |
| Dissatisfied | 9 (41%) |