[Impact of multi-leaf collimator positioning accuracy on quality control of volumetric modulation arc therapy plan for cervical cancer treated with Elekta linear accelerator].

J Liang1,2, F Yu2, J Zhu2, T Song1.   

Abstract

OBJECTIVE: To investigate the influence of positioning accuracy of the multi-leaf collimators (MLC) on the passing rate of the plan dose verification for volumetric modulation arc therapy (VMAT) of cervical cancer using an Elekta linear accelerator.
METHODS: The dose distributions were measured using Sun Nuclear's Mapcheck and Arccheck semiconductors matrix before and after MLC calibration in30 cervical cancer patients undergoing VMAT. Dosimetric comparisons were performed with 2D and 3D gamma passing rates of 3%, 3 mm and 2%, and 2 mm. The 3D gamma distribution was reconstructed with respect to the patient's anatomy using 3DVH software to evaluate the possible influence of MLC positioning accuracy.
RESULTS: Before and after MLC calibration, the gamma passing rates of Mapcheck were (88.80±1.81)% and (99.25 ± 0.53)% under 3% and 3 mm standard, respectively, with an average increase of 10.45%. The corresponding gamma passing rates of Arccheck were (87.61±1.98)% and (98.13±0.99)%, respectively, with an average increase of 10.52%. The gamma passing rates of 3DVH were (89.87±2.28)% and (98.3±1.15)%, respectively, with an average increase of 8.43%.
CONCLUSION: The MLC positioning accuracy is one of the main factors influencing dosimetric accuracy of VMAT for cervical cancer. The application of Autocal software facilitates MLC calibration and improves the accuracy and safety of VMAT delivery for cervical cancer.

Entities:  

Keywords:  Autocal; cervical cancer; multi-leaf collimator; plan verification; volumetric modulation arc therapy

Mesh:

Year:  2022        PMID: 35869775      PMCID: PMC9308871          DOI: 10.12122/j.issn.1673-4254.2022.07.19

Source DB:  PubMed          Journal:  Nan Fang Yi Ke Da Xue Xue Bao        ISSN: 1673-4254


  22 in total

1.  Gantry angle-dependent correction of dose detection error due to panel position displacement in IMRT dose verification using EPIDs.

Authors:  Guang-hua Jin; Jin-han Zhu; Li-xin Chen; Xiao-wu Deng; Bo-tian Huang; Ke Yuan; Xiao-wei Liu
Journal:  Phys Med       Date:  2013-06-17       Impact factor: 2.685

2.  Dose domain regularization of MLC leaf patterns for highly complex IMRT plans.

Authors:  Dan Nguyen; Daniel O'Connor; Victoria Y Yu; Dan Ruan; Minsong Cao; Daniel A Low; Ke Sheng
Journal:  Med Phys       Date:  2015-04       Impact factor: 4.071

3.  Response of the ArcCHECK® device at 6 MV and 15 MV for VMAT and IMRT quality control.

Authors:  Miguel Jiménez-Melguizo; Miguel Espinosa; Joaquín Montes; Damián Guirado; Antonio M Lallena
Journal:  Phys Med       Date:  2020-12-10       Impact factor: 2.685

4.  VMAT QA: measurement-guided 4D dose reconstruction on a patient.

Authors:  Benjamin E Nelms; Daniel Opp; Joshua Robinson; Theresa K Wolf; Geoffrey Zhang; Eduardo Moros; Vladimir Feygelman
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

5.  Comparison of 2D and 3D gamma analyses.

Authors:  Kiley B Pulliam; Jessie Y Huang; Rebecca M Howell; David Followill; Ryan Bosca; Jennifer O'Daniel; Stephen F Kry
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

6.  Racial disparities in brachytherapy administration and survival in women with locally advanced cervical cancer.

Authors:  Stephanie Alimena; David D Yang; Alexander Melamed; Brandon A Mahal; Michael J Worley; Sarah Feldman; Kevin M Elias; Peter F Orio; Larissa J Lee; Martin King
Journal:  Gynecol Oncol       Date:  2019-07-07       Impact factor: 5.482

7.  A novel verification method using a plastic scintillator imagining system for assessment of gantry sag in radiotherapy.

Authors:  Masato Tsuneda; Teiji Nishio; Akito Saito; Sodai Tanaka; Tatsuhiko Suzuki; Daisuke Kawahara; Keiichiro Matsushita; Aya Nishio; Shuichi Ozawa; Kumiko Karasawa; Yasushi Nagata
Journal:  Med Phys       Date:  2018-05-09       Impact factor: 4.071

8.  Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA.

Authors:  Heming Zhen; Benjamin E Nelms; Wolfgang A Tome
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

9.  A comprehensive and clinical-oriented evaluation criteria based on DVH information and gamma passing rates analysis for IMRT plan 3D verification.

Authors:  Xin Yi; Wen-Li Lu; Jun Dang; Wei Huang; Hai-Xia Cui; Wan-Chun Wu; Ying Li; Qing-Feng Jiang
Journal:  J Appl Clin Med Phys       Date:  2020-05-20       Impact factor: 2.102

10.  Pre-treatment 2D and 3D dosimetric verification of volumetric arc therapy. A correlation study between gamma index passing rate and clinical dose volume histogram.

Authors:  Lukasz Szczurek; Robert Juszkat; Jolanta Szczurek; Ilona Turek; Piotr Sosnowski
Journal:  PLoS One       Date:  2019-08-13       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.