Olivier De Castro1, Jean-Nicolas Audinot1, Hung Quang Hoang1, Chérif Coulbary1, Olivier Bouton2, Rachid Barrahma2, Alexander Ost1,3, Charlotte Stoffels1,3, Chengge Jiao4, Mikhail Dutka4, Michal Geryk5, Tom Wirtz1. 1. Advanced Instrumentation for Nano-Analytics, MRT Department, Luxembourg Institute of Science and Technology, 41 rue du Brill, L-4422 Belvaux, Luxembourg. 2. Prototyping, MRT Department, Luxembourg Institute of Science and Technology, 41 rue du Brill, L-4422 Belvaux, Luxembourg. 3. Faculty of Science, Technology and Medicine, University of Luxembourg, 2 Avenue de l'Université, L-4365 Esch-sur-Alzette, Luxembourg. 4. Thermo Fisher Scientific; Achtseweg Noord 5, 5651 GG Eindhoven, Netherlands. 5. Thermo Fisher Scientific; Vlastimila Pecha 12, 627 00 Brno, Czech Republic.
Abstract
The structural, morphological, and chemical characterization of samples is of utmost importance for a large number of scientific fields. Furthermore, this characterization very often needs to be performed in three dimensions and at length scales down to the nanometer. Therefore, there is a stringent necessity to develop appropriate instrumentational solutions to fulfill these needs. Here we report on the deployment of magnetic sector secondary ion mass spectrometry (SIMS) on a type of instrument widely used for such nanoscale investigations, namely, focused ion beam (FIB)-scanning electron microscopy (SEM) instruments. First, we present the layout of the FIB-SEM-SIMS instrument and address its performance by using specific test samples. The achieved performance can be summarized as follows: an overall secondary ion beam transmission above 40%, a mass resolving power (M/ΔM) of more than 400, a detectable mass range from 1 to 400 amu, a lateral resolution in two-dimensional (2D) chemical imaging mode of 15 nm, and a depth resolution of ∼4 nm at 3.0 keV of beam landing energy. Second, we show results (depth profiling, 2D imaging, three-dimensional imaging) obtained in a wide range of areas, such as battery research, photovoltaics, multilayered samples, and life science applications. We hereby highlight the system's versatile capability of conducting high-performance correlative studies in the fields of materials science and life sciences.
The structural, morphological, and chemical characterization of samples is of utmost importance for a large number of scientific fields. Furthermore, this characterization very often needs to be performed in three dimensions and at length scales down to the nanometer. Therefore, there is a stringent necessity to develop appropriate instrumentational solutions to fulfill these needs. Here we report on the deployment of magnetic sector secondary ion mass spectrometry (SIMS) on a type of instrument widely used for such nanoscale investigations, namely, focused ion beam (FIB)-scanning electron microscopy (SEM) instruments. First, we present the layout of the FIB-SEM-SIMS instrument and address its performance by using specific test samples. The achieved performance can be summarized as follows: an overall secondary ion beam transmission above 40%, a mass resolving power (M/ΔM) of more than 400, a detectable mass range from 1 to 400 amu, a lateral resolution in two-dimensional (2D) chemical imaging mode of 15 nm, and a depth resolution of ∼4 nm at 3.0 keV of beam landing energy. Second, we show results (depth profiling, 2D imaging, three-dimensional imaging) obtained in a wide range of areas, such as battery research, photovoltaics, multilayered samples, and life science applications. We hereby highlight the system's versatile capability of conducting high-performance correlative studies in the fields of materials science and life sciences.
A common
fact about a large
number of technological and scientific fields is that, in order to
answer emerging critical questions, new and innovative characterization
tools need to be developed. An essential element of this framework
consists of combining structural and chemical information at the nanometer
scale. The application areas of such characterization instruments
span a wide range of fields and include the imaging of features in
highly complex electronic devices at high lateral resolution (e.g.,
dopant distributions, grain structures, and boundaries) as well as
the precise recording of subcellular chemical information in biological
specimens to obtain better knowledge of the ongoing processes at the
physiological level. Key characteristics that are required are highest
spatial resolution, excellent chemical sensitivity, high dynamic range
(for the detection and mapping of elemental concentrations varying
over several orders of magnitude), and isotopic selectivity.One particular type of instrument, based on the combination of
a high-resolution electron beam column (with spot sizes in the nanometer
range) and a focused ion beam (FIB) column,[1−3] is being used
extensively in this context. The main benefit of this is that the
user has the ability to perform secondary electron microscopy (SEM)
imaging in combination with FIB sample preparation and nanofabrication
tasks in a single tool. Such instruments are typically referred to
as FIB-SEM or dual-beam instruments.The well-established technique
for compositional analysis deployed
on such instruments is energy-dispersive X-ray spectroscopy (EDX).[4−6] Nevertheless, EDX is not able to distinguish between isotopes or
detect trace elements (poor detection limit of ∼1000 ppm[7]). Furthermore, in the low Z-element range (below
boron), the detection of hydrogen is impossible, and the detection
of Li very challenging.[8,9]An analytical technique
offering higher sensitivity (detection
limits down to the ppm range) is secondary ion mass spectrometry (SIMS),
which also enables the detection of all elements from H to U while
differentiating between isotopes and offering a high dynamic range.[10,11] The effort to offer SIMS capability on single-column FIB instruments
goes back to the 1980s, as demonstrated by the works of Levi-Setti
et al.,[12,13] and has been extended in more recent years
to FIB-SEM instruments.One option to perform SIMS on FIB-SEM
instruments is based on the
use of a time-of-flight (TOF) spectrometer as the SIMS spectrometer.[14] More specifically, such a setup is often based
on an orthogonal TOF (o-TOF) configuration in which the secondary
ion (SI) beam is pulsed while working with a continuous primary ion
beam.[15] While such TOF systems have the
benefit of being compact in size and having quasi-parallel mass detection,
they suffer from a low duty cycle (while the sample is constantly
eroded by the DC beam, only fractions of the SIs are pulsed into the
TOF systems), which in addition is mass-dependent. This results in
a reduced overall sensitivity, which also varies as a square root
function with respect to the mass.[15]Another type of SIMS spectrometer being used on FIB-based instruments
is quadrupole spectrometers.[16] Chater et
al. reported a single-column FIB instrument equipped with two quadrupole
spectrometers, to detect both positive and negative SIs at the same
time.[17] Nevertheless, these systems work
in serial acquisition (meaning that ions of different masses are detected
sequentially rather than simultaneously) and are not suitable for
fast chemical imaging.A third kind of SIMS spectrometer is
magnetic sector-based spectrometers.
These operate in DC mode (no pulsing and hence no losses due to duty
cycles) and have high overall transmission of SIs, conferring on them
important advantages in terms of sensitivity and speed over TOF spectrometers.
Furthermore, in contrast to quadrupole spectrometers, they offer parallel
mass detection and are hence optimized for fast imaging tasks. This
function is specifically achieved in a Mattauch-Herzog configuration,
which is characterized by an elongated straight focal plane covering
the full mass range of the spectrometer.[18] Finally, high mass resolution can be achieved when magnetic sector
SIMS systems are operating in double-focusing condition. While magnetic
sector SIMS systems used to be bulky and heavy, modern systems can
be compact and considerably lighter thanks to advances in charged
particle optics, electromagnet coil concepts, and mechanical assembly
design.[19,20]We report here on an FIB-SEM instrument
incorporating a newly developed
compact double-focusing magnetic sector SIMS spectrometer. The instrument
offers key capabilities, such as (1) nanoscale lateral resolution
imaging in SEM mode; (2) nanoscale, high sensitivity, and high dynamic
range chemical/elemental analysis in SIMS mode, including sub-20 nm
imaging resolution and sub-5 nm depth resolution; (3) isotopic analysis;
(4) correlation of SIMS data with other data sets, such as secondary
electron (SE) images, backscattered electron (BSE) data, and EDX data.
The SIMS system can be operated in several modes: recording of mass
spectra, two-dimensional (2D) chemical imaging, three-dimensional
(3D) chemical imaging by recording, and stacking the multiple 2D chemical
maps obtained when removing the sample material layer-by-layer, and
depth profiling (following selected mass signals while milling the
sample). All these capabilities pave the way for a large number of
application possibilities, ranging from analytics needing high sensitivity,
SIMS imaging at highest lateral resolution (15–20 nm), and
FIB nanofabrication (milling/patterning) with in situ process control,
to multimodal approaches correlating SIMS data with SE, BSE, and EDX
data obtained from the same instrument.
Instrument Design
The instrument described here is based on a commercially available
high-vacuum FIB-SEM platform called Scios from Thermo Fisher Scientific.[21,22] As shown in Figure , the instrument is equipped with a vertically orientated SEM column
based on a field emission source. The SEM-column allows electron beam
landing (impact) energies from 200 eV to 30 keV, offering a lateral
resolution of 1–2 nm at 1 keV. The instrument is equipped with
various subsystems for electron beam-induced SE signal detection as
well as for BSE signal detection. A Ga liquid metal ion source (LMIS)-based
FIB column (Ga-FIB) is installed at 52° with respect to the vertical
SEM axis. The Ga-FIB offers an ion beam landing energy range from
500 eV to 30 keV, with probe currents spanning from 1.5 pA to 65 nA.
The Ga+ primary beam can be used for ion beam-induced SE
imaging as well as for ion beam milling/fabrication and analytical
purposes, including SIMS. The smallest probe size of 3 nm is achievable
at 30 keV. Optionally, specific gas injection systems (GIS) can be
installed to offer, for example, the in situ platinum coating of sample
surfaces or injection of precursors for other fabrication tasks. The
eucentric beam coincidence point of the instrument is located at the
FIB working distance (WD) of 19 mm and the SEM WD of 7 mm. When one
positions the region of interest (ROI) of the sample at this precise
location, it is possible to perform sample (stage) tilt movements
without any lateral displacements of the ROI under either primary
beam. This allows one to do nanofabrication with the ion beam while
performing quasi-simultaneous SEM imaging of the exact same ROI.
Figure 1
Schematic
layout of the FIB-SEM-SIMS instrument.
Schematic
layout of the FIB-SEM-SIMS instrument.To perform SIMS on the Scios instrument, we developed a dedicated
compact magnetic sector SIMS spectrometer that can be installed as
an add-on attachment, allowing us to do an SIMS analysis with the
Ga-FIB while not interfering with or reducing the performance of the
standard SEM and FIB operations. Because of the nature of the dual-beam
operation, there are several challenges in the development and integration
of such a dedicated SIMS system, including, in particular, the design
of the extraction and transfer optics. First, the SIMS extraction
optics need to be integrated into an extremely limited space between
the SEM column, the FIB column, and the sample while not disturbing
the operation of other detectors (i.e., the SE detector). Second,
the SIMS needs to be operated with the sample at the eucentric beam
coincidence point, allowing us to perform both SIMS and SEM analyses
of the same ROI. Third, the integration of the SIMS optics needs to
be compatible with available existing ports on the Scios and the orientation
of the FIB column. Finally, apart from the hardware integration challenges
mentioned above, the SIMS system should not degrade the probe size
of the FIB column in the SIMS mode beyond the dimensions of the collision
cascades triggered within the sample upon ion impacts to enable nanoscale
resolution. Moreover, it should have a high extraction efficiency
to maintain high sensitivity even when working with nanoscale sputtered
voxels (from all sputtered particles only ∼1% leave the sample
as SIs, but this fraction significantly depends on the considered
element and the sample composition (matrix effect)). The choice of
the primary ion species has an influence on the SI signal and hence
on the sensitivity. The relatively high mass of Ga (as compared to
light species such as He or Ne used on the helium ion microscope)
results in high sputter yields and therefore also increases the SI
signals. In combination with the residual oxygen in the analysis chamber
when working at a typical pressure of 10–7 mbar,
Ga also allows for good ionization yields in the positive SI mode.
The yields can be further increased by increasing the partial oxygen
pressure in the chamber via an O2 leak. For negative SI
detection, the use of very electropositive primary ions such as Cs
would lead to even higher ionization yields.[19]Addressing all of the above challenges, a dedicated SIMS system,
as schematically presented in Figure , was developed to integrate it into the Scios (a photograph
of the FIB-SEM-SIMS instrument is shown in Figure S1 of the Supporting Information). To maintain a high extraction
efficiency as well as to position the sample at the eucentric point
during the SIMS operation, the ion extraction optics are based on
an extraction box, which was described elsewhere.[19] During SIMS operation, the extraction box is placed in
between the sample and the nose cone of the FIB column, allowing both
the normal incident angle of the primary beam on the sample and the
normal extraction of the SIs from the sample. Furthermore, this chosen
geometry allows the SIMS operation with the sample at the eucentric
point independently of the strength of the extraction field (or sample
bias). As shown schematically in Figure , this extraction box has to be carefully
tailored to fit into the extremely limited space between the FIB column
and the SEM column. The overall height of the extraction box is 16.5
mm. Note that the design also allows the SEM operation to be used
through an aperture of 0.5 mm in diameter integrated at 52° with
respect to the FIB column axis in the outer grounded electrode of
the extraction box.
Figure 2
Schematic layout of the extraction box region close to
the sample/column
end nozzle area, post acceleration region, and second 52° bending
electrostatic sector.
Schematic layout of the extraction box region close to
the sample/column
end nozzle area, post acceleration region, and second 52° bending
electrostatic sector.During the SIMS operation,
the extraction box is set at 0.5 mm
above the sample, and the sample is biased to ±500 V for positive
and negative SI extraction, respectively. The SIs are extracted by
the electric field created by the sample bias potential and the grounded
first electrode of the extraction box. They are then further transported
by a 90° spherical electrostatic sector. As the orientation of
the FIB column is at 52° with respect to the vertical axis, these
transported ions have to be bent by another 52° in order to be
aligned with the mass spectrometer, which is mounted in the horizontal
plane on the Scios. Therefore, a second spherical electrostatic sector
with a bending angle of 52°, in combination with a postacceleration
tube placed between the two electrostatic sectors, is used as illustrated
in Figure . The acceleration
tube allows us to accelerate the SIs from 500 eV to 3.5 keV in order
to maximize the transmission of the subsequent optics. Note that all
the subsequent optics from the 52° electrostatic sector to the
detector are electrically floated at ±3 kV to maintain the kinetic
energy of the SIs at 3.5 keV throughout the SIMS system. A dedicated
transfer optics consisting of three lenses and four quadrupole deflectors
then transports the SIs further to the mass analyzer placed outside
of the Scios chamber (Figure ). When the SIMS is not used, the complete extraction and
transfer optics can be retracted by means of piezo-positioners to
a storage (parking) position close to the chamber wall, restoring
all the normal operation capabilities of the Scios.The mass
analyzer used in this system is based on a modified Mattauch-Herzog-type
magnetic sector configuration, which consists of a 60° spherical
electrostatic sector and 75° magnetic sector. An electromagnet
sector is used, offering both magnetic field polarities for the positive
and negative SIMS modes. The pole pieces and yoke of the magnetic
sector are carefully designed in order to minimize their size and
volume, with the target of reducing weight while maximizing the magnetic
field strength capacity. This magnetic sector, in combination with
the transfer optics, allows a parallel detectable mass range (Mmax/Mmin = 120; Mmax of 400 amu at a magnetic field of 700 mT)
along its focal plane of 400 mm. A mass resolving power (M/ΔM) above 400 is achieved with an overall
transmission of above 40%.The current system is equipped with
four channeltron detectors.
Three detectors can be moved independently by piezo drives along the
focal plane. The fourth detector is fixed at the highest radius position
of the focal plane. This arrangement of the detection system allows
the flexibility to select four masses to be recorded for SIMS analysis
in parallel. Note that this SIMS system offers the capability to integrate
a continuous full range focal plane detector, allowing full parallel
detection of all the masses and hyperspectral SIMS analysis. This
type of focal plane detector has been developed and reported elsewhere,[23] and it will equip the next-generation SIMS on
this FIB-SEM-SIMS system. Moreover, the system integrates an additional
channeltron detector positioned at the end of the transfer optics
for total ion current (TIC) measurements. This option is beneficial
for optics alignment purposes as well as for nonmass filtered imaging
in both SIMS polarities, giving further complementary contrast imaging
capabilities.
Experimental Section
The BAM-L200
sample consists of a nanoscale stripe pattern for
length calibration and specification of spatial resolution prepared
from a cross-sectioned epitaxially grown layer stack of AlGa1–As and InGa1–As
on a GaAs.[24]For the multilayer sample,
Cu, Al, and Cr were deposited successively
on a silicon wafer by Plasma Vapor Deposition (PVD) (Kurt J. Lesker
Company). Power at the rate of 30 W dc was applied successively on
three targets of aluminum, copper, and chromium (2 in. in diameter
and 1/4 in. thick with a purity of 99.95%), and to homogenize the
deposit, the silicon substrate was rotated at a speed of 5 rpm. This
stacking of six layers was characterized by ultralow impact energy
SIMS (Sc-Ultra, Cameca) using a primary cesium ion beam of 5 nA with
an impact energy of 1 keV (Supporting Information). Each layer thickness, Cu (22.6 nm)/Al (11.7 nm)/Cr (33.7 nm)/Cu
(26.5 nm)/Al (13.8 nm)/Cr (26.8 nm), was determined by profilometry
(Tencor P20).The hybrid organic–inorganic halide lead
perovskite absorber
containing Cs cations was fabricated by the perovskite deposition
of a solution by spin coating on a glass substrate. The solution was
obtained with the desired proportions of the precursor solution of
PbI2(TCI), PbI2 (TCI), CsPbI3 (TCI),
formamidinium iodide (Dyesol), and methylammonium bromide (Dyesol)
dissolved in a 1:4 mixture of dimethyl sulfoxide (DMSO)/dimethylformamide
(DMF). The perovskite solution was spun at 5000 rpm for 30 s using
a ramp of 3000 rpms–1. A final heating at 100 °C
for 50 min was performed to complete the process of the crystallization.As described by Schoppe et al.,[25] the
copper indium gallium selenium (CIGS) film was grown by coevaporation
of the elements with a high-temperature multistage in-line process.
The Rb was introduced into the absorber layer via the deposition of
an RbF layer of ∼300 nm in thickness and annealing in vacuum
at 355 °C for 20 min, allowing a constant Rb diffusion in the
CIGS layer. Residual RbF on the CIGS surface was removed with HCl.
For the FIB-SEM-SIMS analysis, a piece of 2 × 2 cm2 was cut with manual tools.The Al–Li–Cu–Mg
alloy sample material
is prepared as described in Xu et al.[26]Biosample: All chemicals were purchased from Sigma-Aldrich.
Caco-2
cells were cultured and seeded on an insert (Millicell Hanging Inserts,
PET 1 μm, 24-well plate). After 21 d of differentiation, cells
were exposed to 5 μM perfluorooctanoic acid for 24 h. Cells
on an insert were fixed with 5% glutaraldehyde in Dulbecco’s
modified phosphate buffered saline (DPBS) for 12 h at 4 °C and
postfixed with 1% osmium tetroxide in DPBS for 1 h at room temperature.
Afterward, samples were dehydrated with a series of graded ethanol
solution (30%; 50%; 70%; 90% for 10 min; 100% for 2 h) and embedded
in Spurr resin. The resin blocks were sliced into 300 nm semithin
sections (ultramicrotome, Leica) and deposited on silicon wafers (Sil’tronix).
Finally, the sections were coated with a 5 nm gold film by sputtering
(Leica).
Results and Discussion
SIMS Performance: Mass Spectrum Recording
The mass
spectra are recorded while scanning the Ga+ beam over a
surface (e.g., from 2 × 2 μm2 to 100 ×
100 μm2) to average the signal intensity over the
ROI. The mass spectra are generated by either scanning the magnetic
field with the detectors placed at fixed positions along the focal
plane or by moving at least one of the three movable detectors (i.e.,
detector scan mode) at a fixed magnetic field.Figure shows a mass spectrum of a
hybrid organic–inorganic halide lead perovskite absorber containing
Cs cations.[27] The mass spectrum was obtained
from an area of 50 × 50 μm2 using a 30 keV Ga+ beam with a current of 6 pA. The magnetic field was scanned
from 5 to 700 mT in 0.10 mT steps, allowing us to cover a mass range
from 0 to 380 amu (using the fixed detector). The counting time (frame
dwell time) was 250 ms per step. The main elements that constitute
the photovoltaic film were detected as well as some small cluster
ions (such as Cs2, PbI, PbCs). The four lead isotopes, 204Pb, 206Pb, 207Pb, and 208Pb, were detected with an isotopic ratio close to the theoretical
values (1%, 24%, 22%, and 52%, respectively). The mass resolving power
(full width at half-maximum (FWHM)) M/ΔM calculated on the 208Pb peak amounted to 410.
Figure 3
Mass spectrum
(magnetic field scan of 5 to 700 mT) of a hybrid
organic–inorganic halide lead perovskite absorber containing
Cs cations[27] using a Ga+ primary
ion beam. The enlarged spectrum (upper-right) shows the distribution
of the main Pb isotopes.
Mass spectrum
(magnetic field scan of 5 to 700 mT) of a hybrid
organic–inorganic halide lead perovskite absorber containing
Cs cations[27] using a Ga+ primary
ion beam. The enlarged spectrum (upper-right) shows the distribution
of the main Pb isotopes.
SIMS Performance: Depth
Profiling
The depth-profiling
mode is traditionally one of the most frequent uses of SIMS.[28] The best depth resolution can be obtained with
low-impact/beam-landing energies, typically below 1 keV.A multilayer
sample, consisting of copper, aluminum, and chromium (two layers each)
was analyzed to create a depth profile from scanned frame integration
(Figure ). The SIMS
detectors were aligned to measure the most abundant isotopes, that
is, 27Al, 52Cr, 63Cu, and 69Ga. The gallium signal was acquired for SI signal normalization purposes.
A 6 pA Ga+ beam with 3 keV landing energy (chosen for both
good signal statistics and depth resolution) was scanned over a surface
of 10 × 10 μm2, at 512 × 512 pixels with
a dwell time of 1 ms/pixel. Figure shows the depth profiles. The total counts of Cu,
Al, and Cr were summed up for the scanned area and were normalized
with respect to the Ga signal to compensate for any instrumental drift
of the SI beam alignment, the sample stage position, or the FIB column
parameters, due to the long acquisition time (in total 12 h). The
time scale was converted to a depth scale (in nm) on the x-axis by imposing the layer thickness determined from measurements
previously obtained on the same sample using a well-calibrated stand-alone
SIMS instrument, the SC Ultra[29] (Supporting Information Figure S2). A depth resolution
(decay length/decade) of 3.8 nm was determined from the raw profile
(average of rising and falling edge for each layer).
Figure 4
Depth profiling on a
multilayer sample: Cu (23 nm)/Al (11 nm)/Cr
(34 nm)/Cu (27 nm)/Al (14 nm)/Cr (27 nm)/silicon substrate. The depth
profile was created by integrating the counts for each element obtained
in each scanned frame (in arbitrary units).
Depth profiling on a
multilayer sample: Cu (23 nm)/Al (11 nm)/Cr
(34 nm)/Cu (27 nm)/Al (14 nm)/Cr (27 nm)/silicon substrate. The depth
profile was created by integrating the counts for each element obtained
in each scanned frame (in arbitrary units).
SIMS Performance: 2D Imaging
SIMS imaging is one of
the main interests in coupling a mass spectrometer to an FIB instrument.
The SIMS imaging resolution was evaluated using a standard BAM-L200
sample;[24,30] see Figure . A 30 keV Ga+ beam was used in two different
current settings to image two different grating period parts of the
sample. Prior to the SIMS imaging, the sample was cleaned over a larger
field of view (FoV) using the Ga+ beam (30 keV, 150 pA)
to remove any surface contamination until the aluminum lines were
clearly seen in the SE imaging mode.
Figure 5
Lateral resolution determination of magnetic
sector SIMS imaging
on the certified reference sample BAM-L200 using the Ga-FIB on the
FIB-SEM-SIMS instrument. (left) Al+ SIMS map of the grating
period region from P4 to P8 with an overlay of a 10-pixel integrated
intensity profile line-scan (Ga-FIB: 30 keV, 10 pA; cut-out of FoV
3.5 × 3.5 μm2 imaged with 512 × 512 pixels
(6.8 nm/pixel); dwell time of 0.5 ms). (center) Certified values of
grating period and single line sizes.[24] (right) Al+ SIMS map of the grating period region from
P9 to P14 with an overlay of a 10-pixel integrated intensity profile
line-scan (Ga-FIB: 30 keV, 1.5 pA; cut-out of FoV 5.2 × 5.2 μm2 imaged with 1024 × 1024 pixels (5.1 nm/pixel); dwell
time of 0.5 ms).
Lateral resolution determination of magnetic
sector SIMS imaging
on the certified reference sample BAM-L200 using the Ga-FIB on the
FIB-SEM-SIMS instrument. (left) Al+ SIMS map of the grating
period region from P4 to P8 with an overlay of a 10-pixel integrated
intensity profile line-scan (Ga-FIB: 30 keV, 10 pA; cut-out of FoV
3.5 × 3.5 μm2 imaged with 512 × 512 pixels
(6.8 nm/pixel); dwell time of 0.5 ms). (center) Certified values of
grating period and single line sizes.[24] (right) Al+ SIMS map of the grating period region from
P9 to P14 with an overlay of a 10-pixel integrated intensity profile
line-scan (Ga-FIB: 30 keV, 1.5 pA; cut-out of FoV 5.2 × 5.2 μm2 imaged with 1024 × 1024 pixels (5.1 nm/pixel); dwell
time of 0.5 ms).The left-hand side of Figure shows the Al+ signal image obtained with
a Ga+ beam setting of 30 keV and 10 pA, scanned over an
FoV of 3.5 × 3.5 μm2 (only part of the FoV is
shown) representing the grating period region from P4 to P8. The scanning
was performed at 512 × 512 pixels (6.8 nm/pixel) with a dwell
time of 0.5 ms/pixel. Further shown is the intensity profile obtained
across the periods in which the counts from a 10-pixel large line
scan were integrated, as indicated in the image. All periods down
to P8 (48.5 nm) were resolved, and even the single line W8 (38 nm)
can be seen. When performing the rising-edge resolution evaluation
at the imaged grating periods, as was also done on the same sample
by Kollmer et al.[31] and Kim et al.,[32] a lateral resolution of 15 nm can be determined
in SIMS imaging mode (80%–20% maximum intensity drop criterion
was used as proposed in ref (31) here across the rising edges of P5 and P6). The right-hand
side of Figure shows
the second grating period region, namely, P9 down to P14, this time
using a 30 keV, 1.5 pA Ga+ beam. It represents again a
cutout of an FoV of 5.2 × 5.2 μm2, taken with
1024 × 1024 pixels (5.1 nm/pixel) and a dwell time of 0.5 ms/pixel.
When one considers the line scan, the period down to P12 (31 nm) is
resolved, and P13 (23 nm) is still partially resolved. Here, even
a faint signal from line W9 (3.6 nm) and a more pronounced signal
from line W10 (14.2 nm) can be seen. Performing again the rising-edge
resolution determination at the periods P9 and P10, a lateral resolution
value of 15 nm is confirmed. These images were obtained with a careful
choice of FoV size, pixel matrix size, and counting time per pixel
for the different beam current settings used, as the sample, and especially
the Al lines, were very easily degraded with an incorrect acquisition
setting.
SIMS Performance: 3D SIMS Volumetric Reconstruction
To demonstrate the instrument’s performances in terms of spatial
and depth resolution in a single representation, a 3D volume reconstruction
was created using the SIMS image frames acquired from a patterned
cuboid structure presented in the Supporting Information (Figure S3).Thus, the RGB images containing
the acquired 63Cu (red), 52Cr (green), and 27Al (blue) signals were aligned and converted into a point
cloud in a 3D space in MATLAB. The correct thickness of each layer
was attributed to the corresponding volume in the point cloud. Figure shows the reconstructed
cuboid in a side view. With this 3D reconstruction, we demonstrate
the capability of the Scios-SIMS to create full 3D volume reconstructions
of specimens consisting of different compositions.
Figure 6
Side view of the 3D SIMS
volume reconstruction of a multilayered
cuboid structure using 180 SIMS image planes corrected for the thickness
of each layer.
Side view of the 3D SIMS
volume reconstruction of a multilayered
cuboid structure using 180 SIMS image planes corrected for the thickness
of each layer.
Applications in Materials
Science and Life Sciences
The novel FIB-SEM-SIMS instrument
has been applied to different topics
in order to address scientific questions and to demonstrate the instrument’s
strong analytical potential. Here we will focus on applications in
the field of materials science for energy conversion and storage,
a new alloy for aerospace applications, and an overview of the potential
of the FIB-SEM-SIMS in the field of life sciences.
Solar Cells
In recent years, for photovoltaic conversion,
many studies have been undertaken to allow a significant increase
in the conversion efficiency as well as the stability of such devices
with time. For copper indium gallium selenide solar cells, it has
been demonstrated that treatments during the elaboration of CIGS thin
films could allow one to obtain higher yields. An alkali fluoride
(e.g., RbF) postdeposition treatment is applied after the growth of
the CIGS layer, on the order of 1 wt %, allowing one to obtain a solar
cell efficiency of 22.6%.[25] The investigation
of the rubidium distribution in the grains and their boundaries requires
high lateral resolution and high sensitivity. The localization determination
of the alkali metal was previously performed by Atom Probe Tomography
(APT) and high-resolution synchrotron-based X-ray fluorescence analysis
(nano-XRF). In the first case, APT requires a complex sample preparation,
without allowing the entire film to be analyzed, although it provides
atomic resolution. In the second case, nano-XRF is a quantitative
technique but with a limited spatial resolution (few μm). In
these previous studies, rubidium was detected only in the grain boundaries,
with a detection limit of 200 ppm. Using the FIB-SEM-SIMS instrument,
we were able to detect Rb in the grain boundaries (Figure ) with a detection limit that
is 10 times better than with the synchrotron technique.[25] Furthermore, with SIMS, we confirmed the absence
of Rb inside the grains after we performed acquisitions on various
ROIs (from 20 × 20 μm2 to 5 × 5 μm2) analyzed in just a few minutes (20 min per ROI using a pixel
size of 12 nm), without complex sample preparation.
Figure 7
CIGS solar cell after
a rubidium fluoride (RbF) postdeposition
treatment was applied. (a) SE image (1536 × 1092 pixels; electron
beam at 30 keV, 0.8 nA); (b–e) SIMS images (512 × 512
pixels, dwell time per pixel of 5 ms; Ga+ beam at 30 keV,
3 pA; FoV (b, c) 20 × 20 μm2/(d, e) 5 ×
5 μm2); (f) line profiles (10 pixels in width as
indicated by arrows in (d, e)), showing that rubidium is accumulated
only in the grain boundary.
CIGS solar cell after
a rubidium fluoride (RbF) postdeposition
treatment was applied. (a) SE image (1536 × 1092 pixels; electron
beam at 30 keV, 0.8 nA); (b–e) SIMS images (512 × 512
pixels, dwell time per pixel of 5 ms; Ga+ beam at 30 keV,
3 pA; FoV (b, c) 20 × 20 μm2/(d, e) 5 ×
5 μm2); (f) line profiles (10 pixels in width as
indicated by arrows in (d, e)), showing that rubidium is accumulated
only in the grain boundary.
Batteries
The prominence of lithium ion batteries for
mobile and stationary use requires the improvement of their performance
to meet increasingly demanding requirements for energy storage, by
the optimization of suitable electrodes and electrolytes.Unlike
analytical techniques based on X-rays, that is, EDX, SIMS allows the
detection of lithium at a very low detection limit, together with
alkali and alkaline earth elements. To investigate degradation at
the interfaces of batteries, which is reducing the storage performance,
a cycled LiMnNiCo battery cathode was extracted from disassembled
cells and analyzed on the FIB-SEM-SIMS (Figure ). The side previously in contact with the
electrolyte was imaged to observe both morphological and chemical
modifications. The elements recorded were Li, Mn, Co, and Ni. The
major isotopes were selected, except for lithium, for which the 6Li isotope (7.5% abundance) was chosen in order not to saturate
the corresponding detector. In this study, we can see the effect of
the degradation of the cathode previously in contact with the electrolyte.
On top of a granular morphology, while manganese is almost homogeneous,
nickel and cobalt present a more heterogeneous distribution. Various
hypotheses related to the degradation of the cathode can be listed.
The one that seems most likely is a diffusion of nickel and cobalt
in the electrolyte to form a solid electrolyte interphase.
Figure 8
LiNiCoMn cathode
material for Li ion batteries. (a) SE image (1536
× 1092 pixels, dwell time per pixel 5 μs; Ga+ beam 30 keV, 10 pA); (b–e) SIMS images (FoV 20 × 20
μm2, 512 × 512 pixels, dwell time per pixel
2.5 ms; Ga+ beam 30 keV, 3 pA). The color intensity scale
is from 0 to 20 (black to red). (f) The image represents an overlay
of lithium (green) and nickel (red).
LiNiCoMn cathode
material for Li ion batteries. (a) SE image (1536
× 1092 pixels, dwell time per pixel 5 μs; Ga+ beam 30 keV, 10 pA); (b–e) SIMS images (FoV 20 × 20
μm2, 512 × 512 pixels, dwell time per pixel
2.5 ms; Ga+ beam 30 keV, 3 pA). The color intensity scale
is from 0 to 20 (black to red). (f) The image represents an overlay
of lithium (green) and nickel (red).To verify this, it would be necessary to analyze the whole cell
in section, that is, the complete cathode/electrolyte/anode system.
One of the advantages of the FIB-SEM system is that it is possible
to do an in situ analysis, minimizing the risk of contamination (high
reactivity of Li with air) and allowing us to obtain the flattest
possible analysis surface.[33] We note that
the samples can be also prepared in a glovebox and transferred to
the instrument using an inert gas atmosphere shuttle. Recently, our
FIB-SEM-SIMS instrument was equipped with a load-lock for rapid sample
introduction and docking station for the shuttle (Ferrovac, CH).
Alloys
Thanks to their mechanical performance and lightness,
aluminum–lithium (Al–Li) alloys are extensively used
for applications in the aerospace industry.The latest Al–Li
alloys contain, in particular, copper, magnesium, manganese, and zinc
as elements, in the range of 0.2–2 wt %. The microstructure
of these materials typically includes finely distributed nanoscale-strengthening
precipitates. The size and distribution of these precipitates have
a significant impact on the mechanical performance of these alloys.
However, the analysis of Li and the low concentration of the inclusions
is challenging, as many techniques, such as EDX, have low elemental
sensitivity and/or insufficient spatial resolution (Supporting Information).First, the same ROI previously
analyzed by EDX, on another dual-beam
instrument (Helios, Thermo Fisher), was localized and imaged in SEM
mode at normal incidence (Figure a). This acquisition was done while the SIMS extraction
system was retracted, allowing a quick approach to search for ROIs.
Second, the SIMS extraction optics was inserted, and 7Li, 55 Mn, 56Fe, and 63Cu ions were mapped
in SIMS mode on the same ROI (Figure b–e).
Figure 9
Al–Li alloys. (a) SE image (1536 ×
1092 pixels; electron
beam of 5 keV, 10 pA); (b–e) SIMS images (FoV 25 × 25
μm2, 512 × 512 pixels, dwell time per pixel
of 2.5 ms; Ga+ beam of 30 keV, 10 pA); (f) IHS image fusion[34] of the SE image (shown in (a)) with an overlay
of the two SIMS images of (c) and (e).
Al–Li alloys. (a) SE image (1536 ×
1092 pixels; electron
beam of 5 keV, 10 pA); (b–e) SIMS images (FoV 25 × 25
μm2, 512 × 512 pixels, dwell time per pixel
of 2.5 ms; Ga+ beam of 30 keV, 10 pA); (f) IHS image fusion[34] of the SE image (shown in (a)) with an overlay
of the two SIMS images of (c) and (e).The observations of intermetallic phases and the main phase are
in accordance with the previous results.[26] An FIB-SIMS analysis reveals segregation of Li at grain boundaries,
which can be confirmed by BSE data.[26] The
identification of the different phases (Al2CuLi, T1), as
described by Xu et al.,[26] is represented
in the Supporting Information Figure S4. Magnesium was detected in high-resolution images, despite its low
concentration (0.2%). Xu et al.[26] also
estimate the grain boundaries to have a width of ∼90 nm. The
composition of the main phase is much better spatially resolved in
the FIB-SIMS results here than in EDX, showing that they are not homogeneous
in composition but, rather, are made up of nanodomains (Figure c–e). To further push
the data treatment, SE and SIMS images were correlated; that is, the
raw intensities from SIMS images were combined with the SE image to
form a new image. To do this, a series of corresponding landmarks
was identified in both images. These were used to calculate the parameters
for an affine transformation, applicable either to the SE or to the
SIMS images. This overlay can be done by the addition and transparency
of the images or by data pixel fusion. In this context, Vollnhals
et al.[34] proposed different SE-sharpening
methods (Laplace fusion, Intensity–Hue–Saturation) to
provide a high-quality correlative SE-SIMS image. The image presented
in Figure f is obtained
by the Intensity-Hue-Saturation (IHS or HIS) sharpening technique
operating at the level of spatial frequencies and by including the
local information present in the high-resolution (SE) image into the
lower-resolution one (SIMS). In this study, the registered SIMS images
were fused with the SE image, using ImageJ and red (copper) and blue
(manganese) colormaps applied to the different SIMS data sets. Hence,
the nanograins are better resolved with respect to the signal dynamics
obtained by EDX (Supporting Information Figure S5) and are thus more helpful for the data interpretation.
Life Sciences
The FIB-SEM-SIMS enables one to do an
in situ investigation of both the morphology and the chemical composition
of biological samples, such as tissues and cells. This is particularly
interesting in toxicology and pharmacology. For example, metal nanoparticles,
halogenated pollutants, metal-based drugs, or halogenated drugs can
be localized at cellular and subcellular levels. Alternatively, molecules
of interest labeled with a metal, halogen, or less abundant isotope
can be traced by detecting the label. Consequently, the mechanisms
of action of these chemicals, and therefore their potential effect
on the body, can be investigated.Halogenated persistent organic
pollutants, such as brominated flame retardants or fluorinated substances,
are toxic chemicals related to many human diseases. Given its sensitivity
and spatial resolution, FIB-SEM-SIMS is perfectly suited to assessing
the fate of these pollutants.The BSE image in Figure a displays a cellular contrast,
well-known in biological microscopy,[35] due
to contrast agent staining. The SIMS images
in Figure b,c enable
us to localize a fluorinated environmental pollutant (in red) and
the cells (in green). The fused image in Figure d provides information about the uptake
and localization of the pollutant inside the cells. As the compound
is mainly localized in the cytoplasm of the cells, we could assume
it has an impact on their cytosol functions, such as signal transduction
between the cell membrane and the nucleus and organelles, metabolite
transport activities, cytokinesis, and metabolism. Obviously, complementary
toxicity tests are necessary to evaluate the toxicology of this compound
and the possible mechanisms of action. However, SIMS imaging reveals
useful information about localization from tissular to subcellular
levels that none of the usual toxicity tests could provide, which
makes it very attractive.
Figure 10
Intenstinal cells exposed to fluorinated toxicant:
(a) BSE image
(FoV of 70 × 70 μm2, 6144 × 4096 pixels,
dwell time per pixel 5 μs; electron beam 5 keV, 0.2 nA); (b,
c) SIMS images (FoV of 40 × 40 μm2, 512 ×
512 pixels, dwell time per pixel 4 ms; Ga+ beam 30 keV,
0.1 nA). (d) Overlay image.
Intenstinal cells exposed to fluorinated toxicant:
(a) BSE image
(FoV of 70 × 70 μm2, 6144 × 4096 pixels,
dwell time per pixel 5 μs; electron beam 5 keV, 0.2 nA); (b,
c) SIMS images (FoV of 40 × 40 μm2, 512 ×
512 pixels, dwell time per pixel 4 ms; Ga+ beam 30 keV,
0.1 nA). (d) Overlay image.One critical step for SIMS imaging of biological samples is their
preparation. As they are investigated under vacuum conditions, common
protocols for electron microscopy are usually applied. As for electron
microscopy, biological specimens should be processed in a way that,
when analyzed, they resemble their original living state as much as
possible. However, because of the numerous processing steps involved,
such as fixation, dehydration, and embedding, chemical information
might be altered (washout and/or redistribution) when the molecules
of interest are not reactive toward fixatives. Thus, the interpretation
of SIMS images requires the careful attention of the investigator.
In some cases, cryo-preservation can be favored over traditional resin
embedding, but this will undoubtedly induce other challenges. One
advantage in life science is that thin sections are usually analyzed,
which overcome topographic issues encountered for other applications
in SIMS.
Conclusion
A magnetic-sector double-focusing
SIMS system was developed and
successfully integrated into a dual-beam instrument as an add-on analytical
tool. In the chosen configuration, the SIMS extraction optics are
retractable and are inserted for SIMS operation directly above the
sample surface, the latter one being at the FIB-SEM instrument’s
eucentric point. This configuration allows for highest secondary ion
extraction efficiencies and, hence, the highest sensitivity, while
being able to access the region of interest in the sample with both
the ion and the electron beam. The FIB-SEM-SIMS system can be used
to generate mass spectra, 2D images, 3D images, and depth profiles.
Mass spectra can be recorded on a mass range from 1 to 400 amu with
a mass resolving power of above 400 and high dynamic range, while
allowing isotopic selectivity. The SIMS imaging results demonstrated
the possibility of recording high-resolution chemical maps with a
lateral resolution of 15 nm. These SIMS images can be stacked into
a 3D volume reconstruction, giving a full representation of the analyzed
volume. When using the FIB-SEM-SIMS for depth profiling, a good depth
resolution can be achieved when lowering the landing energy of the
Ga+ beam. A depth resolution of ∼4 nm was demonstrated
at 3 keV on a multilayer sample.Different materials science
application results have demonstrated
that this newly developed FIB-SEM-SIMS system can be used for high-resolution
high-sensitivity correlative investigations in current topics related
to energy storage and energy production devices (i.e., batteries,
solar cells). Moreover, the possibility of performing 3D volume SIMS
imaging with an excellent lateral and depth resolution opens interesting
prospects for investigating complex 3D architectures, such as in microelectronic
devices. Beyond materials science, the instrument can further be used
to address scientific questions in earth or life sciences.To
summarize, installing a high-performance magnetic sector SIMS
system on FIB-SEM instruments offers significant added value for a
large variety of applications, including highly sensitive analytics,
highest-resolution SIMS imaging (resolution of 15 nm) and depth profiling
(depth resolution of a few nm), in situ process control during patterning/milling,
and direct correlation of SIMS data with other analytical or imaging
data obtained on the same instrument, such as high-resolution SE images
or EDX spectra/data.
Authors: Songkil Kim; Artem Trofimov; Fouzia Khanom; Lewis Stern; William Lamberti; Robert Colby; David Abmayr; Alex Belianinov; Olga S Ovchinnikova Journal: Anal Chem Date: 2019-09-09 Impact factor: 6.986