Endodontic working length measurements of premolars and molars in high-resolution dental MRI: a clinical pilot study for assessment of reliability and accuracy.

Mousa Zidan1, Franz S Schwindling2, Alexander Juerchott1, Johannes Mente3, Holger Gehrig3, Mathias Nittka4, Zahra Hosseini5, Johann M E Jende1, Sabine Heiland1,6, Martin Bendszus1, Tim Hilgenfeld7.   

Abstract

OBJECTIVES: To prospectively assess the reliability and accuracy of high-resolution, dental MRI (dMRI) for endodontic working length (WL) measurements of premolars and molars under clinical conditions.
MATERIALS AND METHODS: Three-Tesla dMRI was performed in 9 subjects who also had undergone cone-beam computed tomography (CBCT) (mean age: 47 ± 13.5 years). A total of 34 root canals from 12 molars (4/8, upper/lower jaw; 22 root canals) and 11 premolars (2/9 upper/lower jaw; 12 root canals) were included. CBCT and dMRI datasets were reconstructed to visualize the root canal in one single slice. Subsequently, two radiologists measured the root canal lengths in both modalities twice in blinded fashion. Reliability and accuracy for both modalities were assessed using intraclass correlation coefficients (ICCs) and Bland-Altman analysis, respectively.
RESULTS: Reliability (intra-rater I/II; inter-rater) of dental MRI measurements was excellent and comparable to CBCT for premolars (0.993/0.900; 0.958 vs. 0.993/0.956; 0.951) and for molars (0.978/0.995; 0.986 vs. 0.992/0.996; 0.989). Bland-Altman analysis revealed a mean underestimation/bias (95% confidence interval) of dMRI measurements of 0.8 (- 1.44/3.05) mm for premolars and 0.4 (- 1.55/2.39) mm for molars. In up to 59% of the cases, the accuracy of dMRI for WL measurements was within the underestimation margin of 0 to 2 mm short of the apical foramen AF.
CONCLUSIONS: In vivo demonstration and measurement of WL are feasible using dMRI. The reliability of measurements is high and equivalent to CBCT. Nonetheless, due to lower spatial resolution and longer acquisition time, the accuracy of dMRI is inferior to CBCT, impeding its current use for clinical treatment planning. CLINICAL RELEVANCE: dMRI is a promising radiation-free imaging technique. Its reliability for endodontic working length measurements is high, but its accuracy is not satisfactory enough yet.
© 2022. The Author(s).

Entities:  

Keywords:  CBCT; Dmri; EAL; Endodontic working length; Magnetic resonance imaging; Working canal length

Year:  2022        PMID: 35861757     DOI: 10.1007/s00784-022-04636-1

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.606


  32 in total

1.  Roentgenographic and direct observation of experimental lesions in bone: I. 1961.

Authors:  I B Bender; Samuel Seltzer
Journal:  J Endod       Date:  2003-11       Impact factor: 4.171

2.  Accuracy of cone-beam computed tomography and periapical radiography in detecting small periapical lesions.

Authors:  Patrick Tsai; Mahmoud Torabinejad; Dwight Rice; Bruno Azevedo
Journal:  J Endod       Date:  2012-05-18       Impact factor: 4.171

Review 3.  Outcome of primary root canal treatment: systematic review of the literature -- Part 2. Influence of clinical factors.

Authors:  Y-L Ng; V Mann; S Rahbaran; J Lewsey; K Gulabivala
Journal:  Int Endod J       Date:  2007-10-10       Impact factor: 5.264

4.  Absorbed and effective doses from cone beam volumetric imaging for implant planning.

Authors:  T Okano; Y Harata; Y Sugihara; R Sakaino; R Tsuchida; K Iwai; K Seki; K Araki
Journal:  Dentomaxillofac Radiol       Date:  2009-02       Impact factor: 2.419

5.  A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results.

Authors:  P Mozzo; C Procacci; A Tacconi; P T Martini; I A Andreis
Journal:  Eur Radiol       Date:  1998       Impact factor: 5.315

6.  A Comparison of 2- and 3-dimensional Healing Assessment after Endodontic Surgery Using Cone-beam Computed Tomographic Volumes or Periapical Radiographs.

Authors:  Tom Schloss; David Sonntag; Meetu R Kohli; Frank C Setzer
Journal:  J Endod       Date:  2017-05-17       Impact factor: 4.171

Review 7.  The importance of cone-beam computed tomography in the management of endodontic problems: a review of the literature.

Authors:  Tadas Venskutonis; Gianluca Plotino; Gintaras Juodzbalys; Lina Mickevičienė
Journal:  J Endod       Date:  2014-10-03       Impact factor: 4.171

8.  Roentgenographic interpretation of experimentally produced bony lesions. I.

Authors:  S F Schwartz; J K Foster
Journal:  Oral Surg Oral Med Oral Pathol       Date:  1971-10

9.  The Recommendation of Cone-beam Computed Tomography and Its Effect on Endodontic Diagnosis and Treatment Planning.

Authors:  Sami Chogle; Maan Zuaitar; Ramzi Sarkis; Manal Saadoun; Anthony Mecham; Yihong Zhao
Journal:  J Endod       Date:  2019-12-11       Impact factor: 4.171

10.  Cone beam CT for dental and maxillofacial imaging: dose matters.

Authors:  Ruben Pauwels
Journal:  Radiat Prot Dosimetry       Date:  2015-03-23       Impact factor: 0.972