Jacob S Buatti1, Kyle J Gallagher1, Isaac Bailey2, Thomas Griglock2, Malcolm Heard1.
Abstract
PURPOSE: The purpose of this study was to evaluate similarities and differences in quality assurance (QA) guidelines for a conventional diagnostic magnetic resonance (MR) system and a MR simulator (MR-SIM) system used for radiotherapy.Entities:
Keywords: MR simulator; MR-SIM; TG-284; external beam radiation therapy; magnetic resonance imaging; magnetic resonance simulation; quality assurance
Mesh:
Year: 2022 PMID: 35851720 PMCID: PMC9359023 DOI: 10.1002/acm2.13730
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
Description of MR imaging studies for American College of Radiology (ACR) annual quality assurance (QA)
|
|
|
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|---|---|---|
| ACR Sagittal Localizer | 90° Spin Echo | 200 | 20 | 25 | 1 | 20 | NA | 1 | 256×256 | 55.8 | 0:53 |
| ACR Axial T1 | 90° Spin Echo | 500 | 20 | 25 | 11 | 5 | 5 | 1 | 256×256 | 55.8 | 2:10 |
| ACR Axial T2 | 90° Spin Echo | 2000 |
20 80 | 25 | 11 | 5 | 5 | 1 | 256×256 | 55.8 | 8:32 |
| Axial T1 Brain Scan | 90° Spin Echo | 500 | 10 | 23 | 11 | 5 | 5 | 1 | 256×190 | 49.8 | 0:59 |
| Axial T2 Brain Scan | 90°Fast Spin Echo | 3000 | 80 | 23 | 11 | 5 | 5 | 1 | 420×288 | 113.7 | 0:39 |
Repetition time (TR), bEcho time (TE), cField of view (FOV) (cm), dNumber of excitations (NEX), eBandwidth (BW) (kHz/pixel).
A Quality assurance testing frequency recommendations for MR technologists
| Frequency | |||
|---|---|---|---|
| Quality assurance (QA) tests for MR technologists | TG‐284 | ACR | |
| 1 | Functionality of patient communication and monitoring | D | Wv |
| 2 | Emergency cart or emergency couch release | D | Wv |
| 3 | Safety signage | D | Wv |
| 4 | Check bore for presence of foreign metal objects | D | Wv |
| 5 | External laser agreement with imaging plane | D | * |
| 6 | Transmit gain central frequency | D | W |
| 7 | Basic coil signal‐to‐no SNR check | D | W |
| 8 | Basic spatial fidelity check | D | W |
| 9 | High‐contrast spatial resolution | *P | W |
| 10 | Low‐contrast detectability | *P | W |
| 11 | Visual checklist | * | W |
| 12 | Artifact evaluation | *P | W |
*Test not identified in report; D, daily; Wv, weekly test included in ACR visual checklist (Table 2B) ; W, weekly. PTest performed by physicists for TG‐284 monthly QA (Table 2C).
**Test not identified in report; D, daily; W, weekly. PTests performed by physicists for TG‐284 monthly QA (Table 2C). Note: TG‐284 contains a visual checklist adapted from the ACR in the appendix of the report but does not identify the checklist as required component of QA or with a stated frequency.
***Test not identified in report; M, monthly; V, visual checklist ‐ tested weekly (Table 2B); A, annually.
A Image quality quality assurance (QA) testing tolerances from American College of Radiology (ACR) and TG‐284 using the ACR Large MRI Phantom
| Tolerances | ||
|---|---|---|
| QA tests | TG‐284 | ACR |
|
| ||
| Geometric accuracy | ±2 mm across 25‐cm field of view (FOV) with phantom >30 cm in diameter | ±2 mm using Large ACR Phantom |
| High contrast spatial resolution | ≤1.0 mm | ≤1.0 mm |
| Low contrast detectability |
21 (0.3 T) to 36 (1.5 T) total spokes, 40 spokes (3 T) |
≥7 spokes, 9 spokes preferred (1.5 T), ≥37 spokes (3 T) |
| Artifact evaluation | No observable artifacts | No observable artifacts |
| Percent image uniformity (head coil) |
≥87.5% for 1.5 T, ≥82.0% for 3.0 T |
≥87.5 % for 1.5 T, ≥82.0 % for 3.0 T |
| Percent signal ghosting | ≤2.5% | ≤2.5 % |
| Central frequency | Vendor‐specified | Vendor‐specified (i.e. ± 500 Hz |
| Transmitter gain | ±5% from baseline | Vendor‐specified (i.e. ± 0.05 dB |
| Flexible RF coil testing | Individual elements, exceeds minimum vendor‐specified threshold | Vendor‐specified |
Tolerance for Philips Ingenia 3.0 T system used for testing at our institution.
Tolerances are specified over all slices and for all 40 spokes per the ACR's measurement method for annual testing .
Vendor‐specified tolerance for Philips Ingenia 3.0 T system used for testing at our institution.
Abbreviations: DSV, diameter spherical volume; NA, not applicable.
Low‐contrast detectability measurements and evaluation with American College of Radiology (ACR) and TG‐284 quality assurance (QA) tolerances
| Image series | Slice number | Total number of spokes | ACR | TG‐284 |
|---|---|---|---|---|
| ACR T1 | 8‐11 | 40 | PASS | PASS |
| ACR T2 | 8‐11 | 38 | PASS | FAIL |
ACR passing criteria for LCD were 37 or more total spokes.
TG‐284 LCD passing criteria for a 3.0 T scanner were 40 total spokes.
Transmitter gain (TG) measurement and evaluation with American College of Radiology (ACR) and TG‐284 quality assurance (QA) tolerances
| Baseline TG Measurement (dB) | Annual TG Measurement (dB) | Relative Difference (%) | Absolute Difference (dB) | ACR | TG‐284 |
|---|---|---|---|---|---|
| 0.8252 | 0.8668 | 5.0 % | 0.0416 | PASS | PASS |
ACR transmitter gain tolerance was ±5.0 dB from the baseline.
TG‐284 transmitter gain tolerance was ±5 % from the baseline.