Literature DB >> 35845544

A novel clinical dynamic stereopsis assessment based on autostereoscopic display system.

Yiyao Wang1, Jing Zhong1, Mengyi Cheng2, Jijing Li1, Ke Ma1, Xiaoqing Hu1, Naiyang Li3, Haowen Liang2, Zhengyuan Zhu4, Jianying Zhou2, Jin Yuan1.   

Abstract

Background: Although it is recognized that dynamic stereopsis is vital in daily life, there is still room for improvement in assessment methods. A novel clinical dynamic stereopsis assessment method was created based on an autostereoscopic display system that did not require additional auxiliary glasses. This study evaluated the optical parameters and clinical performance of the autostereoscopic display system for clinical dynamic stereopsis assessment.
Methods: The autostereoscopic dynamic stereopsis assessment device was based on a directional backlight technology. Experiment 1 was performed under the same environmental conditions to compare luminance, crosstalk, and spectrum between the autostereoscopic dynamic stereopsis assessment device and the conventional dynamic random-dot stereopsis measuring instrument. Experiment 2 was an observational, analytic, cross-sectional study involving 135 healthy participants, each of whom was asked to complete measurements on both the autostereoscopic and conventional devices in random order. Stereo acuity, operating time, acceptance, and visual fatigue scores were recorded for clinical evaluation.
Results: The autostereoscopic device had brighter luminance (139 and 140 cd/m2 for 2 eyes, respectively), lower crosstalk (4.50% for both eyes), and higher color restoration degree than those of the conventional instrument. Clinically, the novel dynamic stereopsis assessment was as accurate as the traditional method [170" (0.00") and 170" (0.00") respectively; P=0.317], and with more efficiency (166±58.9 and 298±116 s, respectively; P<0.001), higher acceptance (3.36±0.93 and 2.02±0.59 points, respectively; P<0.001), lesser fatigue (0.27±0.46 and 0.73±0.66 points, respectively; P<0.001). The autostereoscopic dynamic stereopsis assessment device with brighter luminance, lower crosstalk, and higher color restoration degree was more effective than the traditional instrument at displaying dynamic clues for clinical dynamic stereopsis assessment. Furthermore, its high-quality image and user-friendly interface provided accurate assessment results in all 3 dynamic stereopsis assessment task conditions, with a higher level of acceptance and lesser visual fatigue, than the traditional assessment method. Conclusions: The autostereoscopic device has excellent functions in both optical parameters and clinical performance, and therefore has the potential to be applied and popularized in future assessments. 2022 Annals of Translational Medicine. All rights reserved.

Entities:  

Keywords:  Dynamic stereopsis; autostereoscopy; stereo acuity assessment

Year:  2022        PMID: 35845544      PMCID: PMC9279808          DOI: 10.21037/atm-21-6700

Source DB:  PubMed          Journal:  Ann Transl Med        ISSN: 2305-5839


Introduction

Stereopsis is the perceptual ability to accurately judge the three-dimensional (3D) spatial positions of objects. It relies on precise coordination of the visual pathway, including binocular stimulus, fusion function, and visual perception by brain neural networks (1). A reduced level of stereopsis negatively impacts an individual’s ability to complete daily tasks (2). Therefore, stereopsis assessment plays an integral role in the ophthalmic examination. Furthermore, it is important to assess stereo acuity both statically and dynamically (3,4). Clinical stereopsis testing involves 2 methods. One method uses static stimuli with a given static parallax (a depth cue that can be perceived by binoculars); the other uses dynamic stimuli, which include the movement or depth change of an object. Deficient performance in the observation of depth and motion could represent an impairment in stereopsis (5). A classical dynamic stereopsis program was generated by random-dot stereograms and kinematographs (6). In our previous study, this computer-programmed software for random-dot stereopsis measurements was displayed on a 13.3-inch color monitor with a 1.6-GHz dual-core Intel Core i5, 256 GB (MacBook Air®; Apple, Cupertino, CA, USA) using JAVA programming language (7). Stereopsis can also be assessed by measuring the depth change rate function in response to dynamic random-dot stimuli (8). In 2015, Allen et al. (9) developed a novel approach in which the investigators tested stereopsis by the change detection based on parallax and the interocular velocity difference. Although methods to assess dynamic stereopsis have been constantly updated, instruments typically require glasses (red-blue or polarized) to allow binocular viewing of separate colored or polarized images, which could generate uncomfortable and unnatural visual experiences for the patient (10,11). Furthermore, the accuracy of stereopsis examination is influenced by several instrumental parameters such as luminance, crosstalk, and spectrum (12,13). Improving the testing instruments’ hardware is one way to optimize stereoscopic measurements. In recent years, near-eye 3D displays have experienced rapid growth and shown great promise in a variety of applications, such as entertainment, education, and health care (14). Presenting separate images with parallax in front of the viewer’s eyes without additional auxiliary equipment is the classical principle of computer stereoscopy (10). In our previous studies, we developed an interactive, autostereoscopic screening tool based on a directional backlight technology and evaluated its feasibility for visual function assessments (15). In this study, we combined the autostereoscopic display system based on directional backlight technology and dynamic stereopsis assessment. The research hypothesis was that the autostereoscopic display system can better display dynamic clues for clinical dynamic stereopsis assessment and has the accuracy equivalent to the conventional dynamic random-dot stereopsis measuring instrument. We present the following article in accordance with the STROBE reporting checklist (available at https://atm.amegroups.com/article/view/10.21037/atm-21-6700/rc).

Methods

The autostereoscopic dynamic stereopsis assessment device

The autostereoscopic dynamic stereopsis assessment device was based on the directional backlight technology, which presented the left or right eye parallax image in separate periods on a liquid crystal display (LCD) (16,17). This 24-inch LCD panel had a resolution of 1,920*1,080 pixels and a refresh rate of 120 Hz. Since human eyes cannot distinguish intervals less than 30 Hz, both the left and right eyes could view different images at the same time with this device, and a stereo viewing effect was formed in the brain ().
Figure 1

The schematic diagram of the directional backlight autostereoscopic display. (A) Graph of directional backlight technology. Different images were presented to the left or right eye in separate periods. (B) Diagram of eye-tracking technology. Eye-tracking cameras helped locate the participant’s eyes for accurate assessment. LCD, liquid crystal display.

The schematic diagram of the directional backlight autostereoscopic display. (A) Graph of directional backlight technology. Different images were presented to the left or right eye in separate periods. (B) Diagram of eye-tracking technology. Eye-tracking cameras helped locate the participant’s eyes for accurate assessment. LCD, liquid crystal display. There are three patterns in this autostereoscopic dynamic stereopsis assessment device including motion-only, depth-only, and motion-depth. The test picture consisted of black (0, 0, 0) dots and white (255, 255, 255) dots (red-green-blue relative intensity format). The test target of each pattern was a block composed of 60×50 pixels with a movement speed of 10 fps (frames per second), therefore each eye saw a change in location every 1/10 of a second. In addition, an intelligent eye-tracking camera in the device tracked human eyes to provide an accurate directional projection of light () (18,19). Mobile phones were connected via Bluetooth or wirelessly and used as the operating terminal for participants to complete the tests.

Dynamic stereopsis stimulus and assessment task

Slightly different left and right images presented by LCD could cause binocular parallax, which generated the effect of stereopsis. The parallax calculation formula was as follows: where p was the parallax, n was the pixel difference between the 2 eyes images, L was the width of the display screen, and q was the horizontal pixels of the display screen (). The dynamic stereo acuity calculation formula was as follows:
Figure 2

Graph of calculating dynamic stereo acuity. LCD, liquid crystal display.

Graph of calculating dynamic stereo acuity. LCD, liquid crystal display. where α was the dynamic stereo acuity and d was the viewing distance between the participant and the screen (). This novel autostereoscopic dynamic stereopsis assessment device could present 4 dynamic stereo acuity levels at 170", 340", 510", and 680". Theoretically, at least 2 possible methods of generating dynamic stereopsis were available: one mechanism that detected binocular disparity at different times (depth only) and another one that detected interocular velocity differences (motion only) (20). Therefore, 3 patterns were involved in our novel autostereoscopic dynamic stereopsis assessment device: motion-only, depth-only, and motion-depth, from which the depth and motion cues could be included or omitted independently. Motion-only pattern: the 4 blocks had an equal depth, in which 3 blocks performed the same horizontal displacement movement, while the fourth box performed a Z-shaped movement ().
Figure 3

Drawing of 3 assessment patterns in dynamic stereopsis by autostereoscopic display system: (A) motion-only pattern; (B) depth-only pattern; (C) motion-depth pattern.

Drawing of 3 assessment patterns in dynamic stereopsis by autostereoscopic display system: (A) motion-only pattern; (B) depth-only pattern; (C) motion-depth pattern. Depth-only pattern: the 4 blocks executed the same Z-shaped movement, where 3 blocks had the same depth, while the fourth seemed further (). Motion-depth pattern: the 3 blocks with the same depth executed the same horizontal displacement movement, and the fourth performed a Z-shaped block movement and a different depth ().

Optical parameter measurements

Luminance measurement

All luminance measurements were operated using a calibrated PR655 luminance meter (Photo Research, Inc., Chatsworth, CA, USA). The brightness filtered by red and blue glasses was measured in the conventional instrument. The procedure was repeated in the left and right viewing zones on the autostereoscopic device.

Crosstalk detection

The crosstalk was measured by a calibrated PR655 luminance meter (Photo Research, Inc., USA). The ratio of brightness filtered by blue glasses in red versus blue image mode was the crosstalk of blue glass; the crosstalk of red glass was obtained similarly. In the autostereoscopic device, the crosstalk was the ratio of crosstalk light and main light in the left or right viewing zone.

Spectrum test

Spectrum was assessed using a calibrated spectrograph (USB2000+; Ocean Optics, Shanghai, China). It was measured through the red and blue glasses respectively in the conventional instrument. The same procedure was repeated in the left and right viewing zones on the autostereoscopic device.

Participants

A total of 135 healthy participants at the Zhongshan Ophthalmic Center were recruited in this observational, analytic, cross-sectional study. The criteria of inclusion were as follows: patients with the best-corrected visual acuity of <0.1 logarithms of the minimum angle of resolution (logMAR) measured with the Early Treatment Diabetic Retinopathy Study (EDTRS) chart, and without ongoing or previous history of ocular surgery. The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013) and approved by the Ethics Committee of Zhongshan Ophthalmic Center, Sun Yat-sen University (No. 2016MEKY032). Informed consent was obtained from all participants. This study was registered at clinicaltrials.gov (ID NCT04152122).

Dynamic stereopsis assessment for participants

The specific procedure for the autostereoscopic dynamic stereopsis assessment device began when the participant sat in front of the device and started the software. Eye-tracking cameras identified their eyes and determined whether she/he was in an accurate position (). After entering their basic information, the participant clicked “Start” on a mobile phone connected to the device (). The test then ran automatically in 3 patterns. The participant chose the object that was different from the others and pressed the corresponding button on the phone (). This interactive software could tell the participant whether his/her answer was correct (). The stereo acuity was shown on the mobile phone after the completion of all tests ().
Figure 4

The instruction of the dynamic stereopsis assessment. (A) Diagram of identifying whether the participant was in the accurate position. (B) The mobile phone user interface: entering the participant’s basic information. (C) The mobile phone user interface: selecting the corresponding answer. (D) An example of stereopsis assessment when the answer is right. (E) An example of stereopsis assessment when the answer is wrong. (F) The mobile phone user interface: the result of dynamic stereopsis.

The instruction of the dynamic stereopsis assessment. (A) Diagram of identifying whether the participant was in the accurate position. (B) The mobile phone user interface: entering the participant’s basic information. (C) The mobile phone user interface: selecting the corresponding answer. (D) An example of stereopsis assessment when the answer is right. (E) An example of stereopsis assessment when the answer is wrong. (F) The mobile phone user interface: the result of dynamic stereopsis. The conventional instrument we chose was a computer-programmed random-dot stereopsis measuring software displayed on a 13.3-inch color monitor with a 1.6-GHz dual-core Intel Core i5, 256 GB (MacBook Air®; Apple, USA) using JAVA programming language. The specific procedure was performed as described in our previous study (7). Test protocols for the 2 devices were the same. Participants were required to select the correct answer continuously 10 times at 1 stereoscopic level; those participants with 8 correct answers (80%) could be allowed to move to the next level. The examination started at 680" and raised the stereoscopic difficulty from 680" to 170" until the participant was unable to advance further.

Clinical repeatability evaluation

Ten randomly selected participants were enrolled in dynamic stereopsis assessments twice (with intervals of 24 hours) with the autostereoscopic device by a single examiner. The repeatability was evaluated by comparing the 2 testing results.

Clinical accuracy assessment

All 135 participants were enrolled to compare the screening accuracy between the autostereoscopic dynamic stereopsis assessment device and the conventional dynamic random-dot stereopsis measuring instrument. Each participant was asked to complete measurements on both devices in random order. A 10-minute washout period was included between the measurements.

Clinical acceptability judgment

The acceptance was evaluated by a 5-point grading questionnaire among the 135 participants. The higher the score, the higher the degree of favor, where 0 point meant “refuse it”, 1 points meant “do not refuse it”, 2 points meant “kind of like it”, 3 points meant “pretty much like it” and 4 points meant “love it”.

Clinical asthenopia rating

The asthenopia was also evaluated by a 5-point grading questionnaire among the 135 participants. The higher the score, the higher the degree of fatigue, where 0 point meant “no fatigue”, 1 points meant “mild fatigue”, 2 points meant “sub-moderate fatigue”, 3 points meant “moderate fatigue”, and 4 points meant “severe fatigue”.

Clinical efficiency calculation

The testing time spent in the process between the novel clinical dynamic stereopsis assessment based on an autostereoscopic display system and the conventional dynamic random-dot stereopsis measuring instrument was assessed for work efficiency among the 135 participants. The less time spent, the more efficient the device.

Statistical analysis

All statistical analyses were performed with the software SPSS 22.0 (IBM Corp., Armonk, NY, USA). The Shapiro-Wilk test was used to examine the distribution of the data. When the data met a normal distribution (such as efficiency, acceptability, and asthenopia), a paired t-test was performed to evaluate the differences between the 2 measurements. If the data did not meet a normal distribution (such as a dynamic stereoscopic score), a paired Wilcoxon test was performed to evaluate repeatability and accuracy between novel versus conventional devices. The test level was set as a =0.05, and P<0.05 demonstrated statistical significance. Cases with missing data were excluded from the analyses.

Results

The optical parameter comparison between the autostereoscopic device and the conventional instrument

We focused on comparing luminance, crosstalk, and the spectrum between the autostereoscopic dynamic stereopsis assessment device based on a directional backlight technology system and the conventional dynamic random-dot stereopsis measuring instrument under the same environment. The autostereoscopic device projected a luminance of 139 cd/m2 to the left eye and 140 cd/m2 to the right eye. In the conventional instrument, the luminance dropped to 3.72 and 4.32 cd/m2, respectively ().
Table 1

Comparison of luminance and crosstalk between the autostereoscopic device and the conventional instrument for measuring dynamic stereopsis

Optical parameterAutostereoscopic dynamic stereopsis assessment deviceConventional dynamic random-dot stereopsis measuring instrument
Luminance for left eye (cd/m2)1393.72
Luminance for right eye (cd/m2)1404.32
Crosstalk for left eye4.50%27.8%
Crosstalk for right eye4.50%10.0%

cd/m2, candela/m2.

cd/m2, candela/m2. The crosstalk felt by both eyes was 4.50% in the autostereoscopic device; in the conventional instrument, the crosstalk felt by the left eye was 27.8% and the crosstalk felt by the right eye was 10.0% (). The wavelength of the spectrum revealed no changes in the color distribution in the absence of filtering glasses with the autostereoscopic device (). With the conventional instrument, participants were only sensitive to the light at around 450 nm for one eye and around 625 nm for the other eye, which could lead to chromatic aberration in binocular vision ().
Figure 5

Comparison of spectral measurements between the autostereoscopic device and the conventional instrument. (A) Spectrum of the autostereoscopic device. (B) The spectrum of the conventional instrument.

Comparison of spectral measurements between the autostereoscopic device and the conventional instrument. (A) Spectrum of the autostereoscopic device. (B) The spectrum of the conventional instrument.

The clinical evaluation between the autostereoscopic device and the conventional instrument

The clinical evaluation included a total of 135 healthy participants (103 females and 32 males) whose average age was 28.2±5.27 years (range: 18.0 to 45.0 years). The logMAR visual acuity of the participants, measured with the EDTRS chart was 0.04±0.02.

Clinical repeatability evaluation

Ten participants were selected to undergo a repeatability test. The stereo acuities from the first and the second tests were 170" in 3 patterns ().
Figure 6

Comparison of the 1st and 2nd autostereoscopic dynamic stereopsis assessment in (A) motion-depth pattern, (B) motion-only pattern, and (C) depth-only pattern.

Comparison of the 1st and 2nd autostereoscopic dynamic stereopsis assessment in (A) motion-depth pattern, (B) motion-only pattern, and (C) depth-only pattern.

Clinical accuracy assessment

The stereo acuities (MD) measured with the autostereoscopic device and the conventional instrument were 170" (0.00") and 170" (0.00"), respectively. The difference was not statistically significant (Z =−1.000; P=0.317). As for the motion-only pattern (M), the stereo acuities were 170" (0.00") with the autostereoscopic device and 170" (0.00") with the conventional instrument (Z =−0.447; P=0.656). In the depth-only mode, stereo acuities were 170" (0.00") with the conventional instrument and 170" (0.00") with the conventional instrument. Again, the test results had no significant difference between the 2 devices (Z =−0.846; P=0.400) ().
Figure 7

The stereo acuities distribution comparison of the autostereoscopic device and the conventional instrument includes (A) motion-depth pattern, (B) motion-only pattern, (C) depth-only pattern in 135 healthy participants. Comparison of scores between the autostereoscopic device and conventional instrument for (D) acceptability, (E) asthenopia, and (F) efficiency in the 135 healthy participants. Three asterisks *** indicate a significance level of P<0.001.

The stereo acuities distribution comparison of the autostereoscopic device and the conventional instrument includes (A) motion-depth pattern, (B) motion-only pattern, (C) depth-only pattern in 135 healthy participants. Comparison of scores between the autostereoscopic device and conventional instrument for (D) acceptability, (E) asthenopia, and (F) efficiency in the 135 healthy participants. Three asterisks *** indicate a significance level of P<0.001.

Clinical acceptability judgment

The autostereoscopic device had an acceptability score of 3.36±0.93 points, indicating that most subjects were in favor of this device. The conventional instrument only had a score of 2.02±0.59 points, meaning that this type of instrument was only acceptable for most subjects. The difference was statistically significant (P<0.001), indicating that the novel device was more popular than the conventional device ().

Clinical asthenopia rating

Asthenopia was evaluated based on the questionnaire. The score of the novel device was 0.27±0.46 points, and the conventional instrument was 0.73±0.66 points (P<0.001). This result suggested that the autostereoscopic device did not induce considerable visual fatigue ().

Clinical efficiency calculation

The testing times were 166±58.9 s for the autostereoscopic device and 298±116 s for the conventional instrument (P<0.001) ().

Discussion

In this study, we compared the optical parameters and clinical indexes between the novel clinical dynamic stereopsis assessment based on an autostereoscopic display system and a conventional dynamic random-dot stereopsis measuring instrument. The findings showed that the autostereoscopic device had excellent functions in both physical condition and clinical performance, suggesting the potential for future screening applications and popularization. Optical parameters including luminance, crosstalk, and spectrum of screening equipment have a huge impact on displaying dynamic clues for clinical dynamic stereopsis assessment (12,13). Existing research shows that stereopsis is diminished over a range of illumination from photopic to scotopic light (21). In our study, the luminance of the autostereoscopic device was clearly within the photopic range, which was classified as good interior lighting (250 cd/m2) (22). However, when participants wore filter glasses, the perceived luminance dropped dramatically, with a 97.33% decrease in the left eye and a 96.91% decrease in the right eye, which caused a loss of stereopsis effects. An early study by Mueller and Lloyd (23) reported that the threshold of stereopsis decreased with changes in the field luminance from 1 cd/m2 to the lowest values (3×10-4 cd/m2). Many investigators now believe that stereopsis mainly depends on the level of illumination. Reduced luminance produces a temporal low-pass of neural response, resulting in secondary amplitude reduction and delay, which are particularly disruptive for stereopsis (24). The autostereoscopic device did not suffer a luminance decrease as the conventional instruments did, which demonstrated that the autostereoscopic device had a better ability for dynamic stereopsis assessment. Screen performance of devices were compared: crosstalk de-camouflages test images. High levels of crosstalk cause a loss of fidelity in stereoscopic images. Therefore, it is crucial to maintain lower crosstalk in the stereo acuity assessment (25,26). Previous research revealed that 45° tilting of the head by participants wearing filter glasses resulted in 100% crosstalk in a random-dot stereopsis test (27). Degradation of stereopsis test accuracy by crosstalk may be associated with a violation of the fundamental principle of the stereo acuity assessment. Therefore, the examinations must exclude monocular cues (e.g., texture gradient, relative size) that would enable participants to perform correctly in the absence of stereo acuity. Crosstalk de-camouflages test images that should be detected only via stereopsis (28,29). In our study, the crosstalk of the new autostereoscopic device was significantly lower than that of the conventional instruments, suggesting that test results with the new approach would be more realistic. The binocular chromatic aberration caused by filter glasses can also cause serious problems in stereo acuity assessment (30). Conventionally, the right eye observes test images through a red filter which appears red, while the left eye observes the images through a blue filter which appears blue. Although chromatic information could assist in stereopsis (31,32), previous studies have suggested that similarity, rather than difference, of color would extend the range of perceived stereopsis (33). The anti-correlations in color contrast degrades stereopsis (34). Kingdom et al. (35) concluded that it was unnecessary to pre-filter the image into separate color maps before the image underwent stereopsis processing for the visual system. Reports also tend to agree that visual features are somehow given a “label” based on their chromaticities and that stereoscopic matches with the same label are favored over those without (33,36). The conventional instruments had an obsolete chromatic stereopsis mechanism with limited stereo acuity (37). In contrast, the novel autostereoscopic device had a high color restoration degree that presented the dynamic stereopsis as closer to the actual state, which enabled more accurate dynamic assessments of stereopsis. In the previous research, there was a dynamic stereoscopic measurement based on an autostereoscopic device called Nintendo’s 3DS (38). Compared with our device, there are slight differences in the technical principles of dynamic stereoscopic presentation. The Nintendo’s 3DS is a parallax-barrier autostereoscopic device with no detection floor of high-grade stereo. But would degrade in half the resolution and brightness of the dynamic image received by each eye. Our device is based on directional backlight technology. It ensures that each eye receives the full resolution and brightness of the dynamic image, which could better display dynamic clues for clinical dynamic stereopsis assessment. Clinically, our autostereoscopic device demonstrated excellent accuracy and repeatability in a three-pattern evaluation. Moreover, according to the stereo acuity distribution comparison, 26 of 135 healthy participants with an abnormal range of stereo were recruited in this study. For this group of people, the accuracy of the device remained stable at a high level. Therefore, we could conclude that this autostereoscopic device had satisfactory precision in both normal stereo acuity and reduced stereo acuity, which was satisfied the requirements of dynamic stereopsis variability in the population assessment. In terms of subjective evaluations, the autostereoscopic device had a lower fatigue score. We believe that this is mainly due to the advantages of optical parameters compared to conventional instruments. The brighter luminance, lower crosstalk, and higher color restoration degree of directional backlight technology provide separate images to each eye which allows visual representation to be closer to the human senses (39), offering a better stereopsis performance and lessening visual fatigue. Furthermore, under the same testing protocol, besides higher acceptance and efficiency, the conventional instrument had a sizeable standard deviation of testing time (SD =116 s), indicating that there were substantial differences in adaptation of this examination among participants. In contrast, the autostereoscopic device had a lower standard deviation (SD =58.9 s), suggesting that participants could better understand and master the examination process. We assume that this difference would be mainly caused by the optimization of a user-friendly screening interface. The autostereoscopic device is equipped with interactive software and an eye-tracking camera. Participants can complete the assessments independently without the examiner's guidance. This not only saves time but adds more fun to the examination procedure, thus increasing efficiency and acceptance. The novel clinical dynamic stereopsis assessment based on an autostereoscopic display system with excellent optical parameters, such as brighter luminance, lower crosstalk, and higher color restoration degree, could better display dynamic clues for clinical dynamic stereopsis assessment while maintaining equal accuracy as the conventional dynamic random-dot stereopsis measuring instrument does. Moreover, the novel assessment is of higher patient acceptance and lesser visual fatigue compared with the conventional assessment. As for the limitations of this study, we believe that the optical parameters, such as resolution, could be further optimized to improve the screening ability such as presenting finer grade stereo and narrower spaced options. A thinner and lighter device that is portable can improve the dynamic stereopsis assessment experience either.

Conclusions

The novel clinical dynamic stereopsis assessment based on an autostereoscopic display system with brighter luminance, lower crosstalk, and higher color restoration contributes to the accurate presentation of dynamic stereopsis. Furthermore, its high-quality image and user-friendly interface enable the production of accurate screening results in all 3 dynamic stereopsis assessment task conditions with a higher level of acceptance and reduced visual fatigue. We believe that the autostereoscopic device, with excellent functions in both physical condition and clinical performance, has the potential for future screening applications and popularization. The article’s supplementary files as
  33 in total

1.  On the apparent collapse of stereopsis in random-dot-stereograms at isoluminance.

Authors:  F A Kingdom; D R Simmons; S Rainville
Journal:  Vision Res       Date:  1999-06       Impact factor: 1.886

Review 2.  Single vision during ocular deviation in intermittent exotropia.

Authors:  Ignacio Serrano-Pedraza; Michael P Clarke; Jenny C A Read
Journal:  Ophthalmic Physiol Opt       Date:  2011-01       Impact factor: 3.117

3.  Stereopsis and mean luminance.

Authors:  Alexandre Reynaud; Jiawei Zhou; Robert F Hess
Journal:  J Vis       Date:  2013-09-03       Impact factor: 2.240

4.  High-quality autostereoscopic display with spatial and sequential hybrid control.

Authors:  Jiahui Wang; Haowen Liang; Hang Fan; Yangui Zhou; Peter Krebs; Jianbang Su; Yutao Deng; Jianying Zhou
Journal:  Appl Opt       Date:  2013-12-10       Impact factor: 1.980

5.  Binocular depth-from-motion in infantile and late-onset esotropia patients with poor stereopsis.

Authors:  M Maeda; M Sato; T Ohmura; Y Miyazaki; A H Wang; S Awaya
Journal:  Invest Ophthalmol Vis Sci       Date:  1999-11       Impact factor: 4.799

6.  Stereopsis and positional acuity under dark adaptation.

Authors:  M S Livingstone; D H Hubel
Journal:  Vision Res       Date:  1994-03       Impact factor: 1.886

7.  Optimal Combination of the Binocular Cues to 3D Motion.

Authors:  Brian Allen; Andrew M Haun; Taylor Hanley; C Shawn Green; Bas Rokers
Journal:  Invest Ophthalmol Vis Sci       Date:  2015-11       Impact factor: 4.799

8.  Measurement of crosstalk in stereoscopic display systems used for vision research.

Authors:  Daniel H Baker; Milena Kaestner; André D Gouws
Journal:  J Vis       Date:  2016-12-01       Impact factor: 2.240

Review 9.  Stereopsis: are we assessing it in enough depth?

Authors:  Anna R O'Connor; Laurence P Tidbury
Journal:  Clin Exp Optom       Date:  2018-01-27       Impact factor: 2.742

10.  Detection of Static and Dynamic Stereopsis after Femtosecond Laser Small Incision Lenticule Extraction for High Myopia.

Authors:  Aiqun Xiang; Chu Hang; Xiaoying Wu; Yewei Yin; Yanyan Fu; Ying Lu; Kaixuan Du; Tu Hu; Li Yan; Dan Wen
Journal:  J Ophthalmol       Date:  2021-06-11       Impact factor: 1.909

View more
  1 in total

1.  The impact of autostereoscopic screens on assessing depth perception.

Authors:  Robert W Arnold
Journal:  Ann Transl Med       Date:  2022-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.