| Literature DB >> 35845540 |
Yimin Shen1, Dongmei Liu1, Yong Wang1, Jun Cao1, Shuaishuai Zhang1, Hui Wen1, Qiuqiu Dong1, Dong Zheng1, Jun Qiu1.
Abstract
Background: The accuracy of low-value platelet (PLT) testing is a key reference for clinical decision-making regarding PLT transfusion or surgery. This study aimed to evaluate the capability to detect low-value PLT of the 8 times optical platelet counting (PLT-8X) mode of the BC-6800Plus auto hematology analyzer.Entities:
Keywords: 8 times optical platelet counting (PLT-8X); BC-6800Plus; performance evaluation; platelet counting reference method (PLT counting reference method); thrombocytopenia
Year: 2022 PMID: 35845540 PMCID: PMC9279826 DOI: 10.21037/atm-22-1501
Source DB: PubMed Journal: Ann Transl Med ISSN: 2305-5839
Figure 1Precision profiles of PLT-I and PLT-8X on BC-6800Plus. CV%, coefficient of variation; PLT-I, the impedance method for platelet counting; PLT-8X, 8 times optical platelet counting.
Precision of different range of low-value PLT
| PLT range | PLT-8X | PLT-I | P value | |||||
|---|---|---|---|---|---|---|---|---|
| N | SD | CV% | N | SD | CV% | |||
| 0 to 10×109/L | 10 | 0.238 | 5.23% | 12 | 0.724 | 23.66% | 0.0021 | |
| 10×109/L to 20×109/L | 7 | 0.569 | 3.87% | 9 | 1.429 | 9.63% | 0.0040 | |
| 20×109/L to 30×109/L | 6 | 0.646 | 2.73% | 4 | 1.082 | 4.68% | 0.0439 | |
| 30×109/L to 50×109/L | 17 | 0.959 | 2.36% | 15 | 2.241 | 6.02% | <0.0001 | |
PLT, platelet; PLT-8X, 8 times optical platelet counting; PLT-I, the impedance method for platelet counting; SD, standard deviation; CV%, coefficient of variation.
Figure 2Scatter plot of BriCyte E6 reference method for flow PLT counting, and methodological comparison plot between Passing-Bablok and Bland-Altman. (A) Scatter plot of BriCyte E6 reference method for flow PLT counting. Where P1 area indicates the RBC count and its proportion, P2 area indicates the PLT count and its proportion, and P3 area indicates the number of particles in the overlapping area and its proportion. The RBC/PLT ratio (R) = (P1 + P3)/(P2 + P3). PLT = RBC × 1,000/R. (B) Plot of Passing-Bablok regression analysis. (C) Plot of Bland-Altman method comparison. FITC, fluorescein isothiocyanate; FSC, forward scattered; CI, confidence interval; LoA, limit of agreement; PLT, platelet; RBC, red blood cell.
Figure 3Linear data graph of PLT-8X in the low-value ranges of PLT. (A) The PLT-8X linear analysis of PLT in the range of 0 to 100. (B) The PLT-8X linear analysis of PLT in the range of 0 to 30. X indicates the true concentration of platelets prepared; Y indicates the result of the platelet test from BC-6800Plus. PLT-8X, 8 times optical platelet counting; PLT, platelet.
Figure 4Frequency distribution diagram of LoD data of PLT-I and PLT-8X. The frequency distribution of (A) PLT-I and (B) PLT-8X. LoB, limit of blank; LoD, limit of detection; PLT-I, the impedance method for platelet counting; PLT-8X, 8 times optical platelet counting.