| Literature DB >> 35837608 |
Luigi Pannone1, Ivan Eltsov1,2, Robbert Ramak1, David Cabrita2, Marc Verherstraeten3, Anaïs Gauthey1, Antonio Sorgente1, Cinzia Monaco1, Ingrid Overeinder1, Gezim Bala1, Alexandre Almorad1, Erwin Ströker1, Juan Sieira1, Pedro Brugada1, Mark La Meir4, Gian-Battista Chierchia1, Carlo de Asmundis1.
Abstract
Background: New technologies for ablation procedures are often produced by different companies with no cross-compatibility out of the box. This is not a negligible clinical problem since those separately developed devices are often used together. The aim of this study was to develop a bench-testing method to assess compatibility between the DiamondTemp ablation system (DTA) and the Rhythmia electroanatomic mapping system (EAM).Entities:
Keywords: DiamondTemp ablation system; Rhythmia electroanatomic mapping system; cardiac arrhythmias; catheter ablation; universal compatibility
Year: 2022 PMID: 35837608 PMCID: PMC9274945 DOI: 10.3389/fcvm.2022.917218
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
Figure 1DiamondTemp and Rhythmia connectivity configurations. (A) Setup 3.1. An EPT GCB and Rhythmia Maestro GCB (Maestro GCB, native Rhythmia setup). (1) 14-pin twist connector to RFG via the GCC; (2) 9-pin quick connector to DTC via the GCC to the DTC to RFG cable; (3) DTA RFG; (4) DTA GCC; (5) DTC; (6) DTC to RFG cable; (7) IC out cable; (8) 9-pin quick connector of the Maestro GCB; and (9) Connector from the Maestro GCB to the Rhythmia MS. (B) Setup 3.2. This configuration shows the best accuracy results—The catheter connection is connected to Rhythmia GCB using Genconnect cable and then to MDT GCB as shown in the picture. (1) 14-pin twist connector to RFG via the GCC; (2) 9-pin quick connector to DTC via the GCC to the DTC to RFG cable; (3) DTA RFG; (4) DTA GCC; (5) DTC; (6) DTC to RFG cable; (7) IC out cable; (8) 9-pin quick connector of the Maestro GCB; and (9) Connector from the Maestro GCB to the Rhythmia MS. (C) Setup 3.3. The MDT GCB-E and IC out via the MDT GCB-E directly to the Rhythmia IC in at box 1 (pin A61 to A64) (1) 14-pin twist connector to RFG via the GCC; (2) 9-pin quick connector to DTC via the GCC to the DTC to RFG cable; (3) DTA RFG; (4) DTA GCC; (5) DTC; (6) DTC to RFG cable; (7) IC out cable; and (8)Rhythmia IC out to the EP recording system.
Figure 2Accuracy testing setup “in vitro”. Dedicated testing phantom with all EAM cables connected, IntellaMap Orion catheter, and DiamondTemp ablation catheter submerged in the saline within the phantom.
Figure 3Representation of both Orion and DT catheter in the 3D map on the EAM. This snapshot is showing catheter shift during RF application with configuration 3.1 and distance measurement of 19 mm for configuration 3.3.
Functional and safety parameters with different setups.
|
|
|
| |||
|
|
|
|
|
|
|
| Difference between the power output indicated on RFG and the power measured at the tip of the DTC with the electrosurgery analyzer during ablation (W) | 3 | 2 | 1 | 1 | 1 |
|
| |||||
|
|
|
| |||
|
|
|
|
|
|
|
| Difference between the power output indicated on RFG and the power measured at the tip of the DTC with the electrosurgery analyzer during ablation (W) | 3 | 3 | 1 | 1 | 1 |
|
| |||||
|
|
|
| |||
|
|
|
|
|
|
|
| Difference between the power output indicated on RFG and the power measured at the tip of the DTC with the electrosurgery analyzer during ablation (W) | 4 | 2 | 1 | 1 | 1 |
| Maximum variation in current measured at the tip of the DTC using the electrosurgery analyzer between the 3 measurements performed (mA) | 4 | 0 | 0 | 0 | 0 |
|
| |||||
|
|
|
| |||
|
|
|
|
|
|
|
| Difference between the power output indicated on RFG and the power measured at the tip of the DTC with the electrosurgery analyzer during ablation (W) | 4 | 3 | 1 | 1 | 1 |
| Maximum variation in the current measured at the tip of the DTC using the electrosurgery analyzer between the 3 measurements performed (mA) | 1 | 0 | 0 | 0 | 0 |
|
| |||||
|
|
|
| |||
|
|
|
|
|
|
|
| Difference between the power output indicated on RFG and the power measured at the tip of the DTC with the electrosurgery analyzer during ablation (W) | 4 | 3 | 1 | 1 | 1 |
| Maximum variation in the current measured at the tip of the DTC using the electrosurgery analyzer between the 3 measurements performed (mA) | 4 | 0 | 0 | 0 | 0 |
|
|
|
| |||
|
| |||||
|
|
|
| |||
|
|
|
|
|
|
|
| Difference between the power output indicated on RFG and the power measured at the tip of the DTC with the electrosurgery analyzer during ablation (W) | 4 | 3 | 1 | 1 | 1 |
| Maximum variation in the current measured at the tip of the DTC using the electrosurgery analyzer between the 3 measurements performed (mA) | 4 | 0 | 0 | 0 | 0 |
|
|
|
| |||
DTC, DiamondTemp catheter; RFG, radio frequency generator.
DiamondTemp catheter location after RF1, RF2 or RF3 for all setups tested.
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| Distance to baseline points before any other testing variables (mm) | NA | NA | NA | NA | NA |
| Distance to baseline points after reinsertion of the DTC in the phantom (mm) | NA | NA | NA | NA | NA |
| Distance to baseline points after reconnection of the DTC (mm) | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| Distance to baseline points after RF1 was performed (mm) | 0.9 | 1.2 | 1.1 | 1.1 | 1.2 |
| Distance to baseline points after reinserting the DTC following RF1 (mm) | 0.4 | 0.6 | 1.0 | 1.4 | 1.2 |
| Distance to baseline points after reconnecting the DTC following RF1 (mm) | 0.4 | 0.6 | 1.0 | 1.4 | 1.2 |
| Distance to baseline points after RF2 was performed (mm) | 1.0 | 1.5 | 1.6 | 1.2 | 1.9 |
| Distance to baseline points after reinserting the DTC following RF2 (mm) | 0.5 | 0.4 | 1.2 | 1.6 | 1.0 |
| Distance to baseline points after reconnecting the DTC following RF2 (mm) | 0.5 | 0.4 | 1.2 | 1.6 | 1.0 |
| Distance to baseline points after RF3 was performed (mm) | 0.9 | 1.2 | 0.9 | 1.4 | 1.2 |
| Distance to baseline points after reinserting the DTC following RF3 (mm) | 0.5 | 0.8 | 1.0 | 1.8 | 1.6 |
| Distance to baseline points after reconnecting the DTC following RF3 (mm) | 0.5 | 0.8 | 1.0 | 1.8 | 1.6 |
| Distance to baseline points after variables tested (mm) | NA | NA | NA | NA | NA |
|
| |||||
|
|
|
|
|
|
|
| Distance to baseline points before any other testing variables (mm) | NA | NA | NA | NA | NA |
| Distance to baseline points after reinsertion of the DTC in the phantom (mm) | NA | NA | NA | NA | NA |
| Distance to baseline points after reconnection of the DTC (mm) | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| Distance to baseline points after RF1 was performed (mm) | 0.5 | 0.5 | 1.2 | 2.5 | 12.4 |
| Distance to baseline points after reinserting the DTC following RF1 (mm) | 1.2 | 1.5 | 3.5 | 4.5 | 9.8 |
| Distance to baseline points after reconnecting the DTC following RF1 (mm) | 1.2 | 1.5 | 3.5 | 4.5 | 9.8 |
| Distance to baseline points after RF2 was performed (mm) | 1.5 | 1.0 | 4.0 | 9.6 | 10.5 |
| Distance to baseline points after reinserting the DTC following RF2 (mm) | 0.4 | 0.6 | 1.4 | 11.8 | 12.0 |
| Distance to baseline points after reconnecting the DTC following RF2 (mm) | 0.4 | 0.6 | 1.4 | 11.8 | 12.0 |
| Distance to baseline points after RF3 was performed (mm) | 2.9 | 2.8 | 4.9 | 10.0 | 10.2 |
| Distance to baseline points after reinserting the DTC following RF3 (mm) | 0.6 | 0.8 | 1.0 | 12.0 | 12.2 |
| Distance to baseline points after reconnecting the DTC following RF3 (mm) | 0.6 | 0.8 | 1.0 | 12.0 | 12.2 |
| Distance to baseline points after variables tested (mm) | NA | NA | NA | NA | NA |
|
| |||||
|
|
|
|
|
|
|
| Distance to baseline points before any other testing variables (mm) | NA | NA | NA | NA | NA |
| Distance to baseline points after reconnection of the DTC (mm) | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| Distance to baseline points after RF1 was performed (mm) | 0.9 | 1.2 | 1.1 | 3.1 | 3.2 |
| Distance to baseline points after reinserting the DTC following RF1 (mm) | 0.4 | 0.6 | 1.0 | 2.4 | 2.2 |
| Distance to baseline points after reconnecting the DTC following RF1 (mm) | 0.4 | 0.6 | 1.0 | 2.4 | 2.2 |
| Distance to baseline points after RF2 was performed (mm) | 1.0 | 1.5 | 1.6 | 3.2 | 3.9 |
| Distance to baseline points after reinserting the DTC following RF2 (mm) | 0.5 | 0.4 | 1.2 | 1.6 | 2.0 |
| Distance to baseline points after reconnecting the DTC following RF2 (mm) | 0.5 | 0.4 | 1.2 | 1.6 | 2.0 |
| Distance to baseline points after RF3 was performed (mm) | 0.9 | 1.2 | 0.9 | 3.4 | 4.2 |
| Distance to baseline points after reinserting the DTC following RF3 (mm) | 0.5 | 0.8 | 1.0 | 1.8 | 2.6 |
| Distance to baseline points after reconnecting the DTC following RF3 (mm) | 0.5 | 0.8 | 1.0 | 1.8 | 2.6 |
| Distance to baseline points after variables tested (mm) | NA | NA | NA | NA | NA |
|
|
|
|
|
|
|
| Distance to baseline points before any other testing variables (mm) | NA | NA | NA | NA | NA |
| Distance to baseline points after reinsertion of the DTC in the phantom (mm) | NA | NA | NA | NA | NA |
| Distance to baseline points after reconnection of the DTC (mm) | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| Distance to baseline points after RF1 was performed (mm) | 17.2 | 15.5 | 21.5 | 24.5 | 25.5 |
| Distance to baseline points after reinserting the DTC following RF1 (mm) | 10.5 | 10.4 | 11.1 | 12.5 | 12.4 |
| Distance to baseline points after reconnecting the DTC following RF1 (mm) | 10.5 | 10.4 | 11.1 | 12.5 | 12.4 |
| Distance to baseline points after RF2 was performed (mm) | 18.5 | 21.0 | 22.0 | 24.7 | 32.5 |
| Distance to baseline points after reinserting the DTC following RF2 (mm) | 10.9 | 10.1 | 11.4 | 11.5 | 13.9 |
| Distance to baseline points after reconnecting the DTC following RF2 (mm) | 10.9 | 10.1 | 11.4 | 11.5 | 13.9 |
| Distance to baseline points after RF3 was performed (mm) | 20.2 | 20.7 | 21.2 | 25.0 | 30.0 |
| Distance to baseline points after reinserting the DTC following RF3 (mm) | 10.6 | 11.2 | 10.9 | 12.0 | 12.0 |
| Distance to baseline points after reconnecting the DTC following RF3 (mm) | 10.6 | 11.2 | 10.9 | 12.0 | 12.0 |
| Distance to baseline points after variables tested (mm) | NA | NA | NA | NA | NA |
DTC, DiamondTemp catheter; RF, radio frequency.
Figure 4Localization accuracy with minimal deviation according to configuration 3.2. The catheter positioned on the reference points is represented on the EAM with minimal difference to the reference sensor-enabled catheter.
Figure 5Configuration 3.2 during RF delivery. Localization accuracy of DTA catheter on the EAM during RF application (50w 30 s). This configuration has shown no catheter shift during RF application.
Distances measured between the baseline points after each variable, for each setup tested.
|
|
|
|
|
|
|---|---|---|---|---|
|
| NA | 0.99 | 1.00 | <0.001 |
|
| 0.99 | NA | 0.99 | <0.001 |
|
| 1.00 | 0.99 | NA | <0.001 |
|
| <0.001 | <0.001 | <0.001 | NA |
|
|
|
|
|
|
|
| NA | 0.99 | 1.00 | <0.001 |
|
| 0.99 | NA | 0.99 | <0.001 |
|
| 1.00 | 0.99 | NA | <0.001 |
|
| <0.001 | <0.001 | <0.001 | NA |
|
|
|
|
|
|
|
| NA | 0.93 | 1.00 | <0.001 |
|
| 0.93 | NA | 0.93 | <0.001 |
|
| 1.00 | 0.93 | NA | <0.001 |
|
| <0.001 | <0.001 | <0.001 | NA |
|
|
|
|
|
|
|
| NA | 0.002 | 0.98 | <0.001 |
|
| 0.002 | NA | 0.93 | 0.002 |
|
| 0.98 | 0.93 | NA | <0.001 |
|
| <0.001 | 0.002 | <0.001 | NA |
|
|
|
|
|
|
|
| NA | <0.001 | 0.98 | <0.001 |
|
| <0.001 | NA | 0.003 | 0.025 |
|
| 0.98 | 0.003 | NA | <0.001 |
|
| <0.001 | 0.025 | <0.001 | NA |