Sameer D Shingte1, Olav Altenburg1, Peter J T Verheijen2, Herman J M Kramer1, Huseyin Burak Eral1,3. 1. Process & Energy Department, Delft University of Technology, Leeghwaterstraat 39, 2628 CA Delft, The Netherlands. 2. Biotechnology Department, Delft University of Technology, 2629 HZ Delft, The Netherlands. 3. Van't Hoff Laboratory for Physical and Colloid Chemistry, Debye Institute, Utrecht University, Padualaan 8, 3584 CH Utrecht, The Netherlands.
Abstract
We present a droplet microfluidic platform mixing the contents of the droplet chaotically in microfluidic induction time measurements, a promising method for quantifying nucleation kinetics with minute amounts of solute. The nucleation kinetics of aqueous potassium chloride droplets dispersed in mineral oil without surfactants is quantified in the presence and absence of chaotic mixing. We demonstrate the ability of the proposed platform to dictate droplet size, to provide a homogeneous temperature distribution, and to chaotically mix the droplet contents. Chaotic mixing in induction time measurements is facilitated by the motion of droplets through serpentine micromixer bends, while the extent of mixing is controlled by how much droplets move. Different nucleation kinetics are observed in experiments where the droplets are static, mixed, and in motion. We hypothesize that the droplet motion induces formation of a thin-liquid Bretherton film surrounding the droplets. The thin film shields droplets from solid boundaries that are more efficient heteronucleant surfaces compared to liquid-liquid interfaces. We observed that repeated microfluidic induction time measurements, particularly with moving droplets, produce significantly distinct cumulative nucleation probability curves, indicating that the measured nucleation kinetics depend strongly on the details of the experimental procedure, which we discuss in detail. Finally, we compare the microfluidic experiments to well-mixed, milliliter volume, turbidity-based measurements in the context of classic nucleation theory.
We present a droplet microfluidic platform mixing the contents of the droplet chaotically in microfluidic induction time measurements, a promising method for quantifying nucleation kinetics with minute amounts of solute. The nucleation kinetics of aqueous potassium chloride droplets dispersed in mineral oil without surfactants is quantified in the presence and absence of chaotic mixing. We demonstrate the ability of the proposed platform to dictate droplet size, to provide a homogeneous temperature distribution, and to chaotically mix the droplet contents. Chaotic mixing in induction time measurements is facilitated by the motion of droplets through serpentine micromixer bends, while the extent of mixing is controlled by how much droplets move. Different nucleation kinetics are observed in experiments where the droplets are static, mixed, and in motion. We hypothesize that the droplet motion induces formation of a thin-liquid Bretherton film surrounding the droplets. The thin film shields droplets from solid boundaries that are more efficient heteronucleant surfaces compared to liquid-liquid interfaces. We observed that repeated microfluidic induction time measurements, particularly with moving droplets, produce significantly distinct cumulative nucleation probability curves, indicating that the measured nucleation kinetics depend strongly on the details of the experimental procedure, which we discuss in detail. Finally, we compare the microfluidic experiments to well-mixed, milliliter volume, turbidity-based measurements in the context of classic nucleation theory.
Crystallization from
solution is a widely used separation and purification
method in the production of crystalline solids ranging from pharmaceuticals
to specialty chemicals.[1,2] Our ability to accurately quantify
nucleation kinetics plays a critical role in the design and control
of industrial crystallization processes while offering mechanical
insights into nucleation pathways. Nucleation is commonly quantified
through induction time experiments[3] where
a large number of identical experiments are performed to account for
the stochastic nature of nucleation.[4] A
large number of identical experiments ensure the statistical significance
of experimental deductions.[5,6] Various methods have
been developed to measure induction times, including pendant drop,
double pulse technique, levitated drop, camera, and turbidity-based
methods.[1,3,7−9]Among these methods, measuring nucleation rates in droplets
has
emerged as a promising method.[3,10−12] The idea of using small droplets for induction time measurements
was presented already in 1959 by White and Frost.[13] White and Frost dripped small droplets of an aqueous solution
in mineral oil and performed induction time measurements. These droplets
had diameters varying between 0.2 and 2 mm. With the advent of micromanufacturing,
droplet microfluidics emerged as a cost-effective method for creating
a large number of droplets with controlled size distribution while
using minute amounts of solute—an important feature while working
with difficult to synthesize or expensive compounds. By balancing
capillary forces dominant on the micrometer scale with viscous forces,[14,15] droplet-based microfluidic platforms can provide a large number
of droplets (100–1000 per device) acting as identical isolated
crystallization reactors to ensure statistical accuracy.[10,11] Particularly, droplet volume, a critical parameter in classic nucleation
theory,[4] can be tightly controlled in droplet
microfluidics.[16] Although other approaches
have been proposed,[3,17−20] in a typical microfluidic induction
time experiment, droplets carrying known concentrations of solute
are produced and then cooled down to reach a given supersaturation.
At this prescribed supersaturation, the nucleation kinetics are quantified
by two different approaches with significantly different mixing behaviors.
In the first approach, the static droplets are stored at a fixed supersaturation,
and all the droplets are observed until the end of the experiment
to quantify the fraction of crystallized droplets as a function of
time.[21−23] This fraction is used to construct the cumulative
nucleation probability function, P(t), correlated to the induction time and nucleation rate.[6,24] In the second approach, moving droplets in a capillary or microfluidic
channel are observed at a given observation time, tobs, after a prescribed supersaturation is reached. The
fraction of crystal-containing droplets at this fixed tobs is quantified optically as droplets flow. Varying
this fixed tobs allows construction of
the cumulative nucleation probability function and quantification
of nucleation kinetics; yet, these measurements are often reported
at a single tobs.[16,25] In both approaches, emergence of the first crystal is detected by
video microscopy. The fundamental difference between these two approaches
is that the droplets are static in the first approach,[21−23] whereas they are moved but not actively mixed in the second one.[16,25] These two microfluidic approaches rarely produce the same nucleation
rates even for identical solutes under identical conditions. Moreover,
microfluidic nucleation rates rarely match turbidity-based measurements.
We hypothesize that hydrodynamics and surface interactions may play
a role in explaining this discrepancy. In one of the earlier studies,
Zheng et al.[21] demonstrated crystallization
in the stagnant droplets at a constant temperature for protein screening
applications. Following this study, microfluidic induction time measurements
of droplets containing various solutes[22,23] and the nature
of the oil–water interface have also been investigated.[26] Ildefonso et al.[26,27] studied heterogeneous
nucleation and the emergence of polymorphs in stagnant microfluidic
droplets. Grossier et al.[28] developed an
approach where nucleation is induced with a sharp tip in micron-scale
droplets through contact or electric fields. More recently, Dos Santos
et al.[16] characterized nucleation and growth
rates in moving droplets by varying the supersaturation and residence
time while accounting for the droplet volume distribution and the
uncertainty in the automated image analysis procedure.The influence
of hydrodynamics in microfluidic induction time measurements
has attracted attention recently. Rossi et al.[29] studied the effect of flow comparing the nucleation in
stagnant and flowing droplets where the contents of droplets are mixed
by symmetric advection rolls. The authors showed a significant increase
in nucleation rates in droplets that were moving compared to static
droplets. The authors concluded that the internal circulation caused
by the movement of the droplets affects the kinetic parameter of the
nucleation expression without affecting the thermodynamic parameters.
Nappo et al.[30] investigated the effect
of shear rate on the nucleation by comparing induction times in stagnant
and moving droplets in a microfluidic device with those in stirred
vials of 2 mL. These authors observed a very strong increase of the
nucleation rate in moving compared to static droplets, while somewhat
lower nucleation rates were found in the stirred vials, despite a
two orders of magnitude higher shear rate in the stirred vials. On
a separate note, Dela Cruz et al.[31] concluded
that changes in the shape of the probability distribution is an indication
of the change in the nucleation regimes, which in most cases is caused
by polymorphism, i.e., because of concomitant nucleation of two polymorphic
forms or by transformation of a metastable into a more stable polymorph.
A more extensive review of microfluidic induction time measurements
can be found in following review articles.[3,10,11]As the nucleation rate is reported
per unit volume, one may be
misled to think nucleation rates measured in microfluidics may be
directly implemented in industrial-scale crystallizers. It is important
to emphasize that industrial crystallization processes take place
in larger volumes not only but also in fundamentally different hydrodynamic
environments.[1] Hydrodynamics inside industrial
crystallizers affects the growth rates as well as other hydrodynamics-induced
phenomena such as secondary nucleation, attrition, and aggregation
rates.[2,32,33] Hence, the
“effective” nucleation rates obtained from microfluidics
that include contributions from both nucleation and highly hydrodynamics
dependent growth cannot be used for designing and controlling industrial-scale
models. Nonetheless, development of a microfluidic platform with precisely
controlled mixing can shed light on the often overlooked role that
these two phenomena play in microfluidic induction time measurements.Although microfluidics shows a great deal of potential for various
applications, it also suffers from a number of drawbacks, e.g., limited
mixing and the dispersion of constituents.[34] Microfluidic flows are laminar; hence, the absence of turbulence
results in poor component mixing in the microchannel. Turbulence can
be induced by increasing the fluid flow rate. This, however, may lead
to excessive consumption of the chemical components, which undermines
one of the key advantages of microfluidics. The other way of inducing
rigorous mixing is introducing chaotic advection through the geometry
of the channel. There are different strategies to induce chaotic advection:
one of them is to design a serpentine geometry for the microchannel,
as reported by Bringer et al.[34] and Harshe
et al.[35] The unique geometry of the microchannel
enhances mixing by successive cycles of stretching, folding, and reorientation
of substance molecules in the fluid.[14] These
successive cycles resemble the baker’s transformation during
dough formation. Therefore, mixing is facilitated as droplets move
along the channel. When the droplet moves in the straight microchannel,
symmetric eddies are generated, but there is no effective mass transfer
taking place between the two halves of the droplet, which leads to
limited mixing. Conversely, asymmetric recirculation flows are developed
inside the droplet moving through the winding microchannel. They give
rise to chaotic advection leading to chaotic mixing. Therefore, enhanced
mixing within the droplets is expected using the device presented
in this work compared to the microfluidic systems reported by Nappo
et al.[30] and Rossi et al.[29]The purpose of this work is to present a microfluidic
platform
for induction time measurements where the contents of the dispersed
droplets are mixed by chaotic advection initiated by the serpentine
shape of the device. We share our experiences including experimental
difficulties and challenges in interpreting observations in a candid
and straightforward manner. The proposed platform offers the possibility
to manipulate the hydrodynamics and mixing conditions inside the droplets
through droplet motion. Consequently, the platform enables studying
the intertwined role of droplet motion and mixing on nucleation kinetics
in microfluidic experiments. Moreover, the device builds upon the
well-documented advantages of microfluidic design, such as ability
to perform high-throughput experiments with minute amounts of material
screening at low material costs (low volume) with excellent control
of process conditions due to high mass and heat transfer rates.The presented microfluidic platform (Figure ) produces droplets carrying prescribed concentrations
of a model solute, potassium chloride, KCl, without the need for surfactants
and stores them for observation at an elevated temperature as an undersaturated
solution. The stored droplets are rapidly cooled to a desired temperature
with an integrated Peltier element, while microscopy images of each
droplet are recorded at fixed time intervals. The microscopy images
are then manually processed to calculate the ratio of droplets that
crystallize as a function of time and the effective nucleation rate
at a prescribed supersaturation. The spatial and temporal temperature
distributions across the microfluidic device are characterized to
ensure accurate induction time measurements. The presented microfluidic
platform ensures mixing of droplets with serpentine micromixers. The
degree of mixing is controlled by the number of bends that droplets
travel through in an experiment. We first characterize the nucleation
kinetics as a function of supersaturation and droplet volume in static
microfluidic experiments. Then, we elucidate the intertwined roles
of droplet motion and mixing in microfluidic induction time measurements
by comparing three different microfluidic induction time experiments
referred to as “static”, “moving”, and
“mixing” showing distinct behaviors. Moreover, we compare
our microfluidic experiments with well-mixed, milliliter volume, turbidity-based
measurements in the context of classic nucleation theory. We observed
that the repeated microfluidic induction time measurements under identical
conditions produced different cumulative nucleation probability curves
particularly for experiments with moving droplets. We attribute this
low reproducibility to details of the experimental procedure particularly
the history of the droplets prior to temperature quench. We hope that
the reported experimental procedures discussed in detail provide a
learning experience for future studies. Our results highlight the
relevance of mixing and droplet motion in microfluidic induction time
measurements and open new alleys of investigation at the intersection
of fundamental crystallization phenomena, wetting physics, and hydrodynamics.
Figure 1
Experimental
setup: (a) two syringe pumps, a syringe pump heater
for the dispersed phase, a thermal bath, and the microfluidic device
mounted on an inverted microscope, which visualized the droplets (red
dashed square). The field of view is systematically translated in X and Y axes to image the whole device
with an automated moving stage enabling the observation of ∼300
droplets in one experiment. (b) Details of the microchannel (0.5 mm
wide and 1 mm or 0.5 mm deep) having a serpentine structure to enhance
mixing with the illustration of droplet formation using a T-junction.
(c) 3D view of the microfluidic device. (d) Bright field microscopy
images of the field of view before and after reaching the target crystallization
temperature. Droplets with crystals are the red circles. (e) Temperature
profile. taq is the video acquisition
time, and t0 is the moment reaching the
target temperature. (f) An optical microscopy image of a droplet with
a crystal inside.
Experimental
setup: (a) two syringe pumps, a syringe pump heater
for the dispersed phase, a thermal bath, and the microfluidic device
mounted on an inverted microscope, which visualized the droplets (red
dashed square). The field of view is systematically translated in X and Y axes to image the whole device
with an automated moving stage enabling the observation of ∼300
droplets in one experiment. (b) Details of the microchannel (0.5 mm
wide and 1 mm or 0.5 mm deep) having a serpentine structure to enhance
mixing with the illustration of droplet formation using a T-junction.
(c) 3D view of the microfluidic device. (d) Bright field microscopy
images of the field of view before and after reaching the target crystallization
temperature. Droplets with crystals are the red circles. (e) Temperature
profile. taq is the video acquisition
time, and t0 is the moment reaching the
target temperature. (f) An optical microscopy image of a droplet with
a crystal inside.
Materials
and Methods
Chemicals
All of the chemicals used in this study were
purchased from suppliers and used without further modification. The
chemical used and the suppliers are mineral oil (CAS: 8042-47-5, Sigma),
KCl (CAS: 7447-40-7 CIK), and fluorescent dye rhodamine (CAS: 81-88-9,
Sigma). All of the KCl solutions were prepared on a weight-by-weight
basis by a dissolving a given amount of KCl in weighted deionized
water. The deionized water was acquired from a Millipore device (ELGA
PURELAB, resistivity: 18.2 MΩ·cm at 23.6 °C).
Microfluidic
Device Design and Manufacturing
The proposed
microfluidic tool consists of two sections: a droplet generation section
with a T-junction and an observation and mixing section with serpentine
passive micromixers as shown in (Figure a,b). The T-junction (Figure b) produces droplets of controlled size,
and the serpentine micromixer ensures that the contents of the droplets
are mixed as they move through the bends via baker’s transformations.[14,34] The device is made from an aluminum plate with the channel grooved
on it with micromachining. An acrylic glass plate is fixed on top
with the help of aluminum screws, thus covering the channel. The device
design ensures that (a) a maximum number of droplets are accommodated
within a standard glass slide area of (25 × 75 mm), (b) the droplets
are cooled down as rapidly and homogeneously as possible because of
the high thermal conductivity of aluminum, (c) moving droplets are
observed with video microscopy and mixed to the desired extent through
asymmetric recirculation flows inside the droplets also known as baker’s
transformations.[14,34] The microfluidic devices used
for most static experiments, which are shown in Figures –5, have a
channel width of 0.5 mm and a channel depth of 1 mm as shown in Figure c. Each bend is 1.7
mm. The microfluidic device used for the mixing experiments, shown
in Figure , is the
same, except for a different channel depth of 0.5 mm. This changes
the volume and shape of the droplets, as well as the interface with
different surfaces. It should be noted that the device consists of
two parts made of different materials, hence having distinct wetting
properties and roughness. The aluminum plate is rougher than the acrylic
glass plate. We hydrophobized both surfaces with a silanization agent
according to the procedure described in Schuurmans et al.[36] to ensure that the continuous phase wets all
the confining surfaces uniformly.
Figure 5
Induction time measurements for the “static and not mixed”
microfluidic experiment. Panel a shows the cumulative probability
distribution function, P(t), for
different droplet volumes at fixed supersaturation, S = 1.15. Panel b shows P(t) for
static microfluidic experiments with different supersaturations at
a fixed droplet volume Vd = 0.84 μL.
Error bars are those given by eq 1 in the Supporting Information.
Figure 6
Influence of droplet
motion and mixing on nucleation kinetics:
Panel a shows cumulative distribution function, P(t) at fixed supersaturation, S = 1.13 for “moving”, mixing 12 bends”, and
“mixing 31 bends” experiments. The error bars are standard
deviation. Panel b shows the droplet movement. xd is the displacement, and qc is
the continuous phase volumetric flow. The arrows and pumping sequence
are coded: blue is forward and green backward motion, induced by alternating
between two pumps. Panel c illustrates the pump connections to the
microfluidic chip used for the droplet motion.
Characterizing the Droplet
Formation
To explore our
ability to control the droplet size in the our device, we measured
the droplet size distribution with optical microscopy for various
dispersed and continuous phase flow rates. For this study, mineral
oil and deionized water are used as a continuous phase and dispersed
phase, respectively. The experimental setup is represented in Figure a, which shows two
microfluidic pumps to infuse continuous and dispersed phase liquids.
At the T-junction, the dispersed liquid is fed perpendicularly with
respect to the continuous phase liquid flow. During experiments, the
relative flow rate of water to oil, Q, defined as Q = qw/qo is varied between 0.5 and 1.5. The oil flow rate, qo, is kept constant at either 1, 4, or 10 mL/h,
and the water flow rate, qw, is changed
to reach a given Q.Controlling
droplet size: the nondimensionalized droplet size, L/w, is presented as a function of relative
flow rate, Q, in panel a and capillary number, Ca, in panel b at different flow rates, qo, of the continuous oil phase. Error bars represent the
standard deviation of about 15 droplets for each case.
Characterizing the Temperature Distribution
To characterize
the spatial and temporal temperature distributions across the microfluidic
device, we utilize thermocouples and a thermal camera (FLIR A655sc).
The inset of Figure a shows the various positions where the thermocouples are placed
on the microfluidic device. The measurements shown in Figure a indicate that the target
temperature of 10 °C is reached within 5 min. Figure b shows thermal camera images
at the start of the experiment at t = 0 (Figure b1), t = 2 min (Figure b2), and t = 5 min (Figure b3). These thermal images also incorporate
the thermocouple results and show that the target temperature is reached
within 5 min. Moreover, they point out that the spatial temperature
distribution is homogeneous.
Figure 3
Characterization of spatial and temporal temperature
distribution:
(a) Temperature variation as a function of time measured with thermocouples
throughout the experiment. Thermal equilibrium is achieved within
5 min. The inset shows the microfluidic chip resting on the Peltier
cooling stage with three thermocouples placed (red circles) at three
different positions on the Al plate. Panels b–d show the spatial
temperature distribution probed by a thermal camera at different times
after cooling taq = 0 (b), t0 – 3 = 2 min (c), and t0 = 5 min (d).
Characterization of spatial and temporal temperature
distribution:
(a) Temperature variation as a function of time measured with thermocouples
throughout the experiment. Thermal equilibrium is achieved within
5 min. The inset shows the microfluidic chip resting on the Peltier
cooling stage with three thermocouples placed (red circles) at three
different positions on the Al plate. Panels b–d show the spatial
temperature distribution probed by a thermal camera at different times
after cooling taq = 0 (b), t0 – 3 = 2 min (c), and t0 = 5 min (d).
Cleaning Procedure for
the Microfluidic Device for Induction
Time Measurements
The aluminum and acrylic glass plates were
sonicated while they were kept immersed in an ethanol bath for 20
min. After being washing with deionized water, both plates were dried
using dry N2 gas. Additionally, the tubing carrying the
dispersed and continuous phases was flushed by ethanol and then by
deionized water and then dried using dry N2 gas.
Milliliter
Volume Induction Time Measurements
Induction
time measurements with 1 mL volume KCl solutions at different supersaturations
were performed using a commercial turbidity-based device, Crystal16.
All induction time experiments are performed at 10.0 ± 0.5 °C.
The temperature profile used is identical to microfluidic experiments
to facilitate fair comparison. Note that the stirring rate was reduced
from 700 rpm during the cooling phase to 200 rpm when the constant
supersaturation was reached. The same stirring conditions were maintained
thereafter during the turbidity measurements.
Microfluidic Induction
Time Measurements
Aluminum plates
with microfluidic elements including the T-junction and serpentine
channel were covered tightly with acrylic glass plates by using pairs
of screws and washers as shown in Figure c. The cleaned microfluidic device was placed
over the Peltier element and held firmly over it. The temperature
of the Peltier element was maintained at 40 °C for 30 min prior
to the droplet formation. A clear solution of KCl with a targeted
supersaturation ratio is prepared based on the solubility of KCl at
10 °C, i.e., 31.2 g of KCl per 100 g of water.[37] The supersaturated solution was not filtered.The
aqueous droplets of KCl dispersed in mineral oil were produced at
the T-junction as shown in Figure b. The dispersed aqueous phase and continuous oil phases
were pumped into the microfluidic device with syringe pumps (Figure a) at the prescribed
oil and KCl solution volumetric flow rates as q0 = 10, 4, and 1 mL/h and qaq-KCl = 1, 2, and 6 mL/h, respectively. The magnitudes and ratio of flow
rates determine the droplet size at T-junction as shown in Figure . The transfer tubing
was immersed in a hot water bath kept at 40 °C to avoid crystallization
during solution transfer or prior to application of the temperature
profile in Figure e. Also the syringe carrying the aqueous KCl solution was kept at
40 °C with a microfluidic heater presented in Figure a. Microfluidic heater is a
heating jacket designed for syringe pumps. After the droplet formation,
the inlets were covered with parafilm to prevent evaporation.
Figure 2
Controlling
droplet size: the nondimensionalized droplet size, L/w, is presented as a function of relative
flow rate, Q, in panel a and capillary number, Ca, in panel b at different flow rates, qo, of the continuous oil phase. Error bars represent the
standard deviation of about 15 droplets for each case.
The microfluidic device, filled with droplets of the required droplet
size and supersaturation, was mounted onto an inverted optical microscope
(Nikon TE) in a reflective imaging mode with the acrylic plate facing
the objective. Then, Peltier element cools the chip from 40 to 10
°C with the rate of 20 °C/min and maintains the target temperature
during the experiments as illustrated in Figure f. The supersaturated droplets in microfluidic
device are monitored as a function of time using the automated microscope
stage. The field of view illustrated with the red square in Figure a is smaller than
the whole chip surface. The field of view is automatically moved across
the chip and composite images of the entire microfluidic device is
acquired automatically. Typical composite images at the start of the
experiment and a later time with nucleated crystals are given in Figure d. In the initial
2 h composite images were taken every 5 min. At longer times, intervals
the acquisition intervals were increased to 30 min and 1 h depending
on the supersaturation. As shown in 1e, image
acquisition was initiated before reaching the target temperature denoted
as taq, whereas t0 is the time when the system reaches the target temperature.
Starting video acquisition at taq allows
omitting of crystals forming before reaching the desired supersaturation
from induction time calculations.Lastly, the acquired composite
images (Figure d)
are processed manually by counting the
number of droplets with crystals in a given time frame. Crystals form
nearly instantaneously, and only a single crystal per droplet has
been seen. They are 0.2–0.3 mm in size. The number of crystal
emerging as a function of time is then converted to a cumulative probability
distribution function, P(t), expressed
as,We have
specifically chosen the wording “total
number of droplets without crystals at taq” in eq to
emphasize the fact that droplets crystallizing prior to taq are not considered in calculating P(t).
Models for Fitting Cumulative Distribution
Functions
The cumulative probability distribution functions
are fitted to two
different models: (i) single exponential (eq ) and (ii) two exponential (eq ) using weighted least-squares curve
fitting procedures implemented in Matlab software. The fitting equations
arewhere a is a prefactor,
and
τ is the induction time in eq . In eq , τ1 and τ2 represent two distinct
nucleation time scales.The parameter estimation procedure also
delivers standard errors and confidence intervals in the estimates
of the parameters, such as τ and a. These are
the commonly obtained results from such procedures on the basis of
a linearization of the model near the estimate. In some cases, approximate
confidence intervals were obtained taking the nonlinearity into account.
Typically, the time constants have errors on the order of 10% or more.The conversion to the specific nucleation rates for these models
are through the time constants, such thatwhere Vd is the
mean droplet volume. Similarly, for the mixture of two-exponential
distributions, J1 and J2 are obtained from τ1 and τ2. The error in J is due to the uncertainty
in τ, on the order of 10% or more, and the volume. The droplet
volume is approximated by the relation given in Musterd et al.[38] for a given trapezoidal channel, whereby the
droplet length is obtained from 10 observations in each case and has
a variation on the order of 3%, which gives an insignificant standard
error of the mean on the order of 1%.
Experimental Quantification
of Mixing
The mixing efficiency
of serpentine passive mixers shown in Figure is quantified according to methodology of
Harshe et al.[35] To this end, a new device
with two dispersed phase inlets (as shown in Figure a) allowing mixing of two aqueous streams
(one containing dye and one without) is designed. Except having the
two inlets for the dispersed phase, the dimensions and geometry of
this device are identical to the device in Figure . In the microfluidic device shown in Figure a,b, the first two
aqueous streams (one carrying a fluorescent dye and the other free
of it) are brought together as shown in microscopy images in Figure b,c. As the viscous
forces dominate at microfluidic scale, the two streams can only mix
through diffusion. Consequently, when they meet they do not mix immediately
but stay as two laminar streams as shown in Figure c before they enter the serpentine mixer.
Once the droplets enter the mixer, they gradually mix after each turn
as shown in 4b,c, indicated by the gradual
distribution of dye inside droplet.
Figure 4
Quantification of droplet motion induced
mixing: (a) Illustration
of the microfluidic device designed specifically for mixing quantification
experiments, (b) bright field microscopy image of droplets in serpentine
micromixers, (c) fluorescent microscopy images of droplets showing
the fluorescent dye homogeneously distributed as the droplets move
through the micromixers, (d) Danckwerts number, θ, versus the
number of bends a droplet moves through indicating complete mixing
after six bends under canonical flow rates.
Quantification of droplet motion induced
mixing: (a) Illustration
of the microfluidic device designed specifically for mixing quantification
experiments, (b) bright field microscopy image of droplets in serpentine
micromixers, (c) fluorescent microscopy images of droplets showing
the fluorescent dye homogeneously distributed as the droplets move
through the micromixers, (d) Danckwerts number, θ, versus the
number of bends a droplet moves through indicating complete mixing
after six bends under canonical flow rates.We quantify the degree of mixing by quantifying the gradual change
in fluorescent intensity distribution proportional to dye concentration
as a droplet travels through bends with fluorescent microscopy. Before
the droplet enters the bends, the fluorescent dye is concentrated
in only on one side of the droplet as shown in Figure c. With each bend the droplets travel through,
the fluorescent intensity appears to be become more homogeneous indicating
mixing. The observed trend in Figure c, i.e., the fluorescent intensity distribution becoming
more homogeneous with each bend, is quantified by the Danckwerts number,
θ, here[35] given by,In eq , M is the number
of pixels in area of interest, I is the intensity of nth
pixel, and I̅ is the average intensity in the
area of interest. In essence, the Danckwerts number shows the deviation
of light intensity from the average fluorescent dye intensity in the
area of interest. The higher the Danckwerts number, the less mixed
the droplet is. At a certain moment, the Danckwerts number is expected
to become really small and constant, hence showing the droplet is
completely mixed.
Microfluidic Induction Time Experiments with
Quantified Mixing
To isolate the effect of droplet motion
and mixing, there are different
microfluidic experiments performed at the same supersaturation, S = 1.13 and with the same droplet volume. The following
experiments are preformed: (i) droplets are kept stagnant, (ii) droplets
are moved less than a bend throughout the experiment, consequently
not mixed, (iii) the droplets are moved over 12 bends, consequently
well-mixed, and (iv) the droplets are moved over 31 bends. These are
referred to respectively as “static”, “moving”,
“mixing 12 bends”, and “mixing 31 bends”.For all experiments numerated above, a potassium chloride solution
with S = 1.13 at 10 °C was prepared. The potassium
chloride was dissolved in water at 40 °C. It was kept at this
temperature for at least 1 h to ascertain all potassium chloride dissolved.
Before droplets are generated, mineral oil and the microfluidic device
were both heated to 40 °C as well. The droplets are generated
in the microfluidic device shown in Figure a as described in microfluidic induction
time measurements. During droplet generation the KCl solution in the
syringe and the microfluidic device were both kept at 40 °C with
a syringe heater and a temperature controller. The microfluidic device
is filled with droplets with an oil flow rate of 66.67 μL/min
and an aqueous flow rate of 22.22 μL/min.Induction time measurements for the “static and not mixed”
microfluidic experiment. Panel a shows the cumulative probability
distribution function, P(t), for
different droplet volumes at fixed supersaturation, S = 1.15. Panel b shows P(t) for
static microfluidic experiments with different supersaturations at
a fixed droplet volume Vd = 0.84 μL.
Error bars are those given by eq 1 in the Supporting Information.Once the droplets were
formed and stored in the microfluidic chip
at 40 °C, pump configuration, i.e., how the pumps are connected
to microfluidic chip, was altered to move the droplets as shown in Figure c. The syringe pumps shown in Figure a were replaced with two syringe pumps carrying
oil. One of the pumps is connected to the oil inlet and one other
one to the outlet in Figure . The aqueous phase inlet is closed. With the program shown
in Figure b, the two
pumps, synchronized to pump in the same direction, move the droplets
back and forward in the channel as shown in Figure c. The back and forward motion was done in
a period of 80 s, which is needed for the pumps to build sufficient
pressure and then move the droplets by covering a certain number of
bends. The droplet motion was stopped for 100 s in every cycle (Figure b) to facilitate
imaging of the droplets. In this total period of 180 s, the droplets
are mostly stagnant and moving only a fraction of that period, which
was roughly estimated. However, the number of bends was unambiguous,
which is why the results of these experiments are expressed in bends
as a measure for the actual mixing time. The actual travel time per
bend was in the range from 0.5 to 3 s.Influence of droplet
motion and mixing on nucleation kinetics:
Panel a shows cumulative distribution function, P(t) at fixed supersaturation, S = 1.13 for “moving”, mixing 12 bends”, and
“mixing 31 bends” experiments. The error bars are standard
deviation. Panel b shows the droplet movement. xd is the displacement, and qc is
the continuous phase volumetric flow. The arrows and pumping sequence
are coded: blue is forward and green backward motion, induced by alternating
between two pumps. Panel c illustrates the pump connections to the
microfluidic chip used for the droplet motion.The microscopy images of the droplets were not taken at equal intervals,
but the frequency of imaging was altered to efficiently use the limited
hard disk space. The observations continued until the droplets started
to coalesce or crystals moved out of the droplets. After a few back
and forth movements, some droplets with grown crystals present started
to block the channel causing coalescence. From this point, observations
were halted, as the mixing was hindered, and the droplets were not
unique single size entities anymore.
Results
Controlling
the Droplet Size
The droplet size in microfluidic
induction time measurements has been previously used to differentiate
between homogeneous and heterogeneous nucleation in microfluidic induction
time measurements.[3,17] Consequently, the droplet size
is a relevant parameter in microfluidic induction time measurements.
We changed the droplet sizes by varying the flow rates of the dispersed
water phase (qw), continuous oil phase
(qo), and the relative flow rate (Q defined as qw/qo). Figure a shows the experimentally measured droplet size as a function of Q along with the theoretical prediction by Garstecki et
al.[39]Figure a illustrates how the droplet length, L, increases with Q, as the rate of droplet
elongation prior to pinch-off is proportional to qw.[39] Furthermore, the rate
of droplet detachment is proportional to qo leading to a shorter L at constant Q. This variation highlights the significance of flow rate magnitudes,
as they influence the hydrodynamics of the droplet formation at the
T-junction. At high oil flow rates, our results agree with theoretical
predictions by Garstecki et al.[39]In Figure b, the
capillary number is defined as Ca = μU/σ, where μ, U, and σ denote continuous
phase viscosity, characteristic speed, and surface tension, respectively.
The characteristic velocity U = Q/(wh) with the cross-sectional area wh is channel width times height. As Ca increases,
the normalized droplet size, L/w, decreases as the Ca number reflects the relative
strength of viscous forces to surface tension. In other words, as
the viscous shear forces increase, the droplets become smaller. The
droplet formation in T-junctions is considered to be a balance between
viscous, inertial, and surface tension forces.[40] At low Reynolds (Re = ρUw/μ with ρ the continuous phase density) and Ca numbers, we expect L/w versus Ca at different Q values to collapse.[39,41] However, our systems deviate from this behavior as shown in Figure b. Our experiments
are conducted at Ca smaller than 5 × 10–4. However, because of the relatively large channel
width, approximately 1 mm, Re is on the order of
50 in our experiments. We attribute the lack of collapse in Figure b to inertial effects.[14]
Spatial and Temporal Temperature Measurements
The calibration
of the temperature controlling system is essential as this work involves
cooling crystallization. Induction time measurements are performed
based on the KCl solubility data at 10 °C, for which the calibration
is done. As shown in the inset in Figure , three thermocouples (their position is
indicated with red circles) are used to measure the temperature on
three different locations of the aluminum plate. From Figure a, we learned that to reach
the target of 10 °C, the Peltier controller element has to be
set to a lower temperature setting of approximately 5 °C. We
attribute this to imperfect contact between the heating stage and
the microfluidic chip. More importantly, both the thermocouple measurements
in Figure a and IR
camera measurements in Figure b–d indicate that the time required for the device
to reach a spatially homogeneous temperature of 10 °C is approximately
5 min. These measurements allowed us to accurately predict the time
required to obtain uniform and constant temperature that corresponds
to t0 as shown in Figure d. To summarize, these spatial and temporal
temperature measurements illustrate the temperature profile used in
microfluidic induction time measurements as indicated in Figure e. It shows that
the desired constant supersaturation is achieved 5 min after the start
of acquisition. Moreover, this point marks the start of the induction
time measurements.
Quantification of Mixing
The internal
mixing in the
droplets is observed after each bend in the microchannel by monitoring
the spatial fluorescent intensity distribution in droplets illustrated
in Figure . These
experiments are performed at different flow rates but with the same
relative flow rate Q = 1/3. The mixing after each
bend is quantified by the Danckwerts number, θ, as defined with eq .Figure d shows the Danckwerts number
values reach a plateau after five bends, indicating that the droplets
are then well-mixed. The Dean number in this case is on the order
of one, which is low and implies that little mixing takes place in
one bend. The experiment then puts a lower limit of five, which is
independent of the flow rate as previously observed in the literature.[35] Five bends correspond to a distance of approximately
8.5 mm.Fluorescent images in Figure b confirm this observation. We attribute
the oscillations
in Figure d to the
fact that the droplets are three-dimensional objects going through
baker’s transformation; yet, we observe only a projection of
the droplet in our frame of view. The intensity measured can decrease,
with dye flowing from the front of the droplet to the back of the
droplet. Similar behavior is also seen with Harshe et al.[35] but to a smaller extent.
Static Microfluidic Induction
Time Measurements
Six
different sets of microfluidic measurements have been performed, which
together give the five different profiles induction time profiles
in Figure a,b. The
channel size during these static measurements was 0.5 × 1 mm2.To model the data, we fitted 10 different models to P(t), namely, single exponential, Weibull,
and two exponential, which are generally applied, and Generalized
Pareto, Gompertz, Gumbel, extreme value, log–logistic, log–normal,
and the Pound–La Mer distribution, which are also quoted in
the literature.[3,16,42] In the Supporting Information is shown
that most often the Pound–La Mer and the two exponential are
among the best fitting ones. Both describe the nucleation process
equally well and have two time constants. However, statistically,
the two cannot be distinguished. The choice was made to present here
the results for the two exponential, as it is the simplest to interpret
as the presence of two separate phenomena with distinct rates J1 and J2. The fitted
values are given in Table . The fast component, J1, only
affects the first part of the curve and is also subject to uncertainties
in the number of nuclei lost before the final temperature is reached.
The slow component, J2, is relatively
better estimated and is therefore more meaningful.
Table 1
Two-Exponential Fit for Figure a,b: Parameters with 95% Confidence
Intervals and Effective Rates of Nucleation, J, with
Combined Uncertainty Originating from Fitting Procedure and Droplet
Volume for Different Volume Static Microfluidic Experiments
τ1 (min)
τ2 (min)
a
J1 (m–3 s–1)
J2 (m–3 s–1)
Vd(μL)
Supersaturation
S1.15, Figure 5a
0.437
5.5
1980
0.07 ± 0.03
70 × 105 ± 180 × 105
19.3 × 103 ± 1.6 × 103
0.842
27
2300
0.42 ± 0.03
7.4 × 105 ± 1.8 × 105
8.5 × 103 ± 1.1 × 103
1.270
73
1200
0.35 ± 0.03
1.8 × 105 ± 0.3 × 105
10.6 × 103 ± 0.6 × 103
S (−)
Supersaturation Vd 0.842, Figure 5b
1.10
87
18000
0.39 ± 0.05
2.3 × 105 ± 0.7 × 105
1.1 × 103 ± 1.0 × 103
1.15
27
2300
0.42 ± 0.03
7.4 × 105 ± 1.8 × 105
8.5 × 103 ± 1.1 × 103
1.20
34
1600
0.60 ± 0.02
5.8 × 105 ± 0.5 × 105
12.1 × 103 ± 2.1 × 103
To characterize the microfluidic
platform, we first report the
static induction time measurements as a function of droplet volume.
More than 100 droplets with a varying volume (0.44 ± 0.02, 0.84
± 0.02, and 1.27 ± 0.04 μL) at supersaturation S = 1.15 are produced to assess the effect of droplet volume
on induction time, as presented in Figure a and the first part of Table . The droplets with the largest
volume show the highest induction probability.The fitted parameters
in Table show that
the nucleation rates, expressed per volume
unit, are not constant as would be expected from homogeneous nucleation
following classical nucleation theory (CNT). Both the fast, J1, component and the slow, J2, component tend to decrease by volume.Next, we
investigate the influence of supersaturation ratio, S, defined as ratio of salt concentration divided by equilibrium
concentration at experimental temperature. As expected, due to the
greater driving force, faster kinetics are observed when S is the highest, as evident in Figure b. To analyze the effect, at least 200 droplets of
fixed volume at varying supersaturation levels (S = 1.1, 1.15, and 1.2) were prepared. The exact number of droplets
for each experiment is given in the Supporting Information.
Nonstatic Microfluidic and Turbidity Induction
Time Experiments
With the setup in Figure , three induction time mixing experiments
are conducted to
elucidate the influence of droplet motion and mixing on the microfluidic
induction time measurements. Channel size during these experiments
was 0.5 × 0.5 mm2. A static experiment was conducted
as a comparison with the device of the previous section (see Figure
S.4 in the Supporting Information; τ1 = 12.7 (min), τ2 = 4700 (min), a = 0.15 ± 0.02). The mixing experiments are performed with identical
supersaturation generation procedures, i.e., with identical temperature
profiles at fixed supersaturation.In the first experiment,
the droplets are moved less than a bend length to induce droplet motion
but to avoid mixing, referred as “moving”. In the second
and third experiments, the droplets are moved over respectively 12
and 31 bends to mix the droplet contents rigorously referred as “mixing”.
A more detailed description is given in the Materials
and Methods.The available parameters of the limited
exponential fit for these
moving and mixing experiments are listed in Table . They are expressed in number of bends that
the droplets have traversed. The duration of one bend is in the range
of 0.5–3 s. This could not be ascertained, but the range does
imply that the typical time constant is in the range of 1 s to 1 min.
Because many droplets did not reach the acquisition stage, the fits
were based on a limited number of points causing large uncertainties.
Table 2
Summary of Estimated Nucleation Parameters
for the Moving and Microfluidic Mixing Experimentsa
experiment
τ (bends)
a
moving
no nucleation observed
in 16 h
mixing
12 bends
12.0 ± 9.1
0.95 ± 0.12
mixing 31 bends
>900
∼1
Vd =
0.356 μL and S = 1.13.
Vd =
0.356 μL and S = 1.13.The experiment denoted as “moving” in Figure a shows an unexpected
result.
After 16 h, none of the droplets nucleated. Our initial expectation
was that this experiment would show kinetics similar to the static
case. Yet, it clearly shows considerably slower kinetics.Additionally,
induction time studies with a turbidity-based measurement
in a well-mixed, one milliliter volume for a similar supersaturation
range using Crystal16 are presented in Figure . When fitted with a two-exponential distribution,
the mean induction time of the fast component proved to be on the
order of 5 s, the sampling time of the observations. Besides, the
cooling down period was on the order of 100 s, which implies that
any process described by the fast component would have been completed
already at the first measurement point. The fit results are listed
in Table .
Figure 7
P(t) conducted at a well-mixed
1 mL volume at different supersaturations.
Table 3
Estimated Nucleation
Parameters for
a Two-Exponential Distribution Applied to the Turbidity Measurements
(Figure )a
S (−)
τ1 (min)
τ2 (min)
a
J1 (m–3 s–1)
J2 (m–3 s–1)
1.10
0.20
1.03
0.235 ± 0.059
0.83 × 105 ± 0.21 × 105
16.3 × 103 ± 1.0 × 103
1.13
0.087
0.51
0.156 ± 0.015
1.92 × 105 ± 0.21 × 105
32.6 × 103 ± 0.5 × 103
1.15
0.028
0.32
0.134 ± 0.020
6.0 × 105 ± 1.5 × 105
52.5 × 103 ± 2.5 × 103
Errors show
95% confidence intervals.
P(t) conducted at a well-mixed
1 mL volume at different supersaturations.Errors show
95% confidence intervals.
Overview
Nucleation Rates
According to the classical
nucleation theory (CNT), the dependence of the nucleation rate J on the supersaturation is given bywhere A and B denote the kinetic parameter and
the thermodynamic parameter for
nucleation[4,6] respectively, which can be obtained from
the straight line fit of ln(J/S)
versus 1/ln2S. These are plotted in Figure for both the fast
and the slow component based on the numbers in Table . It is informative to see that effective
nucleation rate values extracted from three different experimental
procedures follow CNT. The estimated parameters A and B are presented in Table , which also confirms that the thermodynamic
component, B, is positive such that the nucleation
rate increases with supersaturation.
Figure 8
Test of classic nucleation theory: Plot
of ln(J/S) as a function of 1/(ln S)2 for induction probability data on a 1 mL
scale and the static
microfluidic experiments.
Table 4
Parameters A and B in eq Applied
to the Static and Turbidity Measurements
A
B
microfluidic fast
66 × 104 ± 21 × 104
0.0090 ± 0.0073
microfluidic
slow
2.0 × 104 ± 0.8 × 104
0.0205 ± 0.0081
turbidity fast
120 × 104 ± 130 × 104
0.027 ± 0.015
turbidity slow
11.8 × 104 ± 4.4 × 104
0.0204 ± 0.0057
Test of classic nucleation theory: Plot
of ln(J/S) as a function of 1/(ln S)2 for induction probability data on a 1 mL
scale and the static
microfluidic experiments.Figure also allows
us to compare the nucleation rate from the two microfluidic devices
used in this study. Consider only the more accurate slow nucleation
rates, J2, at S = 1.13
(or 1/ln2(S) = 67 in the figure). The
prediction for the 0.84 μL from the device with channel depth
of 1 mm used for Table is slightly less than the single measurement at 0.36 μL for
the device with depth of 0.5 mm used in the mixing experiments (see Supporting Information for graph and data). The
difference is significant but is explained by the factor of two difference
in the volume, which was investigated in Figure a. So, the results of both microfluidic devices
are comparable.
Discussion
The first discussion
point is the deviations of P(t)
curves from single exponential behavior. Classical
nucleation theory predicts that P(t) curves should be faithfully represented with a single exponential
function. However, we find systematically different shapes of the
probability curves which are best described by a sum of two exponential
on the basis of their fit to the data; see Table S.1 in the Supporting Information. Such P(t) curves well-represented with a sum of two exponentials
are considered to be a product of a slow component and a fast component.
The slow component may be attributed to the primary nucleation process,
while the fast component may be interpreted as another process causing
the deviation from the single exponential.For microfluidic P(t) curves,
the fast component could be explained by nucleation of a different
polymorph.[31] As we do not individually
characterize the form of each crystal contributing to P(t), this possibility can not be entirely ruled
out. Another possibility is that the fast component originates from
potent heterogeneous nucleation on the interfaces particularly channel
walls. The third option has to do with the operation prior to t = 0. There, the droplets are moved into the channel with
a Bretherton film surrounding it, and then it is cooled down, but
this film takes time to squeeze out. So there will be a change in
the nucleation rate due to this effect until the film has completely
disappeared. The fourth option is that the volume of the drops havs
a distribution. Dos Santos et al.[16] has
shown that the observed curve approximates a Generalized Pareto distribution,
which can fit the data here as well (see Table S.1 in the Supporting Information). Finally, nonclassical
nucleation could occur in the droplets in which nucleation occurs
in two coupled steps. The formation of an unstructured dense phase
followed by the nucleation of a structured crystal in the dense phase,
however, was not observed in our experiments but might also require
more sophisticated measurement techniques. In the case of the turbidity
experiments, the initial fast components could be caused in our experiments
by the sudden reduction in the stirring rate, so it has more to do
with an operational switching effect.Next, we discuss the effect
of droplet volume on the effective
nucleation rate. CNT, see eq , predicts that with the constant S, the
homogeneous nucleation rate should remain the same irrespective of
the system volume. Our result contradict this expectation based on
the values presented in Table . The results of the experiments presented in Figure a show that the slow component
of the nucleation rate J2 decreases with
a larger droplet volume. A similar trend was reported by Steendam
et al.[32] who focused on the effect of scale
up for well-stirred experiments with volumes ranging between 10 and
660 mL. Steendam et al.[32] showed that as
the volumes went down, the nucleation rates went up. We observed similar
trends for static microfluidic experiments with three droplet volumes
ranging between 0.44 and 1.27 μL. Steendam et al.[32] suggested that a possible explanation could
be difference in the shear rates as the volumes were changed. However,
this explanation can not be used for our case as our measurements
were carried out when the droplets were stationary. Moreover, the
volumes in our experiments are significantly smaller. The large surface
to volume ratio in our experiments might explain that the heterogeneous
nucleation could play a more prevalent role. Moreover, this is in
agreement with the work by Sear[3] and Leisner
et al.[43] as well, who interpreted the dependence
of the nucleation rate on the droplet volume as a sign of heterogeneous
nucleation in the context of the classical nucleation theory. This
interpretation is also in line with the hypothesis we offer for higher
nucleation rates observed in static experiments given Table compared to moving experiments
in Table .The
next discussion point is the CNT analysis given in Figure . In all cases, the
points involved are on a straight line within error bars. This applies
to both the fast and the slow component in both the microfluidic and
the turbidity case. The interesting observation is that the slope
of the lines, represented by the thermodynamic parameter B, are all of the same order with a weighted average of 0.020 ±
0.003. This implies that all processes are possibly influenced by
the same energy differences. Table shows that there is no measurable difference in B between the microfluidic and turbidity observations. This
confirms also the conclusion of Rossi et al.[29] for adipic acid that flowing and internal mixing do not influence
the thermodynamic parameter. Also the results of Nappo et al.,[30] who compared the nucleation rates from their
microfluidic device with those obtained in a turbulent vials of milliliter
size, gave a comparable picture. The static microfluidic droplets
give a lower rate, while the mixed droplets gave much higher nucleation
rates than the turbidity-based experiments, despite the much lower
estimated shear rates in the moving droplets in the microfluidic device.Turbidity measurements have been done here as a reference for the
microfluidic nucleation rates. The J2 of
static microfluidic rates are a factor of seven lower than those from
the turbidity observations over the whole range of supersaturations
based on the fits presented in Figure . The decrease in volume would induce an increase in
the nucleation rate (as in Table ), while the absence of mixing/turbulence due to stationary
droplets would induce a decrease in the nucleation rate as has been
found by Rossi et al.[29] The volume effect
is apparently considerably larger than the droplet motion and hydrodynamic
effect. The J1 cannot be usefully compared
as the uncertainties are too high, and its cause could also be instrumental.
The nucleation rates for mixing are one or two orders of magnitude
higher than the turbidity observations (Table ). Apparently, in this case, the hydrodynamic
effect dominates, which might be explained by the intensive mixing
that the droplets undergo in the serpentine channels during the well-mixed
operation.Another issue worth discussing in Figure is the lower number data points
and shorter
experimental times for the mixing experiment compared to other two
data points. This is because of the two complications developed during
the experiments, droplet coalescence, and crystals left behind by
moving droplets blocking the channel. The first complication of droplet
coalescence emerged due to some droplets getting pinned on imperfections
due to contact angle hysteresis.[44] The
pinned droplets then coalesced with moving droplets during day-long
experiments. These imperfections may be either artifacts of machining
of channels or stretches coming from the handling of microfluidic
setup. The second complication emerges when the nucleated crystals
sediment. The sedimented crystals are left behind as the droplets
move across the bends. These crystals not only change the hydrodynamic
resistance and hence the droplet velocities but also block the channels.
Coalesced droplets or droplets which came in contact with crystals
left behind were excluded from the microfluidic induction time results.
Overall, the experiments involving droplet motion and mixing are harder
to conduct than the static experiment. Because of coalescence, increased
hydrodynamic resistance introduced by crystals formed in the channel,
and higher uncertainties and shorter experimental times are recorded
in Figure .One may also consider what other microfluidic designs might be
recommended to identify the influence of distinct mixing behaviors
on P(t). An interesting design is
a version of our design without serpentine channels. Comparing an
experiment in a version of our design with only straight channels
(no serpentine geometry) where the droplets are moved back and forth
just like the mixing experiments would be an interesting future study.
Such design will allow answering questions such as, Do the flow patterns
of mixing influence the observed nucleation kinetics? As with straight
channels, one would expect to have two symmetric circulating vortexes.
Hence, the influence of mixing patterns might be deconvoluted.The final discussion point is the interpretation of the experiments
presented in Figure . As pointed out in the Results, significantly
different P(t) curves are observed
for “static”, “moving”, “mixing
12 bends”, and “mixing 31 bends” conditions.
The expectation is that any induction time would shorten in this order.
However, the “moving” case shows no nucleation, while
the “static” case does, on the order of minutes or hours.
The “mixing 31 bends” case has slower nucleation than
the “mixing 12 bends” case. So far, the only hypothesis
that could contribute to this is the history of the droplets prior
to cooling down, as—with hindsight—this was not a fixed
protocol nor recorded. Our results support the earlier found conclusion[45] that moving droplets are surrounded by the Bretherton
film. When the channels were filled, the droplets moved into place
with this protecting film. Then, the droplets were kept stationary
for some time during which the Bretherton film was squeezed out slowly.
Subsequently, the mixing experiment starts, and the film builds up
again. The result is that the observed initial induction time is the
effect of mixing and the dynamics of varying Bretherton film. The
latter is also determined by the prior protocol. From these experiments,
we can so far only conclude that the hydrodynamics associated with
mixing significantly reduces the induction time and that these experiments
need a fixed protocol to avoid the dynamics of the Bretherton film.Nappo et al.[30] compared the nucleation
kinetics for static and moving, nonmixed droplets of para-amino benzoic
acid. They observed the enhancement of the nucleation rates by more
than two orders of magnitude with the droplets in motion. They suggested
that the shear rate might be responsible for this result. Moreover,
they observed that, beyond a certain threshold limit, a higher shear
rate adversely affected the kinetics. This could be an explanation
for the results of our mixing experiments. For the mixing case, the
shear rate would be in the optimum window, and hence the highest nucleation
rate was obtained compared to turbidity-based measurements. However,
for the moving case, the shear rate could be out of that optimum window
and as a result nucleation might get hindered. Both Rossi et al.[29] and Nappo et al.[30] hypothesized that the probable cause of an enhanced nucleation rate
could be the improved frequency of the collision among the mesoscale
clusters. Particularly, Rossi et al.[29] postulated
that the attachment frequency was enhanced for the droplets in motion
due to the internal recirculation taking place with the droplet motion.
For recirculation, mixing needs to take place. We claim that movement
in a microfluidic channel only causes a thin oil film in between the
droplet and the channel, as described by Bretherton.[45] This thin film literally lubricates the wall–droplet
interface during movement. When the observations are made, the droplets
are stagnant for some 100 s, but the breakup time is estimated to
be much longer based on the work of Kreutzer et al.[46] We employed a passive mixer so-called the serpentine geometry
of the channel to ascertain whether actual mixing is taking place.
Moreover, the mixing efficiency upon droplet motion through serpentine
channels was confirmed with fluorescent measurements. This micromixing
can improve mass transfer and therefore increase the time of observation
of the crystal that has resulted in a significantly high nucleation
rate. Another important point distinction between our study and other
aforementioned studies[29,30] is the chemical nature of solutes
studied. The solute used in our study is KCl, an ionic salt which
completely dissociates in water, whereas para-amino benzoic acid and
adipic acid, organic compounds, are used in other studies. We expect
that the nature of the interaction between the microfluidic channel
walls will be influenced by this fact.We also compare and contrast
advantages and disadvantages of the
proposed microfluidic device to previously reported experimental systems[29,30] focused on measuring nucleation kinetics under different fluid dynamic
conditions. Rossi et al.[29] and Nappo et
al.[30] used T-junctions and millimeter-sized
capillaries where the supersaturation is created by moving droplets
from one temperature bath to another one at a lower temperature. Moreover,
the nucleated crystals are moved to a third temperature bath to let
crystals grow to reach an observable size, at a temperature very close
to the saturation. This experimental setup is easy to clean due to
large diameter capillaries and modules. Furthermore, the third temperature
bath ensures that no further nucleation occurs, but crystals are grown
to a given size. Compared to this setup, the presented setup is more
difficult to clean, and once the microfluidic device is produced,
it is not possible to alter the design. Yet, the presented setup has
certain advantages. First of all, all the droplets are observed at
all times. This ensures that droplets nucleating before reaching the
desired supersaturation can be detected and taken out of construction
of the P(t) curve. As the cooling
system is integrated into the microscope, the desired temperature
hence supersaturation is quickly achieved within 5 min, and sharp
images of nucleated crystals are acquired as shown in Figure . Our design also enables complete
rapid mixing of the droplets compared to axisymmetric circulation
rolls created in droplets moving in straight tubes used by Rossi et
al.[29] In the setup used by Rossi et al.,[29] liquid droplets move along the channel at a
constant speed, and the fluid within them circulates, giving rise
to counter-rotating vortexes with closed streamlines. Yet, the symmetry
of these counter-rotating vortexes is maintained, and the fluid element
circulating can only exchange across closed streamlines via diffusion.
Moreover, that setup had the limitation of the pressure difference
that can be applied due to the longer coil, which is not the case
in our design. In summary, our device provides better images enabling
analysis of crystal shape and more homogeneous mixing.
Conclusions
We present a microfluidic platform that chaotically mixes droplet
contents using serpentine-shaped channels and an experimental methodology
to independently control droplet motion and mixing in microfluidic
induction time measurements. The temperature stabilization and mixing
in the proposed platform take place on the minute scale as deduced
from dedicated experiments. This is negligible compared with the hour
scale that the nucleation takes place in the proposed platform.Cumulative nucleation probability curves extracted from microfluidic
induction time measurements could be adequately fitted with a two-exponential
model. The long-term component is mostly well estimated and signifies
actual estimation for nucleation rates. Independent of the experiment,
it is found that the thermodynamic component of the classical nucleation
theory is the same, 0.020 ± 0.003. The experiment where the droplets
move but the contents are not mixed did not show nucleation, which
could be due to the suppression of nucleation by liquid–liquid
film shielding droplets from contact with solid channel walls. In
experiments with well-mixed moving droplets, faster nucleation kinetics
is observed, emphasizing the role of mixing in microfluidic induction
time measurements. The results were compared to turbidity experiments,
which are done with near perfect mixing.In future work, we
will focus on the automation of crystal detection,
along with the experimental extraction of size per droplet and better
control to avoid droplet coalescence. Experiments should also include
the mean nucleation time as a function of the fluid velocity during
mixing in order to help deconvolute nucleation associated with kinetics,
movement, and mixing. Hereby, the cause of the fast nucleation component
should be investigated. With these experimental challenges resolved,
more replicates of the experiments, particularly with moving droplets,
will be most informative. Moreover, we will focus on designing experiments
considering the history of droplets prior to the temperature stage.
Finally, a study involving more solutes, and a larger range supersaturation,
would be advisable.
Authors: Michelle R Bringer; Cory J Gerdts; Helen Song; Joshua D Tice; Rustem F Ismagilov Journal: Philos Trans A Math Phys Eng Sci Date: 2004-05-15 Impact factor: 4.226
Authors: Carl C L Schuurmans; Anna Abbadessa; Mikkel A Bengtson; Galja Pletikapic; Huseyin Burak Eral; Gijsje Koenderink; Rosalinde Masereeuw; Wim E Hennink; Tina Vermonden Journal: Soft Matter Date: 2018-08-01 Impact factor: 3.679