Bionic eye system mimicking microfluidic structure and intraocular pressure for glaucoma surgery training.

Toshiro Yamanaka1, Tomonori Niino2, Seiji Omata3, Kanako Harada1,4, Mamoru Mitsuishi1, Koichiro Sugimoto5, Takashi Ueta5, Kiyohito Totsuka5, Tomoyasu Shiraya5, Fumiyuki Araki5, Muneyuki Takao5, Makoto Aihara5, Fumihito Arai1.   

Abstract

Among increasing eye diseases, glaucoma may hurt the optic nerves and lead to vision loss, the treatment of which is to reduce intraocular pressure (IOP). In this research, we introduce a new concept of the surgery simulator for Minimally Invasive Glaucoma Surgery (MIGS). The concept is comprised of an anterior eye model and a fluidic circulatory system. The model made of flexible material includes a channel like the Schlemm's canal (SC) and a membrane like the trabecular meshwork (TM) covering the SC. The system can monitor IOP in the model by a pressure sensor. In one of the MIGS procedures, the TM is cleaved to reduce the IOP. Using the simulator, ophthalmologists can practice the procedure and measure the IOP. First, considering the characteristics of human eyes, we defined requirements and target performances for the simulator. Next, we designed and manufactured the prototype. Using the prototype, we measured the IOP change before and after cleaving the TM. Finally, we demonstrated the availability by comparing experimental results and target performances. This simulator is also expected to be used for evaluations and developments of new MIGS instruments and ophthalmic surgery robots in addition to the surgical training of ophthalmologists.

Entities:  

Mesh:

Year:  2022        PMID: 35816482      PMCID: PMC9273099          DOI: 10.1371/journal.pone.0271171

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.752


Introduction

Eye diseases are worldwidely increasing as the number of elderly people aged 65 and over increases. Among those diseases, glaucoma is one of leading causes of vision impairment [1]. In 2013, the global prevalence of glaucoma was estimated as 3.54% for the adult population aged 40–80 years [2]. The number of people with glaucoma worldwide was estimated as 64.3 million in 2013 and is assumed to increase to 111.8 million in 2040 [2]. Glaucoma may hurt the optic nerves and lead to vision loss, the treatment of which is to reduce intraocular pressure (IOP). Minimally invasive glaucoma surgery (MIGS) is one of the commonly used methods to treat primary open-angle glaucoma as it reduces the IOP. Trabeculotomy ab interno procedure relieves the hydraulic resistance to aqueous flow by cleaving the trabecular meshwork (TM) and the inner walls of Schlemm’s canal (SC) [3]. Ophthalmologists have to train surgical procedures before they are performed on patients, so artificial [4] or animal models are required for the training. As well as the bioethical issues, animal models have a number of issues, such as different anatomical properties from those of humans, stable procurement, characteristic guarantee, and so on. Therefore artificial simulators are considered ideal solutions [5]. Developing simulators using virtual reality (VR) is a significant challenge [6, 7]. However, compared to the VR simulators, simulators with artificial eye models are considered to be very practical because they are easy to reproduce the characteristics of the real human eye and actual surgical instruments can be applied to them. There are available systems simulating various kinds of eye surgery [8-10]. However, these systems lack important anatomical details, such as the SC structure and the spherical shape of the eye, which is necessary for the glaucoma surgery simulation. Previously we succeeded in demonstrating the anterior eye model having TM and SC structure by the blow-molding fabrication [11]. In that fabrication method, several issues remained such as poor yields, limited types of materials, constraints on other design requirements, and so on. In this research, we introduce a new concept of the glaucoma surgery simulator for MIGS. The concept is comprised of an anterior eye model with the SC structure and a fluidic circulatory system. It can mimic high IOP close to glaucoma. First, we define the requirements and target performances for the simulator and design the prototype. Next, we establish a fabrication method of the anterior eye model with TM and SC by molding and bonding flexible polymer materials, and assemble the prototype including the model. Using the prototype, we measure the pressure change before and after cleaving membrane mimicking TM. Finally, we demonstrate the availability by comparing experimental results and target performances.

Materials and methods: Concept description

Requirements for glaucoma surgery training

The structure of the anterior eye is shown in Fig 1 [12]. In the anterior eye, there is a Schlemm’s canal (SC) with an inner diameter of several hundred μm [13] around the limbus cornea. The SC is covered with a thin membrane called trabecular meshwork (TM) with a thickness of 40–132 μm [13]. Aqueous humor in the eye passes through the TM and is drained into the SC [12, 14–18]. The flow rate of aqueous humor is several μL/min [12, 17]. In glaucoma, the IOP becomes higher than 21 mmHg (2.8 kPa) [17]. In “inside-out (ab interno) surgery” [3], the TM is incised using a microhook to reduce the IOP to the normal value of about 10 mmHg (1.3 kPa) (Fig 1) [17]. For the training of the surgery, we require an anterior eye model with the SC structure and a fluid circulating system that generates the pressure conditions.
Fig 1

Anterior eye and minimally invasive glaucoma surgery (MIGS) procedure.

System configuration

Fig 2 shows the system configuration that realizes this concept. It consists of an anterior eye model made of a flexible material such as polydimethylsiloxane (PDMS) and a fluid circulating system. The anterior eye model includes an intraocular (IO) space and an SC with a TM membrane. The system has IO and SC channels connected to a reservoir through hydraulic resistance pipes. The fluid circulating system uses a pump and controller to circulate fluid, such as water, in the IO channel. The IOP is measured by a pressure sensor placed in the IO space. The flow rate of aqueous humor is measured by a flow rate sensor placed between the pump and the anterior eye model. The pump is controlled so that the flow rate is 2–3 μL/min. In the glaucoma state, the IOP is larger than about 3 kPa, depending on the constant flow rate and the hydraulic resistance between the IO space and the reservoir. When the TM is incised as the MIGS procedure, the two channels communicate with each other and the IOP drops to the normal value (about 1 kPa).
Fig 2

System configuration.

Parametric design based on the hydraulic theoretical model

Fig 3 shows the hydraulic equivalent circuit of this system. Hydraulic resistance of the pipe in the channels is generally described as follows [19]: where η, L, D are the viscosity of the fluid, the length and internal diameter of the pipe. D is designed to be the minimum value for each channel to act as the maximum resistance.
Fig 3

Hydraulic equivalent circuit.

Using (1), The IOP p before and after the MIGS is described as follows: where Q is the flow rate of the pump. Therefore the IOP drop is theoretically expected due to the decrease in the total hydraulic resistance after cleaving the TM membrane. Table 1 shows the parameters designed to realize the human condition of glaucoma.
Table 1

Design parameters.

MeaningPartSymbolValue (Units)Comment
ViscosityFluid η 1.000 (mPa s)Water (293 K)
Flow ratePump Q 3.3–5.0 × 10−2 (mm3/s)= 2.0–3.0 (μL/min)
Inner diameterPipe D 63 (μm)-
LengthPipe L 35 (mm)-
Hydraulic resistancePipe R 91 (kPa s/mm3) =128ηLπD4
Hydraulic resistancePipe R s 45 (kPa s/mm3)R/2
IOP(Glaucoma)Anterior Eye p > 2.8 (kPa)= RQ
IOP(after cleaving TM)Anterior Eye p + 1.3–2.8 (kPa) =R1+R/RsQ=RQ/3

Prototype design of glaucoma eye module

Glaucoma eye module

Fig 4 shows an eye module prototype of this concept. This module was assembled by sandwiching an anterior eye model made of PDMS from above and below with cover and base parts made of hard resin such as ABS. The cover and base parts were manufactured by 3D printing. Pipes and a pressure sensor can be attached to the base part.
Fig 4

A glaucoma eye module design.

(a) Assembled state. (b) Configuration. (c) Channel structure.

A glaucoma eye module design.

(a) Assembled state. (b) Configuration. (c) Channel structure. The eye module assembled includes the IO and SC channels as shown in Fig 4(c). In addition, the glaucoma eye module can be mounted to the eye surgery simulator (bionic eye surgery evaluator: Bionic-EyE™ [20]) which was developed by our group.

Anterior eye model with Schlemm’s Canal (SC) structure

The anterior eye model is made of PDMS. PDMS has optical transparency and is often used for building artificial organ models [21-23]. Its elastic modulus of 1.3–3.0 MPa [24], very close to that of real sclera tissue (2.0 MPa) [25]. In addition, its optical transparency facilitates a clear visualization of the SC through the cornea using a goniolens during the surgery simulation [11]. The anterior eye model was fabricated using the process as shown in Fig 5. First, the sclera and cornea parts were molded using PDMS as a material. By dividing the process into two steps, the only sclera part can be also made white-colored. The TM membrane was manufactured by spin-coating dextran as a water-soluble sacrificial layer [26] onto a curved-surface mold and then spin-coating uncured PDMS on the sacrificial layer. Next, those anterior parts and the TM membrane coated on the mold were bonded to each other just after being activated by the O2 plasma. Finally, the anterior eye model was released by dissolving the dextran layer in water.
Fig 5

Fabrication process of the anterior eye model.

Results: Proof of concept

Prototype

Fig 6(a) shows a fabricated anterior eye model using the fabrication process in Fig 5. Fig 6(b) shows a microscopic image of the section around the SC. The cross-section widths of the SC were 170 and 550 μm, which are close to those of the human eye [13]. The TM thickness was about 90 μm under the condition that the rotation speed of the spin-coating was 4000 RPM. The thickness seems valid for the simulator because the human’s TM thickness is 40–132 μm [13]. This process made it possible to manufacture this model with good yield.
Fig 6

MIGS simulator prototype.

(a) Fabricated anterior eye model. (b) A microscopic image of the section around the SC. A bar means 500 μm. The TM thickness was 90 μm and the cross-section widths of the SC were 170 and 550 μm. (c) Configuration of glaucoma eye module. (d) Assembled glaucoma eye module. A bar means 10 mm. (e) Fluid/electrical components on the outside of the glaucoma eye module. (f) The glaucoma eye module in which blue water is injected into the SC. (g) The glaucoma eye module mounted on the Bionic-EyE™ [20].

MIGS simulator prototype.

(a) Fabricated anterior eye model. (b) A microscopic image of the section around the SC. A bar means 500 μm. The TM thickness was 90 μm and the cross-section widths of the SC were 170 and 550 μm. (c) Configuration of glaucoma eye module. (d) Assembled glaucoma eye module. A bar means 10 mm. (e) Fluid/electrical components on the outside of the glaucoma eye module. (f) The glaucoma eye module in which blue water is injected into the SC. (g) The glaucoma eye module mounted on the Bionic-EyE™ [20]. The glaucoma eye module (Fig 6(d)) was assembled by sandwiching the anterior eye model between the cover and base parts (Fig 6(c)). Fig 6(e) shows fluid/electrical components on the outside of the module. The hydraulic resistance pipes made of PEEK were connected between the glaucoma eye module and the reservoir. The designed parameters D, L, R, R of the pipes are shown in Table 1. Those pipes function as dominant fluidic resistances R, R which mimic the gross aqueous outflow resistance composed of the trabecular and nontrabecular pathways [27]. The glaucoma eye module can be mounted to the Bionic-EyE™ [20] (Fig 6(g)). As shown in Fig 6(f), we confirmed that the SC structure was fabricated without any leaks by injecting blue-colored water into it.

Minimally invasive glaucoma surgery (MIGS) training

Using the glaucoma eye module with the fluid circulatory system, we demonstrated the MIGS training. We did the procedure similar to the “inside-out (ab interno) surgery” as one of the MIGS procedures. Fig 7 shows the snapshots. First, the IO channel was filled with water by turning on the pump, and at that time, this module should be in the glaucoma state under the above design. Next, a small hole was made in the cornea by incising it with a medical knife (Fig 7(1)). Observing the TM through a microscope and a goniotomy lens, we were able to cleave the TM with the microhook through the hole (Fig 7(2) and 7(3)). Finally, we closed the hole by gluing it.
Fig 7

Snapshots of the MIGS procedure.

(1) Making a small hole in the cornea by incising it. (2)(3) Cleaving the TM membrane in the eye module.

Snapshots of the MIGS procedure.

(1) Making a small hole in the cornea by incising it. (2)(3) Cleaving the TM membrane in the eye module.

Intraocular pressure (IOP) and aqueous flow rate

Figs 8 and 9 show experimental results of the aqueous flow rate Q and IOP p− in the glaucoma state, respectively. The flow rate Q was stable at 2.1±0.4 μL/min. The IOP p− was 4.18±0.06 kPa. Figs 10 and 11 show experimental results of the Q and p+ after all MIGS procedures in Fig 7, respectively. The flow rate Q was stable at 2.4±0.2 μL/min, similar to the value in glaucoma. The IOP p+ was reduced to 1.37±0.05 kPa compared to the p−.
Fig 8

An experimental result of the flow rate Q when simulating glaucoma.

The sampling interval was 0.07 s. Q was 2.1±0.4 μL/min.

Fig 9

An experimental result of the IOP p− when simulating glaucoma.

The sampling interval was 0.26 s. p− was 4.18±0.06 kPa.

Fig 10

An experimental result of the flow rate Q after cleaving the TM membrane and adhesively closing the small incised part in the cornea.

The sampling interval was 0.07 s. Q was 2.4±0.2 μL/min.

Fig 11

An experimental result of the IOP p+ after cleaving the TM membrane and adhesively closing the small incised part in the cornea.

The sampling interval was 0.26 s. p+ was 1.37±0.05 kPa.

An experimental result of the flow rate Q when simulating glaucoma.

The sampling interval was 0.07 s. Q was 2.1±0.4 μL/min.

An experimental result of the IOP p− when simulating glaucoma.

The sampling interval was 0.26 s. p− was 4.18±0.06 kPa.

An experimental result of the flow rate Q after cleaving the TM membrane and adhesively closing the small incised part in the cornea.

The sampling interval was 0.07 s. Q was 2.4±0.2 μL/min.

An experimental result of the IOP p+ after cleaving the TM membrane and adhesively closing the small incised part in the cornea.

The sampling interval was 0.26 s. p+ was 1.37±0.05 kPa.

Discussion

As shown in Fig 7, we were able to demonstrate the MIGS procedures using the glaucoma eye module, so it is considered that this module can be used for the MIGS training. In the glaucoma state (Figs 8 and 9), the aqueous flow rate Q was in the range 2–3 μL/min and IOP p− was larger than 2.8 kPa. After the MIGS procedures (Figs 10 and 11), Q stayed in that range and p+ was reduced to about p−/3. Those results meet the design in Table 1. Therefore the IOP drop by the MIGS procedure was quantitatively demonstrated similar to the human condition by using the glaucoma eye module. There were some errors in the total hydraulic resistances R, R between design values and experimental results. The resistances R = p−/Q, R = p+/Q and those errors ΔR/R, ΔR/R were experimentally estimated as 119 kPa s/mm3, 35 kPa s/mm3, 31%, and −16%, respectively. The main reason is considered to be size variations of the hydraulic resistance pipes. From Eq (1), the resistance errors are described due to the variations ΔL,ΔD as follows: From Eqs (3) and (4), ΔD/D is more sensitive to the resistance errors than ΔL/L. Assuming ΔL = 1 mm (ΔL/L = 2.9%) and ΔD = −4 μm (ΔD/D = 6.3%), the resistance error ΔR/R is estimated as 34%. Another reason might be the pressure difference by bubbles trapped in the pipe [19]. However, the flow velocity in the pipe is estimated to be 11 mm/s, and the assumed bubble should pass through the pipe in about 3 s. Therefore, the bubble pressure is unlikely to affect the steady resistance errors.

Conclusion

In this research, we designed, manufactured, and evaluated the glaucoma eye module with a fluid circulatory system as the artificial simulator for MIGS surgery. Especially, we quantitatively demonstrated the IOP drop similar to the human condition before and after the MIGS procedure by using the simulator. In the near future, we will realize a more practical glaucoma surgery simulator with a human-like appearance and characteristics by improving it based on the evaluation results of ophthalmologists. This simulator is also expected to be used for evaluations and developments of new MIGS instruments and ophthalmic surgery robots in addition to the surgical training of ophthalmologists.

MIGS procedure.

This video describes the MIGS procedure experiment in the paper, which is a real-time scale (that is, 1x speed). This video corresponds to Fig 7 in the paper. (MP4) Click here for additional data file.
  21 in total

1.  Experimental aqueous perfusion in enucleated human eyes.

Authors:  W M GRANT
Journal:  Arch Ophthalmol       Date:  1963-06

2.  Microscale culture of human liver cells for drug development.

Authors:  Salman R Khetani; Sangeeta N Bhatia
Journal:  Nat Biotechnol       Date:  2007-11-18       Impact factor: 54.908

3.  Clinical measurements of aqueous outflow.

Authors:  W M GRANT
Journal:  AMA Arch Ophthalmol       Date:  1951-08

Review 4.  A review of aqueous outflow resistance and its relevance to microinvasive glaucoma surgery.

Authors:  Nicholas H Andrew; Sarmad Akkach; Robert J Casson
Journal:  Surv Ophthalmol       Date:  2019-08-16       Impact factor: 6.048

5.  A comparison of the elastic properties of human choroid and sclera.

Authors:  T R Friberg; J W Lace
Journal:  Exp Eye Res       Date:  1988-09       Impact factor: 3.467

6.  Water-soluble sacrificial layers for surface micromachining.

Authors:  Vincent Linder; Byron D Gates; Declan Ryan; Babak A Parviz; George M Whitesides
Journal:  Small       Date:  2005-07       Impact factor: 13.281

Review 7.  Global prevalence of glaucoma and projections of glaucoma burden through 2040: a systematic review and meta-analysis.

Authors:  Yih-Chung Tham; Xiang Li; Tien Y Wong; Harry A Quigley; Tin Aung; Ching-Yu Cheng
Journal:  Ophthalmology       Date:  2014-06-26       Impact factor: 12.079

8.  Fabrication of 3D Capillary Vessel Models with Circulatory Connection Ports.

Authors:  Mahmoud Gallab; Kyohei Tomita; Seiji Omata; Fumihito Arai
Journal:  Micromachines (Basel)       Date:  2018-02-28       Impact factor: 2.891

9.  InVERT molding for scalable control of tissue microarchitecture.

Authors:  K R Stevens; M D Ungrin; R E Schwartz; S Ng; B Carvalho; K S Christine; R R Chaturvedi; C Y Li; P W Zandstra; C S Chen; S N Bhatia
Journal:  Nat Commun       Date:  2013       Impact factor: 14.919

10.  Microhook ab interno trabeculotomy, a novel minimally invasive glaucoma surgery.

Authors:  Masaki Tanito
Journal:  Clin Ophthalmol       Date:  2017-12-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.