| Literature DB >> 35816454 |
J John Lucido1, Andrew J Veres1, Shawn M Kehret1, John A Angeli2, Robert D Highet2, Robert L Foote1, Scott C Lester1, Chris L Deufel1.
Abstract
PURPOSE: For total skin electron (TSE) beam radiation therapy, the anterior eye and conjunctiva can be protected with eye shields to prevent keratitis, xerophthalmia, and cataractogenesis. Conventional metal eye shields can reduce patient balance by obscuring vision and thus increasing the risk for falls. We report on the design, fabrication, and clinical use of transparent acrylic eye shields for TSE.Entities:
Keywords: eye-shielding; patient safety; total skin electron beam radiotherapy; transparent
Mesh:
Year: 2022 PMID: 35816454 PMCID: PMC9512336 DOI: 10.1002/acm2.13722
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
FIGURE 1Design and photographs of transparent eye shields for total skin electron (TSE) treatment. Photographs showing (a) front‐ and (b) top‐view of shields marked with measured dimensions, and (c) a fully‐assembled eye‐shield. (d) A photograph of the eye shields placed on an anthropomorphic phantom for in‐phantom testing
FIGURE 2Phantom set‐up for radiation shielding measurements of the total skin electron (TSE) eye shields. Photograph of the set‐up for radiation shielding measurements. The eye shields were mounted on an anthropomorphic head phantom and placed inside the TSE treatment stand with the eye shields are at a height typical for an adult patient
FIGURE 3Evaluation of dose reduction in skin surrounding eye‐shields. (a) Photograph of the phantom wrapped in Gafchromic film used to measure the dose distribution around the shields. (b) Image of the exposed film, with lines referring to the location of the profiles obtained: #1, #2, and #3 representing the lateral, inferior, and superior profiles, respectively. (c) 2D dose distribution surrounding eye‐shields, and (d) the dose profiles as a function of distance from physical edge of shield
Transmitted dose measurements for eye shields
| Eye shield set | Shield | Transmitted dose (%) |
|---|---|---|
| Initial | 1 | 5 |
| 2 | 8 | |
| 3 | 6 | |
| 4 | 7 | |
| After cracks discovered | 1 | 6 |
| 2 | 6 | |
| At retirement | 1 | 7 |
| 2 | 7 | |
| Replacement | 5 | 7 |
| 6 | 7 |
Note: Transmitted dose to the lens with the eye shield in place as a percentage of the dose to the lens without shielding for the simulated patient treatment measured with using Gafchromic film on an anthropomorphic head phantom at treatment position. Four individual eye‐shields were fabricated initially (labeled #1–4), and this table includes the measurements made prior to first use, after the cracks were first found in the support structure for one pair of shields (#1 and #2), as well as after they were taken out of service. The measurements prior to first use of the replacement pair (#5 and #6) are also listed.
FIGURE 4Photograph of the eye shields after cracks developed during routine use. Photograph of the first production units after 6 months (approximately 100 treatments) of use. Cracks had developed in the medial aspect of both units. Film measurements indicated that the cracks did not impact radiation shielding capabilities, but the units were replaced as a precaution against potential structure failure