Literature DB >> 35814312

External quality assessment scheme for HbA1c assays in Thailand: A 5-year experience.

Supaporn Suparak1, Busadee Pratumvinit2, Kanokwan Ngueanchanthong1, Petai Unpol1, Ariya Thanomsakyuth3, Chavachol Setthaudom3, Mongkol Kunakorn3, Archawin Rojanawiwat1, Ballang Uppapong1.   

Abstract

Background: Thailand National External Quality Assessment Scheme (NEQAS) for HbA1c was established to evaluate the quality of HbA1c assays in Thailand in 2016.
Methods: HbA1c results from participating laboratories were compared to the target value assigned by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference system.
Results: The pass rates of participating laboratories during 2016-2020 were72-88%. The mean bias ranged between -0.19 and 0.20% of HbA1c. SD ranged from 0.30 to 1.08% of HbA1c. The overall coefficients of variation ranged from 4.46-15.66%. Conclusions: Performance evaluation using IFCC assigned values indicated that different assay methods had an effect on HbA1c results. Participation in external quality assessment programs for HbA1c analysis is essential for improving laboratory quality and benefiting patient management.
© 2022 The Authors.

Entities:  

Keywords:  Accuracy-based; Diabetes mellitus; External quality assessment; HbA1c; Proficiency testing

Year:  2022        PMID: 35814312      PMCID: PMC9260332          DOI: 10.1016/j.plabm.2022.e00288

Source DB:  PubMed          Journal:  Pract Lab Med        ISSN: 2352-5517


Introduction

The precision and accuracy of hemoglobin A1c (HbA1c) measurements are critical for monitoring and diagnosing diabetes mellitus [1,2]; therefore, clinical laboratories must constantly monitor the performance of their assays. External quality assessment (EQA) programs are a tool for laboratories to verify and evaluate the performance of their HbA1c assays. EQA is an interlaboratory comparison program in which EQA providers send sample panels to participating laboratories for analysis on a regular basis. Individual laboratories compare their results to those of other laboratories in a peer group or to an assigned value [3]. Comparing results with the assigned value is better at reflecting the accuracy of HbA1c testing, which is necessary for clinical diagnosis and monitoring using the HbA1c. This article reveals the experience of setting up an accuracy-based EQA scheme with the help of the European Reference Laboratory for Glycohemoglobin in values assignment with the International Federation of Clinical Chemistry (IFCC) secondary reference measurement procedures. Currently, HbA1c laboratory-based assays are commonly based on five principles: liquid chromatography [high-performance liquid chromatography (HPLC), capillary electrophoresis, affinity binding chromatography, immunoassay, and enzymatic methods [2]. Apart from laboratory-based HbA1c assays, the point-of-care (POC) analyzers were also often used in community setting, point-of-care HbA1c brought evidence-based primary care to villages [4]. However, the performance of POC HbA1c testing devices varies significantly across individual studies, limiting their application for diabetes screening and diagnosis [5]. Each method has limitations in its application, particularly in the presence of interferences; therefore, the HbA1c assay must be standardized to reduce variations between results obtained by various methods. The IFCC and National Glycohemoglobin Standardization Program (NGSP) [6] cooperatively have successfully achieved the HbA1c standardization, significantly reducing differences between results obtained by various commercial methods. In addition to using NGSP and IFCC certified methods to achieve a precise and accurate HbA1c measurement, the assays' performance with EQA materials must be monitored periodically. Previously, because of the unavailable targeted EQA scheme, all HbA1c EQA programs in Thailand were the peer group comparison. However, the National EQA HbA1c program was established in Thailand for the first time in 2016 to assess HbA1c based on target values assigned by the European Reference Laboratory for Glycohemoglobin (ERL) by using 4 IFCC calibrated secondary reference measurement procedures. This ISO/IEC 17043: 2010 certified accuracy-based HbA1c EQA program provides high-quality, affordable materials for clinical laboratories throughout the country [7]. In addition, the ISO 13528:2015 [8] quality standard system was used in the EQA program's statistical method.

Materials and methods

Sample preparation and value assignment

EQA samples were prepared in Ramathibodi Hospital using pooled human ethylene diaminetetraacetic acid residual of patient whole blood samples. The samples were tested for HbA1c values using the turbidimetric inhibition immunoassay method (Cobas c501, Roche Diagnostics, Mannheim, Germany) and HbA1c values were verified using turbidimetric inhibition immunoassay method (Cobas c513) at the Faculty of Medicine Siriraj Hospital. Both the Ramathibodi and Siriraj Hospital laboratories are NGSP level I certified which received yearly certification of traceability to the Diabetes Control and Complications Trial and a quarterly monitoring [9]. The samples were screened for hepatitis B, hepatitis C and HIV-1 viruses by chemiluminescence immunoassay (Architect, Abbott Diagnostics, Abbott Park, IL). The aliquot samples were sent to the ERL in the Netherlands in two shipments per year (2016–2017) and in one shipment per year (2018–2020) to obtain the HbA1c values with four secondary reference methods in duplicate, and the mean was calculated for the assigned value. Secondary reference methods included Roche TQ generation three on Cobas c513 (immunoassay), Tosoh G8 (ion-exchange), Trinity Biotech Premier Hb9210 (affinity chromatography), and Abbott Alinity (enzymatic assay). The samples were aliquoted into 500 ml per tube and stored at −70 ∘C until they were delivered to participating laboratories. This study was approved by the local Institutional Review Board (Ref: MURA2016/27), which waived the requirement for informed consent.

Sample packaging and delivery to participating laboratories

Frozen EQA samples are shipped twice a year (2016–2017) and three times a year (2018–2020), with four to five sample panels in each cycle. The transportation company transported the EQA samples on dry ice and delivered them directly to the laboratory (door to door) within 24 h. During the transportation, a temperature monitoring system was installed. The participating laboratories would be notified of the sample delivery date in advance. The participating laboratories were required to check the samples’ condition as soon as they arrived. If the samples were not analyzed immediately, they were stored at 2–8 °C.

Statistical analysis accuracy assessment

The accuracy performance was evaluated by the percent difference of HbA1c values the participating laboratories and the IFCC-assigned value as follows:% Difference = (Xi– IFCC assigned value/ IFCC assigned value) x 100where Xi = % HbA1c value from a participating laboratory. Acceptable limits were within ±10% (year 2016–2017), ±9% (year 2018–2019), and ±8% (year 2020). The difference of mean HbA1c values between each assay method and the assigned value was compared using paired t-test analysis by using SPSS 26.0 statistical software package (SPSS, inc., Chicago, IL, USA). The P < 0.05 was considered statistically significant.

Precision assessment

The precision was assessed using duplicated test samples and statistical analysis for within and between-laboratory Z-scores, as follows:Between laboratory Z-score = Standardized sumwhere medianSS is median of standardized sum, and NIQRSS is normalized interquartile range of standardized sum, respectively.Within laboratory Z-score = Standardized differencewhere medianSD is median of standardized difference, and NIQRSD is normalized interquartile range of standardized difference, respectively. The acceptable criteria was |Z-score| < 3.00. The mean, mean bias, SD, CV, and relative bias of each peer group as well as total results were calculated as follows:Mean %HbA1c = ΣXi / number of participating laboratoriesMean Bias = Mean% HbA1c - assigned value Where X i = % HbA1c value from participating laboratories% coefficient of variation (%CV) = (SD × 100) / Mean% HbA1c% Relative bias = (Mean % HbA1c – assigned value) x 100 / assigned value

Homogeneity test

Ten EQA samples were randomly selected from each set and analyzedfor within-sample variation by Cochran's range test. If Cochran expected value (Cexp) < Cochran critical value (Ccrit), there was no significant difference in each tube.CWhere D2= (Dup1 - Dup2)2D According to ISO13528:2015, the sample homogeneity was assessed by comparing the between-sample standard deviation (SS) with the maximum permissible error criterion for differences (δE). The proficiency test items are considered adequately homogeneous if Ss ≤ 0.1δE.

Stability test

The isochronous stability of the EQA samples panel was performed. The mean %HbA1c of the samples stored at −70 ∘C () was compared with sample stored at 2–8∘C for two weeks (). The samples were considered to be adequately stableif according to ISO13528:2015.

Results

EQA sample panels

The EQA sample panels were hepatitis B, hepatitis C and HIV-1 viruses free. The percentage difference in HbA1c values was within acceptable limits across all three laboratories (Faculty of Medicine Ramathibodi Hospital, Faculty of Medicine Siriraj Hospital, and IFCC), indicating that there was no variation between them (data not shown). EQA sample panels were adequately homogeneous and stable at 2–8 °C for at least two weeks and at −70 °C for 1 year .

Number of participating laboratories and response rate during 2016–2020

During the 2016–2020 period, the number of participating laboratories, including private and public members, increased from 101 to 236 laboratories. The response rates increased from 92% to 100% and the pass rates increased from 72% to 88% (Fig. 1).
Fig. 1

The number of participating laboratories, response, and pass rates in each cycle during 2016–2020.

The number of participating laboratories, response, and pass rates in each cycle during 2016–2020.

HbA1c instrument used

The manufactures and instruments used for HbA1c analysis was shown in Table 1. The assays included liquid chromatography, capillary electrophoresis, affinity binding chromatography, immunoassay, and enzymatic methods. The methods were increased year after year from 20 in 2016 to53 in 2020. The majority of the assay methods used were immunoassays.
Table 1

Manufacturers and instruments used for HbA1c analysis during 2016–2020.

ManufacturerInstrumentsNumber of participants report
2016
2017
2018
2019
2020
(20 Instruments)(27 Instruments)(34 Instruments)(40 Instruments)(53 Instruments)
Archem DiagnosticDirui CS 300B2543
Beckman Coulter, IncDxC-3001
Beckman Coulter LX20Pro41
Beckman Coulter AU400364
Beckman Coulter AU480374
BioSystemsBioSystems BA4002
Drawbridge HealthOlympus AU480411
Furuno Electric Co. Ltd.Furuno CA-8002
Getein Biotech, Inc.Getein 16002
Guangzhou Wondfo Biotech Co., Ltd.Finecarewonfo53
Finecare™ FIA Meter5
Home Access Health CorporationBeckman Couter AU68032
Ortho Clinical DiagnosticsVitros 4600256
Vitros 560035
Vitros 76001
Randox Laboratories LtdRx Imola2261212
RX modena2
Cobas c5021226
Cobas 800013
Cobas c31175666
Cobas c1111196918
Cobas Integra 8001742
Cobas Integra 400 Plus2733544963
Cobas c501/Cobas 600096131213222275
Cobas c5132549
Cobasc50311
Cobas Pro1
Siemens HealthcareSiemen Dimension EXL20025227
Siemen Dimension RXL2452
DCA Vantage Analysera1
Thermo Fisher Scientific OyKonelab prime 602221
Konelab 20i1
Abbott DiagnosticsArchitect C4000/C8000/Ci4100222323545
Alinity ci-series414
BIOZENXL-640/Cromatest21
DiaSys Diagnostic Systems GmbHSysmex BX-3010/Sysmex BX-4000125394641
JEOL Ltd.BIOMAJESTY JCABM6010/C14172539
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.Mindray BS400237
Mindray BS24012
Mindray BS8008
Mindray BS360E1
Mindray BS4301
Mindray BS4801
Siemens Healthcare Diagnostics IncAdvia 18001
Arkray, Inc.ADAMSTM A1C Lite HA-8380V25
ADAMS A1c HA-81602
ADAMS A1C HA-81801110251410
ADAMS A1c HA-8180V113740
Bio-Rad LaboratoriesBIORAD D102264
Jiangsu Audicom Medical Technology Co., Ltd.Audicom AC66002
Shanghai Huizhong Medical Science and Technology Co. Ltd (China)MQ-2000PT6
Shenzhen Labnovation TechnologiesLD500 HbA1c12
LABNOVATION1
GH series (GH900)21
Lifotronic H91322
Tosoh CorporationTosoh HLC-723GX587
Tosoh HLC-723G81
Greencross Medical ScienceArkray Pocket Chem A1c Advanceda2
EKF Diagnostics GmbHQuo-Lab HbA1Ca21
Quo-Test HbA1ca321
Green Cross Medis Corp.LabonaCheckA1Ca2
CERA-STAT 4000a22
OSANG Healthcare Co.,Ltd.CLOVER A1c TM Selfa3385
HemoCue® HbA1c 501a3
Trinity BiotechPremier Hb9210361
WuxiBiohermesBio&MedicalTechnology Co.,Ltd.A1c check proa29
SebiaCapilarys 3 TERA13

Point of care testing.

Manufacturers and instruments used for HbA1c analysis during 2016–2020. Point of care testing.

Laboratory performance

During 2016–2020, the pass rates by samples were 79.57–95.88%. The mean bias varied from −0.19- 0.20% of HbA1c, while the standard deviation (SD) were 0.30–1.08% of HbA1c. The overall coefficients of variation (%CV) ranged from 4.46 to 15.66% (Table 2).
Table 2

Accuracy performance and overall variability of participants’ during 2016–2020.

RoundSampleNumber of participant reported%AcceptableAssigned IFCC valueMean %HbA1cMean biasSD% CV
A-2016019787.634.984.90−0.080.5010.27
029794.855.975.92−0.050.579.62
039795.886.876.81−0.060.304.46
049594.749.159.09−0.060.495.39
B-2016019690.636.536.47−0.060.467.07
029691.678.828.74−0.080.596.77
039690.635.545.51−0.030.559.90
049691.676.516.47−0.040.395.99
A-20170115082.675.325.22−0.100.468.79
0215093.338.818.78−0.030.495.58
0315084.675.335.23−0.100.458.64
0415088.006.766.68−0.080.659.77
B-20170114888.517.567.39−0.170.597.93
0214885.815.585.44−0.140.6712.28
0314790.486.416.23−0.180.6610.63
0414788.449.729.53−0.191.0511.07
A-20180115588.395.345.30−0.040.499.16
0215589.037.497.47−0.020.658.65
0315591.616.086.06−0.020.487.90
0415588.398.668.750.091.0812.40
0515391.616.406.400.000.426.64
B-20180115392.816.336.29−0.040.568.91
0215286.1810.139.98−0.151.0410.44
0315291.457.477.43−0.040.699.34
0415088.005.365.28−0.080.417.69
0515189.406.086.06−0.020.6110.10
C-20180114891.896.086.01−0.070.325.40
0214991.957.407.34−0.060.435.79
0314989.265.335.31−0.020.397.42
0414191.496.226.10−0.120.335.33
0514187.237.997.87−0.120.455.66
A-20190118688.175.295.350.060.7513.95
0218789.305.996.070.080.8413.86
0318786.636.987.030.050.699.79
04178NA5.996.000.010.6611.05
05179NA8.698.66−0.030.788.98
B-20190118086.115.295.24−0.050.438.18
0217987.155.995.95−0.040.549.14
0318090.567.167.14−0.020.638.85
0418091.675.435.430.000.499.03
0518090.005.995.97−0.020.528.75
C-20190118787.175.305.350.050.8215.40
0218789.845.996.030.040.9115.06
0318789.847.077.120.050.9012.59
0418786.105.405.460.060.8515.66
0518789.305.996.050.060.8614.21
A-20200123086.098.038.030.000.658.10
0223084.356.446.37−0.070.568.81
0323079.575.885.78−0.100.518.80
0423081.306.446.37−0.070.589.09
0523082.615.435.35−0.080.488.90
B-20200123384.986.446.43−0.010.599.14
0223382.837.247.23−0.010.638.65
0323385.846.446.450.010.599.18
0423384.988.628.740.120.768.72
0523388.845.915.87−0.040.427.21
C-20200123192.217.137.250.120.496.76
0223188.9310.3910.590.200.666.23
0323192.216.326.410.090.345.30
0423191.396.326.410.090.548.42
0523188.485.545.630.090.376.57

NA; Not analyzed.

Accuracy performance and overall variability of participants’ during 2016–2020. NA; Not analyzed.

The relative bias and variability classified by assay methods

Comparison with the assigned value by using paired t-test, the relative biases were −3.05 to +4.54% (P = 0.248) in immunoassay methods, −5.97 to +4.23% (P = 0.745) in enzymatic methods had, −11.13 to +1.90% (P < 0.001) in HPLC and −21.25 to +31.19% (P = 0.025) in boronate affinity chromatography. Method-specific, between-laboratory CV ranged from 2.59% to 36.64%. The immunoassay method had a CV of 3.15–15.52%; enzymatic methods had 5.74–21.69%; HPLC had 2.59–12.42%, Low-pressure liquid chromatography (not analyze, N = 1), Boronate affinity chromatography had 3.75–36.64% (Table 3).
Table 3

Comparison of mean HbA1c values between each method and assigned value of EQA samples during 2016–2020 (54 samples).

Assay methodsNMean HbA1c assigned value (%)Mean HbA1c value (%)Relative bias (%)SD% CVP-value
Immunoassay546.696.71−3.05 to 4.540.21–0.943.15–15.520.248
Enzymatic method546.696.69−5.97 to 4.230.35–1.915.74–21.690.745
High performance liquid chromatography546.696.40−11.13 to 1.900.15–0.762.59–12.42<0.001
Low pressure liquid chromatography126.725.41−44.87 to −3.85NANA0.001
Boronate affinity chromatography546.696.90−21.25 to 31.190.24–2.703.75–36.640.025

NA; Not analyzed.

Comparison of mean HbA1c values between each method and assigned value of EQA samples during 2016–2020 (54 samples). NA; Not analyzed. Analysis of the bias of HbA1c testing by assay methods in the samples was divided into 3 groups: 1. HbA1c <6.3%, 2. HbA1c 6.3–6.7% and 3. HbA1c > 6.70% was shown in Fig. 2A and Table 4. The performance of the individual methods was demonstrated in Fig. 3, and found some instrument was out of acceptable criteria within ±8%.
Fig. 2

Bias and variability from the IFCC assigned value cassified by assay methods. Level 1 (N = 25 samples), level 2 (N = 8 samples), level 3 (N = 21 samples).

Table 4

The percentage relative bias and variability classified by assay methods during 2016–2020.

Assigned IFCC valueNumber of participants (Min-Max)Mean %HbA1cRelative biasSD% CV
Immunoassay
<6.30%(83–166)(4.95–6.14)(-3.05 to +2.31)(0.21–0.94)(3.54–15.52)
6.30–6.70%(83–166)(6.3–6.82)(-1.64 to +4.54)(0.24–0.51)(3.72–8.01)
>6.70%(83–166)(6.71–10.62)(-1.40 to +3.84)(0.22–9.39)(3.15–9.39)
Enzymatic method
<6.30%(1–59)(4.97–6.16)(-3.22 to +4.08)(0.35–1.01)(5.74–16.89)
6.30–6.70%(2–59)(6.2–6.5)(-2.05 to +1.31)(0.38–0.85)(5.95–13.44)
>6.70%(1–59)(6.81–10.55)(-5.97 to +4.23)(0.54–21.69)(6.73–21.69)
High Performance Liquid Chromatography
<6.30%(7–31)(4.91–5.95)(-8.88 to +1.44)(0.15–0.45)(2.6–8.87)
6.30–6.70%(7–31)(5.75–6.44)(-10.32 to +1.90)(0.16–0.69)(2.59–11.22)
>6.70%(7–31)(6.12–10.49)(-11.13 to +0.96)(0.18–12.42)(2.77–12.42)
Low Pressure Liquid Chromatography
<6.30%1(4.00–5.00)(-6.19 to −27.8)NANA
6.30–6.70%1(3.60–4.50)(-30.88 to −44.87)NANA
>6.70%1(5.20–8.10)(−3.85 to −41.04)NANA
Boronate affinity chromatography
<6.30%(1–7)(4.56–7.04)(-16.21 to +31.19)(0.24–2.44)(3.75–35.2)
6.30–6.70%(2–7)(6.1–7.37)(-6.30 to +14.44)(0.26–2.7)(3.94–36.64)
>6.70%(1–7)(5.96–10.59)(-21.25 to +15.22)(0.42–2.39)(3.97–27.31)

NA; Not analyzed.

Fig. 3

Bias and Variability from the IFCC Target classified by assay methods in 2020. Target is assigned value with acceptable limit (±8.00%). is mean±2SD of participant using each method.* 2SD = ±5.96%HbA1c ** 2SD = ±7.52%HbA1c.

Bias and variability from the IFCC assigned value cassified by assay methods. Level 1 (N = 25 samples), level 2 (N = 8 samples), level 3 (N = 21 samples). The percentage relative bias and variability classified by assay methods during 2016–2020. NA; Not analyzed. Bias and Variability from the IFCC Target classified by assay methods in 2020. Target is assigned value with acceptable limit (±8.00%). is mean±2SD of participant using each method.* 2SD = ±5.96%HbA1c ** 2SD = ±7.52%HbA1c.

Discussion

HbA1c measurement is important for the diagnosis and monitoring of diabetes. The American Diabetes Association (ADA) recommended a diagnostic cutoff value of 48 mmol/mol (6.5% HbA1c) for diabetes, and the World Health Organization (WHO) has published guidelines for using HbA1c in the diagnosis of diabetes mellitus [[10], [11], [12]]. The National External Quality Assessment Scheme for HbA1c has been established since 2016 [13]. We provided an accuracy-based EQA program using whole blood samples and the value assigned by the IFCC reference system to investigate the performance of HbA1c assays in the country. The EQA samples in our program had HbA1c values ranged from 4.90 to 10.59% covering normoglycemia, prediabetes, and diabetes. This approach allowed the laboratories participating in the proficiency testing to monitor the testing quality of all critical HbA1c values, including diabetes risk detection, diagnosis, and treatment monitoring [14]. The best approach of the EQA program for HbA1c is the use of whole blood samples and comparison to the target values assigned by using the IFCC reference procedure [15,16] and NGSP certified method [2,6,17]. Comparing their results and the targeted values, the participating laboratory could realize their HbA1c assays performance. In addition, our program also provides consultation from reference laboratories to the participants to improve and correct their testing. As the noncommutability bias of lyophilized EQA materials for HbA1c was demonstrated [18], the use of commutable EQA materials improves in the evaluation of analytical performance among participating laboratories. Whole blood used in our EQA program is more commutable than other sample types [19,20]. In addition, to ensure the stability of the fresh blood, a proper logistic arrangement with the cold-chain condition was scheduled in advance, and samples arrived within a time frame of 24 h. For the efficient HbA1c laboratory evaluation, it is essential to tighten the acceptable criteria. In 2016–2017, we assessed the performance of HbA1c assays with acceptance limits of ±10% difference compared to the assigned value and reduced to ±8% in 2021. The application of more stringent acceptable criteria will improve laboratory performance in the HbA1c analysis. In 2007, a College of American Pathologists (CAP) survey in the United States began using accuracy grading with a permissible limit of 15% of a target value; this limit was reduced to 6% in 2020 [21]. This limit varies across Europe, ranging from ±5% in Scandinavia to ±8% in Belgium, the Netherlands, and Luxemburg, and ±18% in Germany [20]. The pass rates of our participating laboratories during 2016–2020 were 72–88%. The assay methods in our EQA program included both certified and noncertified NGSP methods. The overall %CV ranged from 4.46 to 15.66%. The recommended interlaboratory HbA1c CV target was <3.5% [10]. In 2016, EQA in Germany, Belgium and the Netherlands using fresh whole blood samples with IFCC reference system target values showed between-laboratory CV of 4.1% in overall assay methods. The CAP survey in 2020, between-laboratory CV ranged from 0.7% to 5.1% [21]. The high CV in our survey could be attributed to the results of some assay methods having a bias with frozen EQA samples. The POCT has unacceptable values, probably from frozen sample, not commutable [22], the POCT results will compare with peer group. During the 2016–2020 survey, the mean percentage bias based on the total data of all participating laboratories was −0.19 - 0.20%, with a SD of 0.30–1.08. It is likely that the bias and variability of HbA1c testing were due to assay methods bias. The HPLC (P < 0.001) were both significantly biased from the assigned value. The assay methods used in Thailand were mainly similar to those used in the European market [19,23]. Positive bias for HPLC assays was also observed in CAP surveys [24] and other studies [20,25]. The laboratory should avoid using instruments with a high CV and a large bias. Tightening the acceptability limit to ±8% may be reasonable and allow for more accurate identification of poor performing laboratories and diagnostic devices. EQA schemes are a key tool for improving accuracy in individual laboratories, and manufacturers. The limitations of test principles and instruments may result in inaccurate results of HbA1c analysis. Participant laboratories where performance evaluations were not passed should investigate the source of the errors and improve their performance. HbA1c testing errors can occur at any stage, including the pre-analytical, analytical, and post-analytical phases [13]. Participant laboratories that were not passed should investigate the source of the errors and correction; therefore EQA program is a tool for improve laboratory performance.

Conclusion

We organized an EQA program using whole blood to investigate the performance of HbA1c assays across the country. Participating in the EQA program is an effective educational tool for monitoring the quality of testing systems and laboratories. When clinical laboratories, manufacturers, and EQA agencies combine efforts, the analytical performance of HbA1c assays can be significantly improved for the benefit of patients with diabetes mellitus.

Author statement

Copyright Assignment: The undersigned authors transfer all copyright ownership of this manuscript to The Japanese Respiratory Society, in the event the work is published. The undersigned authors warrant the article is original, does not infringe upon any copyright or other proprietary right of any third party, is not under consideration for publication by any other journal, and has not been published previously. The authors confirm that they have reviewed and approved the final version of the manuscript. Institutional Committee Approval: Submissions must comply with the following policies: 1) Research involving human subjects should be conducted in conformity with the Declaration of Helsinki, and should be certified that ethical and humane principles of research have been followed: 2) Freely-given informed consent from the subjects or patients must be obtained: 3) Research involving animals should be conducted in conformity with the various laws about prevention of cruelty to animals. Patients Permission: A letter of permission must be obtained from the subjects and patients no matter if the article appears in the published journal or in the online journal.
  15 in total

1.  The IFCC Reference Measurement System for HbA1c: a 6-year progress report.

Authors:  Cas Weykamp; W Garry John; Andrea Mosca; Tadao Hoshino; Randie Little; Jan-Olof Jeppsson; Ian Goodall; Kor Miedema; Gary Myers; Hans Reinauer; David B Sacks; Robbert Slingerland; Carla Siebelder
Journal:  Clin Chem       Date:  2008-02       Impact factor: 8.327

2.  Standardization and analytical goals for glycated hemoglobin measurement.

Authors:  Federica Braga; Mauro Panteghini
Journal:  Clin Chem Lab Med       Date:  2013-09       Impact factor: 3.694

3.  EurA1c: The European HbA1c Trial to Investigate the Performance of HbA1c Assays in 2166 Laboratories across 17 Countries and 24 Manufacturers by Use of the IFCC Model for Quality Targets.

Authors: 
Journal:  Clin Chem       Date:  2018-06-19       Impact factor: 8.327

4.  Comparison of IFCC-calibrated HbA(1c) from laboratory and point of care testing systems.

Authors:  Susan E Manley; Laura J Hikin; Rachel A Round; Peter W Manning; Stephen D Luzio; Gareth J Dunseath; Peter G Nightingale; Irene M Stratton; Robert Cramb; Kenneth A Sikaris; Stephen C L Gough; Jonathan Webber
Journal:  Diabetes Res Clin Pract       Date:  2014-05-23       Impact factor: 5.602

5.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Clin Chem       Date:  2011-05-26       Impact factor: 8.327

6.  Beware of Noncommutability of External Quality Assessment Materials for Hemoglobin A1c.

Authors:  Vincent Delatour; Noémie Clouet-Foraison; Stéphane Jaisson; Patricia Kaiser; Philippe Gillery
Journal:  Clin Chem       Date:  2020-02-01       Impact factor: 8.327

7.  Consensus statement on the worldwide standardization of the hemoglobin A1C measurement: the American Diabetes Association, European Association for the Study of Diabetes, International Federation of Clinical Chemistry and Laboratory Medicine, and the International Diabetes Federation.

Authors: 
Journal:  Diabetes Care       Date:  2007-09       Impact factor: 19.112

8.  Are hemoglobin A1c point-of-care analyzers fit for purpose? The story continues.

Authors:  Erna Lenters-Westra; Emma English
Journal:  Clin Chem Lab Med       Date:  2020-11-02       Impact factor: 3.694

Review 9.  HbA1c: a review of analytical and clinical aspects.

Authors:  Cas Weykamp
Journal:  Ann Lab Med       Date:  2013-10-17       Impact factor: 3.464

10.  Diagnosis and classification of diabetes mellitus.

Authors: 
Journal:  Diabetes Care       Date:  2010-01       Impact factor: 19.112

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.