Literature DB >> 35813546

A Study of Surgical Accuracy with X-Ray-Based Patient-Specific Instrument (X3DPSI®) vs Conventional Instrument in Total Knee Arthroplasty Surgeries.

Vivek Shetty1, Sajeev Shekhar1, Vikas Karade2, Amit Maurya2, Manu Sankar2, Yash Wagh1.   

Abstract

Introduction: Patient-specific instruments (PSI) have been shown to be a good solution in getting accurate bone cuts in total knee arthroplasty (TKA) in many studies. However, the need for an extra CT/MRI makes the existing PSI method costly and unsafe. X-ray-based PSI can solve these problems, if proven to be accurate. The purpose of this study was to introduce a novel method to measure the performance of X-ray to 3D-based PSI (X3DPSI ® ) in achieving planned bone cuts by comparing with conventional instruments (CI). Materials and
Methods: This was a prospective study of a total of ten patients undergoing TKA surgery. Preoperative full length lower limb scanogram was done with specialized calibration strap-on marker to develop X3DPSI ® which was used intraoperatively for the placement of guiding pins for the cutting jig. Similarly, intraoperative guide pins were placed using CI also. Calibrated intraoperative X-ray images, in AP and LAT view, were taken while the cutting-block guiding pins were placed on the distal femur and proximal tibia. These X-ray images were used to calculate the cutting plane alignments for femur and tibia and to simulate the postoperative hip-knee-ankle (HKA) angle and MAD (mechanical axis deviation), in a virtual Tabplan3D environment for both the methods.
Results: The results showed that there was a significant difference between the two methods in terms of femoral cutting plane angles and in terms of the HKA angles and MAD. There was no significant difference in the tibial cutting plane angles in the coronal plane.
Conclusion: The mechanical axis alignment achieved based on the virtual cuts made using the X3DPSI ® was significantly better than the alignment achieved based on the cuts made using the CI, in terms of both HKA and MAD. This novel method of X-ray-based PSI is a low-cost alternative to CT/MRI-based PSI if found to be accurate in future planned studies. © Indian Orthopaedics Association 2022.

Entities:  

Keywords:  PSI; TKR; X-ray based PSI; Xray-based 3D reconstruction

Year:  2022        PMID: 35813546      PMCID: PMC9232679          DOI: 10.1007/s43465-022-00623-6

Source DB:  PubMed          Journal:  Indian J Orthop        ISSN: 0019-5413            Impact factor:   1.033


  28 in total

Review 1.  Component alignment in total knee arthroplasty.

Authors:  James Benjamin
Journal:  Instr Course Lect       Date:  2006

2.  Fat embolism syndrome following the intramedullary alignment guide in total knee arthroplasty.

Authors:  J T Caillouette; S H Anzel
Journal:  Clin Orthop Relat Res       Date:  1990-02       Impact factor: 4.176

3.  3D femur model reconstruction from biplane X-ray images: a novel method based on Laplacian surface deformation.

Authors:  Vikas Karade; Bhallamudi Ravi
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-07-19       Impact factor: 2.924

4.  Surgeon perceptions regarding custom-fit positioning technology for total knee arthroplasty.

Authors:  Michael A Mont; Aaron J Johnson; Michael G Zywiel; Peter M Bonutti
Journal:  Surg Technol Int       Date:  2010-10

5.  Computer-assisted total knee arthroplasty using patient-specific templating.

Authors:  M A Hafez; K L Chelule; B B Seedhom; K P Sherman
Journal:  Clin Orthop Relat Res       Date:  2006-03       Impact factor: 4.176

6.  Results of an initial experience with custom-fit positioning total knee arthroplasty in a series of 48 patients.

Authors:  Stephen M Howell; Kyle Kuznik; Maury L Hull; Robert A Siston
Journal:  Orthopedics       Date:  2008-09       Impact factor: 1.390

7.  An economic model to evaluate cost-effectiveness of computer assisted knee replacement surgery in Norway.

Authors:  Øystein Gøthesen; James Slover; Leif Havelin; Jan Erik Askildsen; Henrik Malchau; Ove Furnes
Journal:  BMC Musculoskelet Disord       Date:  2013-07-06       Impact factor: 2.362

8.  Moderate varus/valgus malalignment after total knee arthroplasty has little effect on knee function or muscle strength.

Authors:  Justinas Stucinskas; Otto Robertsson; Aurimas Sirka; Aleksej Lebedev; Hans Wingstrand; Sarunas Tarasevicius
Journal:  Acta Orthop       Date:  2015-06-10       Impact factor: 3.717

9.  Effect of Femoral Resection on Coronal Overall Alignment after Conventional Total Knee Arthroplasty.

Authors:  Min-Wei Zhao; Lu Wang; Lin Zeng; Yu-Qing Hu; Jin-Xin Hu; Hua Tian
Journal:  Chin Med J (Engl)       Date:  2016-11-05       Impact factor: 2.628

Review 10.  Patient-Specific or Conventional Instrumentations: A Meta-analysis of Randomized Controlled Trials.

Authors:  Yipeng Lin; Wufeng Cai; Baoyun Xu; Jian Li; Yuan Yang; Xuelin Pan; Weili Fu
Journal:  Biomed Res Int       Date:  2020-03-10       Impact factor: 3.411