Han Cui1,2, Andrew Glidle2, Jonathan M Cooper2. 1. Beijing Key Lab for Precision Optoelectronic Measurement Instrument and Technology, School of Optics and Photonics, Beijing Institute of Technology, Beijing 100081, China. 2. Division of Biomedical Engineering, James Watt School of Engineering, University of Glasgow, Glasgow G12 8LT, United Kingdom.
Abstract
The determination of molecular diffusion across biomaterial interfaces, including those involving hydrogels and tissues remains important, underpinning the understanding of a broad range of processes including, for example, drug delivery. Current techniques using Raman spectroscopy have previously been established as a method to quantify diffusion coefficients, although when using spontaneous Raman spectroscopy, the signal can be weak and dominated by interferences such as background fluorescence (including biological autofluoresence). To overcome these issues, we demonstrate the use of the stimulated Raman scattering technique to obtain measurements in soft tissue samples that have good signal-to-noise ratios and are largely free from fluorescence interference. As a model illustration of a small metabolite/drug molecule being transported through tissue, we use deuterated (d7-) glucose and monitor the Raman C-D band in a spectroscopic region free from other Raman bands. The results show that although mass transport follows a diffusion process characterized by Fick's laws within hydrogel matrices, more complex mechanisms appear within tissues.
The determination of molecular diffusion across biomaterial interfaces, including those involving hydrogels and tissues remains important, underpinning the understanding of a broad range of processes including, for example, drug delivery. Current techniques using Raman spectroscopy have previously been established as a method to quantify diffusion coefficients, although when using spontaneous Raman spectroscopy, the signal can be weak and dominated by interferences such as background fluorescence (including biological autofluoresence). To overcome these issues, we demonstrate the use of the stimulated Raman scattering technique to obtain measurements in soft tissue samples that have good signal-to-noise ratios and are largely free from fluorescence interference. As a model illustration of a small metabolite/drug molecule being transported through tissue, we use deuterated (d7-) glucose and monitor the Raman C-D band in a spectroscopic region free from other Raman bands. The results show that although mass transport follows a diffusion process characterized by Fick's laws within hydrogel matrices, more complex mechanisms appear within tissues.
Understanding mechanisms
for molecular diffusion across biological
interfaces plays an important role in formulating many procedures
e.g., deepening our understanding of cell signaling or drug delivery
for the treatment of diseases (including for example the development
of hydrogel patches for transdermal drug delivery or for insertion
into wound sites to enhance healing[1]).
In such cases, bioactive agents are often preloaded within hydrogels
and subsequently diffuse across a biomaterial interface e.g., into
tissue. Effective characterization of mass transfer, through either
convective or non-convective mechanisms, provides a key predictor
of the transport process kinetics and may be important in informing
new biomaterials’ formulations and formats.[2]Current methods to quantify the rates of diffusion
often involve
labeling, include the use of fluorescent markers with confocal imaging
in a z-stack,[3] nuclear magnetic resonance
(NMR),[4] dynamic light scattering (DLS),[5,6] fluorescence recovery after photobleaching (FRAP),[7] and neutron transmission.[8] However,
tagging with such labels is not always practical and may influence
the molecule’s interaction with tissues or cells. In some cases
where the label is large, it may also influence transport properties.Previously, Raman spectroscopy has often been proposed as an effective
method that can be used with non-labeled molecules to observe diffusion
processes.[9,10] The methodology enables quantitative analysis
of a “fingerprint spectra” resulting from the characteristic
molecular vibrations of the different components in a sampled volume.
As a label-free and non-invasive detection technique,[9,10] Raman microspectroscopy has emerged as a useful tool to probe molecular
movement and species identification in complex samples, thereby yielding
spatial concentration profiles.[11−13]In one example, Chuchuen
et al.[13] measured
the spatiotemporal concentration distribution of the antiviral drug
tenofovir in a clinical gel with confocal Raman spectroscopy. Aside
from biomedical applications, the technique has also been used broadly
in materials science. For example, Geisler et al.[11] performed in situ detection at the colloidal interface
of a dissolving glass slide to obtain real-time spatial and temporal
insights into the reaction and transport process occurring, while
Peters et al.[12] realized measurements in
low-volume chip based systems, where Raman microspectroscopy combined
with microfluidics was used to measure diffusion within multicomponent
liquid phase systems.Although the characteristic fingerprint
of a molecule’s
Raman spectrum can simplify its detection, the relative weakness of
the Raman signature (with only 1 in 108 photons being Raman-scattered
following non-resonant, spontaneous excitation) has always proven
problematic. As a consequence, fluorescence (including the autofluorescence
of biological molecules) or other background scattering signals (e.g.,
from the molecule of interest or the surrounding matrix) can obscure
a sample’s signature spectrum, hindering the application of
spontaneous Raman spectroscopy.[14]To address these problems, methods of generating or collecting
Raman spectra have been developed in which these undesirable scattering
features are eliminated from the measured signals. Among these, time-resolved
Raman spectroscopy and stimulated or coherent techniques such as the
pump-probe-based stimulated Raman scattering (SRS) spectroscopy are
most notable.[15]In the case of SRS,
the Raman effect relies upon the coherence
of two stimulating laser sources with a mechanism of photon-vibration
energy transfer taking place, in which energy is transferred between
a “pump” and a “Stokes” laser beam.[16,17] This transfer results in the change in intensity of one or other
of the beams (a change in the pump intensity is called a stimulated
Raman loss or SRL, while a change in the Stokes intensity is a stimulated
Raman gain or SRG). Typically, the method of measuring the Raman signals
simply involves monitoring the SRL or SRG as a photocurrent generated
in a simple silicon diode.Apart from the advantages of being
free from fluorescent and non-resonant
backgrounds,[18] SRS also amplifies Raman
signals by several orders of magnitude due to the coherent excitation
of molecular vibrations with the ultrashort laser pulses, enabling
real-time chemical imaging of non-labeled species. Together, these
advantages overcome many of the problems associated with the complex
sample matrices found in biomedical applications.[19−23] For example, Chiu et al.[19] explored the diffusion of pure solvents into the dense keratinous
material of human fingernails, showing that this penetration was both
related to the molecular size and limited to tens of micrometers even
after 24 h for the larger solvent molecules. Dong et al.[20] used polarization-sensitive stimulated Raman
scattering from fingerprint C=C stretching vibration to visualize
amphotericin B (AmB) in single fungal cells, while Ji et al.[21] applied SRS microscopy to image amyloid plaques,
one of the key pathological features of Alzheimer’s disease
(AD), in the brain tissue of an AD mouse model (demonstrating SRS
as a rapid, label-free tool to differentiate misfolded from native
proteins).In the work below, we now describe the use of stimulated
Raman
scattering to systematically investigate the diffusion behavior of
deuterated glucose (as a readily available model small molecule) in
a hydrogel/tissue structure to mimic the drug delivery process. In
contrast to the diffusion studies above, in our paper, we explore
how the technique could be used to probe the mass transport of small
molecules in a carrier solvent (where the concentrations of active
species are lower) into hydrated and porous material. To illustrate
the motivation for using stimulated (rather than spontaneous) Raman
scattering in studies that involve naturally occurring biomaterials
(e.g., tissue), Supplementary Figure S1 shows the difficulty of resolving spontaneous Raman peaks when a
high, broad-band, scattering background is present: Supplementary Figure S1 shows that although the C–D
peak of deuterated glucose can be seen clearly in samples with low
background scattering (such as transparent hydrogels), when the glucose
is imbibed into a highly scattering sample such as a tissue slice,
the C–D peak is significantly harder to discern from other
background and fluorescence scattering features, even when the glucose
concentration is very high (e.g., 1 M).A further advantage
of the SRS methodology is that a lock-in detection
method can be used to modulate the intensity of the pump laser source,
significantly improving the sensitivity. As a consequence, for example,
detection of the C–D band of deuterated glucose gives a signal
that is easily distinguished from the background scattering signals
of, for example, the tissue components of a sample. Importantly too,
in SRS measurements, it has been shown that the signal detected is
proportional to the concentration of the species being probed[24] (see also Supplementary Figure S2). This, coupled with the rapid readout times from
the silicon detector, means that the technique is highly suited to
tracing and visualizing, for example, glucose diffusion in real time.
This allows data to be collected from highly scattering samples with
good temporal resolution, enabling the determination of whether the
observed mass transport follows the classical Fick’s law mechanism
or other models and, where appropriate, extracting the corresponding
diffusion coefficients.
Methods and Materials
Stimulated Raman Scattering System
The stimulated Raman Scattering (SRS) system shown as a schematic
in Figure is based
upon an Yb solid-state oscillator with 1043 nm central wavelength
as the excitation source. A proportion of the light used as the Stokes
beam passes through the beamsplitter, etalon, and delay system, while
the remainder is reflected by the beamsplitter to enter the fiber-feedback
optical parametric oscillator (OPO) (Stuttgart Instruments, Germany),
which is subsequently frequency-doubled using a periodically poled
lithium niobate crystal to give second-harmonic generation (SHG) as
the pump beam. The Stokes beam and pump beam are converged by a notch
filter and then focused on the sample by a lens, with the Raman signal
collected by a photodiode placed on the other side of the sample as
described in our previous paper.[25]
Figure 1
Schematic experimental
setup. An Yb solid-state oscillator allows
synchronous generation of the Raman Stokes and pump beams. Using an
etalon, the narrowband Stokes beam with picosecond pulse duration
is obtained, while the Raman pump beam is generated by pumping an
optical parametric oscillator (OPO), which is subsequently frequency-doubled
in a periodically poled crystal employing the effect of spectral compression.
SHG: second-harmonic generation, DM: dichroic mirror.
Schematic experimental
setup. An Yb solid-state oscillator allows
synchronous generation of the Raman Stokes and pump beams. Using an
etalon, the narrowband Stokes beam with picosecond pulse duration
is obtained, while the Raman pump beam is generated by pumping an
optical parametric oscillator (OPO), which is subsequently frequency-doubled
in a periodically poled crystal employing the effect of spectral compression.
SHG: second-harmonic generation, DM: dichroic mirror.In the current SRS system constructed here, a pulse
laser based
on an Yb solid-state oscillator (M-FEMTO-LAB-Yb, Montfort Laser, Austria)
was used as the excitation source, which provides 6 W of average output
power at a 1043 nm central wavelength with ca. 390 fs pulses and a
75 MHz pulse repetition rate. A 50× objective (0.55 NA Leica,
Germany) was used as the focus objective, and a 40× objective
(0.55 NA, Zeiss, Germany) was used as the collection lens. The photodiode
(PDA100A2, Thorlabs, USA) sensor was used as the signal detector,
and a lock-in amplifier (MFLI, Zurich instruments, Switzerland) works
as both modulation tool for the pump signal and for weak signal amplification
to pick up the detected SRS signal.The use of these objectives,
with numerical apertures of ∼0.55
result in a focused spot of the order of 4 μm in diameter as
verified by monitoring the signal intensity due to the C–H
band in a hydrogel while scanning the 40× collection objective
in directions perpendicular to the laser beams using motorized stages.
Similarly, the thickness of the sample voxel being probed in the SRS
measurement was estimated by translating the collecting objective
along the axis of the laser beams; this indicated that the length
of voxel was ∼20 μm. i.e., the SRS system probes voxels
of ∼4 μm × 4 μm × 20 μm.
Sample Holder
In order to measure
the transport of glucose into hydrogel and tissue samples, a glucose
solution was added to the top of a thin layer cuvette constructed
from glass coverslips that contained the hydrogel or tissue slice
(Figure ).
Figure 2
Sample holder
and tissue embedded in hydrogel: (a) Thin-layer sample
holder formed by adhering two coverslips together on three sides;
(b) arrangement of the tissue and hydrogel samples contained within
the thin-layer sample holder covered with a deuterated glucose solution
(purple crosses denote glucose molecules); an exaggerated perspective
is shown for clarity, and the side view illustrates how the tissue
sample is tightly sandwiched between the two coverslips. The red line
indicates the path of the SRS laser beams when the sample is scanned
to probe the glucose concentration at different depths from the solution
interface, i.e., it is midway in both the lateral (side-side) direction
and in the sample thickness direction so as to minimize edge effects.
Sample holder
and tissue embedded in hydrogel: (a) Thin-layer sample
holder formed by adhering two coverslips together on three sides;
(b) arrangement of the tissue and hydrogel samples contained within
the thin-layer sample holder covered with a deuterated glucose solution
(purple crosses denote glucose molecules); an exaggerated perspective
is shown for clarity, and the side view illustrates how the tissue
sample is tightly sandwiched between the two coverslips. The red line
indicates the path of the SRS laser beams when the sample is scanned
to probe the glucose concentration at different depths from the solution
interface, i.e., it is midway in both the lateral (side-side) direction
and in the sample thickness direction so as to minimize edge effects.The sample holders used in this study (Figure ) were bespoke cuvettes
comprising two #1.5
coverslips spaced apart by using either a third coverslip (of ∼0.2
or ∼0.4 mm thickness) or a microtomed tissue slice to separate
them during the adhesive process required to make a cuvette that was
sealed on three sides. The internal path length of the cell and the
outer coverslip thicknesses were accurately determined using a micrometer.
The direction of the SRS lasers were perpendicular to the front face
shown in the perspective views of Figure , with the detector placed behind the cuvette
as in a transmission measurement configuration. In addition to the
tissue in hydrogel sample shown in Figure , samples for individual hydrogel or tissue
measurements were prepared in a similar fashion.For those cases
involving the study of hydrogels, the sample was
polymerized within the cuvettes by exposure of the pre-polymer solution
to UV light (see below for details). The three different types of
sample comprised either a polymer-based hydrogel, a microtomed avian
(chicken) breast muscle model, or a sample in which the avian tissue
was embedded in the hydrogel (with the hydrogel only being in contact
with the edge of the tissue slice; Figure ).Once prepared and the cuvette’s
assembly and adhesion was
complete, samples were filled with 0.2 M phosphate-buffered saline
(PBS) and left overnight to elute any unreacted monomer species and
ensure the tissue slices were fully hydrated prior to the experiments
involving per-deuterated glucose (d-glucose-1,2,3,4,5,6,6-d7, 552003, Sigma-Aldrich, Germany) solutions.Note that by using the above method of cuvette assembly, in which
the tissue is sandwiched between two glass coverslips, there could
be concerns that when the test glucose solutions are added to the
cuvette from above the tissue, the solution could simply run down
a microscopic gap at the interface between the glass and tissue. This
would provide a rapid transport path to parts of the tissue that were
far from the bulk of the glucose solution. Thus, to prevent this from
happening, the tissue sample is held in the sandwich using the pressure
of firmly wound micrometer anvils during the gluing process.Nevertheless, to ensure that the concentrations measured by the
SRS probe were not influenced by “edge” effects at the
coverslip interfaces, the cuvette was translated along the axis of
the laser beam while monitoring the signal from the O–H band
of water. This signal was present in the tissue as well as the hydrogel
and glucose solution but not in the glass coverslips. The signal,
which was effectively zero when the laser beam focus was in either
of the two coverslip glass walls, rose to a maximum when the focus
was half way into the center of the ∼200 μm thick samples,
also served to confirm the absence of a thin layer of water between
the coverslips and tissue or hydrogel material. The subsequent C–D
measurements were taken at the translation position corresponding
to the maximum O–H signal.
Diffusion Measurements and Data Acquisition
As indicated above, the uniquely assignable C–D band of
per-deuterated glucose provides a distinct Raman marker; nevertheless,
in order that data acquisitions in which small changes in glucose
concentration could be detected with short data acquisition times
(10–100 ms), a high concentration of glucose was used.Fluorescein was added to the glucose solution and served, from a
pragmatic point of view, to be a convenient visible marker to monitor
the progress of the glucose mass transport (since it is of a similar
size) and thereby authenticate the deuterated glucose concentration
profiles obtained from the SRS measurements in these model experiments.Transmission SRS measurements were acquired by passing the laser
beams through the cuvette mounted in a motorized vertical translation
stage. A Labview program controlled the stage movement. This program
cyclically scanned the cuvette up and down in a series of 150 μm
steps (e.g., along the red line shown in Figure b), pausing after each 150 μm movement
to acquire the SRS signal. Typically, 17–25 150 μm height
translation steps were used in a single top-to-bottom scan to probe
the magnitude of the C–D band in each of the solution above
the hydrogel or tissue sample and at different distances from the
solution/sample interface into the sample (effectively probing the
glucose concentrations at different lateral depths into the sample).The start of the height (z) scan started with
the laser beams passing through the deuterated glucose solution at
a height of around 1 mm above the hydrogel (or tissue) interface and
finishing around 2.5 mm below the interface. At each height position,
a series of 10–100 ms SRS signal acquisitions were made, and
these were collected for 15 s to monitor any transient changes in
response. Typically, this resulted in a top-to-bottom scanning time
of ∼400 s and thus determined the temporal resolution of any
diffusion or kinetic data (note that smaller numbers of height steps
and shorter SRS signal accumulation times could have been used if
the processes occurring were found to be faster).The top-to-bottom
height scanning was cycled for around 15 h to
quantify the diffusion processes over an extended period. During this
period, the SRS signal was collected for the C–D band at 2100
cm–1. Data were also being collected at 1990, 2260,
2520, 2950, and 3390 cm–1 both before and after
the 15 h scanning period. These spectral regions corresponded to the
“background” regions at either side of the C–D
band, the C–H band of the hydrogel or tissue, and the O–H
water band, respectively. Importantly, collection of this data was
used in the data processing routines to verify that the changes in
C–D band signal were due to the ingress of C–D species
and not to changes in the sample background/instrument signal levels.
Furthermore, they also served to monitor any changes in the water
and polymer concentrations that might arise as the glucose diffused
into the hydrogel or tissue sample and occupied void space in these
matrices.
3D PEG Hydrogel Scaffold Fabrication
Polyethylene glycol (PEG) hydrogels were prepared by mixing a photoinitiator
(Irgacure 2959, Sigma-Aldrich) with a 4-Arm PEG-Acrylate (molecular
weight: 10 kDa, Creative PEGWorks) solution. Hydrogel scaffolds were
manufactured by photopolymerization of a solution comprising concentrations
of 10% (w/v) PEG and 0.05% w/v Irgacure 2959 in phosphate-buffered
saline (PBS, Sigma-Aldrich). The mixed solution was poured into a
sample holder for irradiation with UV light (320–390 nm, 5
mW/cm2 with 200 s exposure to achieve the polymerization
throughout the gel).
Data Processing
For the purposes
of data analysis, the “t = 0” time
of each experiment was defined by the moment when the mixture of deuterated
2 M glucose and 0.1% fluorescein solution was added to the cuvette
to form a solution layer above the gel or surrounding the tissue (Figure ). Data processing
was implemented with Matlab (version 2018b, MathWorks, USA) in order
to extract the data corresponding to the discrete height steps from
a long time series. The sample scanning process was controlled and
synchronized with data acquisition by an in-house written Labview
program. As indicated above, the C–D band signal was corrected
for the background signal from the SRS system and the sample cuvette
by subtracting the signals at 1990 and 2260 cm–1 at either side of the C–D peak (typically, these background
levels were only a few percent of the C–D signal), again using
Matlab routines.
Results and Discussion
Diffusion Process in the Hydrogel
The PEG hydrogels used provide hydrophilic polymer networks in many
biomedical engineering studies often have a water content of over
50% v/v and consequently have been studied for applications in several
fields including drug delivery, with one of the key features being
the capacity for the mass transport of small molecules within these
gels. Here, to explore the dynamics of this process, we used tracking
of the transmission Raman signal corresponding to the C–D band
of per-deuterated glucose as the molecules move through the polymer
network of the hydrogel.Signals were collected while scanning
the sample vertically over a distance of several mm and over a period
of ∼15 h to quantify the evolution of concentration profiles
over a range of distances from the solution/sample interface (Figure a). Here, the top
black line corresponds to the relative SRS Raman intensity of the
C–D band in the aqueous layer above the hydrogel (designated
as 100% intensity). The colored lines show that the C–D band
intensity within the gel is lower than in the solution, reflecting
the lower concentrations of glucose within the gel (albeit that they
continuously increase with time as mass transport progresses).
Figure 3
Results of
deuterated glucose particle diffusion in hydrogel: (a)
Diffusion profile of deuterated glucose molecules at all steps down
to 2.7 mm below the interface in pure hydrogel sample as measured
by SRS signal. (b) Comparison of SRS signal from deuterated glucose
(red line) and fluorescence (black line) 15 h after addition of the
combined glucose/fluorescein solution. (c) Variation of measured C–D
band intensity with depth into the sample for selected times after
addition of the glucose/fluorescein solution, along with variations
predicted by a Fickian model using diffusion coefficients calculated
in Table (dotted
lines). (d) Log–log plot of the diffusion depth (y axis) vs time for depths at which the glucose concentration has
reached 70% of that found in the bulk solution above the hydrogel
sample (the x axis time data is calculated from the
time taken for the concentrations at these points to reach 70% of
that found in the solution phase).
Results of
deuterated glucose particle diffusion in hydrogel: (a)
Diffusion profile of deuterated glucose molecules at all steps down
to 2.7 mm below the interface in pure hydrogel sample as measured
by SRS signal. (b) Comparison of SRS signal from deuterated glucose
(red line) and fluorescence (black line) 15 h after addition of the
combined glucose/fluorescein solution. (c) Variation of measured C–D
band intensity with depth into the sample for selected times after
addition of the glucose/fluorescein solution, along with variations
predicted by a Fickian model using diffusion coefficients calculated
in Table (dotted
lines). (d) Log–log plot of the diffusion depth (y axis) vs time for depths at which the glucose concentration has
reached 70% of that found in the bulk solution above the hydrogel
sample (the x axis time data is calculated from the
time taken for the concentrations at these points to reach 70% of
that found in the solution phase).
Table 1
Estimated Diffusion Coefficients with
Time
time (h)
0.5
1.0
1.5
2.0
2.5
3.0
8.0
14.5
estimated diffusion coefficient ×10–6 (cm2/s)
3
2.5
2.5
2.5
2.5
2.5
2.3
2
In common with diffusion processes described by
Fick’s laws,
mass transport is fast in the early stages at all depths (bearing
in mind that in the absence of glucose, the relative signal at t = 0 is 0%) and after longer periods of immersion in the
glucose solution, the intensity plateaus corresponding to the glucose
concentration in the gel reaching a steady state. As indicated earlier,
in these exploratory experiments, to check that the C–D band
SRS Raman measurements tracked the glucose concentration accurately,
we added a low concentration of fluorescein to the glucose solution
so that this could be imaged with a gel reader (Syngene PXi Touch)
at the end of the experiments. Figure b shows that the fluorescence intensity and C–D
Raman intensity track each other well when the measurements from the
different instruments are suitably scaled, i.e., the measurements
of the visible fluorescein marker support the proposition that the
invisible glucose molecules have also penetrated deep into the tissue.Figure a also shows
that for depths up to ∼500 μm, the plateau of C–D
intensity corresponds to 90–95% of that for the solution layer
above the gel, indicating that this uppermost layer of the gel is
very highly hydrated. However, at deeper positions into the gel, the
C–D intensity (or glucose concentration) only reaches 80% of
the solution value even after a period of 15 h. The lower values attained
deeper into the gel could be due to the slowness of the diffusion
process or a difference in the hydrogel polymer density at these depths.
Such a difference in density could manifest itself in both a change
in the glucose diffusion coefficient and the loading capacity of the
gel. To explore and better understand these possibilities, we examined
how well the glucose mass transport data fit to Fick’s descriptions
of diffusion.If we assume that the mass transport process in
the hydrogel is
a non-steady-state diffusion step, this process can be described by
Fick’s second law, as given in eq :Here, C is
the concentration, t is the time, D is the diffusion coefficient, and x is the diffusion
depth.When the diffusion coefficient D is
constant, eq can be
simplified to eq .The solution for eq is shown as eq .Here, C is the concentration of the moving species in the bulk
aqueous phase, C is the initial concentration
in the hydrogel, C(x, t) is the concentration at time t and distance x from the interface, and erf is the error
function, with the following definition as eq .Based
on these equations, we calculated an estimated diffusion
coefficient for a number of different times using the relative glucose
concentrations at different depths. The corresponding calculated diffusion
coefficients are shown in Table . With these estimated values,
predicted concentrations (expressed as relative C–D intensity
values) at different depths and after different durations are plotted
as the dashed line in Figure c.The comparison between the measured diffusion process
lines (the
solid line) and the predicted lines (the dash line) of Figure c was good for all combinations
of depth and time except for deep depths at early times. Here, the
measured data suggest that the glucose concentration is higher at
deeper depths than would be the case based on the average diffusion
coefficient estimated for the early time measurements. This higher
value could indicate that the diffusion is faster at depths further
away from the polymerized gel’s interface, possibly corresponding
to a difference in the degree of crosslinking cf. the formation of
a “skin-like” gel structure at the air/polymer interface
during the photopolymerization process.How well the concentration–depth–time
data fit Fick’s
diffusion laws can be confirmed by plotting the combinations of distance
and time that result in a particular concentration of glucose within
the gel, viz., following eq , the time-distance pairs that correspond to a certain value
for C(x,t), the
right hand side of eq should be a constant:Consequently, it follows
from the requirement for eq to be a constant that diffusion
depth is proportional to the square root of time, as in eq .To examine whether
the data fit this model, as an example, Figure d records combinations
of time and diffusion depth for which the concentration of glucose
is 70% of the bulk value. When plotted as a log–log plot and
considering all the data in the plot, the slope of the best fit line
is 0.44, while the theory predicts that it should be 0.5. This lower
value is attributable to the diffusion distance estimated for the
first few time points, and if these points are excluded, the gradient
increases to 0.48.Aside from the possibility that the mass
transport is not purely
described by Fick’s assumptions, the removal of the first few
time points leading to an increase in the above gradient could reflect
that either the t = 0 time is inaccurate or a small
amount of mixing occurs between the glucose solution and a thin water
layer on top of the gel when the glucose solution is added to the
cuvette (leading to a small, temporary, reduction in the glucose concentration).In summary, although the C–D Raman band SRS data for hydrogel
samples do not fit the Fick diffusion model perfectly, the agreement
is sufficiently close to suggest that, if the material is close to
homogeneous, SRS measurements are a reliable way to track the movement
of untagged molecular species.
Diffusion Process in Tissue
Since
an aim of many drug delivery strategies using hydrogels is to be able
to deliver the active molecules deep into the subcutaneous tissue,
we also explored the mass transport process within and at the interface
of microtomed muscle tissue slices. Thus, Figure shows the results of measurements by SRS
monitoring of the C–D band, with Figure a showing the movement of glucose from solution
to tissue in a heat map format.
Figure 4
Results of deuterated glucose particle
diffusion in tissue: (a)
Diffusion profile of deuterated glucose molecules at height steps
spaced 150 μm steps starting just above the tissue/solution
interface and progressing down to 2.55 mm below the interface in tissue
sample surrounded by glucose solution (cf. Figure without the hydrogel). (b) Evolution of
the glucose concentration at depths corresponding to positions A,
B, and C of (a). (c) Comparison of the glucose concentrations at selected
sample depths in hydrogel and tissue samples. Note that the curves
in (c) are offset for clarity of presentation, with the plateaus at t = 15 h having a normalized intensity of 1.0.
Results of deuterated glucose particle
diffusion in tissue: (a)
Diffusion profile of deuterated glucose molecules at height steps
spaced 150 μm steps starting just above the tissue/solution
interface and progressing down to 2.55 mm below the interface in tissue
sample surrounded by glucose solution (cf. Figure without the hydrogel). (b) Evolution of
the glucose concentration at depths corresponding to positions A,
B, and C of (a). (c) Comparison of the glucose concentrations at selected
sample depths in hydrogel and tissue samples. Note that the curves
in (c) are offset for clarity of presentation, with the plateaus at t = 15 h having a normalized intensity of 1.0.As before, the glucose concentration was measured
at different
depths into the sample up to distances of ∼2.55 mm from the
tissue/solution interface. Whereas for the hydrogel, the glucose concentration
decreases monotonically with depth for a given time (and increases
with time for a given depth; Figure ), Figure a shows that as the depth being probed increases from step
to step, there is some variability in the concentrations, indicating
that any attempt to analyze the data in terms of simple single Fick
diffusion model over the macro scale could be misleading.Instead,
if the sample is considered as a non-homogeneous structure
in which dense regions of tissue are interspersed in a network of
conducting channels, it is possible to envisage that for some parts
of the tissue slice, glucose would be transported there more rapidly
than if it had only to traverse the dense regions of tissue that were
between the particular point being measured and the physically closest
part of the tissue/solution interface. Indeed, the observation that
for this particular sample, there is a region of ∼300 μm
at the solution/tissue interface that is devoid of glucose (the 150
and 300 μm slices) and yet the glucose clearly penetrates into
much deeper depths within the sample indicates the presence of multiple
transport routes. Similarly, there is a second region of the sample
deeper in the slice (∼1650 μm) into which no glucose
penetrates yet beyond which there are regions of significant glucose
concentrations. As would be expected for inhomogeneous slices, it
was found that the presence and positions of regions of low glucose
concentration varied from sample to sample. This could, in future
be probed in greater detail through multidimensional (xyz) SRS mapping. Notwithstanding this, it is notable that for a given
position within the tissue, the glucose concentration always increased
monotonically with time, indicating that whatever mass transport routes
are in operation for particular regions of the tissue, they do not
change during the course of these measurements. This is further illustrated
in Figure b, which
shows the time course of the glucose profile for the depths indicated
by labels A, B, and C in Figure a.A further insight into the differences in
internal mass transport
networks of the hydrogel and tissue samples is obtained by comparing
the rate of mass transport in the tissue with that in the hydrogel
(Figure c). Here,
the concentration–time data for the two systems at similar
distances from the solution interface are overlaid. From this data,
it can be seen that the mass transport in the tissue sample is faster
than in the hydrogel, suggesting that regions of the tissue are more
porous than the hydrogel on a macroscopic scale even though the hydrogel
is very hydrated (80–90% water based on the polymerization
conditions).
Diffusion Process in Tissue Embedded in Hydrogel
Finally, to mimic a biomedical application, we examined a third
example consisting of tissue embedded into the hydrogel (Figure ), in which the deuterated
glucose solution is added to the top of the hydrogel layer. As before,
19 depth steps were performed, this time focusing on traversing the
hydrogel/tissue interface, with 6 × 150 μm steps being
in the hydrogel. A representative spatiotemporal distribution map
of the concentration found for hydrogel/tissue samples is shown in Figure a. The most noticeable
difference between the profiles for the hydrogel and tissue parts
of the sample is that at long times, the concentration of glucose
within the tissue appears to be lower than in the hydrogel. This may
reflect the lower water content of the tissue, restricting the amount
of glucose that can be taken up.
Figure 5
Results of deuterated glucose particle
diffusion in sandwich sample.
(a) Diffusion profile of deuterated glucose molecule at all the steps
down to 2.85 μm below the interface in tissue embedded in hydrogel
sample by SRS. (b) 2D view of (a). (c) Comparison of the diffusion
profile at a distance of 450 μm below the solution/hydrogel
interface or the hydrogel/tissue interface for the pure hydrogel and
the tissue in hydrogel samples, respectively. (d) Comparison of the
diffusion profile in the tissue at 2100 μm below the hydrogel/tissue
or solution/tissue interface in the tissue in hydrogel or pure tissue
in glucose solution samples. Note that these two depths were chosen
as being relatively close and far from the source of the d7-glucose, i.e., the solution or hydrogel phase.
Results of deuterated glucose particle
diffusion in sandwich sample.
(a) Diffusion profile of deuterated glucose molecule at all the steps
down to 2.85 μm below the interface in tissue embedded in hydrogel
sample by SRS. (b) 2D view of (a). (c) Comparison of the diffusion
profile at a distance of 450 μm below the solution/hydrogel
interface or the hydrogel/tissue interface for the pure hydrogel and
the tissue in hydrogel samples, respectively. (d) Comparison of the
diffusion profile in the tissue at 2100 μm below the hydrogel/tissue
or solution/tissue interface in the tissue in hydrogel or pure tissue
in glucose solution samples. Note that these two depths were chosen
as being relatively close and far from the source of the d7-glucose, i.e., the solution or hydrogel phase.If the time course plots for mass transport into
the gel and into
the tissue are separately normalized to their long-exposure time (15
h) values, as in Figure c,d, it can be seen that the relative rate of increase in glucose
concentration is generally faster in the tissue slice compared to
the hydrogel layer, as expected from the preceding results.
Conclusions
In this paper, we demonstrate
that stimulated Raman scattering
is an effective way for measuring transport properties of molecules
in thick and highly scattering tissue samples. Unlike spontaneous
Raman and many other traditional methods, this approach is neither
subject to fluorescence interference nor reliant on labeling samples
with large “tags”. Instead, the method exploits Raman
spectroscopy’s feature of being a non-invasive method for identifying
and quantifying molecules in a mixed environment. For the simple model
systems studied here, the proposed method provides fundamental insights
into the mass transport processes occurring, illustrating that such
SRS-based methods offer future promise in the study of more complex
systems. We also believe that in the future, this technique opens
up the possibility of using more sophisticated fingerprint identifications
to probe evolving concentration profiles of macromolecular structures
within tissue samples in vivo. The technique also has the potential
to demonstrate ligand binding where a molecule may bind to particular
features in the tissue (at which point diffusion would stop), including
for example antibody binding to cells.
Authors: Wing Sin Chiu; Natalie A Belsey; Natalie L Garrett; Julian Moger; M Begoña Delgado-Charro; Richard H Guy Journal: Proc Natl Acad Sci U S A Date: 2015-06-08 Impact factor: 11.205
Authors: Christian W Freudiger; Wei Min; Brian G Saar; Sijia Lu; Gary R Holtom; Chengwei He; Jason C Tsai; Jing X Kang; X Sunney Xie Journal: Science Date: 2008-12-19 Impact factor: 47.728
Authors: Minbiao Ji; Michal Arbel; Lili Zhang; Christian W Freudiger; Steven S Hou; Dongdong Lin; Xinju Yang; Brian J Bacskai; X Sunney Xie Journal: Sci Adv Date: 2018-11-16 Impact factor: 14.136