Reduced Collateral Tissue Damage Using Thermal-Feedback-Based Power Adaptation of an Electrosurgery Inverter.

Congbo Bao1, Sudip K Mazumder1.   

Abstract

Well-selected power with accurate delivery is of importance in electrosurgery to generate proper temperature at the cutting site, and thus, reduce undesired collateral tissue damages. Conventional electrosurgery generator (ESG) targets tracking a preset power, manually set by surgeons per their experience before the surgery, with high accurate delivery. It is possible that this fixed power setting is not at the optimal point and, thus, increases the possibility of added-collateral biomedical tissue damage. To eliminate the potential negative impact of the fixed and ill-suited power setting, a real-time feedback control scheme is outlined in this article to adjust the preset power of the ESG to create an adaptive power reference, which is then tracked using an experimental high-frequency inverter (HFI) that enables electrosurgery with a fundamental (sinusoidal) output frequency of 390 kHz. Subsequently, experiments using the gallium nitride (GaN)-based HFI are carried out to demonstrate the efficacy of the new variable-power approach over the conventional fixed power approach in terms of collateral tissue damage reduction.

Entities:  

Keywords:  Adaptation; collateral damage; electrosurgery; feedback; inverter; power; temperature; thermal

Year:  2022        PMID: 35782315      PMCID: PMC9249024          DOI: 10.1109/tpel.2022.3179301

Source DB:  PubMed          Journal:  IEEE Trans Power Electron        ISSN: 0885-8993            Impact factor:   5.967


Introduction

Electrosurgery has been used for around a century [1]. It applies HF voltage across conductive biomedical tissue along with the current dictated by the tissue impedance to elicit a clinical effect, such as cutting and coagulation [2]. The mechanism of biotissue incision or removal roots in the Joule energy converted from the applied electrical active power. The tissue liquid is rapidly heated up by the energy to the point of vaporization and then the tissue disperses in form of smoke and stream [1], [3]. The cutting effect on a certain tissue is tightly related to the total energy delivered to it. As is known, energy is the integration of power and time, therefore, both cutting speed and cutting power impose an impact on the final cutting effect [1]–[5]. Mismatched cutting speeding or poorly-regulated cutting power either is not able to generate desired clinical effect or may cause undesired added tissue damage, such as charring, thermal spread, dragging, and so on [4]–[6]. ESG cutting speed is exclusively controlled by surgeons according to their clinical experience and expertise. Therefore, traditional ESG aims at delivering the power, manually set by surgeons before the surgery, as accurately as possible regardless of the tissue variation. It is worth mentioning that, conventionally, this power is maintained the same during the entire electrosurgery until it is manually updated by the surgeon. As a result, the interruption of the time-sensitive surgery inevitably occurs. Furthermore, El-Kebir et al. [5] showed that different ESG power settings have an impact on cutting effects and thermal response, but El-Kebir et al. [5] incorporated no thermal-feedback-based real-time power adaptation. Therefore, there is a chance that the power is not optimally orderly set and leads to increased tissue damage or undesired cutting effects. The literature [6]–[9] either focuses on improving the power tracking accuracy or pursues a prompt response to the power setting. In [10], infrared sensing is employed to record liver surface-temperature distribution under a single power setting, and it turns out that a small radial region surrounding the electric scalpel (ES) has the highest temperature during cutting. Moreover, it is reported that thermography-based sensing can be utilized to identify solar cell aging, hotspots, and partial shading faults [11]. However, thermal sensor is employed in existing work simply for temperature measurement. None of them links the thermal feedback with real-time power adaptation for electrosurgery. As such, ill-suited power setting issue persists and its resolution via power adaptation is of interest for an ESG. To tackle the ill-suited power setting challenge and reduce collateral tissue damage, a novel real-time thermal-feedback-based closed-loop control scheme is pursued, as illustrated in Fig. 1, in which a tissue surface temperature measurement from the electrosurgery incision site is used to adjust the ESG preset power to create an adaptive power reference, which is then tracked by a power-controller in an experimental HFI. Section II outlines how an ill-suited ESG setting can cause added tissue damage while Section III outlines the thermal-feedback-based power adaptation scheme. Section IV provides experimental results while Section V concludes this article.
Fig. 1.

Illustration of a thermal-feedback-based monopolar electrosurgery using an ESG. The current- and voltage-sensing transformers are denoted as CT and PT, respectively. The HF transformer (T) boosts the output voltage (Vout (t)). The tissue is placed between the ES and return pad. The thermal camera monitors tissue surface temperature.

Illustration of Impact of Ill-Suited Power Setting

More than 70% weight of soft tissue comprises water and the tissue is removed when the applied energy vaporizes the water at 100 °C. Tissue vaporization is accompanied by another phenomenon, namely, tissue denaturation when tissue cell temperature is between 60 and 100 °C [12]. Physical properties of tissues among individuals exhibit differences associated with gender, age, size, etc. [10], and those differences escalate the extent of tissue denaturation when the constant power is indiscriminately applied. Tissue denaturation occurs due to undesired thermal spread, and it should be minimized by adjustment in power reference Pref(t), by modulating it with ΔP(t) around the Ppreset, when cutting speeding is fixed [4]. We illustrate this point in Fig. 2 with two scenarios. In the first scenario, as illustrated in Fig. 2(a), Pref(t) is not varied but kept fixed at Pref(t) = Ppreset. Because ΔP(t) = 0 for this scenario, if Ppreset is ill-suited and say higher than what is really needed for safe electrosurgery, then, in the absence of power adaptation, more energy is delivered in a time interval. Consequently, the temperature of a larger volume of tissues reaches the vaporization and denaturation range, dictated by the overall specific heat capacity. As a result, unnecessary tissue removal and denaturation occur, leading to collateral damage even in the presence of a power-control loop since the latter will only ensure that . In the second scenario, as illustrated in Fig. 2(b), Pref(t) has the ability to vary with time due to the power adaptation ΔP(t) that is guided by the thermal feedback from the incision site. As such, Pref(t) = Ppreset + ΔP(t), which ensures that a desired Pref(t) is obtained to minimize the collateral tissue damage.
Fig. 2.

Illustration of impact of ill-suited power setting.

Power Control Schemes: Constant Versus Adaptive

The full-bridge and bandpass filter-based HFI was initially introduced in [7] and detailed in [8] and it is redrawn in Fig. 3 with the integration of thermal-feedback-based power adaption.
Fig. 3.

Illustration of output power control for the HFI. The power controller tracks the power reference ( ()). The thermal-feedback-based power adaption control autonomously modifies the preset power () via a power adaption (Δ()). ADC is the analog to digital converter module. Digital signal processor (DSP) used here is TMS320F28379D. Thermal sensor is of 24 × 32 pixels. () is an array of dimension 24 × 32 that represents the spatio-temporally sensed tissue surface temperature while is the maximum of () in each sensing refresh frame. Data transmission delay (τ) is negligible.

The fundamental (sinusoidal) output frequency and switching frequency of the HFI are set to be 390 kHz. Referring to Fig. 3, the ideal sensed output voltage of the abovementioned HFI (V(t)) is captured where Vinand f are the input voltage and the full-bridge switching frequency, respectively. n is the transformer turn ratio, α is the phase shift angle between the diagonal switches in the full-bridge [8], and vscaling represents the voltage-sensor scaling. The corresponding output power P(t) for a linear load, with a load angle of θ, is given by where V is the peak of V(t) and I is the peak of i(t), the scaled output current of the HFI. ρscaling is the coefficient that maps sensed V(t) and i(t) back to actual output power. As seen, the first item of the P(t) is a constant component, which is the average power over an HFI fundamental-output cycle while the second term represents an ac perturbation term that evolves at double the fundamental-output frequency. This article aims to properly control the output voltage, and thus, current, such that is well-regulated and follows the adaptive power reference to reduce collateral tissue damage.

Constant Power Control

The control block diagram in Fig. 3 achieves constant power control by setting ΔP(t) = 0 and yielding Pref(t) = Ppreset. The constant power reference is then compared with feedback and the error is fed to a proportional-integral-based (PI) controller, which ensures the realization of the desired power. In actual electrosurgery, surgeons select Ppreset according to their clinical needs and professional experience. However, owing to physical properties variations among patients, Ppreset may be improperly set, which may cause erroneous incision-site temperature due to a mismatch between the fixed power reference set for the electrosurgery (i.e., Pref(t) = Ppreset) and the Pref(t) that is actually needed. Consequently, undesired cutting effects and collateral tissue damages are elicited, such as thermal spread, carbonization, and burns.

Thermal-Feedback-Based Power-Adaptation Control

To mitigate the collateral tissue damage due to nonoptimality in Ppreset, the proposed thermal-feedback-based power adaptation control monitors tissue surface temperature using an infrared sensor and feeds it back at a frequency of 8 Hz. Based on the sensed temperature data, the power-adaptation controller feeds in real time a power adaption term (ΔP(t)) to the Pref(t) yielding the following, as captured in Fig. 3 In (3), Ttissue(t) is the tissue surface temperature, max (Ttissue(t)) is the maximum of Ttissue(t), and Tnom is the nominal tissue temperature that ensures safe electrosurgery with minimal/no collateral damage. By following (4), any time the max (Ttissue(t)) exceeds Tnom, Pref(t) is so adjusted such that the incision-site tissue temperature is brought back close to Tnom thereby mitigating collateral tissue damage.

Experimental Results

A gallium nitride field effect transistor (GaN-FET)-based hardware prototype, as shown in Fig. 4, is used to examine the cutting effects of the proposed thermal-feedback-based power adaptation control. As shown in the top left corner in Fig. 4, MLX90640 infrared sensor is mounted together with ES such that they move together without relative movement. Consequently, tissue surface temperature around ES is always monitored, and thus, the maximum temperature of entire tissue surface is captured during electrosurgery. The sensed thermal data are processed by Raspberry Pi 4 Model B and then transmitted to the DSP controller, which, using this feedback, implements the power adaptation and subsequently, generates the PWM signals for the GaN FETs (i.e., GS66508B from GaN System) of the full-bridge converter of the HFI, as shown in Fig. 3. Fresh pork with 16 millimeters (mm) thickness is placed between the ES and return pad as biomedical load of the HFI. The detailed experimental parameters are summarized in Table I.
Fig. 4.

Experimental setup of the HFI. Programmable Emile3 3-axes robotic gantry is employed to hold the ES and move it in a specified direction and at a uniform speed for the repeatability of the biomedical tissue cutting experiments.

TABLE I

Parameters for the Experimental Hardware Setup

ParametersValues
Input voltage (Vin)110 V
Scaling factorvscaling = 0.00415, ρscaling = 154.60
HFI parametersThe same as in [8]
ES moving speed5 mm/s
PI parametersKp = 0.035, Ki = 160 000
Using this setup, first, the efficacy of HFI operating, under constant power control with Pref(t) = Ppreset = 50 W (i.e., ΔP(t) = 0) is shown in Fig. 5 with experimental results. This scenario emulates a conventional electrosurgery, where the required output power is preset by the surgeons and the setting remains the same till the surgeons manually change it again.
Fig. 5.

Statistical distribution of output power . Symbol represents the overall average of but spread over all of the 200 samples.

Next, in Fig. 6, the scenario when Pref(t) = Ppreset + ΔP(t) is explored. Five test scenarios are pursued as captured in Table II: test scenarios 1 and 2 correspond to ΔP(t) = 0 and Ppreset set at 50 W and 60 W, respectively; test scenarios 3–5 correspond to ΔP(t) ≠ 0 and instead, ΔP(t) is generated for each of the three scenarios based on three values of nominal temperature (Tnom) of 50 °C, 55 °C, and 65 °C, respectively.
Fig. 6.

Capture of numbered test scenarios. (a) Top view. (b) Front view.

TABLE II

Operating Conditions for the Five Test Scenarios

SettingsTrialsTest Scenarios
#1#2#3#3′#4#5
P preset 50 W60 W60 W60 W60 W60 W
ΔP(t)00Using (3), Tnom = 50 °CUsing (3), Tnom = 50 °CUsing (3), Tnom = 55 °CUsing (3), Tnom = 65 °C
Pref (t)50 W60 WUsing (4)Using (4)Using (4)Using (4)
The purpose of the test scenarios 1 and 2 is to provide an illustrative approach to the choice of Tnom for determining ΔP(t) using (3) [to obtain Pref(t) using (4)] in the last three scenarios. As evident in Fig. 6 and Table III, the test scenario 1 corresponding to Pref(t) = Ppreset = 50 W yields less cutting gap and thermal spread than test scenario corresponding to Pref(t) = Ppreset = 60 W. Based on these two tests, an estimated Tnom range was conjectured to be between 50 and 65 °C guided by the outcomes of test scenarios 1 and 2.
TABLE III

Evaluation of Experimental Cutting Effects for the Five Test Scenarios

MetricsTrialsTest Scenarios
#1#2#3#3′#4#5
Gap at L10.660.060.100.040.090.06
Gap at L20.020.400.050.050.100.19
Spread L10.900.820.950.981.371.51
Spread L20.021.990.990.901.161.70
Gap difference0.640.340.050.010.010.13
Spread difference0.881.170.040.080.210.19

Note: All measurements are gauged in mm.

Using that range, the test scenarios 3–5, following Table II, were pursued and the results are captured in Table III. The latter shows that test scenario 3 together with its repeated trial 3’ yields overall the best results and this is evident in Fig. 6 as well. It illustrates that, compared to test scenario 1, which yielded comparable cutting temperature but with a larger temperature ripple, the results are better for test scenario 3 because of the power reference adaptation, as shown in Fig. 7(a), which also adjusts the maximum tissue surface temperature accordingly with smaller temperature ripple, as shown in Fig. 7(b). The superior result for test scenario 3 over test scenarios 4 and 5, as evident in Table III and Fig. 7, is attributed to the fact that the Tnom for the latter two cases are set to be higher, leading to higher averaged cutting site temperature and thus wider cutting gap and thermal spread. It worth mentioning that the thermal spread repeatability of test scenario 3 are demonstrated by test 3’ and can be extended to other scenarios as well. Furthermore, the cutting gap uniformity with power adaption control is also repeatable, as verified by Table III.
Fig. 7.

(a) Adaptive power reference that comes into effect at 0 where the maximum tissue surface temperature () exceeds 40 °C (slightly higher than normal body temperature) and ends at 1. (b) Maximum of sensed tissue surface temperature (). Cutting period starts at and ends at . The temperatures of test scenarios 1 and 2 fluctuate with a larger ripple due to lack of power adaption. However, test scenarios with power adaption, namely test scenarios 3–5, feature smaller temperature ripples and, thus, contribute to improved cutting uniformity. The hottest part of the ES is buried inside the tissue during cutting process and thermal sensor, under nominal condition, does not have access to it till the end of cutting. ES tip buried during cutting is directly exposed to thermal sensor after the cutting is completed, and thus, a temperature spike occurs in the plot, which gradually decays afterward.

Conclusion

This article outlines a thermal-feedback-based power adaption control to reduce the collateral tissue damage for electrosurgery. The impact of the ill-suited power setting is illustrated, and the power adaption scheme is elaborated. A GaN-based HFI setup is used to examine the cutting effects of constant power and thermal-feedback-based power adaption control in terms of cutting gap and thermal spread. The experiment results show that the cutting test with a higher constant power features a larger cutting gap and wider thermal spread. Moreover, with a constant power setting, the tissue is not evenly cut with visible cutting gap and thermal spread difference at two-incision locations. With the proposed power adaption control, the cutting gap and thermal spread are tangibly reduced with the proper choice of nominal temperature. Moreover, it is also found that thermal sensor location and its resolution have an impact on the accuracy of sensing the surface temperature of the tissue and needs careful attention. Further, the gap and thermal difference at different incision sites are found to be reduced compared to results obtained using a fixed power setting. In other words, cutting uniformity with power adaption is improved in terms of both cutting gap and thermal spread. It is evident that power adaptation in the vicinity of accurate nominal temperature is the key to reduced collateral damage. In practical electrosurgery, this estimate can be obtained more accurately given the repeatability of reliable cutting performance and the availability of a much larger electrosurgery database.
  6 in total

1.  Electrosurgery: history, principles, and current and future uses.

Authors:  Nader N Massarweh; Ned Cosgriff; Douglas P Slakey
Journal:  J Am Coll Surg       Date:  2006-03       Impact factor: 6.113

2.  Electrosurgery with cellular precision.

Authors:  Daniel V Palanker; Alexander Vankov; Philip Huie
Journal:  IEEE Trans Biomed Eng       Date:  2008-02       Impact factor: 4.538

3.  A new dual current-mode controller improves power regulation in electrosurgical generators.

Authors:  Daniel A Friedrichs; Robert W Erickson; James Gilbert
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2012-02       Impact factor: 3.833

4.  Energy Dissipation in Ex Vivo Porcine Liver During Electrosurgery.

Authors:  Wafaa Karaki; Ali Akyildiz; Suvranu De; Diana-Andra Borca Tasciuc
Journal:  IEEE Trans Biomed Eng       Date:  2016-07-27       Impact factor: 4.538

5.  Stabilization of Electrosurgical Cutting Performance Based on Electrode Speed.

Authors:  Vlad Bluvshtein; Lori Lucke; Matt Widule
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2019-07

6.  Multiresonant-Frequency Filter for an Electrosurgery Inverter.

Authors:  Congbo Bao; Sudip K Mazumder
Journal:  IEEE Trans Power Electron       Date:  2021-12-23       Impact factor: 6.153

  6 in total