Multiresonant-Frequency Filter for an Electrosurgery Inverter.

Congbo Bao1, Sudip K Mazumder1.   

Abstract

This paper presents a multi-resonant-frequency (MRF) filter for a high-frequency inverter (HFI) used in electrosurgery. The fundamental (sinusoidal) output frequency of the HFI is 390 kHz and is the same as the switching frequency of the HFI. The MRF is designed to extract the fundamental frequency of the tri-state bipolar waveform, generated by the HFI operating with phase-shift control. The structure and operation of the MRF are outlined. An experimental 300 W GaN-FET-based HFI prototype is developed to validate the feasibility of the proposed MRF under closed-loop control.

Entities:  

Year:  2021        PMID: 35300016      PMCID: PMC8923619          DOI: 10.1109/tpel.2021.3137525

Source DB:  PubMed          Journal:  IEEE Trans Power Electron        ISSN: 0885-8993            Impact factor:   6.153


  1 in total

1.  Electrosurgery: history, principles, and current and future uses.

Authors:  Nader N Massarweh; Ned Cosgriff; Douglas P Slakey
Journal:  J Am Coll Surg       Date:  2006-03       Impact factor: 6.113

  1 in total
  1 in total

1.  Reduced Collateral Tissue Damage Using Thermal-Feedback-Based Power Adaptation of an Electrosurgery Inverter.

Authors:  Congbo Bao; Sudip K Mazumder
Journal:  IEEE Trans Power Electron       Date:  2022-05-31       Impact factor: 5.967