Differences in the Measurement of Functional Residual Capacity Between Body Plethysmographs of Two Manufacturers.

Peter Alter1, Jan Orszag2,3, Emiel F M Wouters4,5, Claus F Vogelmeier1, Rudolf A Jörres2.   

Abstract

Body plethysmography is a fundamental method for the assessment of static lung volumes. Although equipment is provided by several manufacturers, there are no established cross-vendor tools for quality control. In the process of quality control and analysis of data on hyperinflation from the large COPD cohort COSYCONET, hints appeared that plethysmographs of different vendors might yield different values for static lung volumes. Functional residual capacity (FRC) differed about 0.67 litres between plethysmographs of the manufacturer Vyaire and Ganshorn. Absolute differences of residual lung volume (RV) and total lung capacity (TLC) were similar. It appears undeniable that differences of this magnitude have an impact on clinical interpretation. Thus, device harmonization seems to be required.
© 2022 Alter et al.

Entities:  

Keywords:  body plethysmography; functional residual capacity; lung function; static lung volumes

Mesh:

Year:  2022        PMID: 35774592      PMCID: PMC9239195          DOI: 10.2147/COPD.S363493

Source DB:  PubMed          Journal:  Int J Chron Obstruct Pulmon Dis        ISSN: 1176-9106


Introduction

Body plethysmography is a major tool in the inventory of lung function measurements and suited for the assessment of static lung volumes, especially functional residual capacity (FRC), residual volume (RV) and total lung capacity (TLC). Its physiological basis and practical implementation have been described in detail.1 Equipment is provided by a number of manufacturers. There seem to be no independent devices for quality control to be used by all manufacturers. In view of the intricacies of the method, one has to consider the possibility that plethysmographs of different manufacturers show systematic differences in measured values. In the multicenter COPD cohort COSYCONET (COPD and Systemic Consequences - Comorbidities Network), body plethysmography is part of assessments.2 In the process of quality control and analysis, more and more hints appeared that plethysmographs of different manufacturers located at different participating study centers might yield different values for FRC and derived parameters.3,4 This raised the question, whether suspected differences between devices could be verified, whether it reached relevant magnitude, and which factors might be responsible. If such differences should exist and exceed a certain magnitude, they might have consequences with regard to the applicability of reference values, irrespective of the use of ECSC (European community for coal and steel)5 or more recent equations.6,7

Methods

Data from COSYCONET were used. Anthropometric, clinical and functional data including spirometry (forced expiratory volume in 1 s, FEV1; forced vital capacity, FVC) and body plethysmography (FRC, RV, TLC) were obtained following standardized protocols in 30 study centres throughout Germany.2 The study protocol was approved by the Ethical Committees of all centres, and all participants gave their written informed consent. The study was performed in accordance with the declaration of Helsinki. ClinicalTrials.gov: NCT01245933. The centres using a plethysmograph of the manufacturer Vyaire (CareFusion, Jaeger; Höchberg, Germany) provided data from 1988 COPD patients of GOLD grades 1–48 as described previously (38.1% female, mean (SD) age 65.1±8.4 years, GOLD grades 1/2/3/4: 9.4/42.6/38.5/9.5%).4 Additionally, we included 156 patients (51.9% female, mean (SD) age 64.1±9.2 years, GOLD grades: 1/2/3/4: 8.3/46.8/33.3/11.5%) from one centre using a Ganshorn (Niederlauer, Germany) plethysmograph, applying the same criteria of data validity and completeness.4 Thus, the present analysis included 2144 patients. Analysis of baseline characteristics and comparison among both groups was performed (Table 1). As we did not measure the same patient in both plethysmographs, we than performed linear regression analyses of static lung volumes (FRC, RV, TLC), either as % predicted9 or as absolute values in litres, as well as of RV/TLC %. As previously,4 predictors were FEV1 and FVC, both as % predicted,10 sex, age, height and body mass index (BMI), and introducing an indicator variable for the type of plethysmograph. The magnitude of the respective regression coefficient would then indicate the difference between devices after adjustment for anthropometric and spirometric characteristics.
Table 1

Baseline Characteristics

Jaeger n = 1988Ganshorn n = 156p value
GOLD grades 1/2/3/4187/847/766/188 (9.4/42.6/38.5/9.5%)13/73/52/18 (8.3/46.8/33.3/11.5%)0.485
Age [years]65.1 ± 8.464.1 ± 9.20.184
Height [cm]171 ± 9171 ± 100.840
BMI [kg/m2]26.6 ± 5.226.5 ± 4.90.791
Sex [male/female]1231 (61.9%)/757 (38.1%)75 (48.1%)/81 (51.9%)<0.001
Baseline Characteristics

Results

Analysis of the group of 156 patients showed that the relationship between predictors, especially FEV1 and FVC, and RV/TLC % was the same as in the larger group,4 thus patients did not appear to be qualitatively different and the pooled analysis was valid. Regarding RV/TLC %, it revealed a significantly (p<0.001 each) smaller average value in the 156 patients (Δ=−6.9; 95% CI: −7.9; −5.8). The same was true for % predicted values of FRC (Δ=−20.9; 95% CI: −25.4; −16.4), RV (Δ=−34.0; 95% CI: −40.5; −27.6), and TLC (Δ=−10.2; 95% CI: −12.7; −7.7), and for the absolute values in litres of FRC (Δ=−0.67; 95% CI: −0.82; −0.53; Figure 1), RV (Δ=−0.77; 95% CI: −0.92; −0.63), and TLC (Δ=−0.61; 95% CI: −0.77; −0.46). Details are shown in Table 2.
Figure 1

Difference in FRC between body plethysmographs. The horizontal axis shows FEV1 % predicted (GLI), the vertical axis FRC in litres measured by either the Vyaire (blue circles, n=1988) or the Ganshorn (red diamonds, n=156) plethysmograph. Additionally, LOESS fit lines are shown for both groups in the respective colors, indicating a difference of more than 0.5 litres between FRC values at a given value of FEV1 % predicted. When residual terms after adjustment for FEV1, FVC, sex, age, height and BMI were used, essentially the same picture was obtained, with the difference that the dependence on FEV1 disappeared. This demonstrated that the difference shown in the graph was not due to residual confounding but a robust result even in unadjusted data.

Table 2

Results from Multivariate Regression Analyses of the Dependent Variables Given in the Top Row and the Predictors Given Below

PredictorsRV/TLC [%]FRC %PredictedRV %PredictedTLC %PredictedFRC [l]RV [l]TLC [l]
FEV1 %pred−0.2 [−0.2; −0.2]−1.4 [−1.5; −1.3]−1.5 [−1.7; −1.4]−0.6 [−0.6; −0.5]−0.05 [−0.05; −0.04]−0.04 [−0.04; −0.03]−0.04 [−0.04; −0.03]
FVC %pred−0.2 [−0.3; −0.2]0.5 [0.4; 0.6]−0.1 [−0.2; 0.1]0.5 [0.5; 0.6]0.02 [0.02; 0.02]0.00 [0.00; 0.00]0.03 [0.03; 0.04]
Age [years]0.3 [0.2; 0.3]−0.1 [−0.3; 0.0]−1.0 [−1.2; −0.8]−0.2 [−0.3; −0.2]0.00 [0.00; 0.01]0.01 [0.01; 0.02]−0.02 [−0.02; −0.01]
Height [cm]−0.1 [−0.2; −0.1]0.3 [0.1; 0.5]0.2 [−0.1; 0.4]−0.2 [−0.3; −0.1]0.04 [0.04; 0.05]0.03 [0.02; 0.04]0.07 [0.07; 0.08]
BMI [kg/m2]−0.2 [−0.2; −0.1]−1.4 [−1.7; −1.2]−1.3 [−1.6; −1]−0.4 [−0.6; −0.3]−0.05 [−0.05; −0.04]−0.03 [−0.04; −0.02]−0.03 [−0.03; −0.02]
Sex [female vs male]3.1 [2.3; 3.8]15.2 [11.9; 18.5]14.2 [9.4; 18.9]7.8 [5.9; 9.7]−0.43 [−0.53; −0.32]−0.24 [−0.35; −0.13]−0.81 [−0.93; −0.70]
Plethysmograph [Ganshorn vs Jaeger]−6.9 [−7.9; −5.8]−20.9 [−25.4; −16.4]−34.0 [−40.5; −27.6]−10.2 [−12.7; −7.7]−0.67 [−0.82; −0.53]−0.77 [−0.92; −0.63]−0.61 [−0.77; −0.46]

Notes: Non-standardized regression coefficients and their 95% confidence intervals are given. % predicted refers to GLI reference values. Please note that the differences between plethysmographs regarding the absolute volumes are of similar magnitude and that their confidence intervals overlap, indicating that the difference regarding FRC is that which determined those regarding RV and TLC as derived parameters.

Abbreviations: FRC, functional residual capacity; RV, residual volume; TLC, total lung capacity.

Results from Multivariate Regression Analyses of the Dependent Variables Given in the Top Row and the Predictors Given Below Notes: Non-standardized regression coefficients and their 95% confidence intervals are given. % predicted refers to GLI reference values. Please note that the differences between plethysmographs regarding the absolute volumes are of similar magnitude and that their confidence intervals overlap, indicating that the difference regarding FRC is that which determined those regarding RV and TLC as derived parameters. Abbreviations: FRC, functional residual capacity; RV, residual volume; TLC, total lung capacity. Difference in FRC between body plethysmographs. The horizontal axis shows FEV1 % predicted (GLI), the vertical axis FRC in litres measured by either the Vyaire (blue circles, n=1988) or the Ganshorn (red diamonds, n=156) plethysmograph. Additionally, LOESS fit lines are shown for both groups in the respective colors, indicating a difference of more than 0.5 litres between FRC values at a given value of FEV1 % predicted. When residual terms after adjustment for FEV1, FVC, sex, age, height and BMI were used, essentially the same picture was obtained, with the difference that the dependence on FEV1 disappeared. This demonstrated that the difference shown in the graph was not due to residual confounding but a robust result even in unadjusted data.

Discussion

In the present analysis we addressed the question, whether body plethysmographs of two manufacturers showed systematic differences in the measured values of static lung volumes. Indeed, there were such differences. In direct comparison, the Ganshorn device yielded lower and the Vyaire devices higher values of static lung volumes, especially functional residual capacity (FRC). These findings were obtained via regression analysis, adjusting for patient characteristics known to affect lung volumes.3 Notably, they were consistent regarding both absolute values and values in % predicted. The absolute differences of FRC, RV and TLC between plethysmographs were similar, with large overlap of confidence intervals, indicating that the 0.67 L difference in FRC translated into parallel differences in RV and TLC. As usual, these parameters were derived from FRC via consecutive breathing maneuvers involving the expiratory reserve volume (ERV) and the inspiratory vital capacity (VCin). It is clear that % predicted values differed as they depended on the volumes used for reference; they reached average values of about 21% for FRC, 34% for RV and 10% for TLC. It seems difficult to deny that differences of this magnitude are relevant, both as absolute values and as % predicted. Apart from the variability arising from the choice of reference equations, such differences are likely to affect the interpretation of plethysmographic results, potentially leading to marked over- or underestimation of hyperinflation depending on the device used. It is unlikely that the observed differences are confined to the equipment of just the two manufacturers chosen. Their occurrence and magnitude could depend on factors involving hardware, software and the manner the shutter manoeuvre is performed. There are no obligatory calibration devices to be used by all manufacturers. We consider differences in software to be more likely and better amenable to future standardization. The shutter manoeuvre is usually performed at the end of normal tidal expiration, or shortly thereafter, with correction for the inspired volume. To our knowledge, in most plethysmographs, FRC is taken as that volume, while in others half of the tidal volume is added. While this might be adequate regarding the computation of airway resistance (Raw) from specific airway resistance (sRaw) as a value averaged over the breathing cycle, it implies a difference in FRC, and consequently in the computed Raw.1 According to our experience, actual tidal volume may be as large as 1.0 or 1.5 L, despite the fact that more shallow breathing is recommended, thus a numerical difference of 0.67 L may easily be introduced. We also were informed by manufacturers on differences in the advice given to patients prior to the shutter manoeuvre which could affect FRC by promoting changes in lung volumes. In the present analysis, we cannot resolve these issues but the fact of large differences in FRC and derived volumes is not affected. It also has to be noted that the measurements in all study centers were performed did not involve any unusual procedures and were performed in accordance with both guidelines and what we know from clinical routine procedures. Thus, they have impact for clinical practice not just the cohort study from which the data used in this analysis were taken. Our findings reveal a source of variation in body plethysmographic results that should ameliorated by harmonization between manufacturers. Variations of the observed magnitude hamper the usefulness of body plethysmography, and we believe that clinicians and researchers should be aware of this in order to exploit the full potential of this powerful lung function tool.

Take Home Message

Body plethysmographs of two manufacturers showed marked differences in the measured values of static lung volumes that likely to have an impact on clinical interpretation. Some harmonization seems needed.
  9 in total

Review 1.  Body plethysmography--its principles and clinical use.

Authors:  C P Criée; S Sorichter; H J Smith; P Kardos; R Merget; D Heise; D Berdel; D Köhler; H Magnussen; W Marek; H Mitfessel; K Rasche; M Rolke; H Worth; R A Jörres
Journal:  Respir Med       Date:  2011-02-26       Impact factor: 3.415

2.  Global Strategy for the Diagnosis, Management, and Prevention of Chronic Obstructive Lung Disease 2017 Report: GOLD Executive Summary.

Authors:  Claus F Vogelmeier; Gerard J Criner; Fernando J Martinez; Antonio Anzueto; Peter J Barnes; Jean Bourbeau; Bartolome R Celli; Rongchang Chen; Marc Decramer; Leonardo M Fabbri; Peter Frith; David M G Halpin; M Victorina López Varela; Masaharu Nishimura; Nicolas Roche; Roberto Rodriguez-Roisin; Don D Sin; Dave Singh; Robert Stockley; Jørgen Vestbo; Jadwiga A Wedzicha; Alvar Agusti
Journal:  Eur Respir J       Date:  2017-03-06       Impact factor: 16.671

3.  Static lung volumes and airway resistance reference values in healthy adults.

Authors:  Beate Koch; Nele Friedrich; Henry Völzke; Rudolf A Jörres; Stefan B Felix; Ralf Ewert; Christoph Schäper; Sven Gläser
Journal:  Respirology       Date:  2013-01       Impact factor: 6.424

4.  The German COPD cohort COSYCONET: Aims, methods and descriptive analysis of the study population at baseline.

Authors:  Annika Karch; Claus Vogelmeier; Tobias Welte; Robert Bals; Hans-Ulrich Kauczor; Jürgen Biederer; Joachim Heinrich; Holger Schulz; Sven Gläser; Rolf Holle; Henrik Watz; Stephanie Korn; Nina Adaskina; Frank Biertz; Charlotte Vogel; Jørgen Vestbo; Emiel F M Wouters; Klaus Friedrich Rabe; Sandra Söhler; Armin Koch; Rudolf A Jörres
Journal:  Respir Med       Date:  2016-03-11       Impact factor: 3.415

Review 5.  Lung volumes and forced ventilatory flows. Report Working Party Standardization of Lung Function Tests, European Community for Steel and Coal. Official Statement of the European Respiratory Society.

Authors:  P H Quanjer; G J Tammeling; J E Cotes; O F Pedersen; R Peslin; J C Yernault
Journal:  Eur Respir J Suppl       Date:  1993-03

6.  Prediction of air trapping or pulmonary hyperinflation by forced spirometry in COPD patients: results from COSYCONET.

Authors:  Peter Alter; Jan Orszag; Christina Kellerer; Kathrin Kahnert; Tim Speicher; Henrik Watz; Robert Bals; Tobias Welte; Claus F Vogelmeier; Rudolf A Jörres
Journal:  ERJ Open Res       Date:  2020-07-27

7.  Multi-ethnic reference values for spirometry for the 3-95-yr age range: the global lung function 2012 equations.

Authors:  Philip H Quanjer; Sanja Stanojevic; Tim J Cole; Xaver Baur; Graham L Hall; Bruce H Culver; Paul L Enright; John L Hankinson; Mary S M Ip; Jinping Zheng; Janet Stocks
Journal:  Eur Respir J       Date:  2012-06-27       Impact factor: 16.671

8.  Lung volume reference values for women and men 65 to 85 years of age.

Authors:  Francisco Garcia-Rio; Ali Dorgham; José M Pino; Carlos Villasante; Cristina Garcia-Quero; Rodolfo Alvarez-Sala
Journal:  Am J Respir Crit Care Med       Date:  2009-09-10       Impact factor: 21.405

9.  Influence of body mass on predicted values of static hyperinflation in COPD.

Authors:  Peter Alter; Klaus F Rabe; Holger Schulz; Claus F Vogelmeier; Rudolf A Jörres
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2018-08-23
  9 in total