Accuracy of pelvic measurements on virtual radiographic projections based on computed tomography scans compared to conventional radiographs pre- and postoperatively.

Dominik Kaiser1, Armando Hoch2, Christoph Stern3, Stefan Sommer3,4,5,6, Reto Sutter3, Patrick O Zingg2.   

Abstract

BACKGROUND: The anteroposterior (ap) radiograph of the pelvis is decisive in the diagnosis of different pathologies of the hip joint. Technical advantages have reduced the radiation dose of pelvic CT to levels comparable to radiographs. The purpose of this study was to validate if standard radiographic parameters (lateral center edge angle, medial center edge angle, acetabular index, acetabular arc, extrusion index, crossover sign and posterior wall sign) can accurately be determined on radiograph-like projections reconstructed from the CT dataset pre- and postoperatively.
METHODS: A consecutive series of patient with symptomatic dysplasia of the hip and a full radiologic workup (radiographs and CT scan pre- and postoperatively) who underwent periacetabular osteotomy were included. Standard radiographic parameters were compared between radiographs and radiograph-like projections by two authors pre- and postoperatively.
RESULTS: A total of 16 hips (32 radiographs/32 radiograph-like projections) were included in the study. No significant difference was found between the radiographs and radiograph-like images for all parameter for both examiners. ICC between radiograph and radiograph-like projections for all investigated parameters showed good to excellent reliability (0.78-0.99) pre- and postoperatively.
CONCLUSION: Radiograph-like projections show comparable results to radiographs with regard to the important investigated parameters (lateral center edge angle, medial center edge angle, acetabular index, acetabular arc, extrusion index, crossover sign and posterior wall sign). Thus, ultra-low-dose CT scans may reduce the need for conventional radiographs in pre- and postoperative analyses of 3-dimensional hip pathologies in the future, as the advantages increasingly outweigh the disadvantages.
© 2022. The Author(s).

Entities:  

Keywords:  Acetabular index; Developmental dysplasia of the hip; Femoroacetabular impingement; Lateral center edge angle

Year:  2022        PMID: 35767038     DOI: 10.1007/s00402-022-04476-4

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  4 in total

1.  [COXOMETRY. MEASUREMENT OF THE BASIC ANGLES OF THE ADULT RADIOGRAPHIC HIP BY A COMBINED PROTRACTOR].

Authors:  M LEQUESNE
Journal:  Rev Rhum Mal Osteoartic       Date:  1963-07

2.  [False profile of the pelvis. A new radiographic incidence for the study of the hip. Its use in dysplasias and different coxopathies].

Authors:  M LEQUESNE
Journal:  Rev Rhum Mal Osteoartic       Date:  1961-12

3.  What are the radiographic reference values for acetabular under- and overcoverage?

Authors:  Moritz Tannast; Markus S Hanke; Guoyan Zheng; Simon D Steppacher; Klaus A Siebenrock
Journal:  Clin Orthop Relat Res       Date:  2015-04       Impact factor: 4.176

4.  The prognosis in untreated dysplasia of the hip. A study of radiographic factors that predict the outcome.

Authors:  S B Murphy; R Ganz; M E Müller
Journal:  J Bone Joint Surg Am       Date:  1995-07       Impact factor: 5.284

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.