Tae Youn Kim1, Yun-Kyoung Jung2, Sun Hwa Yoon2, Sun Ju Kim3, Kyoung-Chul Cha3, Woo Jin Jung3, Young Il Roh3, Soyeong Kim3, Sung Hwa Kim4, Dae Ryong Kang4, Sung Oh Hwang5.
Abstract
This study aimed to assess the trend of the maintenance status and usability of public automated external defibrillators (AEDs). Public AEDs installed in Seoul from 2013 to 2017 were included. An inspector checked the maintenance status and usability of the AEDs annually using a checklist. During the study period, 23,619 AEDs were inspected. Access to the AEDs was improved, including the absence of obstacles near the AEDs (from 90.2% in 2013 to 99.1% in 2017, p < 0.0001) and increased AED signs (from 34.3% in 2013 to 91.3% in 2017, p < 0.0001). The rate of AEDs in normal operation (from 94.0% in 2013 to 97.6% in 2017, p < 0.0001), good battery status (from 95.6% in 2013 to 96.8% in 2017, p = 0.0016), and electrode availability increased (from 97.1% in 2013 to 99.0% in 2017, p < 0.0001); the rate of electrode validity decreased (from 90.0% in 2013 to 87.2% in 2017, p < 0.0001). The overall rate of the non-ready-to-use AEDs and AEDs with less than 24-h usability accounted for 15.4% and 44.1% of the total number of AEDs, respectively. Although most AEDs had a relatively good maintenance status, a significant proportion of public AEDs were not available for 24-h use. Invalid electrodes and less than 24-h accessibility were the main reasons that limited the 24-h usability of public AEDs.Entities:
Mesh:
Year: 2022 PMID: 35750888 PMCID: PMC9232625 DOI: 10.1038/s41598-022-14611-1
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.996
Figure 1Flow chart showing the enrollment process for AEDs. AED, automated external defibrillator.
Locations of AEDs.
| Facility type | Total (N = 23,619) | 2013 (n = 3134) | 2014 (n = 3402) | 2015 (n = 5622) | 2016 (n = 5909) | 2017 (n = 5552) |
|---|---|---|---|---|---|---|
| Residential settings | 11,564 (48.96) | 1778 (56.73) | 1882 (55.32) | 2738 (48.70) | 2640 (44.68) | 2526 (45.50) |
| Multi-use facilities | 2732 (11.57) | 33,810.78) | 280 (8.23) | 644 (11.45) | 674 (11.41) | 796 (14.34) |
| Schools | 2413 (10.22) | 74 (2.36) | 300 (8.82) | 556 (9.89) | 1320 (22.34) | 163 (2.94) |
| Public buildings | 2126 (9.00) | 491 (15.66) | 113 (3.32) | 759 (13.5) | 274 (4.63) | 489 (8.81) |
| Police offices | 1515 (6.41) | 28 (0.89) | 212 (6.23) | 294 (5.23) | 4778.07) | 504 (9.08) |
| Welfare facilities | 929 (3.93) | 161 (5.14) | 258 (7.58) | 11 (0.20) | 148 (2.50) | 351 (6.32) |
| Transportation facilities | 693 (2.93) | 0 | 0 | 306 (5.44) | 11 (0.19) | 376 (6.77) |
| Industrial facilities | 637 (2.70) | 148 (4.72) | 2 (0.06) | 134 (2.38) | 159 (2.69) | 194 (3.49) |
| Public health/medical clinics | 249 (1.05) | 67 (2.14) | 70 (2.06) | 13 (0.23) | 63 (1.07) | 36 (0.65) |
| Hotels and conference venues | 181 (0.77) | 31 (0.99) | 40 (1.18) | 26 (0.46) | 39 (0.66) | 45 (0.81) |
| Religious facilities | 137 (0.58) | 5 (0.16) | 6 (0.18) | 82 (1.46) | 20 (0.34) | 24 (0.43) |
| Prisons | 7 (0.03) | 1 (0.03) | 3 (0.09) | 1 (0.02) | 0 | 2 (0.04) |
| Other | 436 (1.85) | 12 (0.38) | 236 (6.94) | 58 (1.03) | 84 (1.42) | 46 (0.83) |
Data are presented as n and (percentage) from the total number of AEDs per year.
AED automated external defibrillator.
AED management status.
| Parameter | Total (N = 23,619) | 2013 (n = 3134) | 2014 (n = 3402) | 2015 (n = 5622) | 2016 (n = 5909) | 2017 (n = 5552) | |
|---|---|---|---|---|---|---|---|
| Designation of AED manager | 22,434 (94.98) | 2717 (86.7) | 2985 (87.7) | 5551 (98.7) | 5688 (96.3) | 5493 (98.9) | < 0.0001 |
| CPR and AED training for the manager | 21,744 (92.06) | 2808 (89.6) | 2994 (88.0) | 5405 (96.1) | 5637 (95.4) | 4900 (88.3) | 0.2258 |
| A regular check of AED by the manager | 20,079 (85.01) | 2670 (85.2) | 2690 (79.1) | 4824 (85.8) | 5145 (87.1) | 4750 (85.6) | < 0.0001 |
| Absence of obstacles near AED | 22,532 (95.39) | 2826 (90.2) | 3117 (91.6) | 5223 (92.9) | 5862 (99.2) | 5504 (99.1) | < 0.0001 |
| Presence of the AED guide sign | 13,937 (59.00) | 1074 (34.3) | 1092 (32.1) | 1996 (35.5) | 4709 (79.7) | 5066 (91.3) | < 0.0001 |
Data are presented as n and (percentage) from the total number of AEDs per year.
CPR cardiopulmonary resuscitation, AED automated external defibrillator.
Maintenance status, accessibility, and actual use of AEDs.
| Parameter | Total (N = 23,619) | 2013 (n = 3134) | 2014 (n = 3402) | 2015 (n = 5622) | 2016 (n = 5909) | 2017 (n = 5552) | |
|---|---|---|---|---|---|---|---|
| Normal operating status | 22,893 (96.92) | 2947 (94.0) | 3289 (96.7) | 5463 (97.2) | 5776 (97.8) | 5418 (97.6) | < 0.0001 |
| Good battery status | 22,861 (96.79) | 2996 (95.6) | 3282 (96.5) | 5463 (97.2) | 5745 (97.2) | 5375 (96.8) | 0.0016 |
| Availability of electrodes | 23,342 (98.82) | 3043 (97.1) | 3351 (98.5) | 5582 (99.3) | 5868 (99.3) | 5498 (99.0) | < 0.0001 |
| Valid electrodes | 20,578 (87.12) | 2822 (90.0) | 3117 (91.6) | 5113 (91.0) | 4685 (79.3) | 4841 (87.2) | < 0.0001† |
| 24-h accessibility | 15,771 (66.77) | 2070 (66.0) | 2500 (73.5) | 3842 (67.5) | 3840 (65.0) | 3519 (63.4) | < 0.0001† |
| Actual use of the AED | 224 (0.99) | 22 (0.7) | 38 (1.1) | 67 (1.2) | 56 (0.9) | 41 (0.7) | 0.4911† |
Data are presented as n and (percentage) from the total number of AEDs per year.
AED automated external defibrillator.
p for trend test. †Negative slope indicated a decreasing linear trend.
Non-usable AEDs and their causes.
| Parameter | Total (N = 23,619) | 2013 (n = 3134) | 2014 (n = 3402) | 2015 (n = 5622) | 2016 (n = 5909) | 2017 (n = 5552) | |
|---|---|---|---|---|---|---|---|
| Non-ready-to-use AEDsa | 3646 (15.44) | 524 (16.72) | 354 (10.41) | 646 (11.49) | 1302 (22.03) | 820 (14.77) | < 0.0001 |
| Bad battery status | 677 (17.20) | 138 (26.34) | 113 (31.92) | 159 (24.61) | 133 (10.22) | 134 (16.34) | < 0.0001† |
| AED malfunction | 807 (22.13) | 187 (35.69) | 120 (33.90) | 159 (24.61) | 164 (12.60) | 177 (21.59) | < 0.0001† |
| Invalid electrode | 3051 (83.68) | 322 (61.45) | 285 (80.51) | 509 (78.79) | 1224 (94.01) | 711 (86.71) | < 0.0001 |
| Less than 24-h usabilityb | 10,422 (44.13) | 1235 (39.41) | 1096 (32.22) | 2187 (38.90) | 3207 (54.27) | 2717 (48.94) | < 0.0001 |
| Bad battery status | 677 (6.50) | 138 (11.17) | 113 (10.31) | 159 (7.27) | 133 (6.08) | 134 (4.93) | < 0.0001† |
| AED malfunction | 810 (7.77) | 187 (15.14) | 120 (10.95) | 159 (7.27) | 164 (7.05) | 177 (6.51) | < 0.0001† |
| Invalid electrode | 3051 (29.27) | 322 (26.07) | 285 (26.00) | 509 (23.27) | 1224 (55.97) | 711 (26.17) | < 0.0001 |
| Less than 24-h accessibility | 7848 (75.30) | 1064 (86.15) | 902 (82.30) | 1780 (81.39) | 2069 (94.60) | 2033 (74.83) | < 0.0001 |
Includes cases with duplicate causes.
†Negative slope indicated a decreasing linear trend.
aAEDs with bad battery status, malfunction and/or invalid electrode.
bAEDs with bad battery status, malfunction, invalid electrode and/or less than 24-h accessibility.
Figure 2The proportion of non-usable AED and their causes. (A) The proportion of non-usable AED. (B) Causes of non-usable AED. AED, automated external defibrillator.
AEDs with less than 24-h usability according to the installation site.
| Facility type | Total (N = 10,439) | 2013 (n = 1235) | 2014 (n = 1096) | 2015 (n = 2187) | 2016 (n = 3207) | 2017 (n = 2714) |
|---|---|---|---|---|---|---|
| Residential settings | 2835 (27.16) | 289 (23.40) | 255 (23.27) | 400 (18.29) | 1135 (35.39) | 756 (27.86) |
| Multi-use facilities | 1925 (18.44) | 283 (22.91) | 132 (12.04) | 418 (19.11) | 451 (14.06) | 641 (23.62) |
| Schools | 1704 (16.32) | 41 (3.32) | 185 (16.88) | 411 (18.79) | 938 (29.25) | 129 (4.75) |
| Public buildings | 1703 (16.31) | 401 (32.47) | 67 (6.11) | 634 (28.99) | 232 (7.23) | 369 (13.60) |
| Welfare facilities | 705 (6.75) | 106 (8.58) | 194 (17.70) | 5 (0.23) | 115 (3.59) | 285 (10.50) |
| Transportation facilities | 414 (3.97) | 0 | 0 | 150 (6.86) | 1 (0.03) | 263 (9.69) |
| Police offices | 366 (3.51) | 13 (1.05) | 43 (3.92) | 3 (0.14) | 161 (5.02) | 146 (5.38) |
| Industrial facilities | 207 (1.98) | 21 (1.70) | 0 | 66 (3.02) | 67 (2.09) | 53 (1.95) |
| Public health/medical clinics | 197 (1.89) | 49 (3.97) | 40 (3.65) | 11 (0.50) | 63 (1.96) | 34 (1.25) |
| Religious facilities | 110 (1.05) | 2 (0.16) | 5 (0.46) | 77 (3.52) | 11 (0.34) | 15 (0.55) |
| Hotels and conference venues | 35 (0.34) | 19 (1.54) | 5 (0.46) | 4 (0.18) | 3 (0.09) | 4 (0.15) |
| Prisons | 4 (0.04) | 1 (0.08) | 1 (0.09) | 0 | 0 | 2 (0.07) |
| Other | 234 (2.24) | 10 (0.81) | 169 (15.42) | 8 (0.37) | 30 (0.94) | 17 (0.63) |
Data are presented as n and (percentage) from the total number of AEDs per year.
AED automated external defibrillator.