Analysis of automated external defibrillator device failures reported to the Food and Drug Administration.

Lawrence A DeLuca1, Allan Simpson, Dan Beskind, Kristi Grall, Lisa Stoneking, Uwe Stolz, Daniel W Spaite, Ashish R Panchal, Kurt R Denninghoff.   

Abstract

STUDY
OBJECTIVE: Automated external defibrillators are essential for treatment of cardiac arrest by lay rescuers and must determine when to shock and if they are functioning correctly. We seek to characterize automated external defibrillator failures reported to the Food and Drug Administration (FDA) and whether battery failures are properly detected by automated external defibrillators.
METHODS: FDA adverse event reports are catalogued in the Manufacturer and User Device Experience (MAUDE) database. We developed and internally validated an instrument for analyzing MAUDE data, reviewing all reports in which a fatality occurred. Two trained reviewers independently analyzed each report, and a third resolved discrepancies or passed them to a committee for resolution.
RESULTS: One thousand two hundred eighty-four adverse events were reported between June 1993 and October 2008, of which 1,150 were failed defibrillation attempts. Thirty-seven automated external defibrillators never powered on, 252 failed to complete rhythm analysis, and 524 failed to deliver a recommended shock. In 149 cases, the operator disagreed with the device's rhythm analysis. In 54 cases, the defibrillator stated the batteries were low and in 110 other instances powered off unexpectedly. Interrater agreement between reviewers 1 and 2 ranged by question from 69.0% to 98.6% and for most likely cause was 55.9%. Agreement was obtained for 93.7% to 99.6% of questions by the third reviewer. Remaining discrepancies were resolved by the arbitration committee.
CONCLUSION: MAUDE information is often incomplete and frequently no corroborating data are available. Some conditions not detected by automated external defibrillators during self-test cause units to power off unexpectedly, causing defibrillation delays. Backup units frequently provide shocks to patients.
Copyright © 2011 American College of Emergency Physicians. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21872969     DOI: 10.1016/j.annemergmed.2011.07.022

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  3 in total

1.  A simple method for removing initial irregularity of an electrocardiogram during a transient state of a power supply in a defibrillator.

Authors:  Jeong-Han Yi; Ki-Han Kim; Jin-Su Ahn; Hyung-Sik Kim
Journal:  Technol Health Care       Date:  2020       Impact factor: 1.285

2.  Assessment of Canadian Public Automated External Defibrillator Registries.

Authors:  Laurence d'Amours; Daniel Negreanu; Joel Neves Briard; François de Champlain; Valérie Homier
Journal:  CJC Open       Date:  2020-12-26

3.  Trends in maintenance status and usability of public automated external defibrillators during a 5-year on-site inspection.

Authors:  Tae Youn Kim; Yun-Kyoung Jung; Sun Hwa Yoon; Sun Ju Kim; Kyoung-Chul Cha; Woo Jin Jung; Young Il Roh; Soyeong Kim; Sung Hwa Kim; Dae Ryong Kang; Sung Oh Hwang
Journal:  Sci Rep       Date:  2022-06-24       Impact factor: 4.996

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.