| Literature DB >> 35747637 |
Muhammad J Abbas1, Malik E Dancy2, Erick M Marigi2, Lafi S Khalil3, Toufic R Jildeh4, Patrick J Buckley1, Javair Gillett5, William Burgos5, Kelechi R Okoroha2.
Abstract
Purpose: To evaluate the efficacy of an automated pneumatic torniquet pump and its ability to automatically calculate the limb occlusion pressure (LOP), as compared with the manual Doppler ultrasound technique.Entities:
Year: 2022 PMID: 35747637 PMCID: PMC9210466 DOI: 10.1016/j.asmr.2022.04.002
Source DB: PubMed Journal: Arthrosc Sports Med Rehabil ISSN: 2666-061X
Fig 1(A) Photograph of the upper extremity cuff placed distal to the axilla. Cuff is connected to pneumatically to the Automated SmartCuffs PRO for the calculation of the limb occlusion pressure. (B) Photograph of the lower extremity cuff placement was just distal to the gluteal fold. Cuff is connected to pneumatically to the Automated SmartCuffs PRO for the calculation of the limb occlusion pressure.
Patient Demographic Variables (N = 48)
| Variable | Mean (SD) | Range |
|---|---|---|
| Age, y | 37.1 (14.7) | 20-68 |
| Weight, lbs | 163.8 (35.2) | 105-277 |
| Height, inches | 67.1 (4.12) | 60-76 |
| BMI | 25.5 (3.8) | 18.1-33.5 |
| UE circumference, inches | 30.6 (5.2) | 21-44 |
| LE circumference, inches | 54.3 (7.6) | 39-70 |
| Sex | ||
| Male | 33% | |
| Female | 67% | |
| Race | ||
| White | 58% | |
| African American | 22% | |
| Hispanic | 12% | |
| Other | 8% |
BMI, body mass index; LE, lower extremity; UE, upper extremity; SD, standard deviation.
Limb Occlusion Pressure Difference
| Variable | N | Minimum Difference | Maximum Difference | Mean Difference | Standard Deviation | Paired |
|---|---|---|---|---|---|---|
| Upper | 48 | –16 | 34 | 13 | 14 | 0.29 |
| Lower | 48 | –48 | 48 | 9 | 20 | 0.09 |
NOTE. Values are in mm Hg.
Fig 2Bland–Altman plot of limb occlusion pressure (LOP) difference between proposed technique and manual Doppler technique for the upper extremity (UE). Mean difference is shown (zero bias line) plus or minus 2 standard deviations (95% confidence interval).
Fig 3Histogram of limb occlusion pressure (LOP) difference between proposed technique and manual Doppler technique for the upper extremity (UE).
Fig 4Bland–Altman plot of limb occlusion pressure (LOP) difference between proposed technique and manual Doppler technique for the lower extremity (LE). Mean difference is shown (zero bias line) plus or minus two standard deviations (95% confidence interval).
Fig 5Histogram of limb occlusion pressure (LOP) difference between proposed technique and manual Doppler technique for the lower extremity.