Tourniquet safety in lower leg applications.

James A McEwen1, Deborah L Kelly, Theda Jardanowski, Kevin Inkpen.   

Abstract

PURPOSE: To reduce the chance of injury due to pneumatic tourniquet use, the minimum cuff pressure required to maintain a bloodless field should be used. The purpose of this study was to find out if Limb Occlusion Pressure (LOP--the cuff pressure required to occlude arterial flow) is lower with a wide contoured cuff than with a standard width cylindrical cuff at the calf, if cuff pressures based on measured LOP will be lower than the typical 250 mmHg used in lower leg cuffs, and if a new automatic LOP measurement method gives the same results as the standard Doppler stethoscope method. SAMPLE: 16 adult volunteers were tested in a controlled laboratory setting, and 53 clinical cases were reviewed at two centers.
DESIGN: Repeated measures comparison of LOP on volunteers with the two different cuffs and measurement methods, and review of clinical cases.
RESULTS: LOP was lower with the wide cuff on all volunteers (mean reduction 20 mmHg, SD 8.6, range 5-35, p < 0.001). The average difference of 1.2 mmHg between Doppler and automatic LOP readings was not significant (p = 0.43). Based on the volunteer results, using LOP plus a safety margin of 40, 60, or 80 mmHg (for LOP < 130, 131-190, or 190+ respectively) with a standard width cylindrical cuff will lead to an average cuff pressure of 223 mmHg (range 170-299, SD 36), 11% lower than typical practice and up to 80 mmHg (32%) lower on some patients. Using a wide, contoured cuff should further reduce cuff pressures to an average of 195 mmHg (range 160-280, SD 33), 22% lower than typical practice and a reduction of up to 90 mmHg (36%). At two clinics, the wide cuff maintained a bloodless field in 48 out of 53 cases (91%) when used at 200 mmHg.
CONCLUSIONS: Using a wide, contoured cuff at the calf should reduce required cuff pressures compared to a standard cuff. Setting cuff pressure based on LOP should further reduce cuff pressures for most patients compared to typically used pressures. With continued development, the new automatic method may become a viable alternative to the Doppler method and may make LOP measurement more practical in the clinical setting.

Entities:  

Mesh:

Year:  2002        PMID: 12432700     DOI: 10.1097/00006416-200209000-00009

Source DB:  PubMed          Journal:  Orthop Nurs        ISSN: 0744-6020            Impact factor:   0.913


  13 in total

1.  [Prehospital application of tourniquets for life-threatening extremity hemorrhage : Systematic review of literature].

Authors:  B Hossfeld; R Lechner; F Josse; M Bernhard; F Walcher; M Helm; M Kulla
Journal:  Unfallchirurg       Date:  2018-07       Impact factor: 1.000

2.  Aqueous based Povidone-iodine related chemical burn under the tourniquet (a case report) and literature review.

Authors:  Challa Supradeeptha; Sudhir Mahadev Shandilya; Alavalapati Naresh; Jonnalagadda Satyaprasad
Journal:  J Orthop       Date:  2013-07-23

3.  [Update on gunshot wounds to extremities].

Authors:  F von Lübken; G Achatz; B Friemert; M Mauser; A Franke; E Kollig; D Bieler
Journal:  Unfallchirurg       Date:  2018-01       Impact factor: 1.000

4.  Investigation of clinically acceptable agreement between two methods of automatic measurement of limb occlusion pressure: a randomised trial.

Authors:  Luke Hughes; James McEwen
Journal:  BMC Biomed Eng       Date:  2021-05-08

5.  Tourniquet associated chemical burn.

Authors:  Jae-Hyuk Yang; Hyungtae Lim; Jung-Ro Yoon; Hyeon-Il Jeong
Journal:  Indian J Orthop       Date:  2012-05       Impact factor: 1.251

6.  Blood flow restriction pressure recommendations: a tale of two cuffs.

Authors:  Jeremy P Loenneke; Christopher A Fahs; Lindy M Rossow; Robert S Thiebaud; Kevin T Mattocks; Takashi Abe; Michael G Bemben
Journal:  Front Physiol       Date:  2013-09-10       Impact factor: 4.566

7.  Efficacy and safety of a new elastic tourniquet cuff in total knee arthroplasty: a prospective randomized controlled study.

Authors:  O-Sung Lee; Myung Chul Lee; Hyuk-Soo Han
Journal:  Biomed Eng Online       Date:  2017-08-08       Impact factor: 2.819

8.  Elastic pneumatic tourniquet cuff can reduce postoperative thigh pain after total knee arthroplasty: a prospective randomized trial.

Authors:  Jae-Young Park; Sung Eun Kim; Myung Chul Lee; Hyuk-Soo Han
Journal:  BMC Musculoskelet Disord       Date:  2020-08-21       Impact factor: 2.362

9.  Efficacy of modified pressure cuff for thrombolytic treatment on lower extremity deep venous thrombosis.

Authors:  Yan Li; Janesya Sutedjo; Yu-Chen Chen; Jian-Ping Gu
Journal:  Medicine (Baltimore)       Date:  2021-04-30       Impact factor: 1.817

10.  Efficacy of blood flow restriction exercise during dialysis for end stage kidney disease patients: protocol of a randomised controlled trial.

Authors:  Matthew J Clarkson; Steve F Fraser; Paul N Bennett; Lawrence P McMahon; Catherine Brumby; Stuart A Warmington
Journal:  BMC Nephrol       Date:  2017-09-11       Impact factor: 2.388

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.